Hisense LDC monitor models HMD2G21/HMD3G21/HMD5G21

K160347 · Qingdao Hisense Medical Equipment Co., Ltd. · PGY · Mar 3, 2016 · Radiology

Device Facts

Record IDK160347
Device NameHisense LDC monitor models HMD2G21/HMD3G21/HMD5G21
ApplicantQingdao Hisense Medical Equipment Co., Ltd.
Product CodePGY · Radiology
Decision DateMar 3, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2

Intended Use

The 2MP/3MP LCD Monitor (HMD2G21/HMD3G21) are intended to be used in displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners. The 5MP LCD Monitor (HMD5G21) is intended to be used in displaying and viewing digital images for review and analysis by trained medical practitioners, including digital mammography.

Device Story

The Hisense LCD monitor series (HMD2G21, HMD3G21, HMD5G21) are high-resolution medical display systems used by trained medical practitioners in clinical settings for the review and analysis of digital medical images. The devices accept digital video signals via DVI-D or Display Port interfaces. They feature integrated brightness stabilization circuits, front sensors, and ambient light sensors to ensure persistent calibration and stable image brightness. The monitors support both landscape and portrait orientations and utilize antiglare screens to minimize reflections. By providing high-resolution, calibrated visual output, these monitors assist clinicians in diagnostic review workflows. The devices do not come into direct patient contact and do not require sterilization.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

21.3"/21.2" a-Si TFT LCD or IPS-NEO panels with antiglare coating. Native resolutions: 1600x1200 (2MP), 2048x1536 (3MP), 2560x2048 (5MP). Connectivity: DVI-D, Display Port, USB. Power: DC24V. Features: brightness stabilization circuit, front sensor, ambient light sensor. Complies with IEC 60601-1 (electrical/thermal safety) and IEC 60601-1-2 (EMC).

Indications for Use

Indicated for trained medical practitioners to display and view digital medical images for review and analysis. HMD2G21/HMD3G21 exclude digital mammography; HMD5G21 includes digital mammography.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure, emphasizing the organization's name and national affiliation. March 3, 2016 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Qingdao Hisense Medical Equipment Co.,Ltd. % Mr. Liu Zhitang Regulatory Manager Software outsourcing center 3rd floor north wing No.169 Songling Road, Laoshan Qing dao, Shan dong 266101 CHINA Re: K160347 Trade/Device Name: Hisense LCD monitor models HMD2G21/HMD3G21/HMD5G21 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: February 2, 2016 Received: February 8, 2016 Dear Mr. Zhitang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Michael D.'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) Device Name Hisense LDC monitor models HMD2G21/HMD3G21/HMD5G21 Indications for Use (Describe) The 2MP/3MP LCD Monitor (HMD2G21) is intended to be used in displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners. The 5MP LCD Monitor (HMD5G21) is intended to be used in displaying digital images for review and analysis by trained medical practitioners, including digital mammography. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary [As required by 21 CFR 807.92] ## 1. Date Prepared [21 CFR807.92 (a) (1)] Jan 8", 2016 ## 2. Submitter's Information [21 CFR807.92 (a) (1)] | Name of Sponsor: | Qingdao Hisense Medical Equipment Co., Ltd. | |------------------|-----------------------------------------------------------------------------------------------------------------| | Address: | Software outsourcing center 3rd floor north wing, No.169<br>Songling Road, Laoshan Dist. 266101, Qingdao, China | | Contact Name: | Liu Zhitang | | Telephone No .: | +86 ( 0 ) 532-55753811 | ### 3. Trade Name, Common Name, Classification [21 CFR807.92 (a) (2)] | Trade Name: | Hisense LCD monitor models HMD2G21/ HMD3G21/<br>HMD5G21 | | | | |-----------------------|---------------------------------------------------------|--|--|--| | Common Name: | Display system, medical image workstation, and others | | | | | Classification: | 892.2050 system, image processing, radiological | | | | | Product code: | PGY | | | | | Classification Panel: | Radiology | | | | | Device Class: | II | | | | ### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)] The identified predicates within this submission are as follows: BARCO N.V., Nio 2MP(MDNG-2121) has been cleared by FDA through 510(k) No. K091172, BARCO N.V., Nio 3MP(MDCG-3221) has been cleared by FDA through 510(k) No . K131246 BARCO N.V., Nio 5MP(MDNG-5121) has been cleared by FDA through 510(k) No . K062131 {4}------------------------------------------------ ## 5. Description of the Device [21 CFR 807.92(a) (4)] The 2MP/3MP/5MP LCD Monitor (HMD2G21/HMD3G21) is a display system for medical viewing, with high resolution 1600x1200(HMD2G21)/2048 x 1536(HMD3G21)/ 2560 x 2048(HMD5G21), built-in brightness stabilization circuit, front sensor and ambient light sensor, stable brightness and persistent calibration can be guaranteed. The display can support both landscape and portrait mode. The antiglare screen can prevent display from reflection under highlight conditions, make the image and display clearer. ## 6. Intended Use [21 CFR 807.92(a)(5)] The 2MP/3MP LCD Monitor (HMD2G21/HMD3G21) are intended to be used in displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners. The 5MP LCD Monitor (HMD5G21) is intended to be used in displaying and viewing digital images for review and analysis by trained medical practitioners, including digital mammography. ## 7. Technological Characteristics [21 CFR 807.92(a)(6)] | Panel | 21.3", a-Si TFT LCD, antiglare | |--------------------------|---------------------------------------------------------------------------------------------| | Brightness (typ.) | 1900 cd/m² | | CR (typ.) | 1400:1 | | Viewing angle | R/L 176° , U/D 176° Typ. (CR > 10) | | Pixel Pitch | 0.27 mm | | Native resolution | 1600 x 1200 | | Display area | 432.0mm(H)x324.0mm (V) | | Compatible video signals | 640 x 480@60Hz(progressive)<br>1024 x 768@60Hz(progressive)<br>1600x1200@60Hz (progressive) | | Aspect ratio | 4:3 | | Screen size | 21.3" real diagonal | | Power | DC24V/2.2A | | Power consumption | Max. 60 W | | Input signals | DVI-D, Display Port | ### The HMD2G21 {5}------------------------------------------------ | Digital input | LVDS (single) | |-------------------------------------------|--------------------------------------------------------| | Plug and play | VESA DDC 2B | | Dimension | 384.0mm (W) x 492.0mm (H) x70.0mm (D) (without Stand) | | | 384.0 mm (W) x 637.5 mm (H) x 273.5mm (D) (with Stand) | | Weight | 5.5kg (without Stand) | | | 9.9kg (with Stand) | | Operating<br>temperature and<br>humidity: | Temperature: 0°C ~ 40°C | | | Humidity: 20% ~80% | | | | | Storage<br>temperature and<br>humidity: | Temperature: -20°C ~ 60°C | | | Humidity:10% ~90% | | | | ## The HMD3G21 | Panel | 21.2", IPS-NEO, antiglare | |-------------------------------------------|-------------------------------------------------------------| | Brightness (typ.) | 1700 cd/m² | | CR (typ.) | 1400:1 | | Viewing angle | R/L 176° , U/D 176°<br>Typ. (CR > 10) | | Pixel Pitch | 0.21075 mm | | Native resolution | 2048x1536 | | Display area | (H)431.6x(V)323.7(mm) | | Compatible video<br>signals | 640 x 480@60Hz(progressive)<br>2048x1536@60Hz (progressive) | | Aspect ratio | 4:3 | | Screen size | 21.2" real diagonal | | Power | DC24V/2.2A | | Power consumption | Max. 60 W | | Input signals | DVI-D, Display Port | | Digital input | LVDS (single) | | Plug and play | VESA DDC 2B | | Dimension | 384.0mm (W) x 492.0mm (H) x 70.0mm (D) (without<br>Stand) | | | 384.0 mm (W) x 637.5 mm (H) x 273.5mm (D) (with<br>Stand) | | Weight | 6.1kg (without Stand) | | | 10.4kg (with Stand) | | Operating<br>temperature and<br>humidity: | Temperature: 0°C ~ 40°C<br>Humidity: 20% ~80% | | Storage | Temperature: -20°C ~ 60°C | {6}------------------------------------------------ | temperature and<br>humidity: | Humidity: 10% ~ 90% | |------------------------------|---------------------| |------------------------------|---------------------| ### The HMD5G21 | Panel | 21.3", IPS-NEO, antiglare | |-------------------------------------|-----------------------------------------------------------------------------------------------------------------| | Brightness (typ.) | 1200 cd/m² | | CR (typ.) | 1200:1 | | Viewing angle | R/L 176° , U/D 176° Typ. (CR > 50) | | Pixel Pitch | 0.165 mm | | Native resolution | 2560 x 2048 | | Display area | 422.4mm(H)x337.9mm (V) | | Compatible video signals | 640 x 480@60Hz(progressive)<br>1024 x 768@60Hz(progressive)<br>2560 x 2048@60Hz (progressive) | | Aspect ratio | 5:4 | | Screen size | 21.3" real diagonal | | Power | DC24V/3A | | Power consumption | Max. 72 W | | Input signals | DVI-D, Display Port | | Digital input | LVDS (double) | | Plug and play | VESA DDC 2B | | Dimension | 392.0mm (W) x 498.0mm (H) x83.0mm (D) (without Stand)<br>392.0 mm (W) x 632.0 mm (H) x 273.5mm (D) (with Stand) | | Weight | 8.5kg (without Stand)<br>12.8kg (with Stand) | | Operating temperature and humidity: | Temperature: 0°C ~ 40°C<br>Humidity: 20%~80% | | Storage temperature and humidity: | Temperature: -20°C ~ 60°C<br>Humidity: 10% ~90% | ## 8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92] ### 8.1 Intended uses: ### Table 1 Intended Use Comparison of HMD2G21 | ID | Comparison Item | Proposed Device<br>2MP LCD Monitor (HMD2G21) | Predicate Device<br>Nio 2MP(MDNG-2121) | |----|-----------------|----------------------------------------------|----------------------------------------| |----|-----------------|----------------------------------------------|----------------------------------------| {7}------------------------------------------------ | 1 | Intended<br>Use | The HISENSE 2MP LCD<br>Monitor (HMD2G21) is intended<br>to be used in displaying and<br>viewing digital images (excluding<br>digital mammography) for review<br>and analysis by trained medical<br>practitioners. | The Nio 2MP(MDNG-2121)<br>is intended to be used in<br>displaying and viewing digital<br>images for review by trained<br>medical practitioners. These<br>devices must not be used in<br>primary image diagnosis in<br>mammography. | |---|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |---|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| ### Table 2 Intended Use Comparison of HMD3G21 | ID | Comparison<br>Item | Proposed Device<br>3MP LCD Monitor (HMD3G21) | Predicate Device<br>Nio 3MP(MDCG-3221) | |----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1 | Intended<br>Use | The HISENSE 3MP LCD<br>Monitor ( HMD3G21) is intended<br>to be used in displaying and<br>viewing digital images (excluding<br>digital mammography) for review<br>and analysis by trained medical<br>practitioners. | The Nio 3MP(MDCG-3221) is<br>intended to be used in displaying<br>and viewing digital images for<br>review by trained medical<br>practitioners. These devices<br>must not be used in primary<br>image diagnosis in<br>mammography. | #### Table 3 Intended Use Comparison of HMD5G21 | ID | Comparison<br>Item | Proposed Device<br>5MP LCD Monitor (HMD5G21) | Predicate Device<br>Nio 5MP(MDNG-5121) | |----|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1 | Intended<br>Use | The 5MP LCD Display<br>(HMD5G21) is intended to be<br>used in displaying and viewing<br>digital images for review and<br>analysis by trained medical<br>practitioners, including digital<br>mammography. | The Nio 5MP(MDNG-5121) is<br>intended to be used as a tool in<br>displaying and viewing digital<br>images for review and analysis<br>by trained medical practitioners,<br>including digital mammography. | #### 8.2 Comparison table ### Table 4 General Comparison about HMD2G21 and Nio 2MP | ID | Comparison<br>Item | Proposed Device<br>HMD2G21 | Predicate Device<br>Nio 2MP(MDNG-2121) | |-----|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 2 | Performance | | | | 2.1 | Panel Size and Type | 21.3", TFT LCD display | 21.3", TFT LCD display | | 2.2 | Pixel Pitch | 0.27 mm | 0.27 mm | | ID | Comparison<br>Item | Proposed Device<br>HMD2G21 | Predicate Device<br>Nio 2MP(MDNG-2121) | | 2.3 | Native Resolutions | 1600x1200 | 1600x1200 | | 2.4 | Brightness | 1900 cd/m2 | 1650 cd/m2 | | 2.5 | Contrast Ratio | 1400:1 | 850:1 | | 2.6 | Network Interface | USB(1 Up, 2 Downstream) | USB(1 Up, 2 Downstream) | | 2.7 | Active Display Size<br>(HxV) | 432.0mm x324.0mm | 432mm x324mm | | 3 | Physical Specifications | | | | 3.1 | Dimensions<br>(Wx Hx D) | 384mm (W) x 492mm (H) x<br>70mm (D) (without Stand)<br>384 mm (W) x 517.5-637.5<br>mm (H) x 273.5mm (D) (with<br>Stand) | 382x 488 x 114 mm (without<br>Stand)<br>Portrait:<br>382 x 577~637 x 249 mm<br>Landscape:<br>488 x 472~532 x 249 mm<br>(with Stand) | | | Temperature | | | | 3.2 | Operating | 0°C ~ 40°C | 0°C ~ 35°C | | 3.3 | Transport/ Storage | -20°C ~ 60°C | -20°C ~ 60°C | | | Relative humidity | | | | 3.4 | Operating | 20% ~80% | 8% ~80%(non-condensing) | | 3.5 | Transport/ Storage | 10% ~90% | 5% ~85% | | 4 | Power Supply | | | | 4.1 | PowerCapacity | ≤60W | <70W | | 4.2 | Input Voltage | DC24V/2.2A | 100~240V | | 5 | Human factors (operation characteristic) | | | | 5.1 | Usability | Button operation,<br>LED indicator | Button operation,<br>LED indicator | | 5.2 | Mode of operation | Continuous operation | Continuous operation | | 6 | Biocompatibility | | | | 6.1 | Evaluation | The proposed device does not<br>contain any components that<br>come into direct or indirect<br>contact with patients, so the<br>evaluation doesn't be needed. | The proposed device does<br>not contain any components<br>that come into direct or<br>indirect contact with patients,<br>so the evaluation doesn't be<br>needed. | | 7 | Sterility | | | | 7.1 | Sterilization | The proposed device does<br>not need sterilization. | The proposed device does<br>not need sterilization. | | 8 | Electrical & Mechanical safety& Thermal safety | | | | 8.1 | Type of protection<br>against electric shock | Class I | Class I | | 8.2 | Degree of protection<br>against harmful<br>ingress of liquid | Ordinary equipment. | Ordinary equipment. | | 8.3 | Evaluation | The electrical, mechanical and<br>thermal safety evaluation is | The electrical, mechanical<br>and thermal safety | | ID | Comparison<br>Item | Proposed Device<br>HMD2G21 | Predicate Device<br>Nio 2MP(MDNG-2121) | | | | requirements of the standard<br>IEC 60601-1. | evaluation is conducted as<br>per the requirements of the<br>standard IEC 60601-1. | | 9 | Electromagnetic Compatibility | | | | 9.1 | EMC Evaluation | Complying with<br>IEC 60601-1-2 | Complying with<br>IEC 60601-1-2 | {8}------------------------------------------------ {9}------------------------------------------------ ### Table 5 General Comparison about HMD3G21 and Nio 3MP | ID | Comparison Item | Proposed Device<br>HMD3G21 | Predicate Device<br>Nio 3MP(MDCG-3221) | |-----|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------| | 2 | | Performance | | | 2.1 | Panel Size and Type | 21.3", TFT LCD display | 21.3",TFT LCD display | | 2.2 | Pixel Pitch | $0.21075$ mm | $0.2115$ mm | | 2.3 | Native Resolutions | 2048x1536 | 2048x1536 | | 2.4 | Brightness | 1700 cd/m2 | 800 cd/m2 | | 2.5 | Contrast Ratio | 1400:1 | 1700:1 | | 2.6 | Network Interface | USB(1 Up, 2 Downstream) | USB(1 Up, 2 Downstream) | | 2.7 | Active Display Size<br>(HxV) | 431.6mm x323.7mm | 433mm x325mm | | 3 | | Physical Specifications | | | 3.1 | Dimensions<br>(Wx Hx D) | 384mm (W) x 492mm (H) x<br>70mm (D) (without Stand)<br>384 mm (W) x 517.5-637.5<br>mm (H) x 273.5mm (D) (with<br>Stand ) | 378mm x 491mm x 84mm<br>(without Stand)<br>Portrait:<br>378mm x 628~528mm<br>x 235mm<br>Landscape:<br>491mm x 572~472mm<br>x 235mm<br>(with Stand) | | | Temperature | | | | 3.2 | Operating | 0°C ~ 40°C | 0°C ~ 40°C | | 3.3 | Transport/ Storage | -20°C ~ 60°C | -20°C ~ 60°C | | | Relative humidity | | | | 3.4 | Operating | 20% ~80% | 8% ~80%(non-condensing) | | 3.5 | Transport/ Storage | 10% ~90% | 5% ~95% | | 4 | Power Supply | | | | 4.1 | PowerCapacity | ≤60W | ≤40W | | 4.2 | Input Voltage | DC24V/2.5A | 100~240v | | 5 | Human factors (operation characteristic) | | | | 5.1 | Usability | Button operation,<br>LED indicator | Button operation,<br>LED indicator | | 5.2 | Mode of operation | Continuous operation | Continuous operation | | 6 | Biocompatibility | | | | 6.1 | Evaluation | The proposed device does not | The proposed device does | {10}------------------------------------------------ | ID | Comparison<br>Item | Proposed Device<br>HMD3G21 | Predicate Device<br>Nio 3MP(MDCG-3221) | |-----|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------| | | | contain any components that<br>come into direct or indirect<br>contact with patients, so the<br>evaluation doesn't be needed. | not contain any components<br>that come into direct or<br>indirect contact with patients,<br>so the evaluation doesn't be<br>needed. | | 7 | Sterility | | | | 7.1 | Sterilization | The proposed device does<br>not need sterilization. | The proposed device does<br>not need sterilization. | | 8 | Electrical & Mechanical safety& Thermal safety | | | | 8.1 | Type of protection<br>against electric shock | Class I | Class I | | 8.2 | Degree of protection<br>against harmful<br>ingress of liquid | Ordinary equipment. | Ordinary equipment. | | 8.3 | Evaluation | The electrical, mechanical and<br>thermal safety evaluation is<br>conducted as per the<br>requirements of the standard<br>IEC 60601-1. | The electrical, mechanical and<br>thermal safety<br>evaluation is conducted as<br>per the requirements of the<br>standard IEC 60601-1. | | 9 | Electromagnetic Compatibility | | | | 9.1 | EMC Evaluation | Complying with<br>IEC 60601-1-2 | Complying with<br>IEC 60601-1-2 | ### Table 6 General Comparison about HMD5G21 and Nio 5MP | ID | Comparison<br>Item | Proposed Device<br>HMD5G21 | Predicate Device<br>Nio 5MP(MDNG-5121) | |-----|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | | | | 2 | | Performance | | | 2.1 | Panel Size and Type | 21.3", TFT LCD display | 21.3", TFT LCD display | | 2.2 | Pixel Pitch | 0.165 mm | 0.165 mm | | 2.3 | Native Resolutions | 2,560 × 2,048 | 2,560 × 2,048 | | 2.4 | Brightness | 1200 cd/m2 | 700 cd/m2 | | 2.5 | Contrast Ratio | 1200:1 | 800:1 | | 2.6 | Network Interface | USB(1 Up, 2 Downstream) | USB(1 Up, 2 Downstream) | | 2.8 | Active Display Size<br>(HxV) | 422.4mm x337.9mm | 422mm x338mm | | 3 | Physical Specifications | | | | 3.1 | Dimensions<br>(Wx Hx D) | 392mm×498mm×83mm<br>(without Stand)<br>392mm×514-632mm×273.5mm<br>(with Stand) | 408mm x 492mm x 115mm<br>(without Stand)<br>Portrait:<br>408mm x 489~549mm<br>x 250mm<br>Landscape:<br>492mm x 531~591mm | | ID | Comparison<br>Item | Proposed Device<br>HMD5G21 | Predicate Device<br>Nio 5MP(MDNG-5121) | | | | | x 250mm<br>(with Stand) | | 3 | Temperature | | | | 3.2 | Operating | 0°C ~ 40°C | 0°C ~ 40°C | | 3.3 | Transport/ Storage | -20°C ~ 60°C | -20°C ~ 60°C | | 3 | Relative humidity | | | | 3.4 | Operating | 20% ~80% | 8% ~80%(non-condensing) | | 3.5 | Transport/ Storage | 10% ~90% | 5% ~95% | | 4 | Power Supply | | | | 4.1 | PowerCapacity | <80W | 61W | | 4.2 | Input Voltage | DC24V/3A | 100~250V | | 5 | Human factors (operation characteristic) | | | | 5.1 | Usability | Button operation, | Button operation, | | | | LED indicator | LED indicator | | 5.2 | Mode of operation | Continuous operation | Continuous operation | | 6 | Biocompatibility | | | | 6.1 | Evaluation | The proposed device does not<br>contain any components that<br>come into direct or indirect<br>contact with patients, so the<br>evaluation doesn't be needed. | The proposed device does<br>not contain any components<br>that come into direct or<br>indirect contact with patients,<br>so the evaluation doesn't be<br>needed. | | 7 | Sterility | | | | 7.1 | Sterilization | The proposed device does<br>not need sterilization. | The proposed device does<br>not need sterilization. | | 8 | Electrical & Mechanical safety& Thermal safety | | | | 8.1 | Type of<br>protection<br>against electric shock | Class I | Class I | | 8.2 | Degree of protection<br>against harmful<br>ingress of liquid | Ordinary equipment. | Ordinary equipment. | | 8.3 | Evaluation | The electrical, mechanical and<br>thermal safety evaluation is<br>conducted as per the<br>requirements of the standard<br>IEC 60601-1. | The electrical, mechanical<br>and thermal safety<br>evaluation is conducted as<br>per the requirements of the<br>standard IEC 60601-1. | | 9 | Electromagnetic Compatibility | | | | 9.1 | EMC Evaluation | Complying with<br>IEC 60601-1-2 | Complying with<br>IEC 60601-1-2 | {11}------------------------------------------------ #### 8.4 Discussion of Differences: It is reasonable that there are some differences between our new system and its predicate. All of parameters comply with 21CFR1020.33 and related IEC standards. We did not use any new technology in this system, so those differences between our new {12}------------------------------------------------ system and its predicate do not affect the safety and effectiveness (SE). #### About HDM2G21 and Nio 2MP Review of ID 1 - Intended use, both of them are intended to be used in displaying and viewing digital images, excluding digital mammography, for review and analysis by trained medical practitioners. So the SE is not affected. Review of ID 2 - Performance, except two items as below, both are the same, so the SE is not affected. 1.Brightness, The proposed device is 1900cd/m2 and the predicate device is 1650cd/m2, but the 1900 cd/m2 is better than 1650cd/m2 mm in terms of the bright lifetime. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. 2.Contrast Ratio, The proposed device is 1400:1 and the predicate device is 850:1, but the 1400:1 is better than 850:1 in terms of the Contrast Ratio. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. Review of ID 3 - Physical Specifications, Dimensions and Relative humidity are comparable, so the SE is not affected Review of ID 4 - Power Supply, both of them comply with IEC 60601-1 and IEC 60601-1-2. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. Review of ID 5 - Human factors, both are the same, so the SE is not affected. Review of ID 6 - Biocompatibility, both are the same, so the SE is not affected. Review of ID 7 - Sterility, both are the same, so the SE is not affected. Review of ID 8 - Electrical & Mechanical safety& Thermal safety, both are the same, so the SE is not affected. Review of ID 9 - EMC, both are the same, so the SE is not affected. #### About HDM3G21 and Nio 3MP Review of ID 1 - Intended use, both of them are intended to be used in displaying and viewing digital images, excluding digital mammography, for review and analysis by trained medical practitioners. So the SE is not affected. Review of ID 2 - Performance, except four items as below, both are the same, so the SE is not affected. 1.Pixel Pitch, The proposed device is 0.21075 mm and the predicate device is 0.2115 mm, but the 0.21075 mm is better than 0.2115 mm in terms of the image quality. Therefore, {13}------------------------------------------------ they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. 2.Brightness, The proposed device is 1700cd/m2 and the predicate device is 800cd/m2, but the 1700 cd/m2 is better than 800cd/m2 in terms of the bright lifetime. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. 3.Contrast Ratio, The proposed device is 1400:1 and the predicate device is 1700:1, but the Contrast Ratio are comparable, Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. 4.Active Display Size, The proposed device is 431.6mm x323.7mm and the predicate device is 433mm x325mm, the different of the effective display area due to the different size of the pixel pitch, and the smaller the better in terms of the image quality. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. Review of ID 3 - Physical Specifications, Dimensions and Relative humidity are comparable, so the SE is not affected Review of ID 4 - Power Supply, both of them comply with IEC 60601-1 and IEC 60601-1-2. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. Review of ID 5 - Human factors, both are the same, so the SE is not affected. Review of ID 6 - Biocompatibility, both are the same, so the SE is not affected. Review of ID 7 - Sterility, both are the same, so the SE is not affected. Review of ID 8 - Electrical & Mechanical safety& Thermal safety, both are the same, so the SE is not affected. Review of ID 9 - EMC, both are the same, so the SE is not affected. #### About HDM5G21 and Nio 5MP Review of ID 1 - Intended use, both of them are intended to be used in displaying and viewing digital imaqes, including digital mammography, for review and analysis by trained medical practitioners. So the SE is not affected. Review of ID 2 - Performance, except one item as below, both are the same, so the SE is not affected. 1.Brightness, The proposed device is 1200cd/m2 and the predicate device is 700cd/m2, but the different brightness just affect in terms of the bright lifetime. They can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. {14}------------------------------------------------ 2.Contrast Ratio, The proposed device is 1200:1 and the predicate device is 800:1, and the higher the better in terms of the image quality. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. 3.Active Display Size, The proposed device is 422.4mm x337.9mm and the predicate device is 422mm x338mm, the different of the effective display area due to the different size of the pixel pitch. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. Review of ID 3 - Physical Specifications, Dimensions and Relative humidity are comparable, so the SE is not affected Review of ID 4 - Power Supply, both of them comply with IEC 60601-1 and IEC 60601-1-2. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected. Review of ID 5 - Human factors, both are the same, so the SE is not affected. Review of ID 6 - Biocompatibility, both are the same, so the SE is not affected. Review of ID 7 - Sterility, both are the same, so the SE is not affected. Review of ID 8 - Electrical & Mechanical safety& Thermal safety, both are the same, so the SE is not affected. Review of ID 9 - EMC, both are the same, so the SE is not affected. #### 9. Conclusion [21 CFR 807.92(b) (3)] In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Qingdao Hisense Medical Equipment Co., Ltd. concludes that : The subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device. The differences above between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no new risk is raised regarding to effectiveness and safety.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...