K242376 · Balt USA, LLC · QJP · Dec 30, 2024 · Cardiovascular
Device Facts
Record ID
K242376
Device Name
Next Generation Access Platform
Applicant
Balt USA, LLC
Product Code
QJP · Cardiovascular
Decision Date
Dec 30, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Intended Use
The Next Generation Access Platform is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
Device Story
The Next Generation Access Platform is a percutaneous catheter system used to introduce interventional devices into the peripheral, coronary, and neuro vasculature. The system comprises a single-lumen, braid-reinforced, variable-stiffness access sheath, a dilator, an introducer sheath, a rotating hemostasis valve (RHV), and a hemostasis valve adapter. The sheath features a radiopaque distal marker and a hydrophilic coating to reduce friction during navigation. The dilator facilitates atraumatic entry through skin and subcutaneous tissue. The device is intended for single use in a clinical setting, operated by a physician. The physician navigates the sheath to the target vessel; the sheath provides a conduit for subsequent interventional devices. The platform is provided sterile and non-pyrogenic. Performance is verified through bench testing, including mechanical, biocompatibility, and simulated-use assessments.
Clinical Evidence
No clinical data was provided. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993), mechanical performance (ISO 10555-1), and simulated-use validation in clinically relevant anatomical models.
Technological Characteristics
Single-lumen, braid-reinforced, variable-stiffness catheter. Materials: medical-grade plastics and stainless steel. Coating: hydrophilic. Sterilization: Ethylene Oxide (ISO 11135). Connectivity: None. Dimensions: 80-110 cm length, 6F/7F variants. Standards: ISO 10993 (biocompatibility), ISO 10555-1 (catheter performance), ISO 80369-7 (hub), ASTM F1980 (shelf life).
Indications for Use
Indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K191717 — Route 92 Medical Sheath System · Route 92 Medical, Inc. · Jan 13, 2020
K240529 — 8F Modified Sheath System · Route 92 Medical, Inc. · Oct 2, 2024
K211405 — Prelude Guide Sheath · Merit Medical Systems, Inc. · Oct 4, 2021
K032073 — GORE INTRODUCER SHEATH WITH SILICONE PINCH VALVE · W. L. Gore & Associates, Inc. · Aug 13, 2003
K161659 — FlexSeal Introducer Sheath with Hydrophilic Coating · Libramedical, Inc. · Jul 15, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 30, 2024
Balt USA. LLC Alicia Smith Senior Regulatory Affairs Specialist 29 Parker Irvine, California 92618
Re: K242376
Trade/Device Name: Next Generation Access Platform Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY Dated: November 27, 2024 Received: November 29, 2024
Dear Alicia Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242376
Device Name Next Generation Access Platform
Indications for Use (Describe)
The Next Generation Access Platform is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary: K242376
| Applicant: | Balt USA, LLC<br>29 Parker<br>Irvine, CA 92618<br>Registration No.: 3014162263 |
|------------------------|--------------------------------------------------------------------------------------------|
| Contact Person: | Alicia Smith<br>Senior Specialist, Regulatory Affairs<br>Email: alicia.smith@baltgroup.com |
| Date Summary Prepared: | November 27, 2024 |
| Trade Name: | Next Generation Access Platform |
| Common Name: | Catheter, Percutaneous |
| Device Classification: | Class II |
|------------------------|-----------------------------------|
| Predicate Devices: | Ballast 088 Long Sheath (K182918) |
| | Base Camp Sheath System (K191717) |
Neurology, Cardiovascular
QJP, DQY
21 CFR 870.1250
#### Device Description:
Review Panel:
Product Codes:
Regulation Number:
The Next Generation Access Platform consists of the Next Generation Access Sheath, Dilator, Introducer Sheath, Rotating Hemostasis Valve (RHV), and Hemostasis Valve Adapter (HVA). The Next Generation Access Sheath is a single lumen, braid-reinforced, variable stiffness sheath with a radiopaque zone on the distal end and a luer hub on the proximal end. The Next Generation Access Sheath is compatible with introducer sheaths appropriately sized for the outer diameter of the Next Generation Access Sheath.
The Next Generation Access Platform is provided sterile, non-pyrogenic, and is intended for single use only.
The Dilator facilitates the percutaneous entry of the Next Generation Access Sheath by forming an atraumatic transition of the Next Generation Access Sheath through the skin and subcutaneous tissue to the vessel. The distal 40 cm portion of the Next Generation Access Sheath is covered with a hydrophilic coating to aid in reducing friction.
#### Indications for Use:
The Next Generation Access Platform is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
{5}------------------------------------------------
## Comparison of Subject and Predicate Devices:
| Predicate Devices | | Subject Device | | | | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|----------------------------------|-----------------|
| | Ballast 088 Long Sheath (K182918) | Base Camp Sheath System (K191717) | Next Generation Access Platform (K242376) | | | |
| Indications for Use | The Ballast 088 Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. | The Route 92 Medical Sheath System is indicated for the introduction of interventional devices into the peripheral and neuro vasculature. | The Next Generation Access Platform is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. | | | |
| Device Classification Product Code | Class II DQY (Percutaneous Catheter) | Class II DQY (Percutaneous Catheter) | Class II QJP (Percutaneous Catheter, Neurovasculature) DQY (Percutaneous Catheter) | | | |
| Dimensional Specifications | | | | | | |
| | Ballast 088 Long Sheath | Base Camp Sheath System | Next Generation 6F+ Access Sheath | | Next Generation 7F Access Sheath | |
| | | | Sheath | Dilator | Sheath | Dilator |
| Sheath Tip Shape | Straight | Straight | Same as K182918 | Same as K182918 | Same as K191717 | Same as K191717 |
| Overall Length | 80 cm – 105 cm | 90 cm | 80 cm – 110 cm | 95 cm – 120 cm | 80 cm – 90 cm | 95 cm -105 cm |
| Outer Diameter | 8F or 0.106" | 0.122" | 0.108"<br>(2.74 mm) | 0.088" | 0.120"<br>(3.05 mm) | 0.094" |
| Inner Diameter | 0.088" | 0.106" | 0.091"<br>(2.31 mm) | 0.042" | 0.099"<br>(2.51 mm) | 0.042" |
| Length from Radiopaque marker band to distal tip | 0.025"<br>(0.64 mm) | | 0.028"<br>(0.70 mm) | N/A | 0.028"<br>(0.70 mm) | N/A |
| Device Attributes | | | | | | |
| Catheter Materials | Commonly used medical grade plastics & stainless steel | Commonly used medical grade plastics & stainless steel | Same as K182918 | | Same as K191717 | |
| Coating | Hydrophilic | Hydrophilic | Same as K182918 | | Same as K191717 | |
| Packaging Materials | Commonly used medical device packaging materials | Commonly used medical device packaging materials | Same as K182918 | | Same as K191717 | |
{6}------------------------------------------------
| Accessories | Dilator, Introducer<br>Sheath, Rotating<br>Hemostasis Valve,<br>Hemostasis Valve<br>Adapter | Dilator, Rotating<br>Hemostasis Valve | Same as K182918 | Dilator, Introducer Sheath, Rotating<br>Hemostasis Valve, Hemostasis Valve<br>Adapter |
|---------------|---------------------------------------------------------------------------------------------|---------------------------------------|-----------------|---------------------------------------------------------------------------------------|
| Sterilization | | | | |
| How Supplied | Sterile, Single Use | Sterile, Single Use | Same as K182918 | Same as K191717 |
| Method | Ethylene Oxide | Ethylene Oxide | Same as K182918 | Same as K191717 |
{7}------------------------------------------------
### Biocompatibility:
The following biocompatibility testing was conducted for the Next Generation Access Sheath and Dilator. The Introducer Sheath, Rotating Hemostasis Valve, and Hemostasis Valve Adapter are identical in materials and manufacturing to those previously cleared in K182918.
| Test | Test Method Summary | Results | |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity - MEM Elution | Tested in accordance with ISO 10993-5. | The test article is considered non-cytotoxic. | |
| Hemocompatibility - Hemolysis (Direct Contact & Extract Method) | Tested in accordance with ISO 10993-4. | The test article is considered non-hemolytic. | |
| Hemocompatibility - Complement Activation SC5b-9 Assay | Tested in accordance with ISO 10993-4. | The test article performed similarly to the predicate comparator. | |
| Hemocompatibility - Comparative Surface Assessment | The test article and predicate comparator were visually inspected at minimum 40x magnification. | Requirements have been met by the test article. | |
| Hemocompatibility - In Vitro Blood Loop Assay (Next Generation Access Sheath) | Tested in accordance with ISO 10993-4. | The test article performed similarly to the comparator. | |
| Hemocompatibility - In Vitro Blood Loop Assay (Dilator) | Tested in accordance with ISO 10993-4. | Requirements have been met by the test article. | |
| Pyrogenicity - Material-Mediated Rabbit Pyrogen | Tested in accordance with <USP 151>. | Requirements have been met by the test article. | |
| Sensitization – Guinea Pig Maximization Test | Tested in accordance with ISO 10993-10. | The test article did not elicit a sensitization response. | |
| Systemic Toxicity - Acute Systemic Injection | Tested in accordance with ISO 10993-11. | Requirements have been met by the test article. | |
| Irritation - Intracutaneous Reactivity | Tested in accordance with ISO 10993-23. | Requirements have been met by the test article. | |
| Test | Test Method Summary | Results | |
| Dimensional<br>Verification | The catheter outer diameter, inner diameter, usable<br>length, tip length, and coating length were measured. | Pass | |
| Surface<br>Contamination | Visual inspection completed for surface defects. | Pass | |
| Tensile Strength | The peak tensile force was evaluated per ISO 10555-1<br>after preconditioning in a simulated use model. | Pass | |
| Kink Resistance | Kink resistance was evaluated after preconditioning in a<br>simulated use model. | Pass | |
| Liquid Leakage | The device was exposed to a liquid pressure for 30<br>seconds. The device was inspected for leakage per ISO<br>10555-1. | Pass | |
| Air Leakage | The device was tested for air leakage into the hub per<br>ISO 10555-1. | Pass | |
| Dynamic Burst | Tested to verify the device can withstand internal liquid<br>pressure under dynamic flow conditions with the distal<br>end open. | Pass | |
| Torque Strength | The device was evaluated for torque strength by<br>measuring the number of catheter rotations until failure<br>after preconditioning in a simulated use model and<br>compared to the predicate. | Pass | |
| Hub Validation<br>Testing | The device shall meet the established acceptance criteria<br>per ISO 80369-7. | Pass | |
| Particulate Matter | The catheter underwent simulated use testing and<br>particulate testing was conducted including a reference<br>device for comparison. | Particulate generation was<br>comparable to the predicate<br>device. | |
| Tip Buckling | The maximum force to cause catheter tip buckling while<br>constrained at varying distances was measured. | The tip stiffness was<br>comparable to the predicate<br>and other cleared catheters. | |
| Catheter Shaft<br>Stiffness | Stiffness of catheter shaft measured along the length of<br>the shaft. | Catheter shaft stiffness was<br>comparable to the predicate<br>device. | |
| Corrosion | Corrosion tested per ISO 10555-1. | No evidence of corrosion<br>and met requirements per<br>ISO 10555-1. | |
| Static Burst | The distal tip of the catheter was blocked, and fluid was<br>injected into the lumen at increasing pressure until the<br>catheter burst per ISO 10555-1 and the static burst<br>pressure was compared with the maximum pressure<br>generated with manual syringe injection. | Pass | |
| Test | Test Method Summary | Results | |
| Coating Integrity | The coating integrity was inspected before and after preconditioning through a simulated use model. | No evidence of surface damage or coating defects. | |
| | Saline and<br>Contrast Exposure | After the device was used to deliver saline and contrast media, the device was inspected for damage, and dimensional attributes were measured. | No visual evidence of damage or dimensional changes. |
| Radiopacity<br>(Visibility) | | The device was tested to demonstrate acceptable radiopacity. | Marker radiopacity is comparable to the predicate. |
| | | Design Validation<br>/Usability | The subject and predicate devices were prepared in accordance with their respective instructions for use and tested for device usability in a clinically relevant anatomical model. |
{8}------------------------------------------------
# Performance Data - Bench:
The following performance bench testing was conducted to assess the performance of the Next Generation Access Platform:
{9}------------------------------------------------
### Sterilization and Shelf Life:
The Next Generation Access Platform is sterilized using an ethylene oxide sterilization cycle that was verified to a sterility assurance level of 1 x 10 ° in accordance with ISO 11135. Accelerated aging testing for the Next Generation Access Platform based on ASTM F1980 has established that the subject device and packaging remain functional for the labeled expiration date. Aging studies for packaging integrity, seal strength, and device functionality were performed and met the acceptance criteria.
### Performance Data - Animal:
No animal study was conducted because bench performance testing was deemed sufficient for verification and validation and to support substantial equivalence.
### Performance Data - Clinical:
No clinical study was conducted because bench performance testing was deemed sufficient for verification and validation and to support substantial equivalence.
#### Conclusion:
The evidence presented in this 510(k) submission demonstrates substantial equivalence between the subject device and the predicate devices. The subject and predicate devices have the same intended use and indications for use. The differences in technological characteristics do not raise new questions of safety or effectiveness. Nonclinical bench, biocompatibility, and shelf-life testing demonstrates the Next Generation Access Platform meets the device specifications and performs as intended.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.