The Acumed Pelvic Bone Plate System consists of plates, screws and accessories intended to provide fixation during fractures, fusions, and osteotomies of the acetabulum, sacrum, ilium, and entire pelvic ring, as well as treatment of sacroiliac joint dislocations and symphysis pubis disruptions.
Device Story
Acumed Pelvic Bone Plate System provides internal bone fixation for pelvic injuries. System comprises titanium plates (ASTM F67) and titanium alloy screws (ASTM F136) in various sizes. Plates are contoured or trimmed intraoperatively by surgeons to match patient anatomy and injury patterns. Used in clinical settings by orthopedic surgeons to stabilize fractures, fusions, and osteotomies of the acetabulum, sacrum, ilium, and pelvic ring, and to treat sacroiliac joint dislocations and symphysis pubis disruptions. Fixation is achieved via non-locking screw insertion. Implants are provided sterile or non-sterile for use with general surgical instrumentation. Device stabilizes bone segments to facilitate healing.
Clinical Evidence
Bench testing only. Mechanical testing performed in accordance with ASTM F382 (metallic bone plates) and ASTM F543 (metallic medical bone screws).
Indicated for patients requiring bone fixation for fractures, fusions, and osteotomies of the acetabulum, sacrum, ilium, and entire pelvic ring, or treatment of sacroiliac joint dislocations and symphysis pubis disruptions.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
# 510(k) Summary
## Contact Details
Applicant Name: Acumed LLC 5885 NW Cornelius Pass Road, Hillsboro, OR 97124-9432
> Kara Budor, Regulatory Specialist 503-207-1412
Date Prepared: August 20, 2012
### Device Name
Trade Name: Acumed Pelvic Bone Plate System
Common Name: Pelvic System
Classification: 21 CFR 888.3030 Single/multiple component metallic bone fixation appliances and accessories.
Class: II
Product Code: HRS
### Legally Marketed Predicate Device(s)
There are two predicate devices. The comparison is to the Stryker Trauma Pelvic Set and the I.T.S. Pelvic Reconstruction System.
| 510(k) Number | Product Code | Trade Name | Applicant |
|---------------|-----------------|----------------------------------------|--------------------------------------------------|
| K001614 | KTW<br>888.3030 | Stryker Trauma Pelvic Set | Howmedica Osteonics<br>Corp. |
| K063166 | HRS<br>888.3030 | I.T.S. Pelvic Reconstruction<br>System | I.T.S. Implantat-<br>Technologie-Systeme<br>GmbH |
# Device Description
The Acumed Pelvic Bone Plate System consists of plates, screws and accessories intended to provide fixation during fractures, fusions, and osteotomies of the acetabulum, sacrum, ilium,
{1}------------------------------------------------
and entire pelvic ring, as well as treatment of sacroiliac joint dislocations and symphysis pubis disruptions.
Plates are available in a variety of shapes and sizes which can be used "as is" or additionally contoured and/or trimmed intraoperatively to accommodate varying injury patterns and/or patient anatomy. All plates are made of titanium per ASTM F67 and are designed to accommodate the 3.5mm screws included in the system. Plates range in length from 30mm to 250mm, have a width of 4mm to 17mm, and a thickness of 1.5mm to 4mm.
The system includes 3 families of screws with different diameters: 2.7mm, 3.5mm, and 4.3mm. All screw families can be used for fixation independent of the plates. Screws are made of titanium alloy per ASTM F136.
All implants are provided sterile and non-sterile.
## Intended Use/Indications for use
The Acumed Pelvic Bone Plate System consists of plates, screws and accessories intended to provide fixation during fractures, fusions, and osteotomies of the acetabulum, sacrum, ilium, and entire pelvic ring, as well as treatment of sacroiliac joint dislocations and symphysis pubis disruptions.
#### Substantial Equivalence Comparison
The basic comparison between the Acumed Pelvic Bone Plate System and the predicate devices is given in the table below.
| Category | Acumed Pelvic<br>Bone Plate System | I.T.S. Pelvic Reconstruc-<br>tion System K063166 | Stryker Trauma Pelvic Set<br>K001614 |
|------------------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Material | Titanium per ASTM F67<br>and Titanium alloy per<br>ASTM F136 | Commercially pure<br>titanium and 6-4 alloyed<br>titanium | Stainless Steel per ASTM<br>F138 |
| Plate Shape | Straight plates<br>Curved plates<br>Plates for anatomy specific<br>sites | Straight plates<br>Curved plates<br>Plates for anatomy specific<br>sites | Straight plates<br>Curved plates<br>Plates for anatomy specific<br>sites |
| Plate Fixation<br>Method | Screw fixation | Screw fixation | Screw fixation |
| Screw Type | Non-locking | Non-locking and Locking | Non-locking |
| Provided sterile<br>and/or non-<br>sterile | Provided sterile and non-<br>sterile | Unknown | Provided sterile and non-<br>sterile |
| Implants are<br>used with<br>general surgical<br>instrumentation | Yes | Yes | Yes |
{2}------------------------------------------------
Acumed Pelvic Bone Plate System 510(k) Notification
The Acumed Pelvic Bone Plate System, the I.T.S. PRS Low Profile – Multiple Type – Pelvic Plating System, and the Stryker Trauma Pelvic Set are all systems of plates, screws, and instruments which are used to achieve fixation. There are some differences, but none of them raise new issues of safety or effectiveness. The Acumed Pelvic Bone Plate System is substantially equivalent to the predicate devices.
# Non-clinical Testing
The non-clinical testing included in this submission includes testing to ASTM F382 and F543.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure with three wing-like shapes.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
#### Letter Dated: November 19, 2012
Acumed, LLC % Ms. Kara Budor Regulatory Specialist 5885 Northwest Cornelius Pass Road Hillsboro, Oregon 97124
Re: K122538
Trade/Device Name: Acumed Pelvic Bone Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And
Accessories
Regulatory Class: Class II Product Code: HRS Dated: August 20, 2012 Received: August 21, 2012
Dear Ms. Budor: .
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
#### Page 2- Ms. Kara Budor
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. -
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours. Erin I. Keith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation · Center for Devices and Radiological Health
{5}------------------------------------------------
Acumed Pelvic Bone Plate System 510(k) Notification
# Indications for Use
510(k) Number (if known): K122538
Device Name: Acumed Pelvic Bone Plate System
Indications for Use:
The Acumed Pelvic Bone Plate System includes plates, screws, and accessories designed to provide fixation during fractures, fusions, and osteotomies for the acetabulum, sacrum, ilium, and entire pelvic ring, as well as treatment of sacroiliac joint dislocations and symphysis pubis disruptions.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21. CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) .
K. Ath
(Division Sign-Off) Division of Orthopedic Device 510(k) Number K12253
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.