BENCOX Delta Option Heads
K223828 · Corentec Co., Ltd. · LZO · Mar 16, 2023 · Orthopedic
Device Facts
| Record ID | K223828 |
| Device Name | BENCOX Delta Option Heads |
| Applicant | Corentec Co., Ltd. |
| Product Code | LZO · Orthopedic |
| Decision Date | Mar 16, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3353 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
BENCOX Delta Option Heads of BENCOX Total Hip System is intended for Cementless use in total or partial hip arthroplasty in primary or revision surgery for the following conditions: a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis b. Inflammatory degenerative joint disease, such as rheumatoid arthritis c. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques d. Patients with failed previous surgery where pain, deformity, or dysfunction persists e. Revision of previously failed total hip arthroplasty
Device Story
BENCOX Delta Option Heads are prosthetic components for total or partial hip arthroplasty. Device consists of BIOLOX delta ceramic ball head and titanium alloy (Ti6Al4V) sleeve. Sleeve fits onto in-situ metallic hip stem taper; ceramic head fits onto sleeve. Used in orthopedic surgery to restore joint function in patients with degenerative joint disease or fractures. Surgeon-operated in clinical/OR setting. Provides mechanical articulation for hip joint. Benefits include pain relief and restored mobility. No software or electronic components.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing: burst testing, fatigue testing, post-fatigue burst testing, pull-off testing, torque-out testing, range of motion, and corrosion testing. Pyrogen/endotoxin testing performed per USP<161> and ANSI/AAMI ST72.
Technological Characteristics
Materials: BIOLOX delta ceramic (alumina composite per ISO 6474-2 Type X) and Ti6Al4V titanium alloy sleeve. Mechanical assembly: ceramic head on titanium sleeve on metallic hip stem taper. Non-electronic, mechanical prosthesis. Sterilization method not explicitly stated but compliant with sterility guidance.
Indications for Use
Indicated for patients requiring total or partial hip arthroplasty (primary or revision) due to non-inflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis, traumatic arthritis), inflammatory degenerative joint disease (e.g., rheumatoid arthritis), non-union, femoral neck/trochanteric fractures of proximal femur, or failed previous surgery.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
- BIOLOX Delta Ceramic Femoral Heads, BIOLOX Option Ceramic Femoral Head System (K192416)
Reference Devices
- BENCOX Forte & BENCOX Delta (K121665)
- BIOLOX delta Option and Extra-long Heads (K173776)
- Exactech BIOLOX®Delta Femoral Heads and BIOLOX® Option Femoral Heads and Adapters (K103012)
Related Devices
- K040225 — OTI ALUMINA CERAMIC FEMORAL HEAD SYSTEM · Osteoimplant Technology, Inc. · May 6, 2004
- K191399 — BIOLOX CONTOURA Ceramic Femoral Head · Total Joint Othopedics, Inc. · Aug 28, 2019
- K050966 — PIVOT BIPOLAR FEMORAL HEAD · Ortho Development Corp. · Jul 7, 2005
- K121665 — BENCOX FORET & BENCOX DELTA · Corentec Co., Ltd. · Sep 25, 2012
- K051588 — BIOLOX DELTA CERAMIC FEMORAL HEADS · Howmedica Osteonics Corp. · Jul 12, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 16, 2023
Corentec Co., Ltd. Yoorim Bae RA Specialist 12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu Cheonan-si, Chungcheongnam-do 31056 Republic of Korea
Re: K223828
Trade/Device Name: BENCOX Delta Option Heads Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, KWY Dated: December 12, 2022 Received: December 21, 2022
Dear Yoorim Bae:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Limin Sun -S
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223828
Device Name BENCOX Delta Option Heads
#### Indications for Use (Describe)
BENCOX Delta Option Heads of BENCOX Total Hip System is intended for Cementless use in total or partial hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques
d. Patients with failed previous surgery where pain, deformity, or dysfunction persists
e. Revision of previously failed total hip arthroplasty
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
(As required by 21 CFR 807.92)
Date: March 15, 2023
## ADMINISTRATIVE INFORMATION
| Manufacturer: | Corentec Co., Ltd.<br>12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu<br>Cheonan-si, Chungchongnam-do, Rep. of Korea -31056<br>Telephone: +82-41-585-7114; Fax: +82-41-585-7113 |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Yoorim Bae<br>RA Specialist<br>12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu<br>Cheonan-si, Chungchongnam-do, Rep. of Korea -31056<br>Ph: +82-41-410-7116 (Direct) ; Fax: +82-41-585-7113 |
Email: yoorim.bae@corentec.com
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | BENCOX Delta Option Heads |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Common Name: | Femoral Head Prosthesis |
| Regulation and Classification Name: | 21 CFR 888.3353<br>Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis |
| Regulatory Class: | Class II |
| Product Codes: | LZO, KWY |
| Classification Panel: | Orthopedic Products Panel |
| Reviewing Branch: | Orthopedic Devices Branch |
## Legally Marketed Predicate Devices to Which Substantial Equivalence is Claimed:
| Primary Devices Supporting Substantial Equivalence: | | |
|-------------------------------------------------------|--------------------------------------------------------------------------------------|-------------------------------|
| 510(k) Number | Trade or Proprietary Model Name | Manufacturer |
| K192416 | BIOLOX Delta Ceramic Femoral Heads,<br>BIOLOX Option Ceramic Femoral Head System | Zimmer GmbH |
| Reference Devices Supporting Substantial Equivalence: | | |
| 510(k) Number | Trade or Proprietary Model Name | Manufacturer |
| K121665 | BENCOX Forte & BENCOX Delta | Corentec Co., Ltd. |
| K173776 | BIOLOX delta Option and Extra-long Heads | MicroPort<br>Orthopedics Inc. |
| K103012 | Exactech BIOLOX®Delta Femoral Heads and<br>BIOLOX® Option Femoral Heads and Adapters | Exactech, Inc. |
{4}------------------------------------------------
## Indications For Use
BENCOX Delta Option Heads of BENCOX Total Hip System is intended for Cementless use in total or partial hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques
- d. Patients with failed previous surgery where pain, deformity, or dysfunction persists
- e. Revision of previously failed total hip arthroplasty
## Purpose and Device Description
The BENCOX Delta Option Head consists of a delta ball head and a titanium sleeve (Ti6Al4V). The ball heads are made of the BIOLOX® delta ceramic material, which is a high-purity alumina composite material according to ISO 6474-2 Type X, and come in various outer diameters.
The ceramic BIOLOX® delta ball head is assembled with the corresponding titanium sleeve and is then placed over the titanium alloy tapers of an in-situ hip stem prosthesis. The titanium sleeve has an inner taper which fits the dimensions of a metallic hip stem prosthesis, and the BIOLOX® delta ball head has a taper which fits to the dimensions of the outer diameter of the titanium sleeve.
The design features and materials of the subject device, BENCOX Delta Option head are substantially equivalent to BIOLOX® delta Option devices cleared under K192683 and K173776 since the subject device is supplied by CeramTec.
## Summary of Technological Characteristics
Device Comparisons and performance testing indicate that the BENCOX Delta Option Heads are substantially equivalent to the predicates in terms of intended use, indications, design, materials, performance characteristics and operational principles.
## Non-Clinical Studies
The following tests were performed on the BENCOX Delta Option Heads to demonstrate substantial equivalence of safety and efficacy with the predicate devices:
- . Burst Testing
- . Fatigue Testing
- · Post-fatigue Burst Testing
- . Pull-off Testing
- . Torque-out Testing
- . Range of Motion
- . Corrosion Testing and Assessment
{5}------------------------------------------------
Pyrogen testing was conducted in accordance with USP<161>, and ANSI/AAMI ST72 to ensure the proposed BENCOX Delta Option Heads meet recommended limits per FDA's Guidance Document submission and Review of Sterility Information in Premarket (510(k)) Submission for Devices Labeled as Sterile.
Bacterial endotoxin testing (BET) as specified in ANSVAAMI ST72:2011 was used for pyrogenicity testing to achieve an Endotoxin limit of < 20EU/Device.
Clinical Testing: Clinical testing was not required
## Substantial Equivalence Conclusion
Based upon a comparison of intended use, materials, summary of technological characteristics, and preclinical testing, the BENCOX Delta Option Heads are substantially equivalent to the predicate devices identified in this premarket notification.