JUSHA-M23C LCD Monitor is intended to be used in displaying and viewing digital images by trained medical practitioners. JUSHA-M23C LCD Monitor does not support the display of mammography images for diagnosis.
Device Story
JUSHA-M23C is a 21.3-inch monochrome TFT LCD monitor for medical imaging. It accepts digital video inputs (DVI, DisplayPort) and displays images at 1600x1200 resolution. The device features an integrated calibration sensor, ambient brightness adaptation, and presence induction to automatically adjust display parameters for optimal viewing. It utilizes a 10-DICOM and 3-GAMMA lookup table to manage grayscale output (1024 simultaneous shades). Used in clinical environments by medical practitioners to evaluate high-resolution medical images. The device provides visual output to assist clinicians in diagnostic decision-making. It does not support mammography.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via measurement of angular luminance dependency, luminance/chromaticity non-uniformity (per TG18 guidelines), small-spot contrast ratio, temporal response, and luminance stability. Electrical safety and EMC verified per IEC 60601-1 and IEC 60601-1-2.
Indicated for trained medical practitioners to display and view digital medical images. Not for mammography diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a stylized caduceus symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The caduceus is depicted with a modern, abstract design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 12, 2015
Nanjing Jusha Display Technology Co., Ltd. % Mr. Ma Jing Certifical Engineer 301 Hanzhongmen Street, 8F Block A, No. 1 Nanjing International Service Outsourcing Mansion, Nanjing, 210036 CHINA
Re: K151238
Trade/Device Name: JUSHA-M23C LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: March 16, 2015 Received: May 11, 2015
Dear Ma Jing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K151238
Device Name JUSHA-M23C LCD monitor
#### Indications for Use (Describe)
JUSHA-M23C LCD Monitor is intended to be used in displaying digital images by trained medical practitioners. JUSHA-M23C LCD Monitor does not support the display of mammography images for diagnosis.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | May 6, 2015 | |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Submitter: | Nanjing Jusha Display Technology Co., Ltd<br>Add: 8A, Block 1. Nanjing International Service Outsourcing Mansion, No. 301, Hanzhongmen street, Nanjing, China. | |
| Contact Person: | Ma Jing<br>Certification Engineer<br>Nanjing Jusha Display Technology Co., Ltd<br>Tel: +86-25- 83305050<br>Fax: +86-25- 58783271 | |
| Device Trade Name: | JUSHA-M23C LCD Monitor | |
| Common/Usual Name: | 2MP Monochrome LCD Monitor | |
| Classification Name: Product Code: | System, image processing 21CFR 892.2050<br>PGY | |
| Predicate Device(s): | RADIFORCE GX240; K120407 | |
| Device Description: | JUSHA-M23C LCD Monitor is the display system with the high resolution (1600*1200), high luminance (1000cd/m²), and 1024 simultaneous shades of gray out of a palette of 4096, 10 DICOM look up table and 3 GAMMA look up table inside. JUSHA-M23C has ambient brightness adapting and presence induction system, with these this display can automatic adjustment according to different requirements in order to achieve the best results.<br>The product is consisted of the following components:<br>- 21.3inches, Mono-TFT LCD Panel<br>- JUSHA- SMS_19inch motherboard/FR-4/REV:0.1<br>- JUSHA-M23C LCD Monitor software<br>- Power Adapter<br>- Data Cable. | |
| | The LCD Monitor is designed, tested, and will be manufactured in<br>accordance with both mandatory and voluntary standards:<br><br>1. IEC 60601-1Medical equipment medical electrical<br>equipment - Part 1: General requirements for basic safety<br>and essential performance<br>2005+CORR.1(2006)+CORR.2(2007)<br><br>2. IEC 60601-1-2 Edition 3:2007, Medical electrical<br>equipment - Part 1-2: General requirements for basic<br>safety and essential performance - Collateral standard:<br>Electromagnetic compatibility - Requirements and tests. | |
| Intended Use: | JUSHA-M23C LCD Monitor is intended to be used in displaying and<br>viewing digital images by trained medical practitioners.<br>JUSHA-M23C LCD Monitor does not support the display of<br>mammography images for diagnosis. | |
| Technology: | JUSHA-M23C LCD Monitor is the display system with the high<br>resolution monitor (2 megapixels) with electronic capabilities for<br>evaluation of high resolution medical images, high luminance (1000<br>cd/m²) and 1024 simultaneous shades of gray out of a palette of 4096,<br>10 DICOM look up table and 3 GAMMA look up table inside.<br>JUSHA-M23C has ambient brightness adapting and presence<br>induction system, with these this display can automatic adjustment<br>according to different requirements in order to achieve the best<br>results. | |
| Determination of Substantial<br>Equivalence: | Summary of Non-Clinical Tests:<br>The LCD Monitor complies with voluntary standards as following:<br><br>1 IEC 60601-1Medical equipment medical electrical equipment -<br>Part 1: General requirements for basic safety and essential<br>performance 2005+CORR.1(2006)+CORR.2(2007)<br><br>2 IEC 60601-1-2 Edition 3:2007, Medical electrical equipment -<br>Part 1-2: General requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic compatibility<br>- Requirements and tests.<br><br>JUSHA-M23C is substantially equivalent to RADIFORCE GX240.<br>JUSHA-M23C employs the maximum resolution values same as that<br>of RADIFORCE GX240. Comparison table of the principal<br>characteristics of 2 devices is shown in the Attachment 1.<br>Attachment 1 | |
| | development of the system: | |
| | • Risk Analysis | |
| | • Requirements Reviews | |
| | • Design Reviews | |
| | • Raw materials verification | |
| | • Testing on unit level (Module verification) | |
| | • Integration testing (System verification) | |
| | • Final acceptance testing (Validation) | |
| | • Performance testing (Verification) | |
| | • Safety testing (Verification) | |
| | | |
| | | |
| | | |
| | Summary of Clinical Tests: | |
| | | |
| | The subject of this premarket submission, LCD Monitor, did not | |
| | require clinical studies to support substantial equivalence. | |
| | | |
| | The proposed device is Substantially Equivalent (SE) to the predicate | |
| | device which is US legally market device. Therefore, the subject | |
| | device is determined as safe and effectiveness. | |
| Conclusion: | Nanjing Jusha Display Technology Co., Ltd Considers the | |
| | JUSHA-M23C LCD Monitor to be as safe, as effective, and | |
| | performance is substantially equivalent to the predicate device(s). | |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
This comparison identifies the similarities and differences of the proposed JUSHA-M23C LCD Monitor device to the legally marketed predicate EIZO RADIFORCE GX240 LCD Monitor device to which substantial equivalency is claimed.
| Attributes | Predicate Device | Proposed Device |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product | RADIFORCE GX240 | JUSHA-M23C LCD Monitor |
| 510(k) Number | K120407 | / |
| Display Performance/Specifications | | |
| Screen technology | 21.3" Monochrome TFT<br>LCD Panel | 21.3" Monochrome TFT<br>LCD Panel |
| Viewing angle (H, V) | Horizontal 176 °,Vertical<br>176 ° | Horizontal 176 °,Vertical<br>176 ° |
| Resolution | 1600 x 1200/1200 x 1600 | 1600 x 1200/1200 x 1600 |
| Display area | 432.0 (H) x 324.0(V) mm | 432.0 (H) x 324.0(V) mm |
| Contrast Ratio | 1400:1 | 1400:1 |
| Scanning frequency (H; V) | 31~100 kHz;59~61Hz | 52~76 kHz;59~61Hz |
| Recommended Luminance | 500cd/m² | 500cd/m² |
| Pixel Pitch | 0.27x0.27 mm | 0.27x0.27 mm |
| Backlight | LED | LED |
| DICOM LUT | 10-bit (Display Port) : 1024<br>8-bit: 256 | 12-bit:4096 |
| Luminance calibration | Built in calibration sensor provided | Built in calibration sensor provided |
| Input signals | DVI standard 1.0,<br>DisplayPort 1.1a | DVI standard 1.0,<br>DisplayPort 1.1a |
| Input terminational | DVI-D (dual link) x 1,<br>DisplayPort x 1 | DVI-D (dual link) x 1,<br>DisplayPort x 1 |
| Display controller | Off the shelf | Off the shelf |
| Power Requirement | AC 100~240V 50~60Hz | AC 100~240V 50~60Hz |
| Attributes | Predicate Device | Proposed Device |
| Product | RADIFORCE GX240 | JUSHA-M23C LCD Monitor |
| 510(k) Number | K120407 | / |
| Power<br>Consumption/Save Mode | 36W/less than1.6W | 45W/less than 3W |
| Power<br>Management | DVI DMPM<br>DisplayPort 1.1a | DVI DMPM<br>DisplayPort 1.1a |
| USB<br>Ports/standard | 1 upstream (endpoint),<br>2 downstream/ Rev. 2.0 | 1 upstream (endpoint),<br>2 downstream/ Rev. 2.0 |
| Dimensions w/o<br>stand<br>(W x H x D) | Without stand:<br>376mmx505mmx98mm<br>With stand:<br>376mmx599mmx245.5 mm | Without stand:<br>382mm x490mm x75mm<br>With stand:<br>382mm x533mm x238mm |
| Indication for<br>use | RADIFORCE GX240 is<br>intended to be used in<br>displaying and viewing<br>digital images by trained<br>medical practitioners.<br>RADIFORCE GX240<br>Monitor does not support the<br>display of mammography<br>images for diagnosis. | JUSHA-M23C LCD Monitor<br>is intended to be used in<br>displaying and viewing<br>digital images by trained<br>medical practitioners.<br>JUSHA-M23C LCD Monitor<br>does not support the display<br>of mammography images for<br>diagnosis. |
| Attributes | Predicate Device | Proposed Device |
| Product | RADIFORCE GX240 | JUSHA-M23C LCD Monitor |
| 510(k) Number | K120407 | / |
| Applicable<br>standard | 1 IEC 60601-1Medical<br>equipment medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance<br>2005+CORR.1(2006)+COR<br>R.2(2007) | 1 IEC 60601-1Medical<br>equipment medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance<br>2005+CORR.1(2006)+COR<br>R.2(2007) |
| | 2 IEC 60601-1-2 Edition<br>3:2007, Medical electrical<br>equipment - Part 1-2:<br>General requirements for<br>basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility - Requirements<br>and tests. | 2 IEC 60601-1-2 Edition<br>3:2007, Medical electrical<br>equipment - Part 1-2:<br>General requirements for<br>basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility - Requirements<br>and tests. |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
### PERFORMANCE DATA:
The following performance data were provided in support of the substantial equivalence determination.
## Bench testing:
Bench testing was conducted to demonstrate the JUSHA-M23C meets all performance standards as follows:
- Measurement of the angular dependency of luminance response in horizontal, vertical and diagonal directions
- Measurement of the luminance non-uniformity characteristics of the display screen as specified in TGI18 guideline.
- Measurement of the chromaticity non-uniformity characteristics of the display screen as specified in TG18 guideline.
- Measurement of small-spot contrast ratio.
- Measurement of temporal response
- . Performance data on luminance stability
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the JUSHA-M23C. The device complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC.
## Animal and clinical study
The subject of this premarket submission, JUSHA-M23C, does not require animal or clinical studies to support substantial equivalence.
# CONCLUSIONS
JUSHA-M23C Medical Display is substantially equivalent to the predicate device with respect to technical characteristics, performance, application and intended use. The non-clinical data support the safety of the device should perform as intended in the specified use conditions. Nanjing Jusha Display Technology Co., Ltd considers the JUSHA-M23C Medical Display does not raise any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.