The COREN® Total Hip System is intended for use in total or partial hip arthroplasty and in primary or revision surgery for the following conditions: - Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, a. traumatic arthritis - b. Inflammatory degenerative joint disease, such as rheumatoid arthritis - Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal c. femur with head involvement, unmanageable using other techniques - d. Patients with failed previous surgery where pain, deformity, or dysfunction persists - Revision of previously failed total hip arthroplasty. e.
Device Story
Cementless, metal-on-polyethylene total hip prosthesis system; components include femoral stem, femoral head, acetabular cup, UHMWPE insert, and bone screws. Used in total or partial hip arthroplasty; primary or revision surgery. Implanted by orthopedic surgeons in clinical/hospital settings. System provides structural replacement for diseased or fractured hip joints; restores joint function and reduces pain. Components utilize modular tapered interfaces for assembly. Device relies on mechanical fixation; no electronic or software components.
Clinical Evidence
Bench testing only. Performance testing included fatigue testing (ISO 7206-4, ISO 7206-8), wear testing (ISO 14242-1, ISO 14242-2), attachment strength (torque/lever out), and residual free radical content analysis (ESR). Compliance with numerous ASTM and ISO standards for materials and mechanical properties demonstrated.
Technological Characteristics
Materials: Ti-6Al-4V alloy (ASTM F136), Co-Cr-Mo alloy (ASTM F1537), and UHMWPE (ASTM F648). Cementless, porous-coated acetabular cup; modular femoral head/stem with tapered interface. Sterilization via gamma irradiation or ethylene oxide. No software or electronic components.
Indications for Use
Indicated for patients requiring total or partial hip arthroplasty (primary or revision) due to non-inflammatory degenerative joint disease (avascular necrosis, osteoarthritis, traumatic arthritis), inflammatory degenerative joint disease (rheumatoid arthritis), non-union, femoral neck/trochanteric fractures of proximal femur, or failed previous surgery with persistent pain, deformity, or dysfunction.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
510(k) Summary
JUL 2 1 2011
# Corentec Co., Ltd.
# COREN® Total Hip System
## K103431
July 14, 2011
#### ADMINISTRATIVE INFORMATION
Manufacturer Name: Corentec Co., Ltd. 247 Giro-ri, Ipjang-myeon, Seobuk-Gu Cheonan-si, Chungchongnam-do, 331-822 South Korea Telephone: +82-41-585-7114 +82-41-585-7113 Fax: Official Contact: J.S. Daniel Project Manager - QA/RA 11th Chungho Tower, 748-1 Banpo-1 Dong, Seocho Gu, Seoul, South Korea, 137-040 Telephone: +82-2-3445-5475 Fax: +82-2-3445-5497 Email: isdaniel@corentec.com Kevin A. Thomas, PhD Representative/Consultant: Floyd G. Larson PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130 Telephone: +1 (858) 792-1235 Fax: +1 (858) 792-1236 Email: kthomas@paxmed.com
flarson@paxmed.com
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | COREN® Total Hip System |
|----------------------------|-----------------------------|
| Common Name: | Total Hip Prosthesis System |
| Classification Regulation: | 21 CFR 888.3358, Class II |
| Product Codes: | LPH |
| Classification Panel: | Orthopedic Products Panel |
| Reviewing Branch: | Orthopedic Devices Branch |
e. lot of 3
Page 1 of 23
{1}------------------------------------------------
510(k) Summary
#### INTENDED USE
The COREN® Total Hip System is intended for use in total or partial hip arthroplasty and in primary or revision surgery for the following conditions:
- Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, a. traumatic arthritis
- b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal c. femur with head involvement, unmanageable using other techniques
- d. Patients with failed previous surgery where pain, deformity, or dysfunction persists
- Revision of previously failed total hip arthroplasty. e.
#### DEVICE DESCRIPTION
The COREN Total Hip System is a cementless, metal-on-polyethylene hip system for hip arthroplasty. It consists of the following components: Femoral stem - Modified BL Coren Stem; Femoral head - Coren Metal Head; Acetabular system - Coren U Cup, Coren PE Insert and Coren Bone Screw; and Coren THR Instrumentation. The components are manufactured from the following materials: Ti-6A1-4V alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401); Co-Cr-Mo alloy conforming to ASTM F1537 Standard Specification for Wrought Cobalt-28-Chromium-6-Molybdenum Allovs for Surgical Implants (UNS R31537, UNS R31538, and UNS R31539); and ultra-high molecular weight polyethylene conforming to ASTM F648 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants. COREN Total Hip System components are sterilized by gamma irradiation or ethylene oxide.
#### EQUIVALENCE TO MARKETED DEVICES
Corentec Co., Ltd., submits the following information in this Premarket Notification to demonstrate that for the purposes of FDA's regulation of medical devices, the COREN® Total Hip System is substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices:
Zimmer Austin, Inc., CLSTM Spotorno™ Stem, cleared under K042249;
Sulzer Orthopedics, Inc., CLS Varus/CLS 135 Stem, cleared under K010839;
Aesculap, Inc., Excia Total Hip System, cleared under K042344;
Wright Medical Technology, Inc., LINEAGE® A-CLASS™ Poly Liner, cleared under K052026:
Biomet, Inc., Metal on Metal Acetabular System, cleared under K993438; and
Biomet, Inc., M2a™ 32 mm Taper System, cleared under K003363.
{2}------------------------------------------------
510(k) Summary
K103431
The COREN Total Hip System Modified BL Coren Stem and the CLS™ Spotorno™ Stem (K042249) and the CLS Varus/CLS 135 Stem (K010839) are made of the same material, have the same grit-blasted surface finish, have the same neck angle and have similar overall geometries and ranges of sizes.
The COREN Total Hip System Coren U Cup and the Excia Total Hip System (K042344) are both hemispherical acetabular cup made of the same material and both have plasma-sprayed porous coatings with similar pore size, porosity and coating thickness. The subject and predicate acetabular cups also incorporate holes through the cup in similar positions and angulations for optional screw fixation, and both an internal taper mechanism for locking the UHMWPE insert to the cup.
The COREN Total Hip System and the LINEAGE® A-CLASS™ Poly Liner (K052026) both include UHMWPE inserts made of the same cross-linked material, encompassing a similar range of configurations, and both incorporate a taper mechanism for locking to the acetabular cup.
The COREN Total Hip System Coren Metal Head and the Metal Acetabular System (K993438) and M2a™ 32 mm Taper System (K003363) all include modular femoral heads made of the same material and a similar tapered interface to the femoral stem. The subject and predicate modular femoral heads all encompass a similar range of head sizes, neck lengths and femoral offsets, and all have similar surface finish and sphericity.
Any differences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy.
Performance testing was conducted to demonstrate substantial equivalence and included methods described in the following standards: ISO 7206-4, ISO 7206-8, ASTM F543, ISO 14242-1, ISO 14242-2, ASTM F1820, ISO 7206-10, ISO 7206-9, ISO 7206-2, ASTM D638, ASTM F2183, ASTM F648, ASTM D3418, ASTM F2214, ASTM F2102, ASTM F2381, ASTM F1160, ASTM F1044, ASTM F1147, ASTM F1978, ASTM F2582, ISO 7206-6, and ISO 10993-7. The attachment strength between the UHMWPE liner and the acetabular shell was tested by measuring torque out and lever out strengths. Residual free radical content in crosslinked UHMWPE was assessed by electron spin resonance (ESR).
Overall, the COREN Total Hip System has the following similarities to the predicate devices:
- has the same intended use, .
- uses the same operating principles, .
- incorporates the same basic designs, .
- incorporates the same or very similar materials, and .
- has similar packaging and is sterilized using the same materials and processes. .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol of medicine, with three lines representing the branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Corentec Co. Ltd % Kevin Thomas, PhD Vice President and Director of Regulatory Affairs PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, California 92130
JUL 2 1 2011
Re: K103431
Trade/Device Name: COREN® Total Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH Dated: June 15, 2011 Received: June 16, 2011
Dear Dr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Kevin Thomas, PhD
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Emil Keith
for Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
K103431 510(k) Number:
Device Name: COREN® Total Hip System
Indications for Use:
The COREN® Total Hip System is intended for use in total or partial hip arthroplasty and in primary or revision surgery for the following conditions:
- Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, a. traumatic arthritis
- Inflammatory degenerative joint disease, such as rheumatoid arthritis b.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal C. femur with head involvement, unmanageable using other techniques
- Patients with failed previous surgery where pain, deformity, or dysfunction persists d.
- Revision of previously failed total hip arthroplasty. e.
Prescription Use __ X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of __
for m mcleerson
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103431
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.