JUSHA-C43 LCD Monitor is intended to be used in displaying and viewing digital images for review and analysis by trained medical practitioners. It does not support the display of mammography images for diagnosis.
Device Story
JUSHA-C43 is a 30-inch color active matrix TFT LCD monitor; provides high-resolution (2560 x 1600) visual output for medical imaging. Device receives digital image data via DVI/DisplayPort inputs; utilizes internal DICOM look-up tables to ensure accurate grayscale and color representation. Operated by trained medical practitioners in clinical environments; provides visual display for review and analysis of medical images. Built-in calibration sensor maintains luminance stability. Output assists clinicians in diagnostic review; excludes mammography. Benefits include high-resolution visualization of medical data to support clinical decision-making.
Clinical Evidence
Bench testing only. No clinical or animal studies conducted. Bench testing included measurement of angular dependency of luminance, luminance/chromaticity non-uniformity (per TG18 guidelines), small-spot contrast ratio, temporal response, and luminance stability. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
30-inch color active matrix TFT LCD; 2560 x 1600 resolution; 350cd/m² luminance; 12-bit color depth (68.7 billion colors); 1024 simultaneous shades of gray. Connectivity: DVI-D, DisplayPort. Features built-in calibration sensor. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC).
Indications for Use
Indicated for trained medical practitioners to display and view digital medical images for review and analysis. Not for mammography diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 10, 2015
Nanjing Jusha Display Technology Co., Ltd. % Ma Jing Certification Engineer 301 Hnazhongmen Street, 8F Block A, No. 1 Nanjing International Service Outsourcing Mansion, Nanjing Jiangsu 210036 CHINA
Re: K151972
Trade/Device Name: JUSHA-C43 LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: June 16, 2015 Received: July 16, 2015
Dear Ma Jing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151972
Device Name JUSHA-C43 LCD Monitor
Indications for Use (Describe)
JUSHA-C43 LCD Monitor is intended to be used in displaying digital images for review and analysis by trained medical practitioners. It does not support the display of mammography images for diagnosis.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | Jun,15,2015 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Nanjing Jusha Display Technology Co., Ltd<br>Add: 301, Hanzhongmen Street, 8F Block A, No.1, Nanjing<br>International Service Outsourcing Mansion, Nanjing, 210036 China |
| Contact Person: | Ma Jing<br>Certification Engineer<br>Nanjing Jusha Display Technology Co., Ltd<br>Tel: +86-25- 83305050<br>Fax: +86-25- 58783271 |
| Device Trade Name: | JUSHA-C43 LCD Monitor |
| Common/Usual Name: | 4MP Color LCD Monitor |
| Classification Name: | System, image processing ,Radiology |
| Product Code: | PGY |
| Predicate Device(s): | RADIFORCE RX440;K130070 |
| Device Description: | JUSHA-C43 LCD Monitoris the display system with the high<br>resolution(2560 x 1600), high luminance(350cd/m²) and 1024<br>simultaneous shades of gray out of a palette of 4096, 4 DICOM<br>look up table inside, the product is consisted of the following<br>components:<br>- 30inch, Color Active Matrix Liquid Crystal Display<br>- FR4-v0.4 |
| - JUSHA-C43 LCD Monitor software<br>- Power Adapter<br>- Data Cable.<br><br>The LCD Monitoris designed, tested, and will be manufactured<br>in accordance with both mandatory and voluntary standards:<br>1. IEC 60601-1Medical equipment medical electrical<br>equipment - Part 1: General requirements for basic<br>safety and essential performance 2005 + CORR. 1<br>(2006) +CORR. 2 (2007)<br>2. IEC 60601-1-2 Edition 3:2007, medical electrical<br>equipment - Part 1-2: General requirements for<br>basic safety and essential performance - Collateral<br>standard: Electromagnetic compatibility -<br>Requirements and tests. | |
| Intended Use: | JUSHA-C43 LCD Monitoris intended to be used in displaying<br>and viewing digital images for review and analysis by trained<br>medical practitioners. It does not support the display of<br>mammography images for diagnosis. |
| Technology: | JUSHA-C43 LCD Monitor is the display system with the high<br>resolution monitor (4 megapixels) with electronic capabilities for<br>evaluation of high resolution medical images, high<br>luminance(350cd/m²) and 1024 simultaneous shades of gray<br>out of a palette of 4096, 4 DICOM look up table inside |
| Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Tests:<br>The LCD Monitor complies with voluntary standards as<br>following:<br>1 IEC 60601-1Medical equipment medical electrical |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
| | require clinical studies to support substantial equivalence.<br>The proposed device is Substantially Equivalent (SE) to the<br>predicate device which is US legally market device. Therefore,<br>the subject device is determined as safe and effectiveness. |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | Nanjing Jusha Display Technology Co., Ltd Considers the<br>JUSHA-C43 LCD Monitor to be as safe as effective, and<br>performance is substantially equivalent to the predicate<br>device(s). |
{8}------------------------------------------------
This comparison identifies the similarities and differences of the proposed JUSHA-C43 LCD Monitor device to the legally marketed predicate EIZO RADIFORCE RX440 LCD Monitor device to which substantial equivalency is claimed.
| Attributes | Predicate Device | Proposed Device | Discussion of<br>Differences |
|------------------------------------|----------------------------------|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Product | RADIFORCE RX440 | JUSHA-C43 LCD Monitor | |
| 510(k) Number | K130070 | / | |
| Display Performance/Specifications | | | |
| Screen<br>technology | 29.8" Color TFT LCD<br>Panel | 30", Color Active Matrix<br>TFT LCD Panel | The panel of<br>JUSHA-C43 is larger<br>than the predicate<br>device, it can display<br>more<br>information |
| Viewing angle<br>(H, V) | Horizontal 176°,Vertical<br>176° | Horizontal 178°,Vertical<br>178°(CR > 10) | The viewing angle of<br>JUSHA-C43 is wider<br>than the predicate<br>device, it can get<br>greater field of view |
| Resolution | 2560 x1600/1280 x1600<br>x2 | 2560 x 1600/1280 x 1600<br>x2 | Same |
| Display area | 641.2(H) x 400.8 (V) mm | 641.28 (H) x 400.8 (V) mm | Same |
| Recommended<br>Luminance | 300cd/m² | 300cd/m² | Same |
{9}------------------------------------------------
| Attributes | Predicate Device | Proposed Device | Discussion of<br>Differences |
|------------------------------|-----------------------------------------------------------------------------------------------------------------|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product | RADIFORCE RX440 | JUSHA-C43 LCD Monitor | |
| 510(k) Number | K130070 | / | |
| Contrast Ratio | 1100:1 | 1200:1 | The contrast ratio of<br>JUSHA-C43 is higher<br>than the predicate<br>device, it can display<br>image more clarity |
| Scanning<br>frequency (H; V) | 31 - 159 kHz / 29 - 61 Hz<br>(VGA Text: 69 - 71 Hz)<br>Frame synchronous<br>mode: 59 - 61 Hz, 29.5 -<br>30.5 Hz | 103.8kHz;50Hz | The two differences<br>between the two<br>device is caused by<br>the different no<br>display area defined<br>by different<br>manufacturers, but<br>the display area is<br>same(2048*2560),so<br>the same picture<br>displayed on the two<br>devices has the same<br>effect |
| Dot clock | 152MHz | 148MHz | by different<br>manufacturers, but<br>the display area is<br>same(2048*2560),so<br>the same picture<br>displayed on the two<br>devices has the same<br>effect |
| Pixel Pitch | 0.2505 x 0.2505 mm | 0.2505 x 0.2505 mm | Same |
| Backlighting | LED | LED | Same |
{10}------------------------------------------------
| Attributes | Predicate Device | Proposed Device | Discussion of<br>Differences |
|-----------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Product | RADIFORCE RX440 | JUSHA-C43 LCD Monitor | |
| 510(k) Number | K130070 | / | |
| Display Colors | 10-bit, 1.07 billion colors | 12-bit , 68.7billion colors | The JUSHA-C43 LCD<br>Monitor uses a color<br>extension technology<br>to improve image<br>display quality |
| Luminance<br>calibration | Built in calibration sensor<br>provided | Built in calibration sensor<br>provided | Same |
| Video Signal Input | | | |
| Input signals | DVI standard 1.0 | DVI standard 1.0 | Same |
| Input<br>terminational | DVI-D (dual link) x 1,<br>DVI-D (single link) x 1,<br>DisplayPort x 1 | DVI-D (dual link) x 1,<br>DVI-D (single link) x 1,<br>DisplayPort x 1 | Same |
| Display<br>controller | Off the shelf | Off the shelf | Same |
| Power Related Specification | | | |
| Power<br>Requirement | AC 100 - 120 V, 200 - 240<br>V: 50 / 60 Hz | AC 100-240V 50-60Hz | same |
{11}------------------------------------------------
| Attributes | Predicate Device | Proposed Device | Discussion of<br>Differences |
|----------------------------------------|----------------------------------------|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product | RADIFORCE RX440 | JUSHA-C43 LCD Monitor | |
| 510(k) Number | K130070 | / | |
| Power<br>Consumptions/Save Mode | 84W/less than 0.7W | 82W/less than1.5W | The differences<br>caused by different<br>adapter and<br>components used in<br>the LCD Monitor, the<br>JUSHA-C43 has<br>fewer power<br>consumptions, this<br>only shows the power<br>consumption is<br>different, nothing to<br>do with the display<br>function |
| Power<br>Management | DVI DMPM, DisplayPort<br>1.1a | DVI DMPM, DisplayPort<br>1.1a | Same |
| Miscellaneous Features/Specifications | | | |
| USB<br>Ports/standard | 1 upstream ,<br>2 downstream/ Rev. 2.0 | 1 upstream (endpoint), 2<br>downstream/ Rev. 2.0 | Same |
| Dimensions w/o<br>stand<br>(W x H x D) | Without stand:<br>702x108x473 mm | With stand:<br>692x252x523mm<br>Without stand:<br>692x87x453mm | Different housing<br>design due to the<br>different panel size. |
{12}------------------------------------------------
| Attributes | Predicate Device | Proposed Device | Discussion of<br>Differences |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Product | RADIFORCE RX440 | JUSHA-C43 LCD Monitor | |
| 510(k) Number | K130070 | / | |
| Indication for use | RADIFORCE RX440 is<br>intended to be used in<br>displaying and viewing<br>digital images for review<br>and analysis by trained<br>medical practitioners. It<br>does not support the<br>display of mammography<br>images for diagnosis. | JUSHA-C43 LCD Monitor<br>is intended to be used in<br>displaying and viewing<br>digital images for review<br>and analysis by trained<br>medical practitioners. It<br>does not support the<br>display of mammography<br>images for diagnosis. | Same |
{13}------------------------------------------------
| Attributes | Predicate Device | Proposed Device | Discussion of Differences |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product | RADIFORCE RX440 | JUSHA-C43 LCD Monitor | |
| 510(k) Number | K130070 | / | |
| Applicable<br>standard | 1 IEC 60601-1Medical<br>equipment medical<br>electrical equipment - Part<br>1: General requirements<br>for basic safety and<br>essential performance<br>1988+A1 : 1991 +<br>A2:1995<br>2 IEC 60601-1-2 Edition<br>3:2007, Medical electrical<br>equipment - Part 1-2:<br>General requirements for<br>basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility -<br>Requirements and tests. | 1 IEC 60601-1Medical<br>equipment medical<br>electrical equipment - Part<br>1: General requirements<br>for basic safety and<br>essential performance<br>IEC 60601-1:2005 +<br>CORR. 1 (2006) +CORR. 2 (2007)<br>2 IEC 60601-1-2 Edition<br>3:2007, Medical electrical<br>equipment - Part 1-2:<br>General requirements for<br>basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility -<br>Requirements and tests. | The JUSHA-C43<br>safety standard is<br>edition 3.0,<br>RADIFORCE RX440<br>safety standard is<br>edition 2.0. nothing to<br>do with the display<br>function |
{14}------------------------------------------------
## PERFORMANCE DATA:
The following performance data were provided in support of the substantial equivalence determination.
### Bench testing:
Bench testing was conducted to demonstrate the JUSHA-C43 meets all performance standards as follows:
- . Measurement of the angular dependency of luminance response in horizontal, vertical and diagonal directions
- . Measurement of the luminance non-uniformity characteristics of the display screen as specified in TGI18 guideline.
- Measurement of the chromaticity non-uniformity characteristics of the display screen as specified in TG18 guideline.
- Measurement of small-spot contrast ratio.
- Measurement of temporal response
- Performance data on luminance stability
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the JUSHA-C43. The device complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC.
### Animal and clinical study
The subject of this premarket submission, JUSHA-C43, does not require animal or clinical studies to support substantial equivalence.
# CONCLUSIONS
JUSHA-C43 Medical Display is substantially equivalent to the predicate device with respect to technical characteristics, performance, application and intended use. The non-clinical data support the safety of the device should perform as intended in the specified use conditions. Nanjing Jusha Display Technology Co., Ltd considers the JUSHA-C43 Medical Display does not raise any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.