The CX50N(CX50YQS) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. It is specifically designed for digital mammography applications.
Device Story
The CX50N(CX50YQS) is a 5MP high-resolution LCD monitor system used by trained medical practitioners in clinical settings to display digital medical images. The system comprises an LCD monitor and a high-resolution graphic control board installed into a PACS workstation. It receives digital image data from the PACS workstation via DVI-I or DisplayPort interfaces and renders grayscale and color images for diagnostic review. The device does not contact the patient and does not control life-sustaining equipment; it serves as a visualization tool for clinicians to interpret medical imagery. By providing high-resolution display capabilities, it assists practitioners in the analysis of digital mammography images, facilitating clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only. Compliance with IEC 60601-1 and IEC 60601-1-2 standards for safety and electromagnetic compatibility. Performance testing for resolution, luminance, contrast, and noise met all acceptance criteria.
Technological Characteristics
5MP (2560 x 2048) TFT LCD monitor; 21.3" panel size; 0.165mm pixel pitch; 800cd/m² brightness; 800:1 contrast ratio; DVI-I and DisplayPort inputs; 100-240 VAC power supply. Complies with IEC 60601-1 and IEC 60601-1-2. Standalone display system connected to PACS workstation.
Indications for Use
Indicated for trained medical practitioners to display and view digital medical images for review and analysis, specifically for digital mammography applications.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
5MP Color LCD Monitor CCL550i2 (CL21550) (K160326)
Related Devices
K170783 — CX50N(CX50YQS) · Wide Corporation · Mar 31, 2017
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 7, 2016
WIDE Corporation % Mr. YeoJin Yun RA Manager 12 Wongomae-Ro, Giheung-Gu Yongin-Si, Gyeonggi-Do 17086 REPUBLIC OF KOREA
Re: K160351 Trade/Device Name: CX50N(CX50YOS) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: February 5, 2016 Received: February 8, 2016
Dear Mr. Yun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160351
Device Name CX50N(CX50YQS)
Indications for Use (Describe)
The CX50N(CX50YQS) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. It is specifical mammography applications.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92
#### 1. Date Prepared [21 CFR 807.92(a) (1)]
02/05//2016
# 2. Submitter's Information [21 CFR 807.92(a) (1)]
| Name of Sponsor: | WIDE Corporation. |
|-----------------------|--------------------------------------------------------------------------------|
| Address: | 12 Wongomae-Ro, Giheung-Gu, Yongin-Si, Gyeonggi-Do<br>17086, Republic of Korea |
| Contact Name: | YeoJin Yun |
| | Telephone #: +82-31-218-1675 |
| | Fax #: +82-31-218-7400 |
| | Email: yyjin@widecorp.com |
| Registration Number: | 3004082357 |
| Name of Manufacturer: | Same as Sponsor |
# 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
| Model Name: | CX50N(CX50YQS) |
|----------------------|--------------------------------|
| Common Name: | TFT LCD Medical Monitor System |
| Classification Name: | Display, Diagnostic Radiology |
| Regulation Number: | 21 CFR 892.2050 |
| Product Code: | PGY |
| Device Class: | 2 |
| Review Panel: | Radiology |
# 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
| 510(k) Number: | K160326 |
|----------------------|------------------------------------------|
| Applicant: | JVC KENWOOD Corporation |
| Model Name: | 5MP Color LCD Monitor CCL550i2 (CL21550) |
| Common Name: | TFT LCD Medical Monitor System |
| Classification Name: | Display, Diagnostic Radiology |
| Regulation Number: | 21 CFR 892.2050 |
| Product Code: | PGY |
| Device Class: | 2 |
{4}------------------------------------------------
#### 5. Description of the Device [21 CFR 807.92(a) (4)]
CX50N(CX50YQS) is a flat panel hi-resolution(5MP) LCD monitor system for displaying digital medical images. The system consists of a state-of-the-art LCD monitor and a high-resolution graphic control board that connects to a PACS workstation for grayscale and color image display. The WIDE controller board is installed into the PACS workstation computer or other computer system to display PACS medical images.
# 6. Intended Use [21 CFR 807.92(a) (5)]
The CX50N(CX50YQS) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. It is specifically designed for digital mammography applications.
# 7. Technological Characteristics [21 CFR 807.92(a) (6)]
The device is an image display system which consists of computer software and components. The device does not contact the patient, nor does it control any life sustaining devices. A physician or trained medical practitioner provides ample opportunity for competent human intervention to interpret images and information being displayed.
| Parameter | Subject Device | Predicate Device |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Unknown | K160326 |
| Model Name | CX50N(CX50YQS) | 5MP Color LCD Monitor CCL550i2 (CL21550) |
| Manufacturer | WIDE Corporation. | JVC KENWOOD Corporation |
| Common Name | | TFT LCD Medical Monitor System |
| Classification Name | | Display, Diagnostic Radiology |
| Classification Panel | | Radiology |
| Classification Regulation | | 21 CFR 892.2050 |
| Product Code | | PGY |
| Device Class | | Class II |
| Intended Use | The CX50N(CX50YQS) LCD<br>Monitor System is intended to be<br>used in displaying and viewing<br>digital medical images for review<br>and analysis by trained medical<br>practitioners. It is specifically<br>designed for digital<br>mammography applications. | 5MP Color LCD Monitor<br>CCL550i2 (CL21550) are<br>intended to be used in displaying<br>and viewing medical images for<br>diagnosis by trained medical<br>practitioners.<br>They are to be used in digital<br>mammography PACS, modalities<br>including FFDM, and breast<br>tomosynthensis Type of<br>tomosynthensis. |
# 8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92] K131137
{5}------------------------------------------------
| Parameter | Subject Device | Predicate Device |
|------------------|----------------------|----------------------|
| LCD Panel Size | 21.3" | 21.3" |
| Resolution | 2560 x 2048 | 2560 x 2048 |
| Pixel pitch | 0.165mm x 0.165mm | 0.165mm x 0.165mm |
| Brightness | 800cd/m² | 1000cd/m² |
| Contrast Ratio | 800 : 1 | 1300 : 1 |
| Input Signal | DVI-I, DisplayPort | DVI-I, DisplayPort |
| Power Supply | 100~240 VAC, 50/60Hz | 100~240 VAC, 50/60Hz |
| Color/Monochrome | Color | Color |
When compared to the predicate devices (K160326), the CX50N(CX50YQS) presented in this submission has the same of the followings:
- Technological characteristics
- LCD Panel Size
- Resolution
- Pixel pitch
The two devices share the similar performance as the following:
- Intended Use
- CX50N(CX50YQS) does not support the display of breast tomosynthensis for diagnosis.
- · Brightness
- Contrast Ratio
#### 9. Summary of Non-Clinical Data
CX50N(CX50YQS) comply with the following international and FDA-recognized consensus standards:
> Electrical Equipment --- Part - 1: IEC 60601-1: Medical General Requirements For Basic Safety And Essential Performance IEC 60601-1-2: Medical Electrical Equipment - - Part 1-2: General Requirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Compatibility -Requirements And Tests
The tests conducted for resolution, luminance, contrast and noise all met the acceptance criteria specified in the standards.
#### 10. Summary of Clinical Data
No clinical studies were considered necessary and performed.
{6}------------------------------------------------
#### 11. Conclusion [21 CFR 807.92(b) (3)]
Subject Device is substantially equivalent to the currently marketed and predicate devices in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
Additionally, the safety of the subject device was validated through tests including IEC60601-1 and IEC 60601-1-2. The effectiveness of the device was validated through tests for resolution, luminance, contrast and noise.
The results of these tests demonstrate that CX50N(CX50YQS) meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing demonstrates that the device is as safe and effective as the predicate device and performs as well as the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.