K201583 · Philips Medical Systems Nederland B.V. · OWB · Feb 9, 2021 · Radiology
Device Facts
Record ID
K201583
Device Name
SmartCT R1.0
Applicant
Philips Medical Systems Nederland B.V.
Product Code
OWB · Radiology
Decision Date
Feb 9, 2021
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
SmartCT assists physicians during vascular and non-vascular procedures, with diagnosis, treatment planning, interventional procedures and treatment follow-up by creating 3D views from sets of 2D images created during rotational acquisitions. SmartCT provides high-speed and high-resolution 3D visualizations of vasculature, hemorrhages, soft tissue and bone structures. SmartCT provides live image guidance for navigating endovascular devices through vascular structures anywhere in the body. SmartCT helps to assess anatomical information intra-procedurally, such as the estimate of a vessel or lesion size, diameter or volume, and anatomical distances between relevant structures. SmartCT is intended to be used for human patients that have been elected for the procedures as described in these Indications for Use.
Device Story
Software product (Interventional Tool) running on Interventional Workspot platform; used with Philips Interventional X-Ray systems. Inputs: 2D images from rotational acquisitions (3DRA, CBCT, VasoCT). Processing: 3D reconstruction; metal artifact reduction filters; skull removal; volume rendering. Outputs: 3D visualizations of vasculature, hemorrhages, soft tissue, bone; live 2D/3D overlays for navigation; manual/semi-automatic measurements (size, diameter, volume). Used in OR/interventional suite by physicians/nurses/technicians. Benefits: intra-procedural anatomical assessment; improved navigation; intervention planning.
Clinical Evidence
No clinical study performed. Substantial equivalence demonstrated via non-clinical performance testing, including software verification, usability validation in simulated environments with representative users, and expert clinical image assessment to validate marketing claims.
Technological Characteristics
3D visualization/analysis software; runs on Interventional Workspot platform. Features: 3D acquisition guidance, metal artifact reduction, skull removal, 3D volume rendering, vessel analysis tool. Connectivity: DICOM compliant. Standards: IEC 62304, IEC 62366-1, IEC 82304-1, ISO 14971, UL 2900-1 (cybersecurity).
Indications for Use
Indicated for human patients undergoing vascular and non-vascular procedures requiring 3D visualization, diagnosis, treatment planning, interventional guidance, and follow-up.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square. The full name of the agency is "U.S. Food & Drug Administration".
Philips Medical Systems, Nederland B.V. % Maria Silos Viu Regulatory Affairs Specialist Veenpluis 4-6 Best. Noord-Brabant 5684 PC THE NETHERLANDS
February 9, 2021
### Re: K201583
Trade/Device Name: SmartCT Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, JAK, LLZ, Dated: January 12, 2021 Received: January 15, 2021
Dear Maria Silos Viu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K201583
Device Name SmartCT
Indications for Use (Describe)
SmartCT assists physicians during vascular and non-vascular procedures, with diagnosis, treatment planning, interventional procedures and treatment follow-up by creating 3D views from sets of 2D images created during rotational acquisitions.
SmartCT provides high-speed and high-resolution 3D visualizations of vasculature, hemorrhages, soft tissue and bone structures.
SmartCT provides live image guidance for navigating endovascular structures anywhere in the body.
SmartCT helps to assess anatomical intra-procedurally, such as the estimate of a vessel or lesion size, diameter or volume, and anatomical distances between relevant structures.
SmartCT is intended to be used for human patients that have been elected for the procedures as described in the Indications for Use.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K201583
# 510(k) Summary SmartCT
The 510(k) Summary of the proposed SmartCT is provided in the following pages.
{5}------------------------------------------------
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | June 08, 2020 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684 PC Best<br>The Netherlands<br>Establishment Registration Number: 3003768277 |
| Primary Contact<br>Person: | María Silos Viu<br>Regulatory Affairs Specialist<br>Phone: +31 618596430<br>E-mail: maria.silos@philips.com |
| Secondary Contact<br>Person: | Michael Konings<br>Director of Regulatory Affairs<br>Phone: +31 646941461<br>E-mail: michael.konings@philips.com |
| Device: | |
| Trade Name: | SmartCT |
| Classification Name: | Interventional fluoroscopic x-ray system |
| Classification Regulation: | 21CFR §892.1650 |
| Classification Panel: | Radiology |
| Device Class: | Class II |
| Product Code: | Primary Code: OWB<br>Subsequent Codes: JAK, LLZ |
| Primary Predicate<br>Device: | |
| Trade Name: | XperCT |
| Manufacturer: | Philips Medical Systems Nederland B.V. |
| 510(k) Clearance: | K130893 (September 6, 2013) |
| Classification Name: | Interventional fluoroscopic x-ray system |
| Classification Regulation: | 21CFR §892.1650 |
| Classification Panel: | Radiology |
| Device Class: | Class II |
| Product Code: | Primary Code: OWB<br>Subsequent Code: JAK |
{6}------------------------------------------------
| Additional Predicate<br>Devices: | Trade Name: | Allura 3D-RA |
|----------------------------------|----------------------------|-------------------------------------------------|
| | Manufacturer: | Philips Medical Systems Nederland B.V |
| | 510(k) Clearance: | K121772 (March 21, 2013) |
| | Classification Name: | Interventional fluoroscopic x-ray system |
| | Classification Regulation: | 21CFR §892.1650 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | | Product Code: |
| | Trade Name: | 3D Roadmap |
| | Manufacturer: | Philips Medical Systems Nederland B.V |
| | 510(k) Clearance: | K121772 (March 21, 2013) |
| | Classification Name: | Interventional fluoroscopic x-ray system |
| | Classification Regulation: | 21CFR §892.1650 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | Primary Code: OWB<br>Subsequent Codes: JAK, LLZ |
{7}------------------------------------------------
#### Device description: SmartCT is a 3D image visualization and analysis software product (Interventional Tool) intended to provide fast and high-resolution 3D visualization of vasculature, hemorrhages, soft tissue and bone structures, thereby helping the physician to identify pathologies and supporting the physician to define and plan the intervention strategy.
SmartCT runs on a software platform called the Interventional Workspot, and is intended to be used with a Philips Interventional X-Ray System.
SmartCT supports 3D Rotational Angiography (3DRA), Cone Beam CT (CBCT) and VasoCT acquisition protocols, and it includes 3D roadmap functionality. The CBCT and VasoCT protocols are only available for the 20" detector of the Interventional X-Ray system. The 3DRA protocols are available for all detectors.
The 3DRA protocols are available with SmartCT Angio, the CBCT protocols with SmartCT Soft Tissue and the VasoCT protocols with SmartCT Vaso. The 3D roadmap functionality (SmartCT Roadmap) comes in combination with SmartCT Angio, SmartCT Soft Tissue or SmartCT Vaso.
SmartCT includes filters to improve the image quality of the reconstruction by reducing the noise caused by metal objects or other objects that absorb high levels of X-ray radiation.
SmartCT provides overlays of live 2D fluoroscopic images with a 3D reconstruction of the vessel tree.
SmartCT offers tools to manually measure sizes and volumes of anatomical structures such as lesions, aneurysms or vessels. It also offers a vessel analysis tool that provides semi-automatic measurements of the diameter of a segmented vessel.
SmartCT can be controlled from both the control room and the examination room.
SmartCT provides workflow guidance to support the physician in the workflow of acquiring and processing 3D images.
{8}------------------------------------------------
| Indications for Use: | SmartCT assists physicians during vascular and non-vascular procedures, with<br>diagnosis, treatment planning, interventional procedures and treatment follow-up<br>by creating 3D views from sets of 2D images created during rotational<br>acquisitions. |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SmartCT provides high-speed and high-resolution 3D visualizations of<br>vasculature, hemorrhages, soft tissue and bone structures. |
| | SmartCT provides live image guidance for navigating endovascular devices<br>through vascular structures anywhere in the body. |
| | SmartCT helps to assess anatomical information intra-procedurally, such as the<br>estimate of a vessel or lesion size, diameter or volume, and anatomical distances<br>between relevant structures. |
| | SmartCT is intended to be used for human patients that have been elected for the<br>procedures as described in these Indications for Use. |
| | The indications for use of SmartCT are similar to the indications for use of the<br>currently marketed predicate devices XperCT, Allura 3D-RA and 3D Roadmap.<br>Based on the information provided above, SmartCT is considered substantially<br>equivalent to the currently marketed and predicate devices XperCT, Allura 3D-RA<br>and 3D Roadmap in terms of Indications for Use. |
| Technological<br>characteristics: | SmartCT has similar technological characteristics compared to the predicate<br>devices. Similar features are used in the predicate and subject devices, with<br>exception of the following modifications implemented in the SmartCT: 3D acquisition guidance including injector protocols suggestions and acquisition timeline showing the different phases of the 3D acquisition performed via Philips Interventional X-Ray system. Visualization improvements such as skull removal feature, initial viewing settings and 3D volume rendering options. User Interface improvements to enhance the interaction of the user with the Multi-Modality Touch Screen Module at table side and provide workflow guidance, as well as overall usability improvements, including 3D measurements and annotations. Architectural improvements including enhanced interface between SmartCT and Philips Interventional X-Ray system in order to get additional information on specific settings and acquisition protocols required for the enabling the acquisition guidance functionality. Curved and straightened reformat for vessel views and new implementing algorithm on Vessel Analysis tool. |
| | The differences between SmartCT and the predicate devices do not raise any new<br>questions regarding safety or effectiveness. Based on the information provided<br>above, SmartCT is considered substantially equivalent to the currently marketed<br>and predicate devices XperCT, Allura 3D-RA and 3D Roadmap in terms of<br>fundamental scientific technology. |
{9}------------------------------------------------
Summary of Non-Non-clinical performance testing has been performed on SmartCT and Clinical Performance demonstrates compliance with the following FDA recognized consensus Data: standards:
- . IEC 62304 Medical device software – Software life cycle processes (Edition 1.1, 2015-06). FDA/CDRH recognition number 13-79.
- . IEC 62366-1 Medical devices - Part 1: Application of usability engineering to medical devices (Edition 1.0, 2015-02). FDA/CDRH recognition number 5-114.
- . IEC 82304-1 Health software – Part 1: General requirements for product safety (Edition 1.0 2016-10), FDA/CDRH recognition number 13-97.
- . ISO 14971 Medical devices – Application of risk management to medical devices (Edition 2.0, corrected 2007). FDA/CDRH recognition number 5-40.
- ISO 15223-1 Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements (Third Edition, 2016-11-01). FDA/CDRH recognition number 5-117.
- UL 2900-1, Standard For Safety, Standard For Software Cybersecurity . Network-Connectable Products, Part 1: General Requirements, (First Edition, 2017-08), FDA/CDRH recognition number 13-96
- NEMA PS 3.1 - 3.20 Digital Imaging and Communications in Medicine (DICOM) Set (2016-06). FDA/CDRH recognition number 12-300
- IEC 80001-1 Application of risk management for IT-networks . incorporating medical devices - Part 1: Roles, responsibilities and activities (Edition 1.0, 2010-10). FDA/CDRH recognition number 13-38.
Software verification testing of the functional and non-functional requirements as well as performance, reliability and safety has been performed to verify that all the requirements of System Requirements Specification as well as the safety risk control measures from the Detailed Risk Management Matrix and the Privacy and Security requirements and mitigations have been implemented. Results demonstrated that all executed verification tests were passed.
Non-clinical validation testing has been performed to validate that SmartCT conforms to the intended use, claims, user needs, effectiveness of safety measures and instructions for use. The validation consisted of the following activities:
- -Usability validation was performed with representative clinical users (both physicians and nurse/technicians) in a simulated use environment. SmartCT was found to be safe and effective for the intended use, users and use environment
- A simulated use design validation was undertaken with participants who fulfill the intended user profile. The participants executed validation protocols in the form of a clinical workflow to validate user needs. intended use and claims. Results demonstrated that all executed validation protocols were passed.
- A clinical experience (expert opinion) evaluation was performed in order to validate some marketing claims and user needs regarding the clinical
{10}------------------------------------------------
| functionality of SmartCT by assessing clinical images. Results demonstrated that these commercial claims and user needs were successfully validated. |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| All these tests were used to support substantial equivalence of the subject device and demonstrate that <b>SmartCT</b> : |
| complies with the afore mentioned international and FDA-recognized consensus standards, and meets the acceptance criteria and is adequate for its intended use. |
Therefore, SmartCT is substantially equivalent to the predicate devices XperCT, Allura 3D-RA and 3D Roadmap in terms of safety and effectiveness.
Summary of Clinical The proposed SmartCT did not require clinical study since substantial Performance Data: equivalence to the currently marketed and predicate devices XperCT, Allura 3D-RA and 3D Roadmap was demonstrated with the following attributes:
- Indication for use;
- . Technological characteristics;
- . Non-clinical performance testing; and
- Safety and effectiveness. .
Substantial The SmartCT is substantially equivalent to the currently marketed predicate Equivalence devices XperCT (K130893), Allura 3D-RA and 3D Roadmap (K121772) in terms Conclusion: of indications for use, technological characteristics and safety and effectiveness. Additionally, substantial equivalence was demonstrated by non-clinical performance tests provided in this 510(k) pre-market notification. These tests demonstrate that SmartCT complies with the requirements specified in the international and FDA-recognized consensus standards and is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.