DISCOVERY IGS 740

K133278 · GE Medical Systems SCS · OWB · Apr 24, 2014 · Radiology

Device Facts

Record IDK133278
Device NameDISCOVERY IGS 740
ApplicantGE Medical Systems SCS
Product CodeOWB · Radiology
Decision DateApr 24, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2
AttributesPediatric

Intended Use

The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures.

Device Story

Discovery IGS 740 is an interventional fluoroscopic/angiographic X-ray system; utilizes 41cm square digital detector mounted on laser-guided mobile AGV gantry. Used in interventional suites or ORs; operated by physicians/clinicians. System captures X-ray projections; transforms inputs into fluoroscopic and rotational images for visualization. Provides real-time guidance for diagnostic and interventional procedures; supports image-guided surgery via OR table configuration. Collision avoidance system adapted for 41cm detector. Output displayed on monitors for clinical decision-making; assists in precise anatomical targeting; reduces procedural risk; improves patient outcomes through minimally invasive guidance.

Clinical Evidence

No clinical studies required. Substantial equivalence supported by bench testing, including simulated use testing, usability validation, dose verification, image quality verification (uniformity, contrast, detail), and software reliability cycling.

Technological Characteristics

Interventional fluoroscopic X-ray system; 41cm square solid-state digital detector; laser-guided mobile AGV gantry. Complies with IEC 60601-1, 60601-2-43, 60601-1-2, 60601-1-3, 60601-1-6, and IEC 60601-2-46 (for OR table).

Indications for Use

Indicated for patients newborn to geriatric requiring fluoroscopic and rotational imaging for cardiovascular, vascular, non-vascular diagnostic/interventional procedures, and image-guided surgery.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ APR 2 4 2014 Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The border has decorative swirls or flourishes at three points around the circle. Appendices K GE Healthcare K133278, Discovery IGS 740 Supplement # 510(k) Summary In accordance with 21 CFR 807.92 the following summary of information is provided: | Date Summary was revised: | March 4th, 2014 | |----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | GE Healthcare<br>GE MEDICAL SYSTEMS, SCS<br>283 RUE DE LA MINIERE<br>78530 BUC - FRANCE<br>T: +33-(0)1-30-70-47-41 | | Primary Contact Person: | Michel Genuer<br>GE Healthcare (GE MEDICAL SYSTEMS SCS)<br>Regulatory Affairs Leader,<br>283 RUE DE LA MINIERE<br>78530 BUC - FRANCE<br>T: +33-(0)1-30-70-47-41<br>Email: michel.genuer@ge.com | | Secondary Contact Person: | Hubert Welsch<br>Regulatory Affairs Leader,<br>GE Healthcare<br>283 RUE DE LA MINIERE<br>78530 BUC - FRANCE<br>T: +33-(0)1-30-70-49-51<br>Email: hubert.welsch@ge.com | | Device/Trade Name: | Discovery IGS 740 | | Common/Usual Name: | interventional fluoroscopic x-ray system, angiographic x-ray system | | Regulation Description: | Image-intensified fluoroscopic x-ray system. | | Regulation number: | 892.1650 | | Product Code: | OWB, JAA and IZI | | Class: | II | | Predicate Device(s): | K122457: Innova/Innova IGS/Discovery IGS/Optima angiographic,<br>fluoroscopic X-ray Systems with Cathlab Frontiers solutions. | | Device Description: | The proposed device, Discovery IGS 740 is an interventional<br>fluoroscopic x-ray system/ x-ray angiographic system with a 41 cm<br>square digital detector.<br>Discovery IGS 740 is a new product model of the GE Discovery IGS<br>interventional fluoroscopic x-ray system/ x-ray angiographic system | | | platform with a 41cm square digital detector. GE Discovery IGS x-ray<br>angiographic systems are based on the GE laser guided gantry. The<br>41cm square digital detector of the cleared predicate device Innova<br>IGS 540 K122457 is mounted on the GE Discovery IGS gantry. This<br>gantry is the same gantry as the Discovery IGS 730 predicate device,<br>originally cleared with K113403. | | | Discovery IGS 740 is proposed in an Interventional configuration and in<br>an OR configuration (with OR table) for the indications for use of<br>surgery. | | | Discovery IGS 740 has no additional indications for use compared with<br>the predicate device Innova/Innova IGS/Discovery IGS/Optima<br>angiographic, fluoroscopic X-ray Systems with Cathlab Frontiers<br>solutions K122457. | | Intended Use | The angiographic X-ray systems are indicated for use for patients from | | | newborn to geriatric in generating fluoroscopic and rotational images | | Indications for Use: | of human anatomy for cardiovascular, vascular and non-vascular, | | | diagnostic and interventional procedures. | | | Additionally, with the OR table, the angiographic X-ray systems are | | | indicated for use in generating fluoroscopic and rotational images of | | | human anatomy for image-guided surgical procedures. | | | The OR table is suitable for interventional and surgical procedures. | | Technology: | The proposed Discovery IGS 740 is a new combination of the laser | | | guided mobile AGV gantry of the Discovery IGS 730 (K113403) and of<br>the 41cm solid state detector of the Innova IGS 540 (K122457). | | | This new combination is the most important difference. It impacts the<br>collision avoidance system which has been adapted for the 41cm<br>detector. | | | This adaptation does not impact the safety and effectiveness of the<br>proposed device established with the predicate devices (K113403 &<br>K122457). | | Determination of Substantial<br>Equivalence: | Discovery IGS 740 complies with the voluntary and mandatory<br>standards listed in Table 1. | | | Summary of Non-Clinical Tests: | | | Simulated Use Testing ensures the system conforms to user<br>needs and intended uses through simulated clinical workflow<br>using step-by step procedures that would be performed for<br>representative clinical applications. | | | Usability validation testing is conducted to confirm that the product can be used safely and effectively. The participants for these tests are licensed and/or clinically trained healthcare providers or users. Product verification ensures the system conforms to its requirements including hazard mitigations risk management requirements. The verification tests confirm that design output meets design input requirements. Tests are executed at component, software subsystems and system levels. Dose verification, image quality verification and functional testing are part of system level verification. Performance imaging of Innova CT (CA-CBCT) has been confirmed with bench testing for uniformity, contrast and detail imaging performances. Software reliability is verified through intensive cycling of the system. The cycling tests use scenarios that simulate clinical workflow that would be performed for representative clinical applications. Safety testing is performed to confirm that the product meets the requirements of the standards listed in Table 1. | | | The Non-clinical Testing confirms Discovery IGS 740 meets the user requirements and its intended use and meets the design inputs requirements. | | | <b>Summary of Clinical Tests:</b> | | | The subject of this premarket submission, Discovery IGS 740, did not require clinical studies to support substantial equivalence. It does not introduce new indications for use. Substantial equivalence relies on clinical information that is pre-existing on the cleared predicate device (K023178). | | Conclusion: | GE Healthcare considers the Discovery IGS 740 to be as safe and effective as its predicate devices, and its performances to be substantially equivalent to the predicate device(s). | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circle. There are three swirls around the outside of the circle. Appendices K133278, Discovery IGS 740 Supplement {2}------------------------------------------------ # GE Healthcare Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a circular frame. The letters and the frame are in black, contrasting with the white background. The logo is simple, iconic, and easily recognizable. Appendices # K133278, Discovery IGS 740 Supplement {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. There are three decorative swirls equally spaced around the outside of the circle. GE Healthcare Appendices K133278, Discovery IGS 740 Supplement # Table 1 : · | Standards N° | Standards<br>organization | Standards Title | Version | Date | |----------------------------------|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------| | 21 CFR 1020.30-32 | FDA | FDA Federal Performance Standard | 2013 | 2013 | | 60601-1 | IEC | Medical Electrical Equipment – General<br>requirements for basic safety and<br>essential performance | Ed 3 | 2005 | | 60601-2-43 | IEC | Medical Electrical Equipment – Part 2-43:<br>Particular requirements for the safety of<br>X-ray equipment for interventional<br>procedures (2nd edition) | Ed 2 | 2010 | | 60601-1-2 | IEC | Medical Electrical Equipment – Part -1-2:<br>General requirements for safety –<br>Collateral standard: Electromagnetic<br>Compatibility – Requirements and tests | Ed 3 | 2007 | | 60601-1-3 | IEC | Medical Electrical Equipment – Part -1:<br>General requirements for safety, Section<br>1.3 Collateral standard: General<br>requirements for radiation protection in<br>diagnostic X-ray equipment | Ed 2 | 2008 | | 60601-1-6 | IEC | Medical electrical equipment -Part 1-6:<br>General requirements for basic safety<br>and essential performance -Collateral<br>standard : Usability | Ed 3 | 2010 | | IEC60601-2-46<br>(for the table) | IEC | Medical electrical equipment, Part2-46:<br>Particular requirements for the safety of<br>operating tables | Ed 2 | 2010 | {4}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -WO66-G609 Silver Spring, MD 20993-0002 Image /page/4/Picture/11 description: The image shows a partial view of a logo or emblem, featuring bold, stylized lines that resemble a graphic representation of a human figure or abstract design. The lines are thick and black, creating a strong contrast against the white background. To the left of the graphic, there is a curved arrangement of text, which appears to be part of the organization's name or motto. The text is oriented vertically and follows the curve of the emblem, suggesting a circular or semi-circular layout for the complete logo. #### April 24, 2014 GE Medical Systems Scs Michel Genuer 283 Rue De La Miniere Buc, 78530 FRANCE Re: K133278 > Trade/Device Name: Discovery IGS 740 Regulation Number: 21 CFR 192.1650 Regulation Name: Interventional Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB Dated: April 14, 2014 Received: April 16, 2014 Dear Michel Genuer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ #### Page 2 - Michel Genuer Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safcty/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Smh 77) for Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K133278 Device Name Discovery IGS 740 Indications for Use (Describe) The angiographic X-ray systems are indicated for use for patients from newborn to generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Image /page/6/Picture/15 description: The image contains a signature. The signature is illegible, but it appears to be written in cursive. The signature is black and is on a white background. The signature is likely from a document or other official paper. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## -DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the i ne burden inte for this oclosion of millen facts to rees, gather and maintain the data needed and complete time to review the collections, Send commons regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...