The intended use of the I.T.S. Pelvic Reconstruction System (PRS) is to stabilize one or more pelvic bone fractures in the 'pelvic ring' area of the pelvis of an adult or pediatric patient which include the use of the following pelvic fracture fixation systems. Indications for use of the I.T.S. PRS Low Profile - Multiple Type - Pelvic Plating System include: 1) Fracture reconstruction of the acetabulum, sacrum, illium, and entire pelvic ring. 2) Revision surgery of pseoduarthroses, non-unions and mal-unions, 3) Ilio-Iliac distance osteosynthesis, 4) Osteotomies, 5) Arthrodesis. 6) Sacroiliac joint dislocations, and 7) Symphysis pubis disruptures Indications for use of the I.T.S. PRS Sacral Rod System include: 1) Fixation of fractures of the posterior pelvis, 2) Fixation of fractures of the posterior iliac spine, 3) Fixation of fractures of the posterior inferior iliac spine, 4) Dorsal stabilization of the posterior pelvic ring for unstable pelvic ring injuries, 5) Fixation of sacral fractures, and 6) Fracture dislocations of the sacro-iliac joint Indications for use of the I.T.S. PRS 7.3mm Cannulated SIJ Traction Screw & Washer include fracture fixation of pelvic bone fractures where indicated and for periacetabular osteotomies. The system(s) has not been studied in spinal use, and is not intended for use in vertebral column fracture or fusion procedures.
Device Story
Pelvic Reconstruction System (PRS) provides fracture fixation and reconstruction of pelvic ring; includes multiple plate designs, sacral threaded rod system, and cannulated screws/washers. Plates are low-profile; screws are pre-drilling and self-tapping. Sacral rod system uses threaded pin, wedge-shaped washers, and nut/locknut for ilio-iliac posterior pelvic ring stabilization. Cannulated screws used for guided reduction. Used in clinical/surgical settings by surgeons. Components stabilize bone fragments to facilitate healing. Not for spinal use.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Materials: High strength 6-4 Alloyed Titanium with TIODIZE II surface treatment. Components: Plates (2.5mm-4.0mm thickness), cannulated screws (5.9mm-7.3mm), threaded rods, washers, nuts. Mechanical fixation principle. Non-sterile/sterilizable.
Indications for Use
Indicated for adult and pediatric patients requiring stabilization of pelvic ring fractures, acetabular/sacral/iliac reconstruction, revision of non-unions/mal-unions, osteotomies, arthrodesis, sacroiliac joint dislocations, and symphysis pubis disruptions. Contraindicated for spinal use or vertebral column procedures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
I.T.S. Straight Plate with Angular Stability & Screw System (K060156)
Submission Summary (Full Text)
{0}------------------------------------------------
K063/66
### 510(k) Summary of Safety and Effectivenes Section XII:
# SAFE MEDICAL DEVICES ACT OF 1990
510(k) Summary
| NAME OF FIRM: | I.T.S. Implantat-Technologie-Systeme GmbH.<br>Autal 28.<br>Lassnitzhoehe A - 8301<br>AUSTRIA | DEC 22 2006 |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| 510(k) FIRM CONTACT: | Al Lippincott<br>Engineering Consulting Services, Inc.<br>3150 E. 200th St.<br>Prior Lake, MN 55372 | |
| TRADE NAME: | Pelvic Reconstruction System (PRS) | |
| COMMON NAME: | Pelvic Bone Fixation Set | |
| CLASSIFICATION: | Plate, Fixation, Bone (see 21 CRF, Sec. 888.3030),<br>Pin, Fixation, Threaded (see 21 CRF, Sec. 888.3040)<br>Screw, Fixation, Bone (see 21 CFR, Sec. 888.3040),<br>Washer, Bolt, Nut (see 21 CFR, Sec. 888.3030) | |
| DEVICE PRODUCT CODE: | HRS | |
| SUBSEQUENT PRODUCT CODE: | JDW, HWC, HTN | |
| SUBSTANTIALLY<br>EQUIVALENT DEVICES | Synthes 3.5mm Low Profile Pelvic Reconstruction (K031573)<br>Stryker Trauma Pelvic Set (K001614)<br>Zimmer Pelvic Reconstruction Set<br>Synthes Sacral Bar System (K001720)<br>I.T.S. GmbH FR.O.H. Calcaneus Repair System (K051642)<br>Synthes 3.9mm Pelvic Screw (K013044)<br>I.T.S. Straight Plate with Angular Stability & Screw System (K060156) | |
| DEVICE DESCRIPTION: | The I.T.S. Pelvic Reconstruction System (PRS) encompasses a number<br>of fracture fixation subsystems (multiple pelvic plate designs, sacral<br>threaded rod, and cannulated screw & washer) for facture fixation and<br>reconstruction of pelvic ring fractures in the pelvis.<br><br>The I.T.S. PRS Low Profile - Multiple Type - Pelvic Plating System<br>consists of the following plate types: 1) A Straight Plate at a 2.5mm<br>thickness with 10 to 14 hole length sizes, 2) A Curved Plate with a<br>108mm radius at a 2.5mm thickness with 4 to 16 hole length sizes, 3) A<br>Curved Plate with a 88mm radius at a 2.5mm thickness also with 4 to 16<br>hole length sizes, 4) A Symphysis Plate at a 4.0mm thickness in both a<br>4 & 6 hole size, 5) A SacroIlliac-Joint (SIJ) L-Shaped Plate at a 2.5mm<br>thickness in a left and right 5 hole size, and 6) A SIJ Closed Plate at a<br>2.5mm thickness in a 4 hole size.<br><br>All plate designs are low profile in thickness and made from CP | |
{1}------------------------------------------------
of Cancellous (5.9mm std.compression & 5.9mm Locking screws) and Cortical (4.5mm std.compression screw) screw types and length sizes. All bone screws are pre-drilling and self-tapping in design and manufactured from high strength 6-4 Alloyed Titanium. All components (plates & screws) have a TIODIZE II surface treatment preparation. The I.T.S. PRS Sacral Rod System consists of a threaded pin, wedgeshaped washers, and nut/locknut design with one end of the threaded pin having a trocar point. The assembled unit is used to stabilize ilio-iliac posterior pelvic ring disruption injuries. All components are manufactured from high strength 6-4 Alloyed Titanium material and have the TIODIZE II surface preparation. The I.T.S. PRS 7.3mm Cannulated SIJ Traction Screw & Washer_is a pre-drilling, self-tapping, and back-tapping screw design for guided reduction of pelvic bone fractures. Washers (flat & curved) are available for use with the screw. All 7.3mm Cannulated Screws & washers are manufactured from high strength 6-4 Alloyed Titanium and have the TIODIZE II surface preparation. The intended use of the I.T.S. Pelvic Reconstruction System (PRS) is to INTENDED USE: stabilize one or more pelvic bone fractures in the pelvic ring area of the pelvis of an adult or pediatric patient which include the use of the following pelvic fracture fixation systems. Indications for use of the I.T.S. PRS Low Profile - Multiple Type_ Pelvic Plating System include: 1) Fracture reconstruction of the acetabulum, sacrum, illium, and entire pelvic ring. 2) Revision surgery of pseoduarthroses, non-unions and mal-unions, 3) Ilio-Iliac distance osteosynthesis, 4) Osteotomies, 5) Arthrodesis. 6) Sacroiliac joint dislocations, and 7) Symphysis pubis disruptures Indications for use of the I.T.S. PRS Sacral Rod System include: 1) Fixation of fractures of the posterior pelvis, 2) Fixation of fractures of the posterior iliac spine, 3} Fixation of fractures of the posterior inferior iliac spine, 4) Dorsal stabilization of the posterior pelvic ring for unstable pelvic ring injuries, 5) Fixation of sacral fractures, and 6) Fracture dislocations of the sacro-iliac joint Indications for use of the I.T.S. PRS 7.3mm Cannulated SIJ Traction Screw & Washer include fracture fixation of pelvic bone fractures where indicated and for periacetabular osteotomies. The system(s) has not been studied in spinal use, and is not intended for use in vertebral column fracture or fusion procedures. The I.T.S. Pelvic Reconstruction System (PRS) is substantially BASIS OF SUBSTANTIAL equivalent to the Synthes, Zimmer, and Stryker pelvic reconstruction EQIVLAENCE: systems and the I.T.S. GmbH stabilizing bone plating systems. The I.T.S. Pelvic Reconstruction System (PRS) is shown to be safe and SUMMARY OF SAFETY effective for use in 'pelvic ring' bone fracture fixation of the pelvis. AND EFFECTIVENESS:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
I.T.S Implantat-Technologie-Systeme GmbH % Mr. Al Lippincott Engineering Consulting Services, Inc. 3150 E. 200th Street Prior Lake, Minnesota 55372
DEC 2 2 2006
Re: K063166
Trade/Device Name: Pelvic Reconstruction System (PRS) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, JDW, HWC Dated: October 13, 2006 Received: October 18, 2006
Dear Mr. Lippincott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
### Page 2 – Mr. Al Lippincott
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0210 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Barbag Buchnd
to
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with the letters "U.T.S." in a bold, sans-serif font, with a right-pointing arrow extending from the right side of the "S". The letters are arranged horizontally, and the arrow is stylized with a gradient effect. Below the logo, there is a textured or patterned background that appears to be composed of small dots or pixels, creating a visually noisy effect.
# Indications for Use
510(k) NUMBER: K063166
#### DEVICE NAME: PELVIC RECONSTRUCTION SYSTEM (PRS)
The intended use of the I.T.S. Pelvic Reconstruction System (PRS) is to stabilize one or more pelvic bone fractures in the 'pelvic ring' area of the pelvis of an adult or pediatric patient which include the use of the following pelvic fracture fixation systems.
Indications for use of the I.T.S. PRS Low Profile - Multiple Type - Pelvic Plating System include:
- 1) Fracture reconstruction of the acetabulum, sacrum, illium, and entire pelvic ring,
- 2) Revision surgery of pseoduarthroses, non-unions and mal-unions,
- 3) Ilio-Iliac distance osteosynthesis,
- 4) Osteotomies,
- 5) Arthrodesis,
- 6) Sacroiliac joint dislocations, and
- 7) Symphysis pubis disruptures
Indications for use of the I.T.S. PRS Sacral Rod System include:
- 1) Fixation of fractures of the posterior pelvis,
- 2) Fixation of fractures of the posterior iliac spine,
- 3) Fixation of fractures of the posterior inferior illac spine,
- 4) Dorsal stabilization of the posterior pelvic ring for unstable pelvic ring injuries,
- 5) Fixation of sacral fractures, and
- 6) Fracture dislocations of the sacro-iliac joint
Indications for use of the I.T.S. PRS 7.3mm Cannulated SIJ Traction Screw & Washer include fracture fixation of pelvic bone fractures where indicated and for periacetabular osteotomies.
The system(s) has not been studied in spinal use, and is not intended for use in vertebral column fracture or fusion procedures.
Prescription Use X AND/OR Over-The-Counter-Use *_*_
(Per 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Barbara Buchen
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of General. Restorative.
**510(k) Number** K063166
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.