Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments for use with the Direct Superior Approach
Applicant
Stryker Orthopaedics
Product Code
LPH · Orthopedic
Decision Date
Feb 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stryker Orthopaedic Acetabular Shells for use with the Total Hip Instruments are intended for use in total hip arthroplasty and are intended for either primary or revision total hip arthroplasty. The Stryker Orthopaedic Hip Implant Stems for use with the Total Hip Instruments are intended for use in total or hemi-arthroplasty of the hip and are intended for either primary or revision procedures. The Direct Superior Approach does not change the intended use of the subject devices.
Device Story
System comprises acetabular shells, femoral stems, and manual surgical instruments (retractors, cup impactors, drivers, guides, reamers). Used in total or hemi-hip arthroplasty via Direct Superior Approach. Implants provide cementless fixation. Instruments are hand-held, reusable, non-sterile. Surgeon uses instruments to prepare acetabulum and femur for implant placement. Device facilitates hip reconstruction; restores joint function; addresses pain and mobility issues in patients with degenerative or traumatic hip conditions. No change to fundamental scientific technology or materials compared to predicate devices.
Clinical Evidence
No clinical data required; bench testing not required as submission is limited to a labeling modification for a surgical protocol.
Technological Characteristics
Materials: Titanium (Ti-6Al-4V) alloy, commercially pure (CP) titanium plasma spray, CP titanium, and PureFix® hydroxylapatite (HA). Design: Cementless acetabular shells and femoral stems. Instruments: Hand-held, reusable, non-sterile surgical tools. No software or electronic components.
Indications for Use
Indicated for patients with painful, disabling hip joint disease (degenerative, rheumatoid, posttraumatic arthritis, or late-stage avascular necrosis); revision of failed femoral head replacement or cup arthroplasty; clinical management where arthrodesis/reconstructive techniques are less likely to succeed; or poor/inadequate bone stock. Stems are also indicated for noninflammatory degenerative joint disease, correction of functional deformity, nonunions, femoral neck/trochanteric fractures. Constrained liners are indicated for patients at high risk of dislocation due to prior history, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 17, 2015
Stryker Orthopaedics Ms. Denise Daugert Senior Regulatory Affairs Specialist 325 Corporate Drive Mahwah. New Jersev 07430
Re: K143085 Trade/Device Name: Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments for Use with the Direct Superior Approach Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, JDI, KWL, KWY, KWZ, LWJ, LZO, MAY, MBL, MEH Dated: November 11, 2014 Received: November 12, 2014
Dear Ms. Daugert:
This letter corrects our substantially equivalent letter of February 9, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
#### Page 2 - Ms. Denise Daugert
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known): K143085
Device Name: Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments for use with the Direct Superior Approach.
Trident PSL Shell, Trident Tritanium Shell, Trident Hemispherical Shell
- o Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
- . Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- o Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
The Trident shells are intended for cementless fixation within the prepared acetabulum.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
510(k) Number (if known): K143085
Device Name: Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments for use with the Direct Superior Approach.
Tritanium Acetabular Shell System
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, ● rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques ● are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
The Tritanium Acetabular Shell System is intended for cementless use only.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{4}------------------------------------------------
510(k) Number (if known): K143085
Device Name: Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments for use with the Direct Superior Approach.
Accolade II Femoral Stem
The indications for use for total hip arthroplasty with stems include:
- . noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and, ●
- o nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of the Accolade II Femoral Stem with compatible Howmedica Osteonics Constrained Liners:
- When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, ● the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Accolade II Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{5}------------------------------------------------
510(k) Number (if known): K143085
Device Name: Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments for use with the Direct Superior Approach.
Secur-Fit Advanced Hip Stem
The indications for use for total hip arthroplasty with stems include:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular ● necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and, ●
- nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with ● head involvement that are unmanageable using other techniques.
Additional indication specific to use of Secur-Fit Advanced Hip Stems with compatible Howmedica Osteonics Constrained Liners:
- When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, ● the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Secur-Fit Advanced Hip Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{6}------------------------------------------------
510(k) Number (if known): K143085
Device Name: Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments for use with the Direct Superior Approach.
#### Anato Hip Stem
The indications for use for total hip arthroplasty with stems include:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and, ●
- nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of Anato Hip Stem with compatible Howmedica Osteonics Constrained Liners:
- When the stem is to be used with compatible Howmedica Osteonics Constrained Liners. o the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
The Anato Hip Stem is intended for cementless use only and is intended for total and hemiarthroplasty procedures.
Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
AND/OR
Over-The-Counter Use
Over-The-Counter Use
(21 CFR 807 Subpart C)
Over-The-Counter Use
(21 CFR 807 Subpart C)
{7}------------------------------------------------
## 510(k) Summary
| Sponsor | Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Denise Daugert<br>Sr. Regulatory Affairs Specialist<br>Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>P: 201-831-5413<br>F: 201-831 4413 |
| Alternate Contact: | Patricia Setti-LaPerch<br>Manager, Regulatory Affairs<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>P: 201-831-5938<br>F: 201-831-4938 |
| Date Prepared: | October 23, 2014 |
| Proprietary Name: | Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip<br>Instruments for use with the Direct Superior Approach |
| Common Name: | Total Hip Joint Replacement |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis (888.3358)<br>Hip joint Metal/Ceramic/Polymer Semi-Constrained Cemented or<br>Non Porous Uncemented Prosthesis (888.3353)<br>Hip joint metal/polymer semi-constrained cemented prosthesis<br>(888.3350)<br>Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented<br>prosthesis (888.3390)<br>Hip joint metal/polymer constrained cemented or uncemented<br>prosthesis (888.3310)<br>Hip joint femoral (hemi-hip) metallic cemented or uncemented<br>prosthesis (888.3360) |
| Product Codes: | 87LPH, 87 MEH, 87LZO, 87MBL, 87MAY<br>87LWJ, 87KWZ, 87KWY, 87KWI, 87JD1 |
{8}------------------------------------------------
#### Legally Marketed Device to Which Substantial Equivalence is claimed:
Trident PSL Shell (K983382) Trident Tritanium Shell (K010170) Trident Hemispherical Shells (K013676) Tritanium Acetabular Shell System (K081171) Accolade II Femoral Stem (K120578) Secur-Fit Advanced Stem (K122853) Anato Hip Stem (K123604)
#### Device Description:
The devices covered by this submission include Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments. The Instruments consist of Retractors, Cup Impactors, U-Joint Bolt Driver, Alignment Guide, Angled Reamer Handle, Instrument Trays and a Case. All instruments are hand-held surgical instruments used during orthopaedic surgery and are nonsterile, reusable devices. All Class II devices in this submission have been previously determined substantially equivalent in prior 510(k) submissions and are commercially available. The acetabular shells and femoral stems are manufactured from the following materials: titanium (Ti-6Al-4V) allov, commercially pure (CP) titanium plasma spray, commercially pure titanium and PureFix® hydroxylapatite (HA).
#### Intended Use:
The Stryker Orthopaedic Acetabular Shells for use with the Total Hip Instruments are intended for use in total hip arthroplasty and are intended for either primary or revision total hip arthroplasty. The Stryker Orthopaedic Hip Implant Stems for use with the Total Hip Instruments are intended for use in total or hemi-arthroplasty of the hip and are intended for either primary or revision procedures. The Direct Superior Approach does not change the intended use of the subject devices.
{9}------------------------------------------------
Indications for Use (Trident PSL Shell, Tritanium Shell, Trident Hemispherical Shells):
- . Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other ● procedure.
- . Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques . as indicated by deficiencies of the acetabulum.
The Trident shells are intended for cementless fixation within the prepared acetabulum.
### Indications for Use (Tritanium Acetabular Shell System):
- o Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other . procedure.
- o Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques o as indicated by deficiencies of the acetabulum.
The Tritanium Acetabular Shell System is intended for cementless use only.
#### Indications for Use (Accolade II Femoral Stem):
The indications for use for total hip arthroplasty with stems include:
- Noninflammatory degenerative joint disease, including osteoarthritis and avascular . necrosis:
- Rheumatoid arthritis; 0
- Correction of functional deformity; ●
{10}------------------------------------------------
- . Revision procedures where other treatments or devices have failed; and,
- Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with o head involvement that are unmanageable using other techniques.
Additional indication specific to use of the Accolade II Femoral Stems with compatible Howmedica Osteonics Constrained Liners:
- When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, . the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Accolade II Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures.
#### Indications for Use (Secur-Fit Advanced Hip Stem):
The indications for use for total hip arthroplasty with stems include:
- Noninflammatory degenerative joint disease, including osteoarthritis and avascular . necrosis:
- Rheumatoid arthritis; ●
- Correction of functional deformity; ●
- Revision procedures where other treatments or devices have failed; and,
- Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with ● head involvement that are unmanageable using other techniques.
Additional indication specific to use of the Secur-Fit Advanced Hip Stems with compatible Howmedica Osteonics Constrained Liners:
- When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, . the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
{11}------------------------------------------------
Secur-Fit Advanced Hip Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures.
#### Indications for Use (Anato Hip Stem):
The indications for use for total hip arthroplasty with stems include:
- Noninflammatory degenerative joint disease, including osteoarthritis and avascular . necrosis:
- Rheumatoid arthritis; ●
- Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and, ●
- Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with . head involvement that are unmanageable using other techniques.
Additional indication specific to use of the Anato Hip Stem with compatible Howmedica Osteonics Constrained Liners:
- . . When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
The Anato Hip Stem is intended for cementless use only and is intended for total and hemiarthroplasty procedures.
#### Summary of Technological Characteristics:
Device comparisons show that the Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments for use with the Direct Superior Approach is substantially equivalent to its predicates in terms of intended use, indications, design, materials, performance characteristics and operational principles. The intended use and indications for implants used with the Direct Superior Approach are identical to the intended use and indications of the cleared Trident PSL Shell (K983382), Trident Porous Titanium Acetabular Shell (K010170), Trident Hemispherical Shells (K013676). Tritanium Acetabular Shell System (K081171). Accolade II Femoral Stem
{12}------------------------------------------------
(K120578), Secur-Fit Advanced Hip Stem (K122853) and Anato Hip Stem (K123604), used with the Total Hip Instruments. The materials, design features and functionality for Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments are substantially equivalent to the previously, cleared predicate devices. The fundamental scientific technology of the modified device has not changed relative to the predicate devices.
The material and design of the Total Hip Surgery Instruments for use with the Direct Superior approach are identical to those of the predicate devices cleared for use with the Posterolateral Approach and Direct Anterior Approach. The operational principles of the Total Hip Surgery Instruments are similar to those of the predicate devices.
Non-Clinical Testing: Performance testing was not required in support of this submission because this submission covers a labeling modification to introduce an additional surgical protocol. Direct Superior Approach.
Clinical Testing: Clinical testing was not required as a basis for substantial equivalence.
Conclusion: The Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments for use with the Direct Superior Approach are substantially equivalent to the predicate devices identified in this premarket notification.