I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix)
Applicant
I.T.S. GmbH
Product Code
HRS · Orthopedic
Decision Date
Aug 31, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The I.T.S. Pelvic Reconstruction Systems (PRS RX & PRS Phoenix) are indicated to stabilize one or more pelvic bone fractures in the pelvis in an adult patient.
Device Story
System of metallic bone fixation plates, screws, and spikes for pelvic fracture reconstruction. Plates (2.5mm-4.0mm thickness) and screws (3.5mm cortical/locking, 4.2mm cancellous locking, 8.5mm cannulated) provide mechanical stabilization of bone fragments. Used in surgical settings by orthopedic surgeons. Components are non-sterile, single-use, and feature 'DOTIZE' Type II anodized surface treatment. Surgeon uses ancillary instrumentation (drills, guides, clamps) to reduce fractures and secure plates/screws to bone. Output is rigid internal fixation, facilitating bone healing and restoring pelvic stability.
Clinical Evidence
No clinical data submitted. Bench testing only, including performance testing per ASTM F543 (screws) and ASTM F382 (plates), plus FEA analysis simulation.
Technological Characteristics
Plates: CP Titanium (ASTM F67). Screws/Spikes: Titanium Alloy 6Al-4V (ASTM F136). Surface: 'DOTIZE' Type II anodization. Mechanical fixation via locking/standard screws. Non-sterile, single-use. Low-profile design.
Indications for Use
Indicated for adult patients with pelvic bone fractures, including acetabular, pelvic ring, quadrilateral surface, symphysis, ilium, and SIJ fractures; symphysis pubis disruptions, osteotomies, arthrodesis, sacroiliac joint dislocations, and revision surgery for pseudoarthroses, non-unions, and mal-unions. Not for spinal use.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
I.T.S. - Pelvic Reconstruction System - PRS (K063166)
{0}------------------------------------------------
August 31, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
I.T.S. GmbH c/o Jennifer Hadfield Oserve Group US U.S. Agent and Official Correspondent 350 S Main Street, Suite 309 Doylestown, PA 18901
Re: K210935
Trade/Device Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC, HTN Dated: March 15, 2021 Received: March 29, 2021
Dear Jennifer Hadfield:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K210935
#### Device Name
I.T.S. Pelvic Reconstruction Systems (PRS RX & PRS Phoenix)
Indications for Use (Describe)
The I.T.S. Pelvic Reconstruction Systems (PRS RX & PRS Phoenix) are indicated to stabilize one or more pelvic bone fractures in the pelvis in an adult patient.
Indications for use of the I.T.S. Pelvic Reconstruction System (PRS RX) include:
- Fractures of the acetabulum
- Fractures of the pelvic ring
- Fractures of the quadrilateral surface
- Fractures of the sympyhsis
- Fractures of the ilium
- · Fractures of the SIJ
- · Ilio-iliac distance osteosynthesis
- · Symphysis pubis disruptions, osteotomies, arthrodesis and sacroiliac joint dislocations
- · Revision surgery of pseudoarthroses, non-unions and mal-unions
The I.T.S. PRS RX System is not intended for spinal use.
Indications for use of the I.T.S. Pelvic Reconstruction System (PRS Phoenix) include:
- · Fractures involving the Posterior Wall & Posterior Column
- · Fractures involving the Anterior Column of the Acetabulum
- Fractures involving the Quadrilateral Surface
- · Symphyseal Disruptions & Para-symphyseal Fractures
- Fractures of the ilium
- · Fractures of the SIJ
- · Dorsal neutralization plating for posterior pelvic ring fractures
- · Osteotomies, arthrodesis and sacroiliac joint dislocations
- · Revision surgery of pseudoarthroses, non-unions and mal-unions
The I.T.S. PRS Phoenix System is not intended for spinal use.
Indications for use of the I.T.S. Infra-acetabular screw placement include:
· Fractures involving the anterior column, e.g. anterior column plus posterior hemitransverse (ACPH) and associated both column (ABC) fractures
The I.T.S. Infra-acetabular screw is not intended for spinal use.
Indications for use of the I.T.S. 8.5mm Cannulated Screws & Washer include:
- · Pelvic fractures
The I.T.S. 8.5mm Cannulated Screws & Washer are not intended for spinal use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
# In Accordance with 21 CFR 807.92 of the Federal Code of Regulations 510(k) Summary
| NAME OF FIRM: | I.T.S. GmbH<br>Autal 28<br>Lassnitzhoehe, 8301<br>Austria<br>www.its-implant.com |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) FIRM CONTACT: | Jennifer Hadfield<br>Qserve Group US, Inc.<br>350 S Main Street, Suite 309<br>Doylestown, Pennsylvania, 18901, United States<br>Tel. No. 215-6003103<br>e-mail: jennifer.hadfield@qservegroup.com |
| TRADE NAME: | I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) |
| DATE: | August 30th, 2022 |
| COMMON NAME: | Plate, Fixation, Bone; Screw, Fixation, Bone; Washer, Bolt Nut |
| CLASSIFICATION: | Single/multiple component metallic bone fixation appliances and<br>accessories (primary)<br>Class II, (see 21 CFR. Sec. 888.3030)<br>Smooth or threaded metallic bone fixation fastener<br>Class II, (see 21 CFR. Sec. 888.3040) |
### DEVICE PRODUCT CODE: HRS, HWC, HTN
#### SUBSTANTIAL EQUIVALENCE: PRIMARY PREDICATE I.T.S. - Pelvic Reconstruction System - PRS (K063166) ADDITIONAL PREDICATES I.T.S. - FLS Extremity System (K131722) Stryker - Trauma Pelvic Set (K001614) Stryker - Pelvis II Implant System (K131132) Synthes - 3.5mm Low Profile Pelvic Reconstruction Plate (K031573) Synthes - 3.5mm Quadrilateral Surface Plates (K093928) Acumed - Pelvic Bone Plate System (K122538) Synthes - 3.5mm Spring Plate (K061973) DEVICE DESCRIPTION: The I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) encompasses a number of fracture fixation subsystems of multiple plate designs for fracture fixation and reconstruction of pelvic ring fractures in the pelvis. The I.T.S. Pelvic Reconstruction Plating System (PRS RX & PRS Phoenix) consists of the following plate types: 1) A Curved Plate at a 2.5mm thickness with 4 to 16 hole length sizes, 2) A Straight Plate at a 2.5mm thickness with 10 to 14 hole length sizes, 3) A SIJ Closed Plate at a 2.5mm thickness in a 4-hole size, 4) A SIJ L-Plate at a 2.5mm thickness in a 5-hole size, 5) A J-Plate at a 2.5mm thickness with 6 to 16 hole length sizes,
{5}------------------------------------------------
#### I.T.S. GmbH - Pelvic Reconstruction System (PRS RX & PRS Phoenix) - 510(k) Summary
6) A Symphysis Plate at a 4.0mm thickness in a 4 & 6 hole size,
7) A Symphysis Plate Curved at a 4.0mm thickness in a 6 & 8 hole size,
8) A RIM Plate at a 2.5mm thickness in a 10 & 14 hole, Right & Left size.
9) A Posterior Wall Plate at a 2.5mm thickness in a 6 hole, Right & Left size.
10) A Posterior Wall Plate II at a 2.5mm thickness in a 7 & 8 hole, Right & Left size,
11) A Quadrilateral Column Plate at a 2.5mm thickness in a Small & Large, Right & Left size,
12) A Quadrilateral Supporting Plate at a 2.5mm thickness in a Right & Left size.
13) A Posterior Wall Plate Extended at a 2.5mm thickness in a Small & Large, Right & Left size,
14) A Posterior Column Plate at a 2.5mm thickness in a 8-hole, Right & Left size,
All plate designs are low profile in thickness and made from Implant Grade 2 CP Titanium material (to ASTM F67). The PRS RX & PRS Phoenix Plating System encompasses a 3.5mm Cortical Locking and Standard Screw and a 4.2mm Cancellous Locking Screw in various lengths. All bone screws are pre-drilling and self-tapping in design and manufactured from Implant Grade 5 high strength 6-4 Alloyed Titanium (to ASTM F136). A threaded Spike allows inter-operative plate fixation to bone fragments. The Spike is also manufactured from Implant Grade 5 high strength 6-4 Alloyed Titanium material (to ASTM F136). All components (plates, screws, spike) have a anodize 'DOTIZE' Type II surface treatment preparation.
Ancillary instrumentation (Drills, Drill Guides, Insertion Guides, Clamps, Bending Heaver, Measuring Gauges, and Screwdrivers) is made available for bone fragment reduction and plate/screw placement and insertion. All plates, screws, spike are provided Non-Sterile for singleuse.
#### INTENDED USE:
The I.T.S. Pelvic Reconstruction Systems (PRS RX & PRS Phoenix) are indicated to stabilize one or more pelvic bone fractures in the pelvis in an adult patient.
# Indications for use of the I.T.S. Pelvic Reconstruction System
(PRS RX) include:
- · Fractures of the acetabulum,
- · Fractures of the pelvic ring,
- Fractures of the quadrilateral surface,
- Fractures of the symphysis. .
- · Fractures of the ilium,
- · Fractures of the SIJ,
- Ilio-iliac distance osteosynthesis,
Symphysis pubis disruptures, osteotomies, arthrodesis and sacroiliac joint dislocations,
· Revision surgery of pseudoarthrosis, non-unions and mal-unions
The I.T.S. Pelvic Reconstruction System (PRS RX) is not intended for spinal use.
{6}------------------------------------------------
### Indications for use of the I.T.S. Pelvic Reconstruction System
(PRS Phoenix) include:
- · Fractures involving the Posterior Wall & Posterior Column
- · Fractures involving the Anterior Column of the Acetabulum
- Fractures involving the Quadrilateral Surface
- · Symphyseal Disruptions & Para-symphyseal Fractures
- Fractures of the ilium
- · Fractures of the SIJ
- · Dorsal neutralization plating for posterior pelvic ring fractures
- · Osteotomies, arthrodesis and sacroiliac joint dislocations
- · Revision surgery of pseudoarthroses, non-unions and mal-unions
The I.T.S. Pelvic Reconstruction System (PRS Phoenix) is not intended for spinal use.
Indications for use of the I.T.S. Infra-acetabular screw placement include:
· Fractures involving the anterior column, e.g. anterior column, anterior column plus posterior hemitransverse (ACPH) and associated both column (ABC) fractures
The I.T.S. Infra-acetabular screw placement is not intended for spinal use.
Indications for use of the I.T.S. 8.5mm Cannulated Screws & Washer include:
· Pelvic fractures
The I.T.S. 8.5mm Cannulated Screws & Washer are not intended for spinal use.
{7}------------------------------------------------
| BASIS OF SUBSTANTIAL<br>EQIVLAENCE: | The I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) is<br>Substantially Equivalent (SE) to the I.T.S, Stryker, Synthes, and Acumed<br>- Pelvic Reconstruction - Predicate Systems. An Engineering Analysis<br>and Perfomance Testing was provided for plate bending strength and<br>bending stiffness and screw insertion/removal, yield torque, self-tapping<br>torque and axial pullout strength to demonstrate substantial equivalence.<br>Biocompatibility Risk Assessments were also performed. |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CLINICAL TESTING: | Clinical data was not submitted. |
| NON-CLINICAL TESTING: | Performance testing as a consensus standard was performed under<br>ASTM F543 for all bone screws and under ASTM F382 for bone plates<br>and in addition a FEA analysis simulation following ASTM F382. |
| SUMMARY OF TECHNO-<br>LOGICAL<br>CHARACTERISTICS: | The I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) is<br>substantially equivalent in material, geometry, design, indications<br>and operational principles to the following predicate systems legally<br>marketed in the US: |
| PRIMARY PREDICATE | I.T.S. - Pelvic Reconstruction System - PRS (K063166)<br>Differences in technological characteristics:<br>- Geometry / Dimensions / Design: comparable in plate sizes, thickness,<br>width and locking features<br>- Material: identical CP titanium grade 2 for all plate components and<br>identical titanium alloy 6Al-4V for all screw components<br>- Chemical composition: identical CP titanium grade 2 acc. to ASTM F67<br>and identical titanium alloy 6Al-4V acc. to ASTM F136<br>- Manufacturing: identical manufacturing steps<br>- Surface treatment: identical anodization with Type II |
| CONCLUSIONS: | Based on the similarity in material, geometry, design, indications and<br>operational principles, as well as both the Engineering Analysis and<br>Performance Testing, the I.T.S. Pelvic Reconstruction System (PRS RX<br>& PRS Phoenix) has been demonstrated to be substantially equivalent<br>(SE) to the predicate devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.