3MP COLOR LCD MONITOR, RADIFORCE RX340

K113562 · Eizo Nanao Corporation · LLZ · Jul 16, 2012 · Radiology

Device Facts

Record IDK113562
Device Name3MP COLOR LCD MONITOR, RADIFORCE RX340
ApplicantEizo Nanao Corporation
Product CodeLLZ · Radiology
Decision DateJul 16, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2

Intended Use

The RadiForce RX340 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis.

Device Story

RadiForce RX340 is a 3MP color LCD monitor for viewing medical images; excludes mammography. Inputs digital video signals via DVI or DisplayPort; transforms signals into calibrated visual output using internal lookup tables for DICOM GSDF compliance. Features LED backlight for luminance stability; includes integrated sensors: Backlight Sensor (BS) for stabilization, Built-in Front Sensor (IFS) for automatic grayscale calibration, Presence Sensor (PS) for power management, and Ambient Light Sensor (ALS) for ambient light measurement. Used in clinical settings by trained medical practitioners. RadiCS software provides quality control and calibration management. Device benefits include consistent image quality, reduced maintenance via automated calibration, and energy efficiency.

Clinical Evidence

Bench testing only. Performance validated against predicate device regarding image quality characteristics, DICOM GSDF compliance, and calibration accuracy using internal vs. external sensors. No clinical data presented.

Technological Characteristics

3MP color LCD monitor; 1536 x 2048 resolution; 0.21075 mm pixel pitch. LED backlight system. Connectivity: DVI and DisplayPort. Integrated sensors: Backlight Sensor (BS), Built-in Front Sensor (IFS), Presence Sensor (PS), Ambient Light Sensor (ALS). Factory-calibrated lookup tables for DICOM GSDF. Compliant with safety and EMC standards.

Indications for Use

Indicated for trained medical practitioners to display and view digital diagnostic images (e.g., X-ray, MRI). Contraindicated for the display of mammography images.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ .K.113562 JUL 16 2012 EIZO NANAO CORPORATION. 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 Name Department Hiroaki Hashimoto Medical System Standards Telephone Fax E-Mail +81 (76) 274-2468 +81 (76) 274-2484 hiro@nanao.co.jp Date November 28th, 2011 # Traditional 510(k) Summary (in accordance with 21 CFR 807.92) - · 1. Date of Summary November 28th, 2011 - 2. Company EIZO NANAO CORPORATION 153 Shimokashiwano, Hakusan Ishikawa 924-8566 Japan # 3. Authorized Contact Person Hiroaki Hashimoto # 4. Device Information - Trade Name/Model: RadiForce RX340 . - 3MP Color LCD Monitor Common Name: . - System, Image Processing, Radiological Classification Name: . - Classification Number: 21 CFR 892.2050. Product Code LLZ � #### 5. Predicate Device - Color LCD Monitor, RadiForce RX320 (K090045) . ## 6. Device Description The RadiForce RX340 is a color LCD monitor for viewing medical images other than those of mammography. The matrix size (or resolution) of the panel is 1,536 x 2,048 pixels (3MP) with a pixel pitch of 0.21075 mm. Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used. ಗ Section 05 {1}------------------------------------------------ Model variations with cosmetic differences are distinguished by characters attached to the name of the base model "RX340" such as "RX340-AR", a model with an Anti-Reflective coating although the hardware design, components and labeling remain unchanged. RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including RX340 based on several QC guidelines. The RadiCS and its subset, RadiCS LE are included in this 510(k) submission as an accessory to the RadiForce RX340. ## 7. Intended Use The RadiForce RX340 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis. ### 8. Technological Characteristics The RadiForce RX340 can be said to have at least the same display performances as those of the predicate device by default due to the following reasons: - The matrix sizes (1536 x 2048) and the active area sizes (323.7 mm x 431.6 mm) of a. the LCD panels used by the both devices are the same. - The DICOM calibrated luminance values of the both devices are the same (400 b. cd/m²) though the typical maximum luminance values (1000 cd/m²) is higher than that of the predicate device (900 cd/m2). The higher luminance to be maintained constantly was realized by the employment of LED backlight deteriorating more slowly than conventional CCFL backlights. - The LED backlight was newly employed instead of CCFL backlight because it is C. mercury-free, consumes less power and deteriorates more slowly. We have not recognized any adverse effects of the LED backlight on the quality of displayed images. Refer to section 18 "Performance Testing - Bench" where several image quality characteristics of the proposed device are compared with those of the predicate device. - The both devices display images in accordance with DICOM GSDF by default d. utilizing the factory calibrated display mode stored in lookup tables inside of them. - As for input video signal, in addition to the Digital Visual Interface (DVI) also e. supported by the predicate device, RadiForce RX340 supports DisplayPort, another type of digital video interface based on an industrial standard. As far as the both the DVI and the DisplayPort are digital, their differences do not affect the quality of displayed images. As for the maintenance, the same QC software is used for the both devices and the implementation of the Backlight Sensor (BS) stabilizing the backlight is also the same. As for built-in sensors, in addition to BS common to the both devices, RX340 has three kinds of sensors. However, only the Built-in Front Sensor (IFS) has something to do with the {2}------------------------------------------------ maintenance or the calibration; the Presence Sensor (PS) detects the absence of the user to trigger the power saving mode of the monitor and the Ambient Light Sensor (ALS) is used to measure the ambient light by lx. The IFS enables automatic grayscale calibration by measuring the luminance at the screen surface. Without IFS, the grayscale calibration process requires human intervention and the use of an external sensor. The accuracy data of the calibration with external sensors and that with the IFS is included in section 16.9 "Verification and Validation Documentation". The overall design of the RadiForce RX340 was validated in accordance with internationally recognized safety and EMC standards by independent testing facilities and in-house ones. Besides, EIZO NANAO CORPORATION performed a range of system and performance tests to ensure that the RadiForce RX340 performs in accordance with its specifications. None of the tests revealed behaviors inconsistent with the expected performance. #### 9. Conclusion The 3MP Color LCD Monitor, RadiForce RX340 is substantially equivalent to the predicate device with respect to technical characteristics, application and intended use. The specifications of the primary component employed by the proposed device are the same to those of the predicate device and other differences have been independently validated. Any differences between the devices do not affect safety or effectiveness. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird in flight, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. The logo is black and white. # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring. MD 20993-0002 Mr. Hiroaki Hashimoto Manager of Medical System Standards Division EIZO NANAO Corporation 153 Shimokashiwano HAKUSAN ISHIKAWA 924-8566 JAPAN JUL 16 2012 Re: K113562 Trade/Device Name: 3MP Color LCD Monitor, RadiForce RX340. Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 28, 2012 Received: July 5, 2012 # Dear Mr. Hashimoto: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, · Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {4}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for EIZO. The logo consists of a square made up of smaller squares on the left and the word "EIZO" on the right. The square is black and white, and the word "EIZO" is black. There is a registered trademark symbol next to the O in EIZO. # Indications for Use # 510(k) Number (if known): K11.3562 ## Device Name: 3MP Color LCD Monitor, RadiForce RX340 ## Indications For Use: The RadiForce RX340 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR . Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Imhild O'Hara Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety 510K K113562 Indications for Use Statement_RX340: Page 1 of 1 Section 04
Innolitics
510(k) Summary
Decision Summary
Classification Order