Penumbra Smart Coil

K160832 · Penumbra, Inc. · HCG · Apr 20, 2016 · Neurology

Device Facts

Record IDK160832
Device NamePenumbra Smart Coil
ApplicantPenumbra, Inc.
Product CodeHCG · Neurology
Decision DateApr 20, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.5950
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Smart Coil is indicated for the embolization of: - Intracranial aneurysms - Other neurovascular abnormalities such as arteriovenous malformations and arterioyenous fistulae - Arterial and venous embolizations in the peripheral vasculature

Device Story

Endovascular embolization coil system; treats aneurysms and vascular abnormalities. Device consists of platinum/tungsten coil, detachment pusher, and detachment handle. Physician delivers coil through vasculature to target site; packs coil to achieve occlusion. Detachment mechanism allows controlled release. Used in clinical settings; disposable, single-use. Modification involves updated detachment pusher pull wire tensile strength specification to improve manufacturability. Benefits patient by providing reliable occlusion of neurovascular and peripheral vascular lesions.

Clinical Evidence

Bench testing only. Design verification included tensile testing (minimum break force) and simulated use testing in an anatomical model representing neurovascular tortuosity to evaluate delivery and detachment performance. All tests passed.

Technological Characteristics

Materials: Platinum/Tungsten, Nitinol, Polymer, Adhesives, Stainless Steel. Components: Coil, Detachment Pusher, Detachment Handle. Dimensions: Coil diameter 1-18 mm, length 1-60 cm; Pusher length 185 cm. Shapes: Complex, Helical. Sterilization: Ethylene oxide (EO). Single-use, disposable.

Indications for Use

Indicated for patients requiring embolization of intracranial aneurysms, neurovascular abnormalities (arteriovenous malformations, arteriovenous fistulae), and arterial/venous embolizations in the peripheral vasculature.

Regulatory Classification

Identification

A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.

Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, stacked on top of each other, with flowing lines beneath them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 20, 2016 Penumbra, Inc. Mr. Charles Denault Regulatory Affairs Specialist III One Penumbra Place Alameda, California 94502 Re: K160832 Trade/Device Name: Penumbra Smart Coil Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: March 24, 2016 Received: March 25, 2016 Dear Mr. Denault: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Carlos L. Pena -S/A Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K160832 Device Name Penumbra Smart Coil™ Indications for Use (Describe) The Smart Coil is indicated for the embolization of: - Intracranial aneurysms - · Other neurovascular abnormalities such as arteriovenous malformations and arterioyenous fistulae - · Arterial and venous embolizations in the peripheral vasculature Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span style="font-size: 20px;">☑</span> </div> | |----------------------------------------------|------------------------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span style="font-size: 20px;">□</span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 1 510(k) Summary, K160832 (as required by 21 CFR 807.92) Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the Penumbra Smart Coil™. #### 1.1 Sponsor/Applicant Name and Address Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA #### 1.2 Sponsor Contact Information Charles DeNault Regulatory Affairs Specialist III Phone: (510) 748-3302 Fax: (510) 217-6414 Email: cdenault@penumbrainc.com #### Date of Preparation of 510(k) Summary 1.3 April 19, 2016 #### Device Trade or Proprietary Name 1.4 Penumbra Smart Coil™ #### Primary Device Classification 1.5 | Regulatory Class: | II | |-----------------------|-----------------------------------| | Classification Panel: | Neurology | | Classification Name: | Neurovascular embolization device | | Regulation Number: | 21 CFR 882.5950 | | Product Code: | HCG | #### 1.6 Secondary Device Classification | Regulatory Class: | II | |-----------------------|------------------------------| | Classification Panel: | Cardiovascular | | Classification Name: | Vascular embolization device | | Regulation Number: | 21 CFR 870.3300 | | Product Code: | KRD | {4}------------------------------------------------ #### 1.7 Predicate Devices | 510(k) Number | Clearance Date | Name of Predicate<br>Device | Name of<br>Manufacturer | |---------------|----------------|-----------------------------|-------------------------| | K143218 | March 18, 2015 | Penumbra Smart Coil | Penumbra, Inc. | | K151572 | July 10, 2015 | Penumbra Smart Coil | Penumbra, Inc. | #### Predicate Comparison 1.8 | | Predicate: Penumbra Smart Coil | Subject: Penumbra Smart Coil | |-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------| | 510(k) No. | K143218, K151572 | K160832 | | Classification | Class II, HCG, KRD | Same | | Indications for Use | Indicated for the embolization of:<br>• Intracranial aneurysms<br>• Other neurovascular abnormalities such<br>as arteriovenous malformations and<br>arteriovenous fistulae<br>• Arterial and venous embolizations in the<br>peripheral vasculature | Same | | Materials | | | | Coil | Platinum/Tungsten, Nitinol, Polymer,<br>Adhesives | Same | | Detachment Pusher | Stainless Steel, Polymer,<br>Platinum/Tungsten, Nitinol | Same | | Detachment Handle | Plastic, Stainless Steel | Same | | Dimensions/Shape | | | | Coil Secondary<br>Diameter | 1-18 mm | Same | | Coil Length | 1-60 cm | Same | | Coil Secondary Shape | Complex, Helical (Curve) | Same | | Pusher Length | 185 cm | Same | | Other | | | | Packaging<br>configurations | Individual | Same | | Sterilization method | Ethylene oxide (EO) | Same | | Shelf life | 5 years | Same | | Use | Single use, disposable | Same | #### 1.9 Device Description The Smart Coil is an endovascular embolization coil for the treatment of aneurysms or other vascular abnormalities. The Smart Coil functions to selectively embolize targeted segments of the vasculature by packing a sufficient quantity of platinum coils to achieve occlusion. The {5}------------------------------------------------ Smart Coil consists of three components: a Coil, a Detachment Pusher (attached to Coil), and a Detachment Handle (packaged separately). The purpose of this 510(k) pre-market notification is to change the detachment pusher pull wire tensile strength specification to a more relevant value as measured in engineering studies to improve device manufacturability. ## 1.10 Indications for Use The Penumbra Smart Coil System is indicated for the embolization of: - Intracranial aneurysms . - . Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae - Arterial and venous embolizations in the peripheral vasculature . ## 1.11 Summary of Non-Clinical Data Design Verification testing was repeated for all attributes affected by change to device specifications. Results of testing demonstrate that Smart Coil with modified specification is substantial equivalence to predicate Smart Coil. The following tests were performed and all tests passed successfully: | Test | Summary | Results | |---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Tensile | Minimum break force value measured per specification | Pass | | Simulated Use | Simulated use testing of the Smart Coil in an anatomical model<br>which simulated the tortuosity of the neurovasculature. Devices were<br>delivered through the tortuous anatomical model to evaluate the<br>ability of the device to deliver and detach embolic coils. | Pass | # 1.12 Summary of Substantial Equivalence The Smart Coil was found to have a safety and effectiveness profile that is similar to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...