K173311 · W.O.M. World of Medicine GmbH · HRX · May 9, 2018 · Orthopedic
Device Facts
Record ID
K173311
Device Name
ARTHRO-Pump PA304
Applicant
W.O.M. World of Medicine GmbH
Product Code
HRX · Orthopedic
Decision Date
May 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ARTHRO-Pump PA304 is an arthroscopic pump intended to provide fluid distension and irrigation of knee, shoulder, hip, elbow, ankle and wrist joint cavities and fluid aspiration during diagnostic and operative arthroscopic procedures.
Device Story
ARTHRO-Pump PA304 is a microprocessor-controlled single-roller peristaltic pump for arthroscopic procedures. It transports sterile irrigation fluid to joint cavities and provides fluid aspiration. System components include power supply, touch display, roller wheel, pump head, and casing. Used with specific tube sets (irrigation, suction, vacuum). Features RFID transponder technology for tube set recognition and reuse prevention. Includes touch-sensitive display with presets for various joint types. Incorporates a pressure sensor to maintain stable pressure ranges and compensate for pressure changes. Operated by clinical staff in surgical settings. Output provides controlled joint distension and fluid management, assisting surgeons in maintaining visibility and joint access during arthroscopy.
Clinical Evidence
Bench testing only. 36 test runs conducted to verify stable pressure generation and pressure compensation capabilities. No clinical data presented.
Technological Characteristics
Microprocessor-controlled single-roller peristaltic pump. Materials: tube sets validated per ISO 10993-1, -5, -10, -11. Sterilization: Ethylene oxide per ISO 11135-1, ISO 14937, ISO 10993-7. Connectivity: RFID transponder for tube set recognition. Electrical safety: AAMI ANSI ES60601-1, IEC 60601-1-2. Software: IEC 62304 compliant.
Indications for Use
Indicated for patients undergoing diagnostic or operative arthroscopic procedures of the knee, shoulder, hip, elbow, ankle, or wrist joint cavities requiring fluid distension, irrigation, and aspiration.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
May 9, 2018
W.O.M. World of Medicine GmbH Soeren Markworth Head of Regulatory Affairs Salufer 8 10587 Berlin, Germany
Re: K173311
Trade/Device Name: ARTHRO-Pump PA304 Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: April 9, 2018 Received: April 12, 2018
Dear Soeren Markworth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K173311
Device Name
ARTHRO-Pump PA304
Indications for Use (Describe)
The ARTHRO-Pump PA304 is an arthroscopic pump intended to provide fluid distension of knee, shoulder, hip, elbow, ankle and wrist joint cavities and fluid aspiration during diagnostic and operative arthroscopic procedures.
| XJ Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--|
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995.<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | |
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
{3}------------------------------------------------
# 510(k) SUMMARY OF SAFETY & EFFECTIVENESS
## (Content in accordance with 21 CFR §807.92)
### 1. General Information
| Submitter: | W.O.M. WORLD OF MEDICINE GmbH |
|----------------------|----------------------------------------------------------------------------------------|
| | Salufer 8 |
| | 10587 Berlin |
| | Germany |
| Registration Number: | 3001556604 |
| Contact Person: | Dr. Soeren Markworth<br>Head of Regulatory Affairs |
| Date prepared: | Phone: +4930-399 81-594<br>Fax: +4930-399 81-593<br>E-mail: soeren.markworth@wom.group |
| Date prepared: | 6th of April, 2018 |
## 2. Proposed Device
| Trade Name: | ARTHRO-Pump PA304 |
|----------------------|----------------------|
| Common Name: | Pump for Arthroscopy |
| Classification Name: | Arthroscope |
| Regulation Number: | 21 C.F.R. § 888.1100 |
| Regulatory Class: | II |
| Product Code: | HRX |
{4}------------------------------------------------

| | K173311 |
|-------------------------------|---------|
| ARTHRO-Pump PA304 | |
| 510(k) Premarket Notification | |
### 3. Predicate Device
| Trade Name: | Arthroscopy Pump A107 |
|---------------------|-----------------------|
| 510(k) Number: | K030402 |
| Classification Name | Arthroscope |
| Regulation Numer | 21 C.F.R. § 888.1100 |
| Regulatory Class | II |
| Product Code | HRX |
#### 4. Reference Device
| Trade Name: | LAP-Pump PP110 |
|---------------------|-----------------------------------------|
| 510(k) Number: | K163320 |
| Classification Name | Gynecologic laparoscope and accessories |
| Regulation Numer | 21 C.F.R. § 884.1720 |
| Regulatory Class | II |
| Product Code | HET |
#### 5. Device Description
The ARTHRO-Pump PA304 is a microprocessor controlled single roller pump that functions according to the peristaltic principle. It transports sterile irrigation fluid to distend cavities and provides fluid aspiration for arthroscopic procedures. The pump consists of the following main components: a power supply, a power switch, a touch display, a connection for the suction function, a roller wheel, a pump head and a casing. The pump has to be used with the following tube sets:
- . Standard Irrigation Tube Set (also referred to as "ST261")
- . Suction Tube Set (one connection, "I"-shape; also referred to as "ST287")
- Suction Tube Set (two connections, "Y"-shape; also referred to as "ST282") .
- Vacuum Tube Set (also referred to as "ST291"). .
{5}------------------------------------------------
### 6. Intended Use
The ARTHRO-Pump PA304 is an arthroscopic pump intended to provide fluid distension and irrigation of knee, shoulder, hip, elbow, ankle and wrist joint cavities and fluid aspiration during diagnostic and operative arthroscopic procedures.
### 7. Comparison of Technological Characteristics
The Arthroscopy Pump A107 is the predicate device for the ARTHRO-Pump PA304. Both pumps are designed, developed and manufactured by W.O.M. WORLD of MEDICINE GmbH. They have the same intended use and incorporate the same basic design. Specifically, both the ARTHRO-Pump PA304 and the Arthroscopy Pump A107 are pumps for use during arthroscopic procedures to provide fluid irrigation and aspiration. They are both roller pumps that function according to the peristaltic principle and are to be used with specially designed tube sets.
The differences in the technological characteristics of both the proposed device ARTHRO-Pump PA304 and the predicate device Arthroscopy Pump A107 are minor and do not raise new questions of safety and effectiveness. The main differences between both devices are the following:
- · The suction function is achieved by different technological means. The ARTHRO-Pump PA304 is a pump with one roller wheel for irrigation and incorporates a suction pump in order to provide fluid aspiration. The Arthroscopy Pump A107 has two roller wheels, one for irrigation and one for aspiration. Both the suction pump of the proposed device and the second roller wheel of the predicate device have the same function, which is to provide negative pressure on the outflow side of the arthroscope.
- . The ARTHRO-Pump PA304 is designed with a radio frequency identification (RFID) transponder technology tube set recognition function that controls the proper position of the Inflow Tube Set and eliminates accidental reuse of an inflow tube set by invalidating the tube set after the irrigation process has been started.
{6}------------------------------------------------
- The ARTHRO-Pump PA304 is designed with a touch sensitive display and offers . presets for knee, hip, shoulder or small joints arthroscopies.
- . Finally, the ARTHRO-Pump PA304 incorporates only one pressure sensor in place of the redundant pressure measurement of the Arthroscopy Pump A107.
#### 8. Performance Data
#### Software
The software was developed, tested, and verified in accordance with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and in accordance with the following standard:
- . IEC62304:2006 - Medical Device Software - Software Life Cycle Processes.
Design verification testing of the ARTHRO-Pump PA304 demonstrates that the device performs as intended and that the performance does not raise new questions of safety and effectiveness.
#### Electrical safety and electromagnetic compatibility
Electrical safety and electromagnetic compatibility testing was performed in accordance to the below standards:
- . AAMI ANSI ES60601-1: 2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance;
- IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General Requirements . for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbances - Requirements and Tests;
#### Biocompatibility
Biocompatibility testing was performed on the tube sets of both the proposed device and predicate device in accordance with:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for WOM, which is a medical device company. Below the logo is the text "ARTHRO-Pump PA304". The text "510(k) Premarket Notification" is below the model number. The logo is a circle with a stylized shape inside, with the bottom portion of the shape colored in pink.
- . ISO 10993-1 - Biological evaluation of medical devices- Evaluation and testing within a risk management system;
- . ISO 10993-5 - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity; and
- . ISO 10993-10 - Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization;
- . ISO 10993-11 - Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity.
#### Sterilization
In addition, sterilization validation on the tube sets of both devices has been performed in accordance with:
- · ISO 11135-1:2015 Sterilization of health care products Ethylene oxide Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices;
- ISO 14937:2009 Sterilization of health care products General criteria for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices; and
- ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals.
Residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) data shows that the limit of EO < 4 mg and ECH < 5 mg after 10 days of aeration (gas release) that remain on the tube set will not be exceeded. The sterility assurance level (SAL) was ≤ 10t%. Package and product integrity of the tube sets were tested in accordance with ISO11607-1 - Packaging for terminally sterilized medical devices and ASTM-F- 1980:2002 - Standard for accelerated aging of sterile medical device packages.
{8}------------------------------------------------
#### Performance Testing - Bench
Finally, bench testing was performed to demonstrate that the proposed device ARTHRO-Pump PA304 is safe and effective. In total 36 test runs have been conducted. The test results demonstrate that the ARTHRO-Pump PA304 generates a stable pressure in a predefined pressure range and to compensate for changes in pressure.
#### 9. Conclusion
Based on the similar intended use, the same basic technological characteristics and performance testing, the ARTHRO-Pump PA304 is substantially equivalent to the predicate device Arthroscopy Pump A107 (K030402). The differences between the proposed device and the predicate device do not raise new questions of safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.