LYMA Laser

K210823 · Lyma Life , Ltd. · OHS · Mar 18, 2022 · General, Plastic Surgery

Device Facts

Record IDK210823
Device NameLYMA Laser
ApplicantLyma Life , Ltd.
Product CodeOHS · General, Plastic Surgery
Decision DateMar 18, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The LYMA Laser is an Over the-Counter (OTC) light based device intended for the use in treating full-face wrinkles.

Device Story

LYMA Laser is an OTC light-based device for treating full-face wrinkles. Device incorporates near-infrared (808 nm) and red (620 nm) LED light sources. Optical lens diffuses light to produce narrowband, divergent distribution across an 8 cm² treatment area. Energy output is 62.5 mW/cm². Device is intended for patient self-use. Patient applies device to skin for 3 minutes per treatment area. Mechanism involves Low Level Light Therapy (LLLT) to provide beneficial treatment for wrinkles. Device is handheld; no complex software or algorithmic processing described.

Clinical Evidence

No clinical data provided. Evidence consists of bench testing (electrical safety per AAMI/ANSI ES60601-1, EMC per IEC 60601-1-2, laser safety per IEC 60825-1, and biocompatibility per ISO 10993-1) and a usability study. Usability study (n=26) demonstrated 88% success rate for self-selection, labeling comprehension, and proper device operation, meeting the 80% success threshold.

Technological Characteristics

Light-based device using 808 nm (near-infrared) and 620 nm (red) LEDs. Energy output 62.5 mW/cm². Optical lens for light diffusion. Treatment area 8 cm². Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, and IEC 60825-1. Biocompatibility evaluated per ISO 10993-1.

Indications for Use

Indicated for the treatment of full-face wrinkles in adult patients via over-the-counter use.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 18, 2022 LYMA Life Ltd. % Allison Komiyama Principal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Avenue, Suite B-257 San Diego, California 92110 Re: K210823 Trade/Device Name: LYMA Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: February 9, 2022 Received: February 11, 2022 Dear Allison Komiyama: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210823 Device Name LYMA Laser Indications for Use (Describe) The LYMA Laser is an Over the-Counter (OTC) light based device intended for the use in treating full-face wrinkles. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # L Y M A # 510(k) Summary K210823 #### DATE PREPARED March 17, 2022 #### MANUFACTURER AND 510(k) OWNER LYMA Life Ltd 93 Gloucester Place Marylebone, London W1U 6JQ, United Kingdom Telephone: +44 (0)7950 396247 Official Contact: Simon Goff, CEO ## REPRESENTATIVE/CONSULTANT Allison C. Komiyama, Ph.D., R.A.C. Matthieu Kirkland, M.S. RQM+, LLC Telephone: +1 (619) 458-9547 Email: akomiyama@RQMplus.com, mkirkland@RQMplus.com #### DEVICE INFORMATION | Proprietary Name/Trade Name: | LYMA Laser | |------------------------------|----------------------------------------------------------| | Regulation Name: | Laser surgical instrument for use in general and plastic | | | surgery and in dermatology | | Regulation Number: | 21 CFR 878.4810 | | Class: | II | | Product Code: | OHS | | Review Panel: | General & Plastic Surgery | | Premarket Review: | Surgical and Infection Control Devices (OHT4) | | | General Surgery Devices (DHT4A) | #### PREDICATE DEVICE IDENTIFICATION The LYMA Laser is substantially equivalent to the following predicates: | 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate | |---------------|------------------------------------------------------|-------------------| | K120775 | LightStim for Wrinkles / LED Intellectual Properties | ✓ | | K120560 | Trinity Wrinkle Remover / Carol Cole Company | | The predicate devices have not been subject to a design related recall. {4}------------------------------------------------ ## DEVICE DESCRIPTION The LYMA Laser is an Over-the-Counter (OTC) light based device intended for the use in treating full-face wrinkles. The LYMA Laser incorporates a near-infrared LED light source within the optimum Therapeutic Optical Window at 808 nm, optimized for Low Level Light Therapy, and a red LED light source at 620 nm. An optical lens that diffuses the light to produce narrowband, divergent distribution across an 8 cm² treatment area. The energy output of the 808 nm and 620 nm wavelengths is 62.5 mW/cm² which produces a beneficial treatment for full-face wrinkles. # INDICATIONS FOR USE The LYMA Laser is an Over-the-Counter (OTC) light based device intended for the use in treating full-face wrinkles. # COMPARISON OF TECHNOLOGICAL CHARACTERISTICS LYMA Life believes that the LYMA Laser is substantially equivalent to the predicate devices based on the information summarized here: - The subject device has the same intended use as the predicates (i.e., the treatment of full-face wrinkles) - The subject device has a similar output as the predicates (i.e., 62.5 mW/cm²) ● - The subject device utilizes a similar near-infrared treatment wavelength and a similar red light wavelength as the predicates (i.e., 808 nm and 620 nm) - . The subject device utilizes the same treatment duration as the predicates (i.e., 3 minutes per treatment area) - The subject device utilizes a similar irradiation, or treatment, area as the predicates (i.e., 8 cm²) We believe we have identified appropriate predicate devices with the same intended use, same or similar technological characteristics and the same treatment parameters. Therefore, this information supports the determination of substantial equivalence in safety and effectiveness between the subject and predicate devices. ## SUMMARY OF NON-CLINICAL TESTING No FDA performance standards have been established for the LYMA Laser. The following tests were performed to demonstrate safety based on current industry standards: Electromagnetic Compatibility and Electrical Safety: The subject device was tested in compliance to AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2(2010)/(R)2012 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance (which complies with IEC 60601-1); and IEC 60601-1-2 Medical electrical equipment — Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests. {5}------------------------------------------------ Additionally, testing to IEC 60825-1 Safety of laser products - Part 1: Equipment classification and requirements and an evaluation per ISO 10993-1 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process were provided. ## SUMMARY OF USABILITY TESTING To demonstrate that lay users can correctly operate the LYMA Laser, a usability study was completed. The study was split into two Parts. Part A examined the participants' abilities correctly self-identify whether they are appropriate users of the device based on the outer packaging (also known as the box label). Part B evaluated how well participants could comprehend the instructions for use and correctly operate the subject device. This usability study evaluated the suitability of the LYMA Laser for over-the-counter (OTC) distribution. The ability of participants to correctly self-identify whether they are appropriate users of the device, comprehend labeling, and safely and effectively use the subject device was assessed in two independent parts of the study. The number of participants required to answer all questions correctly in each part was 21 out of 26, for an 80% success rate. The results of the usability testing (passing rates of 88% for the Self-Selection as well as the Labeling Comprehension and Proper Use of the Device sections) demonstrate that the LYMA Laser is suitable for OTC distribution and that lay users can correctly self-select themselves as users of the subject device. #### CONCLUSION Based on the testing performed and usability testing, it can be concluded that the subject device does not raise different questions of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed LYMA Laser are assessed to be substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...