GUGLIELMI DETACHABLE (GDC)

K991139 · Boston Scientific, Target · HCG · Dec 22, 1999 · Neurology

Device Facts

Record IDK991139
Device NameGUGLIELMI DETACHABLE (GDC)
ApplicantBoston Scientific, Target
Product CodeHCG · Neurology
Decision DateDec 22, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5950
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Guglielmi Detachable Coil (GDC) is intended for embolizing certain intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be: a)very high risk for management by traditional operative techniques, or, b) inoperable, and for embolizing other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neuro vasculature. The GDC is also intended for arterial and venous embolizations in the peripheral vasculature. Fibered GDC is intended for the embolization of vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neuro vasculature. Fibered GDC is also intended for arterial and venous embolizations in the peripheral vasculature.

Device Story

GDC system comprises battery-operated power supply, connecting cables, patient return electrode, and detachable platinum occlusion coils. Used in neurosurgical/interventional procedures to treat intracranial aneurysms and vascular malformations. Physician navigates coil to target site; power supply initiates electrolytic detachment of coil from delivery wire via controlled DC current. System provides real-time monitoring of current, voltage, and elapsed time; includes indicators for poor patient connection and successful detachment. Enables minimally invasive embolization for patients at high risk for or ineligible for traditional surgery.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Battery-operated (two 9-volt) DC power supply unit. Components: platinum wire coils, stainless-steel core delivery wire with Teflon jacket. Features: current/voltage/time displays, connection check indicator, detachment indicator. Connectivity: wired connection to patient electrode and coil delivery wire.

Indications for Use

Indicated for use with Guglielmi Detachable Coils for embolization of intracranial aneurysms, neuro/peripheral vascular malformations (AVMs, AV fistulae), and arterial/venous embolizations in peripheral vasculature. Prescription use only.

Regulatory Classification

Identification

A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.

Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 12/22/99 K991139 #### Section 2 - 510 (k) Summary of Safety and Effectiveness and Class III Certification and Summary #### a. Summary Of Safety And Effectiveness Contact Person Jim Leathley Senior Regulatory Affairs Specialist Boston Scientific / Target 47900 Bayside Parkway Fremont, CA. 94538 Trade Name Guglielmi Detachable Coil (GDC), Class III Common Name Occlusion Coil Classification Name Artificial Embolization Device (21 CFR Section 882.5950) Predicate Devices | Number | Description | Predicate for | Clearance<br>Date | |-------------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------| | K962503<br>(Boston<br>Scientific /<br>Target) | Guglielmi Detachable Coil | All GDC devices which<br>are the subject of this<br>premarket notification | 20 Sept. 1996 | | K971395<br>(Boston<br>Scientific /<br>Target) | Shelf Life, Various Class II<br>and Class III Devices for | For shelf-life validation<br>method and subsequent<br>placement of shelf-life<br>information on device<br>label | 14 July 1997 | | K914786<br>(Boston<br>Scientific /<br>Target) | Platinum Fibered<br>Occlusion Coils | For Dacron / Polyester<br>fiber of Fibered GDC | April 1992 | | K930738<br>(Deknatel,<br>Division of<br>Howmedica,<br>Inc.) | Surgical Sutures | Suture material for<br>Stretch Resistant GDC | 26 July 1994 | 223 510(k) Notification, Boston Scientific/Target, GDC Version 4 March 1999 Section 2, Page 2 of 47 {1}------------------------------------------------ ### Intended Use The Guglielmi Detachable Coil (GDC) is intended for embolizing certain intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be: a)very high risk for management by traditional operative techniques, or, b) inoperable, and for embolizing other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neuro vasculature. The GDC is also intended for arterial and venous embolizations in the peripheral vasculature. Fibered GDC is intended for the embolization of vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neuro vasculature. Fibered GDC is also intended for arterial and venous embolizations in the peripheral vasculature. #### Device Description The GDC system consists of - · GDC power supply - · GDC occlusion coil attached to a delivery wire - · set of GDC connecting cables - · patient return electrode - · two 9-volt batteries each of which is sold separately. The occlusion coil is detached by electrolytically dissolving a small portion of the delivery wire upon desired placement of the coil in the anatomy. GDC occlusion coils are manufactured from platinum wire which is first wound into a primary coil and then formed into a secondary helical shape. Coils are attached to a delivery wire, which consists of a ground stainlesssteel core wire with a stainless-steel coil welded at the distal end and a Teflon® outer jacket. The delivery wire is the same as that employed for the predicate GDC cleared under K962503. 334 {2}------------------------------------------------ The GDC Power Supply is a self-contained, battery-operated unit designed to initiate and control detachment of a GDC coil. It has an on/off button, connections for the GDC cables, and the following controls and displays: · Current setting switch and display - Allows for selection of the different current settings. The selected current (mA) is briefly displayed after which the unit displays the current flowing through the system. · Voltage diplay and indicator - Displays the DC output voltage. • Time display - Displays the elapsed time that the current has been flowing through the GDC system. • Check indicator - Flashes on and off at the beginning of a detachment procedure if there is a poor connection to the patient. · Detach indicator - Flashes on and off when the power supply has detected detachment. Detachment is accompanied by five beeps. · Battery indicator - Illuminates when the voltage of the internal 9-volt batteries falls below specification. 225 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows a logo with a stylized bird in flight, enclosed within a broken circle. The bird is represented by three curved lines that suggest movement and feathers. The circle is incomplete, giving the impression of an open or dynamic form. The logo is rendered in black and white, with a simple and clean design. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 2 2 1999 Mr. Jim Leathley Senior Regulatory Affairs Specialist Boston Scientific/Target 47900 Bayside Parkway Fremont, California 94538 Re: K991139 Trade Name: Guglielmi Detachable Coil (GDC) Power Supply Regulatory Class: III Product Code: HCG and KRD Dated: September 29, 1999 Received: October 1, 1999 Dear Mr. Leathley: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ ## Page 2 - Mr. Jim Leathley If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Neil R.P. Ogden James F. Dillard III James E. Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Boston Scientific Target. The words "Boston Scientific" are stacked on top of each other in a bold, sans-serif font. Below "Scientific" is the word "TARGET" in a larger, bolder, sans-serif font. There is a horizontal line separating "Scientific" and "TARGET". # INDICATIONS FOR USE STATEMENT 510(k) Number: K991139 Device Name: GDC (Guglielmi Detachable Coil) Power Supply #### Indications for Use: Boston Scientific/Target's Guglielmi Detachable Coil (GDC) Power Supply is intended for use with all versions of Boston Scientific/Target's Guglielmi Detachable Coils in the embolization of intracranial anewrysms and other vascular malformations of the neuro and ﺴ peripheral vasculature. | | PROOF for JZD | |-----------------------------------------|---------------| | (Division Sign-Off) | | | Division of General Restorative Devices | | | 510(k) Number | K991139 | OR Concurrence of CDRH, Office of Device Evaluation (ODE) × Prescription Use (Per 21 CFR 801.109) Over The Counter Use _________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...