INFX-8000C, V6.20

K152697 · Toshibamedical Systems Corporation · OWB · Jan 15, 2016 · Radiology

Device Facts

Record IDK152697
Device NameINFX-8000C, V6.20
ApplicantToshibamedical Systems Corporation
Product CodeOWB · Radiology
Decision DateJan 15, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2

Intended Use

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Device Story

INFX-8000C, V6.20 is an interventional X-ray system comprising a C-arm, X-ray tube, beam limiter, flat panel detector (FPD), X-ray controller, and patient table. Used in clinical interventional settings by physicians to perform diagnostic and angiographic procedures. The system captures X-ray images of blood vessels; processes data via system software; and displays images on monitors for real-time guidance during procedures. Features include 2D/3D roadmap functions and optional connectivity with CT systems (IVR-CT). Output assists clinicians in navigating catheters/devices and visualizing vascular anatomy, potentially improving procedural accuracy and patient outcomes. Modifications from predicate include improved FPDs (increased scintillator thickness, 16-bit ADC, ASIC upgrades), new C-arm support (CAS-930A), expanded sliding angles, and wireless footswitch capability.

Clinical Evidence

Bench testing only. No clinical data. Performance verified via technical metrics including spatial resolution, low contrast resolution, dynamic range, DQE, MTF, and artifact analysis. Compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-43, 62304, and 62366 standards confirmed.

Technological Characteristics

Digital X-ray system; C-arm architecture. FPDs (TFP-800A/C1, TFP-1216A/C1) with increased scintillator thickness, 16-bit ADC, and ASIC improvements. Liquid metal bearing X-ray tube (3.0 MHU). Connectivity: DICOM/networked, optional CT integration. Software: Moderate level of concern. Standards: 21 CFR Subchapter J, IEC 60601 series, ISO 13485.

Indications for Use

Indicated for patients requiring diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen, and lower extremities.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a circular seal for the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around the perimeter. In the center is the department's logo, which consists of a stylized caduceus symbol with three intertwined figures representing people. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 15, 2016 Toshiba Medical Systems Corporation % Ms. Janine Reyes Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780 Re: K152697 Trade/Device Name: INFX-8000C, V6.20 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA Dated: December 16, 2015 Received: December 18, 2015 Dear Ms. Reyes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152697 Device Name INFX-8000C, V6.20 Indications for Use (Describe) This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities. Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) - SUMMARY OF SAFETY AND EFFECTIVENESS ### 1. CLASSIFICATION and DEVICE NAME: | Classification Name: | lmage-Intensified Fluoroscopic X-Ray System | | |----------------------|------------------------------------------------------------------|--| | Requlation Number: | 21 CFR 892.1650 (Class II) | | | Product Code | OWB - Interventional Fluoroscopic X-ray System (primary) | | | | JAA - System, X-Ray, Fluoroscopic, Image-Intensified (secondary) | | | Trade Proprietary | İnfinix | | | Name: | | | | Model Number: | INFX-8000C, V6.20 | | ### 2. ESTABLISHMENT REGISTRATION: 9614698 ### 3. CONTACT PERSON, U.S. AGENT and ADDRESS: ### Official Correspondent/U.S. Agent: Paul Biggins Director, Regulatory Affairs (714) 730-5000 Fax: (714) 730-1310 paul.biggins@toshiba.com ### Contact Person: Janine Reyes Manager, Regulatory Affairs (714) 669-7853 Fax: (714) 730-1310 janine.reyes@toshiba.com ### Establishment Name and Address: Toshiba America Medical Systems, Inc. (TAMS) 2441 Michelle Drive Tustin, CA. 92780 ### 4. MANUFACTURING SITE Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan Contact: Akinori Hatanaka ### 5. DATE PREPARED: September 18, 2015 {4}------------------------------------------------ ## 6. TRADE NAME(S): INFX-8000C, V6.20 [Infinix CC-i] ## 7. COMMON NAME: Image-Intensified Fluoroscopic X-Ray System Interventional Fluoroscopic X-Ray System System, X-Ray, Fluoroscopic, Image-Intensified ## 8. DEVICE CLASSIFICATION: Class II (per 21 CFR 892.1650) ## 9. PRODUCT CODE / DESCRIPTION: Primary Product Code: OWB - Interventional Fluoroscopic X-Ray System Secondary Product Code: JAA - System, X-Ray, Fluoroscopic, Image-Intensified ## 10. PERFORMANCE STANDARD: This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard]. ## 11. PREDICATE DEVICE: INFX-8000C, w/Spot Fluoroscopy Option (K113052) ## TABLE 1: Predicate Device | Product | Marketed by | 510(k) Number | Clearance Date | |------------------------------------------|------------------------------------|---------------|-------------------| | INFX-8000C, w/Spot<br>Fluoroscopy Option | Toshiba America Medical<br>Systems | K113052 | November 22, 2011 | ## 12. REASON FOR SUBMISSION: Modification of a cleared device ## 13. SUBMISSION TYPE: Traditional 510(k) ## 14. DEVICE DESCRIPTION: INFX-8000C, V6.20, is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm, which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table. ## 15. SUMMARY OF INTENDED USES: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities. {5}------------------------------------------------ ### 16. SUMMARY OF CHANGE(S) This submission is to report the following items have been changed: - a. The current flat panel detectors have been modified for improved DQE performance, model numbers are: - i. Model Number: Old TFP-800A; new TFP-800A/C1 (20cm x 20cm) - ii. Model Number: Old TFP-1216A; new TFP-1216A/C1 (30cm x 40cm) - 1. Scintillator thickness increased - 2. ADC Bus from 14 to 16 bits - 3. ASIC improvements - 4. DQE performance enhancement - b. New C-arm support assembly (CAS-930A) ### TABLE 2: Changes to INFX-8000C since 510(k) Clearance (K113052) | TABLE 2: Changes to INFX-8000C since 510(k) Clearance (K113052) | | | |-----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------| | Change | Predicate Device:<br>INFX-8000C, with Spot Fluoroscopy<br>Option (K113052) | Subject Device:<br>INFX-8000C, V6.20 | | C-arm Support** | CAS-830B/A1 | CAS-830B /A1<br>CAS-930A | | C-arm Sliding<br>Angle** | RAO 90° LAO 50° | RAO 90° LAO 50°<br>RAO 120° LAO 90° | | SID Range* | 90cm to 120cm<br>(30cm x 40cm FPD, 20cm x 20cm FPD) | 90cm to 120cm<br>(30cm x 40cm FPD)<br>90cm to 125cm<br>(20cm x 20cm FPD, 30cm x 30cm FPD) | | X-ray Tube<br>Assembly* | Liquid metal bearing<br>Ball bearing | Liquid metal bearing | | Anode Heat<br>Storage Capacity* | 3.0MHU (Liquid metal bearing)<br>1.8MHU (Ball bearing) | 3.0MHU (Liquid metal bearing) | | Focal Spot Size* | 0.5/0.8 mm (Liquid metal bearing)<br>0.3/0.6/1.0 mm (Liquid metal bearing)<br>0.3/0.5/0.8 mm (Ball bearing)<br>0.3/0.6/1.0 mm (Ball bearing) | 0.5/0.8 mm (Liquid metal bearing)<br>0.3/0.6/1.0 mm (Liquid metal bearing)<br>0.4/0.6/0.9 mm (Liquid metal bearing) | | Beam Limiting<br>Device | Cardiac: 230mm x 230mm Maximum<br>(on the plane 870 mm from the focus)<br>Angiography: 406mm x 406mm Maximum<br>(on the plane 930 mm from the focus) | Cardiac: 340mm x 340mm Maximum<br>(on the plane 900 mm from the focus)<br>Angiography: 400mm x 400mm Maximum<br>(on the plane 900 mm from the focus) | | X-ray Exposure<br>Field* | | | | Detector Size | 20cm x 20cm<br>30cm x 40cm | 20cm x 20cm**<br>30cm x 40cm**<br>30cm x 30cm* | | DQE** | 65% or more (0 lp/mm)<br>(20cm x 20cm, 30cm x 40cm) | 65% or more (0 lp/mm) or 77%±5% (0 lp/mm)<br>(20cm x 20cm, 30cm x 30cm, 30cm x 40cm) | | Fluoroscopic<br>Monitor/ Reference<br>Monitor* | 19.0 inch | 19.0 inch<br>56.2 inch | | Roadmap<br>Function* | 2D Roadmap | 2D Roadmap<br>3D Roadmap | | Footswitch* | Wired | Wired<br>Wireless | | Connection with<br>CT System* | Not Available | Available | | Combined with CT<br>System for IVR-CT* | Not Available | CGBA-033A, Aquilion ONE VISION<br>CGBA-032A, Aquilion PRIME | {6}------------------------------------------------ *Cleared by previous requlatory affairs action **Part of this Pre-Market Notification 510(k) submission ## 17. SUBSTANTIAL EQUIVALENCE: This device is substantially equivalent to the INFX-8000C, w/Spot Fluoroscopy Option (K113052), marketed by Toshiba America Medical Systems. INFX-8000C, V6.20, includes incremental improvements to the cleared flat panel detectors (TFP-800A and TFP-1216A) and a new C-arm support assembly (CAS-930A). The basic system configuration, method of operation, base software and manufacturing process remain unchanged from the cleared device. There are no new indications for use or intended use of the device. ### 18. SAFETY: The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 standards, its collateral standards and particular standards: IEC 60601-1:2005, IEC 60601-1-2:2007, IEC 60601-1-3:2008, IEC 60601-1-6:2010, IEC 60601-2-28:2010, IEC 60601-2-43:2010, IEC 62304:2006, IEC 62366:2007. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR §1020, that apply to this device, will be met and reported via product report. ## 19. TESTING: This submission contains test data that demonstrates the system modifications result in performance that is equal to or better than the predicate system. Testing included spatial resolution, low contrast resolution, dynamic range, DQE, MTF, artifacts/contrast/dynamic range of DS and horizontal streak noise. The results of this testing demonstrate equivalent or increased performance when compared to the predicate device. Testing of the modified system was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and X-ray Systems: IEC 60601-1:2005, IEC 60601-1-2:2007, IEC 60601-1-3:2008, IEC 60601-1-6:2010, IEC 60601-2-28:2010, IEC 60601-2-43:2010, IEC 62304:2006, IEC 62366:2007. Software Documentation for a Moderate Level of Concern, per the FDA quidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission. Additionally, the design controls used for this device included risk management and all known risks were mitigated to an acceptable level. {7}------------------------------------------------ ## 20. CONCLUSION The subject device is substantially equivalent to the INFX-8000C, w/Spot Fluoroscopy Option, which was cleared via Pre-Market Notification 510(k), K113052. The INFX-8000C, V6.20, incorporates modifications to the cleared device which include a new C-arm support assembly and incremental improvements to the cleared flat panel detectors. While these modifications do introduce new technological characteristics, they did not raise different questions of Substantial Equivalence compared to the predicate device. Documentation of the design change process, as well as non-clinical testing data as described in Section 19 "Testing", demonstrate equivalent or improved performance. The modifications incorporated into the INFX-8000C, V6.20, do not alter the indications for use or the intended uses associated with the previously cleared device, as described in the labeling. It is the contention of Toshiba that all new safety issues have been addressed in the design of this change and that adequate evidence of this is presented with this submission.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...