Optima Coil System

K172390 · Blockade Medical, LLC (D.B.A. Balt Usa) · HCG · Feb 18, 2018 · Neurology

Device Facts

Record IDK172390
Device NameOptima Coil System
ApplicantBlockade Medical, LLC (D.B.A. Balt Usa)
Product CodeHCG · Neurology
Decision DateFeb 18, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5950
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.

Device Story

Optima Coil System consists of platinum/tungsten embolization coils attached to a delivery pusher; used for endovascular embolization of intracranial aneurysms and other neurovascular/peripheral vascular abnormalities. Device inserted into vasculature under angiographic visualization; coils placed into aneurysm to create blood stasis and induce thrombosis. Coils detached from delivery pusher via thermal mechanism activated by XCEL Detachment Controller. Used in clinical settings by physicians; output is physical occlusion of target vessel/aneurysm. Benefits include permanent obstruction of blood flow to vascular malformations. System includes various coil shapes, lengths, and sizes.

Clinical Evidence

No human clinical data provided. Evidence consists of bench testing (corrosion resistance, advancement/retraction force, MRI compatibility, detachment temperature characterization, simulated use, usability, particulate characterization, design verification) and an acute porcine animal model study evaluating in vivo performance (introduction, tracking, deployment, reposition, detachment). All tests passed.

Technological Characteristics

Implant: Platinum/tungsten alloy coils. Delivery system: Stainless steel (SSTL) core wire, hypotube, and gold-plated connector. Detachment: Thermal mechanism via XCEL Detachment Controller. Biocompatibility: ISO 10993 compliant. Sterilization: Gamma (coils) and EO (controller). MRI: MR Conditional. Connectivity: None (standalone).

Indications for Use

Indicated for endovascular embolization of intracranial aneurysms, arteriovenous malformations, and arteriovenous fistulae; vascular occlusion of neurovascular blood vessels to obstruct blood flow to aneurysms or malformations; and arterial/venous embolizations in peripheral vasculature.

Regulatory Classification

Identification

A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.

Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The word "Administration" is on the second line. February 18, 2018 Blockade Medical, LLC (d.b.a. Balt USA) Rebecca K. Pine Official Correspondent 18 Technology Drive, Suite 169 Irvine, California 92618 Re: K172390 Trade/Device Name: Optima Coil System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: January 18, 2018 Received: January 19, 2018 Dear Rebecca K. Pine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Image /page/1/Picture/6 description: The image contains the text "Carlos L. Pena -S FDA". The text is arranged horizontally, with "Carlos L. Pena" appearing first, followed by "-S", and then "FDA". The text is in a sans-serif font and is black. The background is white. Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K172390 Device Name Optima Coil System #### Indications for Use (Describe) The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------| | <div> <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### Optima Coil System 510(k) Summary This 510(k) summary for Optima Coil System is submitted in accordance with the requirements of 21 CFR 807.87(h) and 807.92 and following the recommendation outlined in FDA Guidance. The 510(k) Program: Evaluating substantial Equivalence in Premarket Notification [510(k)], dated 28 July, 2014. #### SUBMITTER [807.92(a)(1)] Blockade Medical, LLC (d.b.a. Balt USA) 18 Technology Dr. Ste 169 Irvine, CA 92618 Contact person: Rebecca K Pine Phone: (760) 809-5178 Date prepared: February 12, 2018 ## DEVICE [807.92(a)(2)] Name of the device: Optima Coil System Common of usual name: Neurovascular embolization device Classification name: Neurovascular embolization device Vascular embolization device Regulatory Class: Class II Product Code: HCG KRD Submission Type: Traditional 510(k) Regulation Number: 21 CFR 882.5950 Reviewing Product Branch: Division of Neurological and Physical Medicine Devices (Office of Device Evaluation, CDRH) ### PREDICATE DEVICE [807.92(a)(31)] Barricade Embolization Coil System (K151760) ### DEVICE DESCRIPTION [807.92(a)(4)] The Optima Coil System is a series specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum/tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are thermally detached from the delivery pusher in serial manner until the aneurysm is occluded. {4}------------------------------------------------ ## INDICATIONS FOR USE [807.92(a)(5)] The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE [807.92(a)(6)] The technological characteristics of the Optima Coil System is highly analogous to the technological characteristics of the Barricade Embolization Coil System previously cleared (K151760) version of the device, as shown below: | | Barricade<br>Embolization Coil<br>System (K151760)<br>(Predicate Device) | Optima Coil System<br>(Subject Device) | Effect on<br>substantial<br>equivalence | |--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Principle of Operation | Facilitates<br>endovascular<br>embolization of<br>intracranial<br>aneurysms and other<br>vascular<br>abnormalities | SAME | None. Identical | | System components | Coil (implant)<br>Delivery system<br>Detachment<br>controller | SAME | None. Identical | | Coil delivery mechanism | Pusher | SAME | None. Identical | | Method<br>of<br>supply<br>(coil/delivery system) | Sterile, single use | SAME | None. Identical | | | Coil (implant) | | | | Main Coil Material | Platinum/Tungsten<br>alloy | SAME | None. Identical | | Primary Coil Diameter | 0.010"-0.014" | SAME | None. Identical | | Coil Secondary diameter | 1.0mm - 15mm | 1mm-24mm | None. An<br>increase in<br>diameter range has<br>no effect on the<br>intended<br>performance of the<br>device. | | Coil Wire Diameter | 0.00125"-0.003" | 0.00125"-0.0035" | Minor increase<br>(0.0005") in<br>diameter range has | | | | | no effect on the<br>intended<br>performance of the<br>device | | Secondary Shapes | Complex/Helical | SAME | None. Identical | | Coil Types | Framing, Filling,<br>Finishing | Framing, Filling,<br>Finishing (standard, soft,<br>super soft, 18) | None. Identical.<br>Nomenclature has<br>merely been<br>updated | | Coil length | 1cm – 50cm | 1cm – 65cm | None. An<br>increase in length<br>range has no effect<br>on the intended<br>performance of the<br>device. | | Stretch<br>resistance/attachment<br>thread | .0022” Polyolefin<br>Engage thread<br>.0009” PET thread | .0022” Polyolefin Engage<br>thread | None. Minor<br>change to<br>accommodate<br>thermal<br>detachment<br>mechanism | | Coupler | N/A | 90%/10% Pt/Ir<br>Markerband | None. Minor<br>change to<br>accommodate<br>thermal<br>detachment<br>mechanism | | Detachment Zone | .020" (nominal) | 0.050" (nominal) | None. Minor<br>change to<br>accommodate<br>thermal<br>detachment<br>mechanism | | Delivery System | | | | | Construction/Design | SSTL core wire | Body coil laser welded to<br>hypotube | None. Minor<br>change to<br>accommodate<br>thermal<br>detachment<br>mechanism.<br>Change has no<br>effect on the<br>intended<br>performance of the<br>device. | | Body coil | RO Coil (92/8 Pt/W) | 4-part coil<br>A. Heater coil (92/8 Pt/W)<br>B. Distal Coil (SSTL)<br>C. Radio-opaque (RO) Coil (92/8 Pt/W) | None. Minor<br>change to<br>accommodate<br>thermal<br>detachment | | | | D. Proximal Coil (SSTL) | mechanism.<br>Change has no<br>effect on the<br>intended<br>performance of the<br>device. | | Hypotube | N/A | SSTL hypotube | None. Minor<br>change to<br>accommodate<br>thermal<br>detachment<br>mechanism.<br>Change has no<br>effect on the<br>intended<br>performance of the<br>device. | | Connector | N/A | Gold plated, SSTL<br>hypotube | None. Minor<br>change to<br>accommodate<br>thermal<br>detachment<br>mechanism.<br>Change has no<br>effect on the<br>intended<br>performance of the<br>device. | | Adhesive | Dymax 1128A-M | Dymax 1128A-M-VT | None. Minor<br>change in adhesive<br>viscosity does not<br>affect the intended<br>performance of the<br>device. | | Jacket | PET | Same | None. Identical | | Flouro safe markers | Pad Printed PET<br>Shrink tube | Same | None. Identical | | Epoxy | N/A | Epoxy 353 ND | None. Minor<br>change to<br>accommodate<br>thermal<br>detachment<br>mechanism.<br>Change has no<br>effect on the<br>intended<br>performance of the<br>device. | | Lead wires | N/A | Polyimide coated silver<br>lead wires | None. Minor | | | | accommodate<br>thermal<br>detachment<br>mechanism.<br>Change has no<br>effect on the<br>intended<br>performance of the<br>device. | | | Heater coil coating | N/A | Polyimide coating | None. Minor<br>change to<br>accommodate<br>thermal<br>detachment<br>mechanism.<br>Change has no<br>effect on the<br>intended<br>performance of the<br>device. | | Detachment Controller | | | | | Coil detachment | Electrolytic via<br>detachment controller | Thermal via detachment<br>controlled | None. Change in<br>detachment<br>method does not<br>alter the intended<br>performance of the<br>device. | {5}------------------------------------------------ {6}------------------------------------------------ {7}------------------------------------------------ The implant segment detaches via a thermal detachment mechanism upon activation of the XCEL Detachment Controller. The predicate device, Barricade Coil System implant segment detaches via an electrolytic detachment mechanism upon activation of the detachment controller. The device is designed to be deployed and detached in the neuro and peripheral vasculature to permanently obstruct blood flow. The difference in detachment mechanism does not alter the intended performance of the device; therefore, the change in detachment mechanism does not affect the safety, effectiveness, and benefit/risk profile of the Optima Coil System. ### PERFORMANCE DATA [807.92(b)] Performance Bench Testing and Animal Testing: Results of the performance benching testing and animal testing indicate that Optima Coil System meets established performance requirements, and is substantially equivalent for its intended use. | Performance Bench Testing | | | |-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------| | Test | Test Method Summary | Results | | Corrosion Resistance | The deployed coil was placed in a<br>sodium phosphate buffered saline<br>solution and the resting potential<br>(Er) is recorded for one (1) hour in | All test samples passed testing. | | | an open circuit configuration and allowed to plateau. | | | Advancement and<br>Retraction Force | The test represents the maximum force required to advance and retract the coil through the microcatheter | All test samples passed testing. | | MRI Compatibility | SAR patterns and temperature rises were evaluated under MRI conditions using RF coils at 64-MHz and 128 MHz and to determine the magnetic field interactions, heating, and artifacts. for the Optima Coil System. The SEMCAD software package was used to evaluate surface heating patterns for the Optima Coil when placed inside the gel of the ASTM phamtom under 1.5T and 3-T MRI systems. | All test samples passed testing. MR Conditional. | | MRI Artifact | The MRI artifact was evaluated under predetermined magnetic resonance angiography (MRA) imaging parameters. | All test samples passed testing. | | Detachment<br>Temperature<br>Characterization | The temperature characterization of the Optima Coil System's delivery pusher during the detachment cycle was evaluated by comparing the temperature of the delivery pusher to commercially available thermal detachment coils. | All test samples passed testing. | | Simulated Use testing | Simulated use testing is to demonstrate that the device will meet the requirements of the product specifications and to demonstrate that the device will meet the product specifications requirements when tested in worst case tortuosity vessel. | All test samples passed testing. | | Usability - XCEL<br>Detachment Controller | The clinical results of the usability of XCEL Detachment Controller were evaluated in accordance to IEC 60601-1-6:2010 + AMD1:2013 and IEC 62366-1:2015. | All test samples passed testing. | | Particulate<br>Characterization | Particulate matter in injections of the device were quantified after advancement/retraction procedures. | All test samples passed testing. | | Design Verification and | This test is to evaluate the device | All test samples passed testing. | | Packaging Validation | design and packaging design and<br>to demonstrate that the device will<br>meet the product specification<br>requirements at t=0-year time-<br>point after exposing to 1x Gamma<br>sterilization. Bubble immersion,<br>seal peel strength, simulated use,<br>detachment, SR tensile test and DZ<br>tensile strength testing were<br>performed. | | | Performance Animal Testing | | | | Animal Testing (GLP) | Animal testing is to evaluate the <i>in vivo</i> performance of the Optima<br>Coil System and XCEL<br>Detachment Controller in an acute<br>porcine model. Performance<br>metrics such as Introduction,<br>Tracking, Deployment, Reposition,<br>and Detachment within an | All test samples passed testing. | {8}------------------------------------------------ {9}------------------------------------------------ # Biocompatibility: | Biocompatibility Test | Results | Conclusion | |---------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------| | Optima Coil Implant | | | | Cytotoxicity - MEM Elution<br>(GLP) - 72 hour extract<br><br>ISO 10993-5 | The test article scored '0' at 24,<br>48 and 72±4 hours | Non-cytotoxic | | Sensitization - ISO Guinea Pig<br>Maximization Sensitization Test<br>(GLP-2 extract)<br><br>ISO 10993-10 | None of the extracts of the test<br>article exhibited a sensitization<br>response greater than '0'. | Did not elicit sensitization<br>response | | Irritation or Intracutaneous<br>irritation reactivity - ISO<br>Intracutaneous Irritation Test<br>(GLP-2 Extracts)<br><br>ISO 10993-10 | The difference in the mean test<br>and control scores of the extract<br>dermal observations were less<br>than 1.0. | Non-irritant | | Acute systemic toxicity - ISO<br>Acute Systemic Injection Test<br>(GLP-2 Extracts)<br><br>ISO 10993-11 | None of the test article treated<br>animals were observed with<br>clinical signs consistent with<br>toxicity at any of the<br>observation periods. | No signs of toxicity | | Acute systemic toxicity - ISO<br>Material Mediated Rabbit<br>pyrogen (GLP) | None of the test article extracts<br>had a temperature rise > 0.5 °C<br>at the required observation time<br>points. | Non-pyrogenic | | ISO 10993-11 | | | | Hemocompatibility - ASTM<br>Hemolysis Assay – Direct<br>Contact and Extract Method<br>(GLP) | Based on the validity of the<br>assay, the test article is<br>considered non-hemolytic under<br>the test conditions employed. | Non-hemolytic | | ISO 10993-4 | | | | Hemocompatibility -<br>Prothrombin Time (PT) -GLP | No statistically significant<br>difference was observed<br>between the plasma exposed to<br>the test article, predicate and<br>control. | No adverse effect on<br>prothrombin coagulation | | ISO 10993-4 | | | | Hemocompatibility -<br>Complement Activation SC5b-9<br>Assay with Sponsor-Supplied<br>Comparison Article (GLP) | When compared to the reference<br>control data the test article and<br>comparison article results for the<br>SC5b9 assay showed no<br>statistically significant<br>difference between the results<br>(p<0.05). | Exhibited activation not<br>significant/Passed | | ISO 10993-4 | | | | Genotoxicity - ISO In Vitro<br>Mouse Lymphoma with<br>Extended Treatment (GLP) | No test article mutant frequency<br>was found to be significantly<br>different than concurrent<br>negative controls according to<br>statistical analysis. No test<br>article result exceeded the GEF<br>in any treatment. The test article<br>is considered to be non-<br>mutagenic and non-clastogenic<br>in the test system. | Non-mutagenic and non-<br>clastogenic | | ISO 10993-3 | | | | Genotoxicity ISO Bacterial<br>Mutagenicity Test – Ames assay<br>(GLP-4 Salmonella Strains and<br>1 E. Coli Strain - 2 Extracts) | The test article did not induce<br>substantial increases in<br>reversion rates of the type that<br>are associated with mutagenesis.<br>Furthermore, no substantial<br>test article toxicity was noted that<br>may have interfered with the<br>ability of the test system to detect<br>mutagens. | Non-mutagenic | | ISO 10993-3 | | | | Genotoxicity - ISO In Vivo<br>Mouse Micronucleus Assay<br>(GLP-2 Extracts) | There were no apparent gross<br>manifestations of toxicity nor<br>biologically significant<br>erythropoietic disturbances<br>resulting in delayed<br>mutagenesis. Furthermore, there<br>were no biologically significant<br>increases in mPCE production in<br>the test article treated groups as<br>compared to the concurrent<br>negative controls | Non-mutagenic | | ISO 10993-3 | | | | Implantation - ISO<br>Intramuscular Implantation Test<br>with Histopathology – 4 Week –<br>3 Rabbits (GLP)<br>ISO 10993-6 | The average irritation score of<br>the test article is 6.7 whereas the<br>average irritation score of the<br>control article is 6.3. | When comparing the irritation<br>score of the test article to the<br>control article, the irritation<br>scores are found to be<br>comparable.<br>Pass | | Implantation - ISO<br>Intramuscular Implantation with<br>Histopathology<br>– 13 Week – 4 Rabbits<br>ISO 10993-6 | The average irritation score of<br>the test article is 8.8 whereas the<br>average irritation score of the<br>control article is 9.3. | When comparing the score of<br>the test article to the control<br>article, the irritation scores are<br>found to be comparable.<br>Pass | | Carcinogenicity<br>ISO 10993-18<br>ISO 10993-17 | Toxicological Risk Assessment | Non-carcinogenic | | Subacute/Subchronic Tooxicity<br>ISO 10993-11 | Toxicological Risk Assessment | Pass | | ChronicToxicity<br>ISO 10993-11 | Toxicological Risk Assessment | Pass | | Delivery Pusher | | | | Cytotoxicity - ISO MEM<br>Elution Using L-929 Mouse<br>Fibroblast Cells (GLP)<br>ISO 10993-5 | The test article scored '0' at 24,<br>48, and 72 + 4 hours | Non-cytotoxic | | Sensitization - ISO Guinea Pig<br>Maximization Sensitization Test<br>(GLP-2 extract)<br>ISO 10993-10 | None of the animals challenged<br>with the test article extracts<br>were observed with a<br>sensitization response greater<br>than '0'. | Did not elicit sensitization<br>response | | Irritation or Intracutaneous<br>irritation reactivity - ISO<br>Intracutaneous Irritation Test<br>(GLP-2 Extracts)<br>ISO 10993-10 | The differences in the mean test<br>and control scores of the extract<br>dermal observations were less<br>than 1.0 | Non-irritant | | Acute systemic toxicity - ISO<br>Acute Systemic Injection Test<br>(GLP-2 Extracts)<br>ISO 10993-11 | None of the test article treated<br>animals were observed with<br>clinical signs consistent with<br>toxicity at any of the<br>observation periods. | No signs of toxicity | | Acute systemic toxicity - ISO<br>Materials Mediated Rabbit | None of the e test article extracts<br>had a temperature rise > 0.5 0C | Non-pyrogenic | | ISO 10993-11 | | | | Hemocompatibility<br>Hemolysis Assay - Direct<br>Contact and Extract Method<br>(GLP) | Based on the validity of the<br>assay, the test article is<br>considered non-hemolytic under<br>the test conditions employed. | Non-hemolytic | | ISO 10993-4 | | | | Hemocompatibility<br>Complement Activation - SC5b-<br>9 Assay with Sponsor-Supplied<br>Comparison Article (GLP) | The SC5b-9 assay results for the<br>reference control and the results<br>of test article were not<br>statistically significant ( $p>0.05$ )<br>and is considered satisfactory<br>under the test conditions<br>employed. The SC5b-9 assay<br>results for the comparison article<br>were statistically significantly<br>( $p<0.05$ ) lower than the<br>reference control data. | Exhibited activation not<br>significant | | ISO 10993-4 | | | | Hemocompatibility -<br>Prothrombin Time (PT) – GLP | No statistically significant<br>difference was observed<br>between the plasma exposed to<br>the test article, predicate and<br>control. | No adverse effect on<br>prothrombin coagulation | | ISO 10993-4 | | | | Hemocompatibility -<br>Thromboresistance Evaluation<br>(GLP -4 Hour – 2 Dog) | Implantation of the test and<br>control devices in the jugular<br>veins of two canines resulted in<br>no adverse effects or clinical<br>signs. | Thrombus formation not<br>significant/Passed | | ISO 10993-4 | | | | Hemocompatibility - ISO in<br>Vitro Mouse Lymphoma with<br>Extended Treatment (GLP) | No test article mutant frequency<br>exceeded the GEF or was<br>significantly different from<br>concurrent negative control<br>results at a 95% confidence<br>interval. | Non-mutagenic and non-<br>clastogenic | | ISO 10993-3 | | | | Genotoxicity - ISO Bacterial<br>Mutagenicity Test – Ames assay<br>(GLP-4 Salmonella Strains and<br>1 E. Coli Strain - 2 Extracts) | The test article did not induce<br>substantial increases in<br>reversion rates of the type that<br>are associated with mutagenesis. | Non-mutagenic | | ISO 10993-3 | | | {10}------------------------------------------------ {11}------------------------------------------------ {12}------------------------------------------------ # Sterilization: | Test | Test Method Summary | Results | |---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------| | Sterilization –<br>Optima Coil System | To establish the sterilization dose and a routine sterilization process and to validate the sterilization process to achieve an SAL of 10-6 for the Optima | All test samples passed testing. | {13}------------------------------------------------ | | Coil System. | | |-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------| | | Bioburden, dose verification,<br>Sterility, and B&F testing were<br>performed. | | | Sterilization –<br>XCEL Detachment Controller | To establish the sterilization<br>cycle and a routine sterilization<br>process and to validate the<br>sterilization process to achieve<br>an SAL of 10-6 for the XCEL<br>Detachment Controller.<br><br>Bioburden, EO and ECH<br>residual, and sterility testing<br>were performed. | All test samples passed testing. | | Bacterial Endotoxin (LAL) | Fully packaged devices were<br>assembled and packaged using<br>approved materials defined for<br>the finished product and<br>sterilized per the documented<br>sterilization process. Samples<br>were tested for<br>Inhibition/Enhancement Assay<br>and Kinetic chromogenic LAL<br>Bacterial Endotoxin testing | All test samples passed testing. | # Shelf Life and Packaging: | Test | Test Method Summary | Results | |------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------| | Shelf Life- Optima Coil System | Test samples were subjected to<br>one (1) time Gamma<br>sterilization cycle at the<br>maximum dose range,<br>accelerated aged for the<br>equivalent of 2 years,<br>environmentally conditioned per<br>ASTM F1980-16 and subjected<br>to transportation simulation per<br>ASTM 4169-16. | All test samples passed testing. | | Shelf Life-XCEL Detachment<br>Controller | Test samples were subjected to<br>EO sterilization,<br>environmentally conditioned<br>and subjected to transportation<br>simulation per ASTM 4169.<br>Subsequent to the transport<br>simulation, packaging tests were<br>performed to evaluate the<br>integrity and seal strength of the<br>sterile barrier system | All test samples passed testing. | {14}------------------------------------------------ #### CONCLUSIONS The Optima Coil System met all specified criteria and did not raise new safety or performance questions. Based on the 510(k) summary and information provided herein, we conclude the subject device, Optima Coil System, is substantially equivalent in its intended use, design, material, performance, and the underlying fundamental scientific technology used, to the predicate device, Barricade Coil System (K151760).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...