INFISTITCH FOR I5 DIGITAL X-RAY IMAGING SYSTEM, MODEL I5

K101833 · Infimed, Inc. · OWB · Jul 23, 2010 · Radiology

Device Facts

Record IDK101833
Device NameINFISTITCH FOR I5 DIGITAL X-RAY IMAGING SYSTEM, MODEL I5
ApplicantInfimed, Inc.
Product CodeOWB · Radiology
Decision DateJul 23, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1650
Device ClassClass 2

Intended Use

The InfiMed i5m Digital X-ray Imaging System is a high resolution digital imaging system intended to replace conventional film techniques, or existing digital systems, in multipurpose or dedicated applications specified below. The i5m Digital X-ray Imaging System enables an operator to acquire, display, process, export images to portable media, send images over a network for long term storage and distribute hardcopy images with a laser printer. Image processing algorithms enable the operator to bring out diagnostic details difficult to see using conventional imaging techniques. Images can be stored locally for temporary storage. The i2m Digital X-ray Imaging System has the ability to interface with a variety of image receptors from CCD cameras to commercially available flat panel detectors. The major system components include an image receptor, computer, monitor and imaging software. For the DR application, the InfiMed i500 Digital X-ray Imaging System is intended for use in general radiographic examinations and applications (excluding fluoroscopy, angiography, and mammography). For the RF/DSA application, the InfiMed if TM Digital X-ray Imaging System is intended for use where general fluoroscopy, interventional fluoroscopy or angiography imaging procedures are performed.

Device Story

InfiStitch for i5 Digital X-ray Imaging System is an optional software feature for the i5 Digital X-ray Imaging System. It allows operators to acquire sequential radiographic images and electronically join them to create a single composite electronic image. The system accepts inputs from various X-ray receptors, including CCD cameras and flat panel detectors (e.g., Trixell, Varian, Carestream, Samsung, Toshiba). The system consists of an image receptor, computer, monitor, and imaging software. It is used by clinicians in clinical settings to acquire, process, display, and export images. The stitching feature assists in creating a single view from multiple exposures, aiding diagnostic visualization. The system supports network storage and hardcopy printing.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical verification and validation testing performed in accordance with established protocols and FDA guidance for solid-state X-ray imaging devices.

Technological Characteristics

Digital X-ray imaging system comprising image receptor, computer, monitor, and software. Interfaces with various CCD and flat panel detectors. Features image stitching software for joining sequential radiographic images. Connectivity via network for storage and export. Operates as a digital imaging platform for general radiography, fluoroscopy, and angiography.

Indications for Use

Indicated for general radiographic examinations (excluding mammography) and general fluoroscopy, interventional fluoroscopy, or angiography procedures. Intended for use in multipurpose or dedicated clinical applications as a replacement for conventional film or existing digital systems.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the word "INFIMED" in bold, black letters. To the left of the word is a black, geometric shape that resembles a stylized diamond or square with curved lines. The font of the word is sans-serif, and the letters are closely spaced together. The overall impression is a logo or branding element for a company or product named Infimed. # Section 3: 510(k) Summary Date Prepared: 6/30/10 Contact Person: Catherine Mulcahy Quality Assurance Manager Telephone: Fax: 315-453-4545 x288 315-453-4550 Submitter Name: InfiMed. Inc. 121 Metropolitan Drive Liverpool, NY 13088 Device Trade Name: ·InfiStitch for i5™ Digital X-ray Imaging System Common Name: Digital X-ray Imaging System Classification Name(s): InfiStitch for i5m4 Digital X-ray Imaging System Product Codes: JAA, MQB, oWB 101833 JUL 2 8 2010 ( Predicate Device: iSTM Digital X-ray Imaging System 510(k) Number: K093066 Product Codes: JAA. MOB ## Device Description: The InfiMed i m Digital X-ray Imaging System is a high resolution digital imaging system designed for digital X-ray imaging through the use of an X-ray detector. The InfiMed i-1 Digital X-ray Imaging System is designed to support general radiography (excluding mammography), fluoroscopy, interventional fluoroscopy or angiography imaging procedures through a single common imaging platforms. The modified InfiMed if me Digital X-ray Imaging System consists of an X-ray imaging receptor (any of the following: CCD Camera, Trixell Pixium 3543, Trixell Pixium 4600, Varian PaxScan 4336R, Varian PaxScan 4343R, Carestream Health Detector, Samsung LTX240AA01-A, Toshiba FDX 4343R), computer, monitor, and the digital i maging system. The InfiStitch for i-m Digital X-ray Imaging System is an option which allows the operator to generate sequential radiographic images and electronically jointhum to create a single electronic image. This feature will be available for any InfiMed customer utilizing the i-The Digital X-ray Imaging System DR application on any flat parel detector. #### Intended Use: The InfiMed i5m Digital X-ray Imaging System is a high resolution digital imaging system intended to replace conventional film techniques, or existing digital systems, in {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "INFIMED" in all capital letters. To the left of the word is a black logo. The logo appears to be a stylized design, possibly representing a medical or technological concept. multipurpose or dedicated applications specified below. The i5m Digital X-ray Imaging System enables an operator to acquire, display, process, export images to portable media, send images over a network for long term storage and distribute hardcopy images with a laser printer. Image processing algorithms enable the operator to bring out diagnostic details difficult to see using conventional imaging techniques. Images can be stored locally for temporary storage. The i2m Digital X-ray Imaging System has the ability to interface with a variety of image receptors from CCD cameras to commercially available flat panel detectors. The major system components include an image receptor, computer, monitor and imaging software. For the DR application, the InfiMed i500 Digital X-ray Imaging System is intended for use in general radiographic examinations and applications (excluding fluoroscopy, angiography, and mammography). For the RF/DSA application, the InfiMed if TM Digital X-ray Imaging System is intended for use where general fluoroscopy, interventional fluoroscopy or angiography imaging procedures are performed. ## Technological Characteristics Comparison: The modified device supports the same modalities as the predicate device with the same components or imaging concepts, and delivers equivalent or better image quality as the predicate device. The comparison chart reveals that functions performed by the predicate device are performed by the modified if TM Digital X-ray Imaging System. The modified device, InfiStitch for if Digital X-ray Imaging System, also has the ability to interface with an additional image receptor, Toshiba FDX4343R. Therefore, the modified device is substantially equivalent to the predicate device. ### Non-clinical Tests Discussion: Validation was completed in accordance with the Validation Protocols included with this submission. Protocols were designed, executed and documented according to the Design Validation process with predetermined test methods and corresponding acceptance criteria. In conclusion, all release criteria have been met and the modified i-m Digital Xray Imaging System is as safe and effective as the predicate device. ### Clinical Tests Discussion: Clinical Data submitted is consistent with FDA guidance document "Guidance for Industry and/or for FDA Reviewers/Staff and/or Compliance: Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices" available at the website http://www.fda.gov/downloads/McdicalDevices/DeviceRegulationandGuidance/Guidanc eDocuments/ucm073781.pdf. #### Conclusion: {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the word "INFIMED" in all capital letters. To the left of the word is a black graphic that appears to be a stylized logo. The font of the word is bold and sans-serif. The image is black and white. Based upon the results of Verification and Validation testing, the InfiMed InfiStitch for if TM Digital X-ray Imaging System has no new indications for use, has no simificant technological differences, and is as safe and effective as, and therefore substantially equivalent to the above listed current legally marketed predicate device. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in a bold, sans-serif font and is blue. The text is likely part of a document or presentation. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 InfiMed, Inc. % Mr. Jigar Shah Consultant mdi Consultants, Inc. 55 Northern Blvd., Suite 200 GREAT NECK NY 11021 JUL 30 2012 Re: K101833 Trade/Device Name: InfiStitch for iSTM Digital X-ray Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA, and MQB Dated: June 30, 2010 Received: July 1, 2010 Dear Mr. Shah: This letter corrects our substantially equivalent letter of July 23, 2010. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of {4}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "INFIMED" in bold, black letters. To the left of the word is a black, geometric shape that resembles a stylized diamond or square. The letters are evenly spaced and appear to be a sans-serif font. The overall impression is a logo or brand name. ## Section 2: Indications for Use K101833 510(k) Number (if known): Device Name: InfiStitch for i5m Digital X-ray Imaging System ### Indications for Use: The InfiMed if TM Digital X-ray Imaging System is a high resolution digital imaging system intended to replace conventional film techniques, or existing digital systems, in multipurpose or dedicated applications specified below. The i5m Digital X-ray Imaging System enables an operator to acquire, display, process, export images to portable media, send images over a network for long term storage and distribute hardcopy images with a laser printer. Image processing algorithms enable the operator to bring out diagnostic details difficult to see using conventional imaging techniques. Images can be stored locally for temporary storage. The i2m Digital X-ray Imaging System has the ability to interface with a variety of image receptors from CCD cameras to commercially available flat panel detectors. The major system components include an image receptor, computer, monitor and imaging software. For the DR application, the InfiMed i-714 Digital X-ray Imaging System is intended for use in general radiographic examinations and applications (excluding fluoroscopy, angiography, and mammography). For the RF/DSA application, the InfiMed if TM Digital X-ray Imaging System is intended for use where general fluoroscopy, interventional fluoroscopy or angiography imaging procedures are performed. Prescription Use \ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | OIVD | |------------------------------------------------------------|---------| | | | | | | | (Division Sign-Off) | | | Division of Radiological Devices | | | Office of In Vitro Diagnostic Device Evaluation and Safety | | | 610K. | K101833 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...