XPLUS 35 Series (XPLUS 35, XPLUS 35FD) are mobile digital C-arms designed to provide fluoroscopic and radiographic images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include endoscopy, orthopedic, neurologic, critical care and emergency room procedures. i.e. surgical interventions needing X-ray imaging and/or guidance and interventions inside and outside the operating room.
Device Story
Mobile digital C-arm system; provides fluoroscopic and radiographic imaging for diagnostic, surgical, and interventional procedures. Components include X-ray tube, generator, C-arm with image receptor (I.I. or FPD), monitor cart, and medical monitors. Operated by healthcare professionals in ORs or clinical settings. Input: X-rays penetrating patient anatomy; transformed by image receptor into signals. Output: Live and reference images displayed on medical monitors; images can be edited, saved, and transmitted to PACS. Facilitates real-time visualization for surgical guidance; assists clinicians in diagnostic and interventional decision-making. Benefits include improved anatomical visualization during procedures.
Clinical Evidence
Bench testing only. Software verification testing confirmed functional requirements, safety risk controls, and cybersecurity mitigations. Non-clinical validation testing confirmed conformance to intended use, user needs, and safety measures. No clinical data presented.
Technological Characteristics
Mobile digital C-arm; X-ray tube (stationary anode, 0.5/1.5 mm focal size); Image receptors (I.I. 9/6/4.5" or FPD 9"); kV range 40-110; connectivity to PACS; software-based image processing/management. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-28, 60601-2-43, and 21 CFR 1020.30-32.
Indications for Use
Indicated for patients requiring fluoroscopic and radiographic imaging during diagnostic, surgical, and interventional procedures, including orthopedic, neurologic, vascular, cardiac, angiography, endoscopy, urologic, critical care, and emergency room applications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
November 16, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure, while the FDA part includes the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Gemss Healthcare Co., Ltd. % Mr. Dave Kim RA Consultant Mtech Group LLC 7505 Fannin St. Suite 610 HOUSTON TX 77054
Re: K233200
Trade/Device Name: XPLUS 35 Series (XPLUS 35, XPLUS 35FD) Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA, OXO Dated: September 22, 2023 Received: September 28, 2023
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233200
Device Name XPLUS 35 Series (XPLUS 35, XPLUS 35FD)
Indications for Use (Describe)
XPlus 35 Series (XPLUS 35. XPLUS 35FD) are mobile digital C-arms designed to provide fluoroscopic and radiographic images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include endoscopy, orthopedic, neurologic, critical care and emergency room procedures.
i.e. surgical interventions needing X-ray imaging and interventions inside and outside the operating room.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for GEMSS Healthcare. The word "GEMSS" is written in orange, with each letter connected to the next. Below the word "GEMSS" is the word "HEALTHCARE" in gray. The logo is simple and modern.
1F, 822, Bogwang-ro, Gwangtan-myeon, Paju-si, Gyeonggi-do, 10952 South Korea Tel. +82 31-906-9017 Fax. +82 31-906-9019 sales@gemss-healthcare.com
# 510(k) Summary K233200
This summary of 510(K) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: November 15, 2023
### 1. Applicant / Submitter
- GEMSS HEALTHCARE CO., LTD. ।
- Address: 1F, 822, Bogwang-ro, Gwangtan-myeon, Paju-si, Gyeonggi-do, 10952, Republic of Korea ।
- Tel: +82-31-906-9017 | Fax: +82-31-906-9019 | Email:olivia@gemss.co.kr -
- Official Correspondent (U.S): Dave Kim / Mtech Group LLC । 7505 Fannin St. Suite 610, Houston, TX 77054 / Tel) 1-713-467-2607 / davekim@mtechgroupllc.com
### 2. Device Information
- Trade/Device Name: XPLUS 35 Series (XPLUS 35, XPLUS 35FD) -
- Regulation Number: 21CFR 892.1650 -
- Regulation Name: Image-Intensified Fluoroscopic X-ray System -
- Device Class: Class II -
- । Product Code: OWB
- -Subsequence product code: JAA, OXO
### 3. Predicate Device
- K Number: K091918
- Manufacturer: GENORAY Co., Ltd. -
- Trade Name: ZEN-2090 Pro -
- Regulation Number: 21CFR 892.1650 -
- Regulation Name: Image-Intensified Fluoroscopic X-ray System -
- -Device Class: Class II
- Product Code: OWB ।
- Subsequence product code: JAA, OXO -
### 4. Reference Device
- K Number: K222080 -
- -Manufacturer: SG HealthCare Co. Ltd.
- -Trade Name: Garion
- Requlation Number: 21CFR 892.1650 -
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for GEMSS Healthcare. The word "GEMSS" is in orange, and the word "HEALTHCARE" is in gray. The logo is simple and modern.
- Regulation Name: Image-Intensified Fluoroscopic X-ray System
- Device Class: Class II
- Product Code: OWB ।
## 5. General Description
XPLUS 35 Series (XPLUS 35, XPLUS 35FD), C-arm mobile, is used for providing fluoroscopic image of patient anatomy, especially during diagnostic, surgical and interventional procedures. XPLUS 35 Series (XPLUS 35, XPLUS 35FD) consist of X-ray tube, X-ray generator, C-arm including image receptor, monitor cart, medical monitor and accessories such as foot switch and hand switch.
- 1) C-arm
- X-ray tube: It converts high voltage currents into an X-ray through a generator. -
- -Image receptors: It translates the x-ray, which penetrates the patient's body, into the image signals.
- 2) Monitor cart
Main C-arm units and monitor cart are connected to each other with a cable. There is a main power cable to supply power to the monitor cart and main unit, which is not a battery type.
- Software: The displayed images can be edited and saved, and the images can also be sent to PACS server. Moreover, the patient information and database can be managed through the monitor cart
- Medical montor: It displays the images by the monitor through visualizing the image signals. It displays live image on the left, and reference image on the right.
- 3) Accessories
- Foot switch: It is used for fluoroscopy and can save images. -
- Hand switch: It is used in both radiographic and fluoroscopic modes. It controls X-ray generation.
### 6. Indication for use
XPlus 35 Series (XPLUS 35, XPLUS 35FD) are mobile digital C-arms designed to provide fluoroscopic and radiographic images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include endoscopy, orthopedic, neurologic, critical care and emergency room procedures. i.e. surgical interventions needing X-ray imaging and/or guidance and interventions inside and outside the operating room.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for GEMSS Healthcare. The word "GEMSS" is written in orange, and the word "HEALTHCARE" is written in gray below it. The font is sans-serif and the text is left-aligned.
Comparison of the subject device to the predicate device.
| Descriptive Information | Subject Device | Predicate Device | Reference Device | Note | |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|---------|
| Manufacturer | GEMSS<br>HEALTHCARE CO.,<br>LTD. | GENORAY Co.,<br>Ltd. | SG HealthCare Co.<br>Ltd. | | |
| Device Name | XPLUS 35,<br>XPLUS 35 FD | ZEN-2090 Pro | Garion | | |
| 510(k) number | K233200 | K091918 | K222080 | - | |
| Regulatory Number | 21CFR 892.1650 | 21 CFR 892.1650 | 21 CFR 892.1650 | Same | |
| Product Code | OWB, JAA, OXO | OWB, JAA, OXO | OWB | Same | |
| Regulatory Class | 2 | 2 | 2 | Same | |
| Indications for Use | XPLUS 35 series<br>(XPLUS 35, XPLUS<br>35FD) are mobile<br>digital C-arms<br>designed to provide<br>fluoroscopic and<br>radiographic images<br>of the patient during<br>diagnostic, surgical<br>and interventional<br>procedures.<br>Examples of clinical<br>application may<br>include orthopedic,<br>neurologic, vascular,<br>cardiac,<br>angiography,<br>endoscopy, urologic,<br>critical care and<br>emergency room<br>procedures. i.e.<br>surgical interventions<br>needing X-ray<br>imaging and/or<br>guidance and<br>interventions inside<br>and outside the<br>operating room. | ZEN-2090 Pro is<br>a mobile digital<br>C-arm designed<br>to provide<br>fluoroscopic and<br>radiographic<br>images of the<br>patient during<br>diagnostic,<br>surgical and<br>interventional<br>procedures.<br>Examples of<br>clinical<br>application may<br>include<br>cholangiography,<br>endoscopy,<br>orthopedic,<br>neurologic, stone<br>location, critical<br>care and<br>emergency room<br>procedures<br>i.e. surgical<br>interventions<br>needing X-ray<br>imaging and/or<br>guidance and<br>interventions<br>inside and<br>outside the<br>operating room. | The Garion is<br>intended to be<br>used and<br>operated by:<br>adequately<br>trained, qualified<br>and authorized<br>health care<br>professionals who<br>have full<br>understanding of<br>the safety<br>information and<br>emergency<br>procedures as well<br>as the capabilities<br>and functions of<br>the device. The<br>device is used for<br>radiological<br>guidance and<br>visualization<br>during diagnostic,<br>interventional and<br>surgical<br>procedures on all<br>patients, except<br>neonates (birth to<br>one month),<br>within the limits of<br>the device. The<br>device is to be<br>used in healthcare<br>facilities both<br>inside and outside<br>the operating<br>room, sterile as<br>well as non-sterile<br>environment in a<br>variety of<br>procedures. | Similar | |
| Radiogra | kV range | 40 to 110 kV | 40 to 110 kV | 40 to 125 kV | Same |
| phic | mA Range | 30 to 45 mA | 20 mA | 0.3~ 32 mA | Similar |
| Mode | mAs Range | 1.0 to 200 mAs | 0.4 to 100 mAs | 0.2 ~ 100 mAs | Similar |
| Fluorosco<br>pic Mode | kV range | 40 to 110 kV | 40 to 110 kV | 40 to 110 kV | Same |
| | mA range | 0.2 to 8 mA<br>10 mA (Boost Mode) | 0.2 to 6 mA<br>10 mA (Boost<br>mode) | 0.3 ~ 6.3 mA | Similar |
| | Pulse<br>Fluoro | Yes | Yes | Yes | Same |
| C-arm | Range of<br>C-arm Rail<br>Rotation | I.I Type: 135 ° (90<br>° / 45 °)<br>FPD Type: 165°<br>(120° / 45°) | 120 °<br>(90 ° / 30 °) | 150 ° | Similar |
| | Range of<br>the<br>Horizontal<br>C-arm<br>Movement | 200 mm | 200 mm | 200 mm | Same |
| | Range of<br>the Vertical<br>C-arm<br>Movement | 500 mm | 400 mm | 400 mm | Similar |
| | Range of<br>Swing-arm<br>Movement | + 12.5 ° | + 12.5 ° | + 15 o | Same |
| | Range of<br>Stay-arm<br>Rotation | 225 ° | 180 ° | 180 ° | Similar |
| Image<br>Receptor | Image<br>Intensifier | E5830SD-P4A<br>(Manufactured by<br>TOSHIBA)<br>9″ / 6″ / 4.5″ | E5830SD-P4A<br>(Manufactured<br>by TOSHIBA)<br>9″ / 6″ / 4.5″ | - | Same |
| | Flat Panel<br>Detector | Mercu 0909F<br>(Manufactured by<br>IRay) 9″ | | Mercu 0909F<br>(Manufactured by<br>IRay) 9″ | |
| | Pixel Pitch | 205µm | N/A | 205um | Same |
| | DQE @ 0<br>lp/mm | 77% at 0lp/mm | 65% at 0lp/mm | 77% at 0lp/mm | Same |
| | MTF @<br>1lp/mm | 61% | N/A | 61% | Same |
| | Frame<br>rates (fps<br>values) | N/A for I.I.<br>Up to 60fps for FPD | N/A | Up to 60fps | Same |
| Collimators | | Iris and Leaf | Iris and Leaf | Motor<br>control/rotation | Same |
| Laser Guide | | Yes | Yes | Yes | Same |
| | Model | DF-151SBR<br>(Manufactured by<br>Canon) | DF-151SBR<br>(Manufactured<br>by Canon) | Rotating Anode<br>(200kHU) | Same |
| X-ray<br>Tube | Anode<br>Type | Stationary Anode | Stationary Anode | Rotating anode | Same |
| Focal size | | 0.5/1.5 mm | 0.5/1.5 mm | 0.3/0.6 mm | Same |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for GEMSS Healthcare. The logo consists of the word "GEMSS" in orange, with the word "Healthcare" in gray underneath. The font is sans-serif and the letters are bold.
1F, 822, Bogwang-ro, Gwangtan-myeon, Paju-si, Gyeonggi-do, 10952 South Korea
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for GEMSS Healthcare. The word "GEMSS" is in orange, and the word "HEALTHCARE" is in gray. The logo is simple and modern.
XPLUS 35 Series (XPLUS 35, XPLUS 35FD) is substantially equivalent to the predicate device ZEN-2090 Pro (K091918) and Garion (K222080) These devices are very similar in the intended use, the design principle, the performance and the applicable standards. Some characteristics, for example, different X-ray generators, the appearance, and the user interfaces are different. The Xray generator capacity of the subject device has higher maximum power output compared with the predicate devices. The subject device utilized both solid state X-ray imager (FPD) and image intensifier which are the same as the predicate devices. All detectors used for the subject devices have been already cleared by FDA and do not raise new concerns for safety and effectiveness. However, these small and minor differences do not raise any new questions of safety and effectiveness.
#### 8. Safety, EMC and Performance data comparison to Predicate
- IEC60601-1 (ed3.1, 2012), Medical electrical equipment Part 1: General requirements for basic safety and essential performance (Recognition No: 19-4)
- -IEC60601-1-2: 2014, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances -Requirements and tests (Recognition No: 19-8)
- IEC60601-1-3 (ed 2.1, 2013), Medical electrical equipment Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment (Recognition No: 12-269)
- -IEC60601-2-28 (ed 3, 2017) Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis (Recognition No: 12-309)
- IEC60601-2-43 (ed 2.2, 2019) Medical electrical equipment Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures (Recognition No: 12-329)
In addition, the device complies with the FDA Radiation Safety Performance Standard Performance Standard for Diagnostic X-Ray Systems and Their Major Components (21CFR 1020.30, 1020.31, 1020.32); Small Entity Compliance Guide. In addition, labeling was developed and information provided in accordance with this FDA Guidance Document: Pediatric Information for X-ray Imaging Device Premarket Notifications, Guidance for Industry and Food and Drug Administration Staff. Labeling also includes reference to the Image Gently website (http://www.imagegently.org/). Also we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for GEMSS Healthcare. The logo consists of the word "GEMSS" in orange, with the word "HEALTHCARE" in gray underneath. The font is sans-serif and the letters are bold.
Software verification testing of the functional requirements as well as performance and safety has been performed to verify that all the requirements of System Requirements Specification as well as the safety risk control measures from the Detailed Risk Management Matrix and the Privacy and Security requirements and mitigations have been implemented. FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical devices" to ensure substantial equivalence.
Results demonstrated that all executed verification tests were passed.
Non-clinical validation testing has been performed to validate that XPLUS 35 Series (XPLUS 35, XPLUS 35FD) conforms to the intended use, claims, user needs, effectiveness of safety measures and instructions for use.
As a results, all test results were satisfactory and the result of bench and clinical evaluation indicates that the new device is as safe and effective as the predicate device.
### 9. Conclusion
In reference to the comparison provided in substantial equivalence chart, and the most of functions and electronic features are similar with predicate device. We believe that the XPLUS 35 Series (XPLUS 35, XPLUS 35FD) is safe and effective as predicate device, and has no new indication for use. Therefore, XPLUS 35 Series (XPLUS 35FD) is substantially equivalent to predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.