JUSHA-C33C LCD MONITOR

K141684 · Nanjing Jusha Display Technology Co., Ltd. · PGY · Aug 1, 2014 · Radiology

Device Facts

Record IDK141684
Device NameJUSHA-C33C LCD MONITOR
ApplicantNanjing Jusha Display Technology Co., Ltd.
Product CodePGY · Radiology
Decision DateAug 1, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2

Intended Use

JUSHA-C33C LCD Monitor is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis.

Device Story

JUSHA-C33C is a 21.3" color TFT LCD diagnostic monitor; inputs digital image signals via DVI or DisplayPort; processes signals using Color/Gray Automatic (CGA) function to identify signal type and apply appropriate Look-Up Tables (DICOM for gray, GAMMA 2.2 for color); features ambient brightness adaptation and presence induction for automatic display adjustment; outputs high-resolution (2048x1536) images for clinical diagnosis; used by radiologists/physicians in clinical settings; assists in diagnostic decision-making by providing calibrated, high-luminance (600 cd/m²) visualization of medical images.

Clinical Evidence

No clinical data; substantial equivalence supported by bench testing, including risk analysis, design reviews, module/system verification, and final validation testing.

Technological Characteristics

21.3" Color TFT LCD panel; 2048x1536 resolution; 600 cd/m² luminance; 12-bit color expansion; 8 DICOM LUTs and 3 GAMMA LUTs; ambient brightness/presence sensors; DVI/DisplayPort connectivity; LED backlight; complies with IEC 60601-1 and IEC 60601-1-2.

Indications for Use

Indicated for trained medical practitioners to display and view digital images for diagnosis of X-ray or MRI. Not indicated for mammography image display.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 1. 2014 Nanjing Jusha Display Technology Co., Ltd. % Ms. Li Le Certification Manager 8A. Block 1. Nanjing International Service Outsourcing Mansion, No. 301 Hanzhongmen Street, Nanjing 210036 CHINA Re: K141684 Trade/Device Name: Jusha-C33C LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: June 24, 2014 Received: June 30, 2014 Dear Ms. Le: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ![FDA signature](https://i.imgur.com/9999999.png) for Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) #### K141684 Device Name JUSHA-C33C LCD Monitor Indications for Use (Describe) JUSHA-C33C LCD Monitor is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) FDA This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary In accordance with 21 CFR 807.92 the following summary of information is provided: | Date: | May 6, 2014 | | | |---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | Submitter: | Nanjing Jusha Display Technology Co., Ltd | | | | | Add: 301, 8F Block A, No.1, Nanjing International Service<br>Outsourcing Mansion, Hanzhongmen Street, Nanjing, 210036 China | | | | Contact Person: | Li le | | | | | Certification Manager | | | | | Nanjing Jusha Display Technology Co., Ltd | | | | | Tel: +86-25- 83305050 | | | | | Fax: +86-25- 58783271 | | | | Device Trade Name: | JUSHA-C33C LCD Monitor | | | | Common/Usual Name: | 3MP Color LCD Monitor | | | | Classification Name: Product<br>Code: | Display, Diagnostic Radiology 21CFR 892.2050<br>PGY | | | | Predicate Device(s): | EIZO RX340;K113562 | | | | Device Description: | JUSHA-C33C LCD Monitor is the display system with the high<br>resolution (2048*1536), high luminance (600 cd/m²), and10 bit<br>display colors expanded to 12 bit color expansion technology, 8<br>DICOM look up table and 3 GAMMA look up table inside. In<br>particular, JUSHA-C33C LCD Monitor contains CGA function, it is<br>specially made by JUSHA, it can automatic identify gray and color<br>signals, then gray area calls DICOM LUT and color area calls<br>GAMMA2.2 LUT. JUSHA-C33C has ambient brightness adapting<br>and presence induction system, with these this display can automatic<br>adjustment according to different requirements in order to achieve the<br>best results. | | | | | The product is consisted of the following components: | | | | | - 21.3" Color TFT LCD Panel | | | | | - JUSHA-C33/45T motherboard/FR-4/REV:0.9 | | | | | - JUSHA-C33C LCD Monitor software | | | | | - Power Adapter | | | | | - Data Cable. | | | | | The LCD Monitor is designed, tested, and will be manufactured in<br>accordance with both mandatory and voluntary standards: | | | | | IEC 60601-1Medical equipment medical electrical<br>1.<br>equipment - Part 1: General requirements for basic safety<br>and essential performance 1988+A1 : 1991 + A2:1995 | | | | | IEC 60601-1-2 Edition 3:2007, Medical electrical<br>2.<br>equipment - Part 1-2: General requirements for basic | | | {4}------------------------------------------------ | safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests. | | |-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use: | JUSHA-C33C LCD Monitor is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis. | | Technology: | JUSHA-C33C LCD Monitor is the display system with the high resolution monitor (3 megapixels) with electronic capabilities for evaluation of high resolution medical images, high luminance (600 cd/m²) and10 bit display colors expanded to 12 bit color expansion technology, 8 DICOM look up table and 3 GAMMA look up table inside. In particular, JUSHA-C33C LCD Monitor contains CGA function, it is specially made by JUSHA, it can automatic identify gray and color signals, then gray area calls DICOM LUT and color area calls GAMMA2.2 LUT. JUSHA-C33C has ambient brightness adapting and presence induction system, with these this display can automatic adjustment according to different requirements in order to achieve the best results. | | Determination of Substantial Equivalence: | Summary of Non-Clinical Tests:<br>The LCD Monitor complies with voluntary standards as following:<br>1 IEC 60601-1Medical equipment medical electrical equipment - Part 1: General requirements for basic safety and essential performance 2005+CORR.1(2006)+CORR.2(2007)<br>2 IEC 60601-1-2 Edition 3:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.<br>JUSHA-C33C is substantially equivalent to EIZO RX340. JUSHA-C33C employs the maximum resolution values same as that of EIZO RX340. Comparison table of the principal characteristics of 2 devices is shown in the Attachment 1.<br>The following quality assurance measures were applied to the development of the system:<br>Risk AnalysisRequirements ReviewsDesign ReviewsRaw materials verificationTesting on unit level (Module verification)Integration testing (System verification)Final acceptance testing (Validation)Performance testing (Verification)Safety testing (Verification) | | | <b>Summary of Clinical Tests:</b><br>The subject of this premarket submission, LCD Monitor, did not<br>require clinical studies to support substantial equivalence. | | | The proposed device is Substantially Equivalent (SE) to the predicate<br>device which is US legally market device. Therefore, the subject<br>device is determined as safe and effectiveness. | | Conclusion: | Nanjing Jusha Display Technology Co., Ltd Considers the JUSHA-<br>C33C LCD Monitor to be as safe, as effective, and performance is<br>substantially equivalent to the predicate device(s). | {5}------------------------------------------------ ### 510(k) Premarket Notification Submission ## Attachment 1: | Attributes | Predicate Device | Proposed Device | |---------------------------|-------------------------------------------|-------------------------------------------| | Product | EIZO RX340 | JUSHA-C33C LCD Monitor | | 510(k) Number | K113562 | K141684 | | Screen technology | 21.2" Color TFT LCD Panel | 21.3" Color TFT LCD Panel | | Viewing angle (H, V) | Horizontal 170°, Vertical 170° | Horizontal 176°, Vertical 176° | | Resolution | 2048 x 1536/1536x 2048 | 2048 x1536/1536x 2048 | | Display area | 323.7 (H) x431.6 (V) mm | 324.864 (H) x 433.152 (V) mm | | Contrast Ratio | 1400:1 | 1400:1 | | Scanning frequency (H; V) | 31~127 kHz;29~61.5 Hz | 96.7kHz;60Hz | | Recommended Luminance | 400cd/m² | 400cd/m² | | Pixel Pitch | 0.21075 x 0.21075 mm | 0.2115x0.2115 mm | | Backlight | LED | LED | | Display Colors | 10-bit, 1.07 billion colors | 12-bit, 68.7billion colors | | Luminance calibration | Built in calibration sensor<br>provided | Built in calibration sensor<br>provided | | Input signals | DVI standard 1.0,<br>DisplayPort 1.1a | DVI standard 1.0,<br>DisplayPort 1.1a | | Input terminational | DVI-D (dual link) x 1,<br>DisplayPort x 1 | DVI-D (dual link) x 1,<br>DisplayPort x 1 | | Display controller | Off the shelf | Off the shelf | | Power Requirement | AC 100~240V 50~60Hz | AC 100~240V 50~60Hz | {6}------------------------------------------------ | Power Consumption/Save<br>Mode | 56W/less than 3W | 65W/less than 3W | |-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Power Management | DVI DMPM<br>DisplayPort 1.1a | DVI DMPM<br>DisplayPort 1.1a | | USB Ports/standard | 1 upstream (endpoint),<br>2 downstream/ Rev. 2.0 | 1 upstream (endpoint),<br>2 downstream/ Rev. 2.0 | | Dimensions w/o stand<br>(W x H x D) | Without stand:<br>376mmx505mmx98 mm<br>With stand: | Without stand:<br>382mm x490mm x75mm With<br>stand:<br>382mm x533mm x238mm | | Applicable standard | 1 IEC 60601-1Medical<br>equipment medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance<br>2005+CORR.1(2006)+CORR.<br>2(2007)<br>2 IEC 60601-1-2 Edition<br>3:2007, Medical electrical<br>equipment - Part 1-2: General<br>requirements for basic safety<br>and essential performance -<br>Collateral standard:<br>Electromagnetic compatibility<br>- Requirements and tests | 1 IEC 60601-1Medical<br>equipment medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance<br>2005+CORR.1(2006)+CORR.2(<br>2007)<br>2 IEC 60601-1-2 Edition<br>3:2007, Medical electrical<br>equipment - Part 1-2: General<br>requirements for basic safety<br>and essential performance -<br>Collateral standard:<br>Electromagnetic compatibility -<br>Requirements and tests |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...