The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are not intended for mammography applications. Innova IVUS option: The Innova IVUS software option simplifies the clinical workflow associated with the use of Volcano IVUS products by: (1) automatically synchronizing the patient demographic information (patient name, date of birth, DICOM attributes etc.) from Innova system with an IVUS imaging system, (2) providing a remote access to commonly used IVUS functions from the Innova table side user interface. (3) displaying the IVUS images on the multi-monitor display of the Innova cathlab system.
Device Story
Innova Digital Fluoroscopic Imaging System performs fluoroscopic X-ray examinations using amorphous silicon detectors with cesium iodine scintillators; monoplane or biplane positioner; vascular/cardiac table. Innova IVUS option provides connectivity with third-party intravascular ultrasound (IVUS) devices. System operates in cathlab environment; used by physicians/technicians. IVUS option synchronizes patient demographics (name, DOB, DICOM attributes) between Innova and IVUS systems; provides remote access to IVUS functions via table-side interface; displays IVUS images on cathlab multi-monitor system. Enhances clinical workflow by integrating third-party imaging data into primary display and control environment. Benefits include streamlined data management and centralized visualization during interventional procedures.
Clinical Evidence
No new clinical data provided. Substantial equivalence is supported by previously submitted clinical data for the predicate Innova systems, as the IVUS option does not alter the fundamental imaging performance or diagnostic capabilities of the fluoroscopic system.
Technological Characteristics
Amorphous silicon detector with cesium iodine scintillator; monoplane/biplane positioner; vascular/cardiac table. Compliant with UL 187, UL 2601, and IEC 60601-1. Power: 480 VAC 50/60Hz. Connectivity: Fiber Channel link to acquisition system; DICOM network integration. Software-based IVUS integration.
Indications for Use
Indicated for patients undergoing vascular angiography, cardiology, diagnostic, and interventional procedures requiring fluoroscopic imaging. Not intended for mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
KOGII 63
Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are in black, contrasting with the white background. The logo is a well-known symbol associated with the General Electric company.
GE Healthcare
P.O. Box 414, W-440
.
Milwaukee, WI 53201 USA
## 1. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
JUN - 7 2006
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
| Identification of Submitter: | Larry A. Kroger, Ph.D.<br>Senior Regulatory Programs Manager<br>GE Healthcare<br>Tel. (262) 544-3894<br>Summary prepared: April 25, 2006 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Identification of Product: | Digital Fluoroscopic Imaging System |
| Classification Name: | Fluoroscopic X-ray System |
| Manufacturer: | GE Medical Systems SCS.<br>283, rue de la Minière<br>78530 Buc Cedex, France |
| Distributed by: | GE Medical Systems, LLC, Milwaukee, WI |
| Marketed Devices: | The GE Healthcare Innova 2121 <sup>IQ</sup> , Innova 3131 <sup>IQ</sup> , Innova 4100,<br>Innova 4100 <sup>IQ</sup> , Innova 3100, Innova 3100 <sup>IQ</sup> , Innova 2100 <sup>IQ</sup><br>devices with <i>Innova IVUS</i> option are substantially equivalent to<br>the currently marketed GE Healthcare Innova 2121 <sup>IQ</sup> , Innova<br>3131 <sup>IQ</sup> , Innova 4100, Innova 4100 <sup>IQ</sup> , Innova 3100, Innova<br>3100 <sup>IQ</sup> , Innova 2100 <sup>IQ</sup> devices (K060259, K033244, K052412,<br>K031637, K050489) |
| | This opinion is based on the information contained in the<br>comparison table and the product data sheets. |
| Device Description: | The <i>Innova IVUS</i> is offered as an option for Innova 2121 <sup>IQ</sup> ,<br>Innova 3131 <sup>IQ</sup> (cleared under K060259), Innova 4100 (cleared<br>under K033244), Innova 3100 (cleared under K031637), Innova<br>2100 <sup>IQ</sup> (cleared under K050489), Innova 4100 <sup>IQ</sup> and Innova<br>3100 <sup>IQ</sup> (cleared under K052412). |
| | The Innova Digital Fluoroscopic Imaging Systems are designed<br>to perform fluoroscopic x-ray examinations. The detector is<br>comprised of amorphous silicon with a cesium iodine scintillator. |
Attachment 1 Rev 1
{1}------------------------------------------------
The resulting digital image can be sent through a Fiber Channel link to an acquisition system then to network (in using DICOM) for applications such as post-processing, printing, viewing and archiving. Digital Fluoroscopic Imaging System consists of an a monoplane or biplane positioner, a vascular or cardiac table, an X-RAY system and one or two digital detectors.
The Innova IVUS option provides enhanced connectivity with third party intravascular ultrasound devices.
Materials: All construction and materials are compliant with UL 187 and IEC 60601-1 for the existing parts of the product and with UL 2601 and IEC 60601-1 for the new parts.
Design: The design is validated through Failures Modes Effects Analysis (FMEA) process, which allows managing the risks.
480 VAC 50/60Hz. Energy Source:
Indications for Use: For Innova 2121 10, Innova 3131 10, Innova 4100, Innova 4100'9 Innova 3100, Innova 310010, Innova 210010 devices with Innova IVUS option:
> The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for fluoroscopic images of human anatomy generating for cardiology, diagnostic, and interventional procedures.
> They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are not intended for mammography applications.
Innova IVUS option:
The Innova IVUS software option simplifies the clinical workflow associated with the use of Volcano IVUS products by:
(1) automatically synchronizing the patient demographic information (patient name, date of birth, DICOM attributes etc.) from Innova system with an IVUS imaging system,
(2) providing a remote access to commonly used IVUS functions from the Innova table side user interface.
(3) displaying the IVUS Images on the multi-monitor display of the Innova cathiab system.
{2}------------------------------------------------
Comparison with
The GE Healthcare Innova 2121 19, Innova 3131 19, Innova 4100, Innova 410019, Innova 3100, Innova 310019, Innova 210019 devices with Innova IVUS option are substantially equivalent to the currently marketed GE Healthcare Innova 2121 10, Innova കാശസ്തു അക്കാരം - Ennova 4100", Innova 3100, Innova 3100, Innova
3131 ", Innova 2100" devices (K060259, K033244, K052412, K031637, K050489)
This opinion is based on the information contained in the comparison table and the product data sheets.
## Summary of the Studies:
Innova 31310, Innova 212119, Innova 4100, Innova 410019 Innova 3100. Innova 31000 and Innova 21000 with Innova IVUS option are considered substantially equivalent to the predicates in terms of image quality and diagnostic capabilities. Therefore, previously submitted clinical data are applicable for this submission.
GE Healthcare considers that Innova IVUS option for Digital Conclusions: Fluoroscopic Imaging Systems Innova 31319, Innova 21210 Innova 4100, Innova 410019, Innova 3100, Innova 310019 and Innova 21000 to be equivalent with the predicate devices. The potential hazards, related to the introduction of Innova IVUS options are controlled by a risk management plan including:
- A hazard identification (Attachment 8)
- A risk evaluation (Attachment 8)
- Software Development and Validation Process A (Attachment 7)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo consists of a blue emblem on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The emblem is a stylized representation of a human figure, and the text is in a clear, sans-serif font.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Medical Systems. LLC P.O. Box 414. W-400 MILWAUKEE WI 53201
JUL 30 2012
Re: K061163
Trade/Device Name: Digital Fluoroscopic Imaging Systems-Innova 4100. Innova 41000. Innova 3100, Innova 310010, Innova 210010, Innova 313110, Innova 21210 with Innova IVUS Option
Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB. JAA and IYO Dated: April 25, 2006 Received: April 26. 2006
Dear Dr. Kroger:
This letter corrects our substantially equivalent letter of June 7, 2006.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Janine M. Mdrri Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 2. Indications for Use
510(k) Number (if known):
Device Name:
K061163
Digital Fluoroscopic Imaging Systems - Innova 4100, Innova 4100 °, innova 3100, innova 3100 ° , Innova 2100 ° Innova 3131"0" , Innova 21210 with Innova IVUS option.
Indications for Use:
The Innova systems are Indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures.
They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are not intended for mammography applications.
Innova IVUS option:
The Innova IVUS software option simplifies the clinical workflow associated with the use of Volcano IVUS products by:
(1) automatically synchronizing the patient demographic information (patient name, date of birth, DICOM attributes etc.) from Innova system with an IVUS imaging system, (2) providing a remote access to commonly used IVUS functions from the Innova table side user interface.
(3) displaying the IVUS images on the multi-monitor display of the Innova cathlab system.
Prescription Use
× (Part 21 CFR 801 Subpart D)
Over-The-Counter-Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IS NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Larson
Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K061163
Posted 13 Nov 2001
.
Page 이 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.