K152696 · Toshibamedical Systems Corporation · OWB · Jan 12, 2016 · Radiology
Device Facts
Record ID
K152696
Device Name
INFX-8000V, V6.20
Applicant
Toshibamedical Systems Corporation
Product Code
OWB · Radiology
Decision Date
Jan 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Device Story
INFX-8000V V6.20 is an X-ray system for interventional/diagnostic angiography; comprises C-arm, X-ray tube, beam limiter, receptor, controller, computers, and patient table. System generates X-rays; flat panel detector converts transmitted beams to electrical signals; digital subsystem processes/records images for monitor display. Operated by clinicians in interventional settings. Modifications include: improved Auto Pixel Shift (APS) algorithm for image correction; Spot ROI Fluoroscopy for peripheral visualization outside regions of interest; Clinical Mode providing preset 2D roadmap display settings to streamline workflow. System does not transmit data wirelessly. Output assists clinicians in visualizing vasculature during procedures, potentially improving diagnostic accuracy and procedural efficiency.
Clinical Evidence
Bench testing only. Performance testing conducted per IEC standards for medical devices and X-ray systems. Software documentation provided for Moderate Level of Concern. Testing confirmed improved APS algorithm accuracy is equal to or greater than the predicate.
Technological Characteristics
Digital radiography/fluoroscopy system; C-arm configuration. Materials/components conform to IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-43, 62304, 62366, and 21 CFR 1020. Sensing via flat panel detector (scintillation effect). Network connectivity for image transfer. Software-controlled X-ray conditions and image processing.
Indications for Use
Indicated for diagnostic and angiographic procedures of blood vessels in the heart, brain, abdomen, and lower extremities in patients requiring interventional or diagnostic imaging.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines beneath them, possibly representing water or movement. The text is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 12, 2016
Toshiba America Medical Systems, Inc. % Ms. Janine Reyes Manager, Regulatory Affairs 2441 Michelle Drive TUSTIN CA 92780
Re: K152696
Trade/Device Name: INFX-8000V. V6.20 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA and IZI Dated: December 16, 2015 Received: December 18, 2015
Dear Ms. Reyes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Oolo
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Reset Form
Indications for Use
510(k) Number (if known)
K152696
Device Name
INFX-8000V, V6.20
#### Indications for Use (Describe)
This device is a digital radiography/fluoroscopy system used in a diagnostic interventional angiography configuration. The system is indicated for use in diagnostic and anqioqraphic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Type of Use (Select one or both, as applicable)
P Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
## 510(k) - SUMMARY OF SAFETY AND EFFECTIVENESS
## 1. CLASSIFICATION and DEVICE NAME:
| | Classification Name: Image-Intensified Fluoroscopic X-Ray System | |
|--------------------|--------------------------------------------------------------------|--|
| Regulation Number: | 21 CFR 892.1650 (Class II) | |
| Product Code: | OWB - Interventional Fluoroscopic X-Ray System (primary) | |
| | JAA - System, X-Ray, Fluoroscopic, Image-Intensified (secondary)) | |
| | ZI - System, X-Ray, Angiographic (secondary) | |
| Trade Proprietary | | |
| Name: | Infinix | |
| Model Number: | INFX-8000V, V6.20 | |
### 2. ESTABLISHMENT REGISTRATION: 9614698
## 3. U.S. AGENT and ADDRESS:
#### Official Correspondent/U.S. Agent:
Paul Biggins Director, Regulatory Affairs (714) 730-5000 Fax: (714) 730-1310 paul.biggins@toshiba.com
### Contact Person:
Janine Reyes Manager, Regulatory Affairs (714) 669-7853 Fax: (714) 730-1310 janine.reyes@toshiba.com
### Establishment Name and Address:
Toshiba America Medical Systems, Inc. (TAMS) 2441 Michelle Drive Tustin, CA 92780
### 4. MANUFACTURING SITE:
Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan Contact: Akinori Hatanaka
{4}------------------------------------------------
## 5. DATE PREPARED:
September 18, 2015
## 6. TRADE NAME(S):
INFX-8000V, V6.20 [Infinix CC-i]
## 7. COMMON NAME:
Image-Intensified Fluoroscopic X-Ray System Interventional Fluoroscopic X-Rav System System, X-Ray, Fluoroscopic, Image-Intensified System, X-ray, Angiographic
### 8. REGULATION NAME:
Image-Intensified Fluoroscopic X-Ray System
## 9. DEVICE CLASSIFICATION:
Class II (per 21 CFR 892.1650)
## 10. PRODUCT CODE / DESCRIPTION:
Primary Product Code: OWB - Interventional Fluoroscopic X-Ray System Secondary Product Code: JAA - System, X-Ray, Fluoroscopic, Image-Intensified Secondary Product Code: IZI - System, X-ray, Angiographic (secondary)
### 11. PERFORMANCE STANDARD:
This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard].
### 12. PREDICATE DEVICE:
INFX-8000V, with Wireless Footswitch (K143225)
| Product | Marketed by | 510(k) Number | Clearance Date |
|-----------------------------------------|------------------------------------|---------------|-----------------|
| INFX-8000V, with<br>Wireless Footswitch | Toshiba America<br>Medical Systems | K143225 | January 9, 2015 |
### 13. REASON FOR SUBMISSION:
Modification of a cleared device
### 14. SUBMISSION TYPE:
Traditional 510(k)
### 15. DEVICE DESCRIPTION:
INFX-8000V, V6.20, is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm, which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.
{5}------------------------------------------------
The device software is used to control the system, set X-ray conditions, acquire digital images from the X-ray detector, display images on the monitors, perform image processing and recording. This X-ray system has a wireless footswitch option. This X-ray system does not have wireless transmission of data.
## 16. INDICATIONS FOR USES:
This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
### 17. SUMMARY OF CHANGE(S)
This submission is to report improvements to the Auto Pixel Shift (APS) algorithm, the addition of Spot ROI Fluoroscopy and Clinical Mode.
- i. Auto Pixel Shift (APS) - the correction accuracy of the improved APS algorithm was confirmed to be equal to or greater than the correction accuracy of the current APS algorithm.
- ii. Spot ROI (region of interest) Fluoroscopy - allows for peripheral region visualization outside of the area of interest.
- iii. Clinical Mode - provides preset 2D roadmap display settings to enhance user workflow during common clinical cases.
# 18. SUBSTANTIAL EQUIVALENCE:
This device is substantially equivalent to the INFX-8000V, with Wireless Footswitch, (K143225), marketed by Toshiba America Medical Systems. INFX-8000V, V6.20, includes improvements to the Auto Pixel Shift (APS) algorithm, the addition of Spot ROI Fluoroscopy and Clinical Mode. The basic system configuration, method of operation, base software and manufacturing process remain unchanged from the cleared device. There are no new indications for use or intended use of the device.
| Item | Predicate Device:<br>INFX-8000V with Wireless Footswitch<br>(K143225) | Subject Device:<br>INFX-8000V, V6.20 | Notes |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|
| Intended<br>Use | This device is a digital<br>radiography/fluoroscopy system used in a<br>diagnostic and interventional angiography<br>configuration. The system is indicated for use<br>in diagnostic and angiographic procedures for<br>blood vessels in the heart, brain, abdomen<br>and lower extremities. | This device is a digital<br>radiography/fluoroscopy system used in a<br>diagnostic and interventional angiography<br>configuration. The system is indicated for use<br>in diagnostic and angiographic procedures for<br>blood vessels in the heart, brain, abdomen<br>and lower extremities. | Same |
| Operation<br>Principles | The high voltages output by the X-ray<br>generator are supplied to the X-ray tube, and<br>the X-ray beams are generated. The X-ray<br>beams that have passed through the patient<br>are converted to electrical signals by the flat<br>panel detector based on the X-ray scintillation<br>effect. The signals are then sent to the digital<br>imaging subsystem as digital image signals,<br>and digital images are processed/recorded. In<br>addition, the images are displayed on the<br>monitor and are transferred to the network. | The high voltages output by the X-ray<br>generator are supplied to the X-ray tube, and<br>the X-ray beams are generated. The X-ray<br>beams that have passed through the patient<br>are converted to electrical signals by the flat<br>panel detector based on the X-ray scintillation<br>effect. The signals are then sent to the digital<br>imaging subsystem as digital image signals,<br>and digital images are processed/recorded. In<br>addition, the images are displayed on the<br>monitor and are transferred to the network. | Same |
{6}------------------------------------------------
| Item | Predicate Device:<br>INFX-8000V with Wireless Footswitch<br>(K143225) | Subject Device:<br>INFX-8000V, V6.20 | Notes |
|--------------------------|-----------------------------------------------------------------------|-------------------------------------------|--------|
| Auto Pixel<br>Shift | Available | Algorithm Improvement | Change |
| Spot ROI<br>Fluoroscopy | Not Available | Available | Change |
| Clinical<br>Mode | Not Available | Available | Change |
| Roadmap Function | 2D Roadmap<br>3D Roadmap<br>Multimodality Roadmap | 2D Roadmap<br>3D Roadmap<br>Not Available | Change |
## 19. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 standards, its collateral standards and particular standards: IEC 60601-1:2005, IEC 60601-1-2:2007, IEC 60601-1-3:2008, IEC 60601-1-6:2010, IEC 60601-2-28:2010, IEC 60601-2-43:2010, IEC 62304:2006, IEC 62366:2007. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR §1020, that apply to this device, will be met and reported via product report.
### 20. TESTING:
This submission contains test data that demonstrates that the system modifications result in performance that is equal to or better than the predicate system. Testing of the modified system was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and XR Systems. Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission. Additionally, the design controls used for this device included risk management and all known risks were mitigated to an acceptable level.
### 21. CONCLUSION
The subject device is substantially equivalent to the INFX-8000V, with Wireless Footswitch, which was cleared via Pre-Market Notification 510(k), K143225. The INFX-8000V, V6.20, incorporates modifications to the cleared device which include improvements to the Auto Pixel Shift (APS) algorithm, the addition of Spot ROI Fluoroscopy and Clinical Mode. The changes to this device do not alter the Indications for Use or the intended uses associated with the previously cleared device, as described in the labeling. It is the contention of Toshiba that all new safety issues have been addressed in the design of this change and that adequate evidence of this is presented with this submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.