ACCIN HIP SYSTEM

K073068 · Accelerated Innovation, LLC · LPH · Jan 10, 2008 · Orthopedic

Device Facts

Record IDK073068
Device NameACCIN HIP SYSTEM
ApplicantAccelerated Innovation, LLC
Product CodeLPH · Orthopedic
Decision DateJan 10, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Accin™ Hip System components are for use in total hip arthroplasty as a result of: Hip arthritis caused by rheumatoid disease, non-inflammatory degenerative joint disease, osteoarthritis, and arthritis resulting from biologic or mechanical trauma to the hip Correction of functional deformity Avascular Necrosis Treatment of non-unions, femoral neck and trochanteric fractures of the proximal femur Difficult clinical management problems involving persistent pain and physical impairment where conventional arthodesis is not likely to achieve satisfactory results These components are single use only and are intended for implantation without bone cement.

Device Story

Accin™ Hip System is a total hip arthroplasty prosthesis for uncemented implantation. System components include a plasma spray-coated titanium alloy femoral stem, cobalt chrome femoral head, plasma spray-coated titanium alloy acetabular shell, bone screws, and a polyethylene acetabular insert. Device is intended for surgical replacement of hip joint to address arthritis, deformity, necrosis, or fractures. Implanted by orthopedic surgeons in clinical/hospital settings. Provides structural replacement of hip joint to restore function and reduce pain. Single-use device.

Clinical Evidence

Bench testing only. Testing conducted per FDA guidance documents: "Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants," "Class II Special Controls Guidance Document: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis," and "Non-clinical Information for Femoral Stem Prostheses."

Technological Characteristics

Materials: Titanium alloy (femoral stem, acetabular shell) with CP Titanium plasma spray coating, cobalt chrome (femoral head), polyethylene (acetabular insert). Uncemented, semi-constrained prosthesis. Fixation via bone screws. Single-use.

Indications for Use

Indicated for patients requiring total hip arthroplasty due to rheumatoid arthritis, non-inflammatory degenerative joint disease, osteoarthritis, post-traumatic arthritis, functional deformity, avascular necrosis, non-unions, or proximal femoral fractures (femoral neck/trochanteric) where conventional arthrodesis is insufficient.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K073068 (pg 1/1) ## Summary of Safety and Effectiveness | Submitter: | Michael Kvitnitsky<br>Accelerated Innovation, LLC<br>1033 US Highway 46, Suite A204<br>Clifton, NJ 07103<br><br>JAN 10 2008 | | |---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Date Prepared: | October 10, 2007 | | | Device: | Accin™ Hip System | | | Classification: | 87 LPH - Hip joint metal/polymer/metal semi-constrained uncemented prosthesis,<br>21 CFR 888.3358, Class II and 87 HWC and screw, fixation, bone, 21 CFR<br>888.3040, Class II | | | Predicate Device: | Provident Hip System - K935484, K946371, K960180, K991622 K001745, and<br>K002796 and Stelkast Bone Screws - K934162 | | | Device Description: | The Accin™ Hip System consists of plasma spray coated titanium alloy femoral<br>component, a cobalt chrome femoral head, a plasma spray coated titanium alloy<br>acetabular shell, bone screws and a polyethylene acetabular insert. | | | Intended Use: | The Accin™ Hip System components are for use in total hip arthroplasty as a result of:<br>Hip arthritis caused by rheumatoid disease, non-inflammatory degenerative joint<br>disease, osteoarthritis, and arthritis resulting from biologic or mechanical trauma<br>to the hip<br>Correction of functional deformity<br>Avascular Necrosis<br>Treatment of non-unions, femoral neck and trochanteric fractures of the proximal<br>femur<br>Difficult clinical management problems involving persistent pain and physical<br>impairment where conventional arthodesis is not likely to achieve satisfactory<br>results<br>These components are single use only and are intended for implantation without bone<br>cement. | | | Comparison to Predicates: The Accin™ Hip system consists of cobalt chrome femoral head, titanium alloy femoral stem, acetabular | | | components plasma spray coated with CP Titanium, bone screws and an acetabular polyethylene insert. The device is equivalent to the Stelkast Provident Hip System, which also has the same components manufactured from the same materials. Accin™ has determined that any differences in the proposed device will not impact the safety or effectiveness of the hip system for its intended use. Testing has shown that the proposed device meets the requirements of the current FDA Guidance documents on total hip arthroplasty product and the proposed device is equivalent to the predicate device. ## Synopsis of Test Methods and Results: Tests were performed on the Hip System. The tests performed can be found in the guidance documents entitled "Guidance for Industry on the Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements;" "Class II Special Controls Guidance Document: Hip Joint Metal/Polymer Constrained Cermented or Uncom of the Prosthesis;" and "Guidance for Industry and FDA Staff - Non-clinical Information for Femoral Stem Prostheses." {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in a circular arrangement. Public Health Service JAN 10 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Accelerated Innovation, LLC % Mr. Michael Kvitnitsky 1033 US Highway 46 Suite A204 Clifton, NJ 07103 Re: K073068 Trade/Device Name: Accin™ Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, HWC Dated: December 21, 2007 Received: December 27, 2007 Dear Mr. Kvitnitsky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 -- Mr. Michael Kvitnitsky This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Mark N. Millerson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K073068 (pg 1/1) ## Indications for Use Form 510(k) Number (if known): Device Name: Accin™ Hip System Indications for Use: Indications for Use: The Accin™ Hip System components are for use in total hip arthroplasty as a result of: - Hip arthritis caused by rheumatoid disease, non-inflammatory degenerative joint . disease, osteoarthritis and arthritis resulting from biologic or mechanical trauma to the hip - Correction of functional deformity . - Avascular necrosis . - Treatment of non-unions, femoral neck and trochanteric fractures of the proximal femur . - Difficult clinical management problems involving persistent pain and physical impairment . where conventional arthodesis is not likely to achieve satisfactory results. These components are single use only and are intended for implantation without bone cement. Prescription Use X (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 (Division Sign-Off) **Division of General, Restorative,** **and Neurological Devices** **510(k) Number** K073068
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...