The Progressive Orthopaedic Total Hip System

K151424 · The Progressive Orthopaedic Company, LLC · LPH · Jul 27, 2016 · Orthopedic

Device Facts

Record IDK151424
Device NameThe Progressive Orthopaedic Total Hip System
ApplicantThe Progressive Orthopaedic Company, LLC
Product CodeLPH · Orthopedic
Decision DateJul 27, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Progressive Orthopaedic Total Hip System implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and acetabular portions of the hip that are severely disabled and/or very painful as a result of: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis - Rheumatoid arthritis or traumatic arthritis - Correction of functional deformity - Non-union femoral neck fracture - Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques. The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.

Device Story

The Progressive Orthopaedic Total Hip System consists of orthopedic implant components, specifically additional screw sizes and an apical hole plug, designed for use with the previously cleared Progressive Orthopaedic Total Hip System (K143314). These components are used in cementless hip arthroplasty to reconstruct femoral and acetabular articulating surfaces. The device is intended for patients suffering from severe hip pain or disability due to degenerative joint disease, arthritis, or fractures. Surgeons perform the implantation procedure in an operating room setting. The components provide structural support and fixation for the hip joint, aiming to restore function and alleviate pain. The system is an extension of an existing cleared platform, utilizing established materials and design principles to ensure compatibility with the primary system.

Clinical Evidence

No clinical testing was required. Substantial equivalence is supported by bench testing, including mechanical characterization, torsion, driving torque, and pull-out testing.

Technological Characteristics

Components are manufactured from Ti6Al4V alloy per ASTM 1472. The system includes screws (6.5mm OD x 15-45mm length) and an apical hole plug (8.52mm OD, 5/16 x 24 UNF 2A thread). The device is a cementless, semi-constrained porous-coated hip prosthesis.

Indications for Use

Indicated for patients requiring cementless hip arthroplasty due to severe disability or pain from non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, traumatic arthritis, functional deformity, non-union femoral neck fracture, or proximal femur trochanteric fractures with head involvement. Suitable for primary or revision total hip arthroplasty.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other to create a sense of depth and unity. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 27, 2016 The Progressive Orthopaedic Company, LLC Mr. Thomas Smith Quality/Regulatory Consultant 801 US Highway 1, Suite B North Palm Beach, Florida 33408 Re: K151424 Trade/Device Name: The Progressive Orthopaedic Total Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: June 24, 2016 Received: June 27, 2016 Dear Mr. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K151424 Device Name The Progressive Orthopaedic Total Hip System #### Indications for Use (Describe) The Progressive Orthopaedic Total Hip System implant components are in cementless reconstruction of the articulating surface of femoral and acetabular portions of the hip that are severely disabled and/or very painful as a result of: - · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis - · Rheumatoid arthritis or traumatic arthritis - · Correction of functional deformity - · Non-union femoral neck fracture - · Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques. The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty. | Type of Use (Select one or both, as applicable) | <table><tr><td><div style="display:inline-block;">☒</div> Production (Part 21 CFR 601 Subpart D)</td></tr><tr><td><div style="display:inline-block;">☐</div> For Test Article (21 CFR 601 Subpart E)</td></tr></table> | <div style="display:inline-block;">☒</div> Production (Part 21 CFR 601 Subpart D) | <div style="display:inline-block;">☐</div> For Test Article (21 CFR 601 Subpart E) | |------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | <div style="display:inline-block;">☒</div> Production (Part 21 CFR 601 Subpart D) | | | | | <div style="display:inline-block;">☐</div> For Test Article (21 CFR 601 Subpart E) | | | | > Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for The Progressive Orthopaedic Company. The logo consists of a blue hexagon with a white caduceus symbol inside. To the right of the hexagon is the company name, with "The" above "PROGRESSIVE" and "Orthopaedic Company" below. # 510(k) Summary | Applicant/Sponsor: | The Progressive Orthopaedic Company, LLC.<br>801 US Highway 1, Suite B<br>North Palm Beach, FL, 33408<br>(561) 440-4460 | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Thomas Smith<br>Quality and Regulatory Consulting<br>801 US Highway 1, Suite B<br>North Palm Beach, FL, 33408<br>(203) 641-3936 | | Proposed Trade Name: | The Progressive Orthopaedic Total Hip System | | Common Name: | Hip Joint Prosthesis | | Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-<br>coated uncemented prosthesis per 21 CFR 888.3358. This<br>falls under the Orthopedics panel/87 as a Class II device. | | Device Product Code: | LPH, LZO | | Predicate Devices: | Progressive Orthopaedic Total Hip System (K143314)<br>NovoSource NovoHip Total Hip System (K140701) | # Device Description: The present 510k The Progressive Orthopaedic additional screw sizes and apical hole plug are components to be used with the Progressive Orthopaedic Total Hip System, previously cleared via K143314. | Screw Sizes | 6.5mm OD x 15-45mm Length | |-------------------------|---------------------------| | Material | Ti6Al4V per ASTM 1472 | | Apical Hole Plug OD | $\u00d8$ 8.52mm OD | | Apical Hole Plug Thread | 5/16 X 24 UNF 2A | | Material | Ti6Al4V | The Progressive Orthopaedic Company 801 US Highway 1, Suite B, North Palm Beach, FL, 33408 Office: (561)440-4460 Fax: (304)777-4790 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for The Progressive Orthopaedic Company. On the left side of the logo is a blue hexagon with a white caduceus symbol inside. To the right of the hexagon is the company name, with "The" above "PROGRESSIVE" and "Orthopaedic Company" below. # Intended Use: Hip joint arthroplasty ### Indications for Use: The Progressive Orthopaedic Total Hip System implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and acetabular portions of the hip that are severely disabled and/or very painful as a result of: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis - Rheumatoid arthritis or traumatic arthritis - Correction of functional deformity - Non-union femoral neck fracture - Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques. The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty. ## Substantial Equivalence The Progressive Orthopaedic Total Hip System and NovoSource Total Hip System have similar intended use, indications for use and the same manufacturing materials as the predicate device. The range of sizes of the Progressive Ortho Total Hip System is similar to the predicate devices. ## Non-Clinical Testing Mechanical testing has demonstrated the device's ability to perform under expected conditions. Testing included mechanical characterization testing, torsion, driving torque and pull out. In all testing, the subject device shows equivalent properties when compared to predicate products. Testing has determined that the device is substantially equivalent to the predicate device in material, construction, and performance characteristics. ## Clinical Performance Data Summary No clinical testing was required. The Progressive Orthopaedic Company 801 US Highway 1, Suite B, North Palm Beach, FL, 33408 Office: (561)440-4460 Fax: (304)777-4790
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...