CerusEndo Microcatheter (027)

K213314 · Cerus Endovascular, Inc. · QJP · Mar 15, 2022 · Cardiovascular

Device Facts

Record IDK213314
Device NameCerusEndo Microcatheter (027)
ApplicantCerus Endovascular, Inc.
Product CodeQJP · Cardiovascular
Decision DateMar 15, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Intended Use

The CerusEndo Microcatheter (027) is intended to deliver therapeutic devices within the neurovasculature.

Device Story

The CerusEndo Microcatheter (027) is a percutaneous, neurovascular catheter designed to deliver therapeutic devices (e.g., occlusion coils) to discrete locations within the neurovasculature. The device features a polymer tapered shaft, stainless steel reinforcement, a Teflon-lined inner lumen, and a hydrophilic coating. The distal segment is flexible for navigation through tortuous anatomy, and the distal tip is steam-shapeable by the physician at the point of use. The device is operated by a physician in hospital interventional neuroradiology suites; it is advanced over a guidewire through a guide catheter. The device is supplied sterile for single use. By providing a trackable, shapeable conduit, the catheter facilitates the precise placement of therapeutic implants, potentially improving clinical outcomes in neurovascular interventions.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including tensile strength, shaft flexibility, shape retention, kink resistance, static burst pressure, simulated use, particulate assessment, coating durability, and torque testing. Biocompatibility was confirmed per ISO 10993-1, including cytotoxicity, hemocompatibility (hemolysis, complement activation), and canine carotid thrombogenicity studies.

Technological Characteristics

Percutaneous catheter; 150/160 cm lengths; 0.027 in ID. Materials: Pebax outer/inner layers, stainless steel braid/coil, PTFE liner, platinum/iridium marker, nylon hub. Features: hydrophilic coating, steam-shapeable distal tip. Sterilization: Ethylene oxide. Biocompatibility: ISO 10993-1 compliant. Connectivity: None (manual).

Indications for Use

Indicated for patients undergoing vascular procedures requiring delivery of therapeutic devices within the neurovasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ March 15, 2022 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that. Cerus Endovascular, Inc. Lori E. Adels, Ph.D. Chief Compliance Officer 47757 Fremont Boulevard Fremont, California 94538 Re: K213314 Trade/Device Name: CerusEndo Microcatheter (027) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP Dated: February 11, 2022 Received: February 14, 2022 Dear Lori Adels: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K213314 Device Name CerusEndo Microcatheter (027) Indications for Use (Describe) The CerusEndo Microcatheter (027) is intended to deliver therapeutic devices within the neurovasculature. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| |X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K213314 This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 8807.92. #### I. SUBMITTER Submitter Name: Cerus Endovascular, Inc. - 47757 Fremont Boulevard, Address: Fremont, CA 94538 Phone Number: (510) 651-4000 - Fax Number: (510) 405-8356 - Contact Person: Lori Adels, PhD Chief Compliance Officer Date Prepared: March 15, 2022 #### II. DEVICE Name of Device: CerusEndo Microcatheter (027) Common Name: Catheter, Percutaneous, Neurovasculature Classification Name: Percutaneous Catheter 21 CFR 870.1250 Regulatory Class: Class II Product Code: QJP #### III. PREDICATE DEVICE K142449: Headway 27 Microcatheter manufactured by MicroVention Inc. The predicate device has not been subject to a design-related recall. #### IV. DEVICE DESCRIPTION The subject device is a microcatheter available in two versions: a 150 cm length microcatheter with a 34 cm length distal segment; and a 160 cm length microcatheter with a 34 cm length distal segment. The distal segment of the catheter is flexible to facilitate access into tortuous anatomy, and the distal tip of the catheter is formable, {4}------------------------------------------------ allowing the physician to shape it according to the needs of the procedure at the point of use. The CerusEndo Microcatheter (027) has a hydrophilic coating, radiopaque marker, Luer hub on the proximal end, polymer tapered shaft construction, stainless steel reinforced shaft, and Teflon lined inner lumen. The subject device is controlled by user manipulation to access discrete locations within the vascular anatomy. It is intended to deliver therapeutic devices through its inner lumen. It is designed to be used in neurovascular locations. #### V. INDICATIONS FOR USE The CerusEndo Microcatheter (027) is intended to deliver therapeutic devices within the neurovasculature. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The subiect device has similar design, dimensions, materials, intended use, and technological characteristics to the legally marketed predicate cleared under K142449. Non-clinical testing has been performed to demonstrate that any differences in technological characteristics do not raise new questions of safety and effectiveness. A table comparing the intended use and technical characteristics of the proposed device and the legally marketed predicate is provided in Table 1. | Manufacturer | Subject Device<br>CerusEndo Microcatheter (027)<br>(Cerus Endovascular) | Predicate Device K142449<br>Headway 27 Microcatheter<br>(MicroVention) | Comparison | |------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Clinical Attributes | | | | | Indications for<br>Use | The CerusEndo Microcatheter (027)<br>is intended to deliver therapeutic<br>devices within the<br>neurovasculature. | The Headway Microcatheter is intended<br>for general intravascular use, including<br>the peripheral, coronary and neuro<br>vasculature for the infusion of diagnostic<br>agents, such as contrast media, and<br>therapeutic agents, such as<br>occlusion coils. | Same | | Environments of<br>Use | Hospital interventional<br>neuroradiology suites | Hospital interventional<br>neuroradiology suites | Identical | | Patient<br>Population | Patients undergoing vascular<br>procedures | Patients undergoing vascular<br>procedures | Identical | | Contraindications | No known contraindications | No known contraindications | Identical | | Functions | To facilitate introduction of<br>therapeutic devices into the<br>vasculature | To facilitate introduction of diagnostic<br>and therapeutic devices into the<br>vasculature | Similar | | Patient Access | Device access is gained using an<br>introducer sheath inserted into the<br>vasculature and is advanced over a<br>guidewire and through a guide<br>catheter | Device access is gained using an<br>introducer sheath inserted into the<br>vasculature and is advanced over a<br>guidewire and through a guide catheter | Identical | | | | Table 1 - Comparison Table of Proposed Device and Predicate Device | |--|--|--------------------------------------------------------------------| | | | | {5}------------------------------------------------ | Manufacturer | Subject Device<br>CerusEndo Microcatheter (027)<br>(Cerus Endovascular) | Predicate Device K142449<br>Headway 27 Microcatheter<br>(MicroVention) | Comparison | |----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------| | Intraoperative<br>Use | Yes | Yes | Identical | | Technological Attributes | | | | | General<br>Description | Percutaneous Catheter<br>Intravascular Catheter | Percutaneous Catheter Diagnostic<br>Intravascular Catheter | Same | | Device<br>Configuration | Proximal Luer hub, hydrophilic<br>coating, radiopaque marker,<br>polymer tapered shaft construction,<br>stainless steel reinforced shaft, and<br>lined inner lumen | Proximal Luer hub, hydrophilic<br>coating, radiopaque marker,<br>polymer tapered shaft construction,<br>stainless steel reinforced shaft, and<br>lined inner lumen | Same | | Catheter Body<br>Materials | Pebax outer layer, stainless steel<br>braid/coil, Pebax inner layer, PTFE<br>liner | Polyurethane outer layer, Pebax<br>inner layer, stainless steel<br>braid/coil, PTFE/polyolefin liner | Similar | | Marker | Platinum/iridium | Platinum/iridium | Identical | | Hub | Nylon | Nylon | Identical | | Strain Relief | Pebax | Pebax | Identical | | Introducer | Not applicable | Pebax | The CerusEndo<br>Microcatheter (027)<br>is not provided with<br>an introducer | | Shaping Mandrel | Stainless Steel | Stainless steel | Identical | | Proximal ID/OD | 0.027 in min / 0.040 in | 0.0278 in min / 0.040 in | Similar | | Distal ID/OD | 0.027 in min / 0.036 in | 0.027 in min / 0.035 in | Similar | | Effective Length | 150 cm, 160 cm | 150 cm, 156 cm | Similar | | Coating | Hydrophilic coating | Hydrophilic coating | Same | | Tip Configuration | Straight - Steam Shapeable by<br>physician prior to use | Straight - Steam Shapeable by<br>physician prior to use | Identical | | Guidewire<br>Compatibility | $\leq$ 0.018 in | 0.014 / 0.016 / 0.018 / 0.021 in | Similar | | Accessories | Shaping mandrel | Shaping mandrel, introducer | The CerusEndo<br>Microcatheter (027) is<br>provided with a<br>shaping mandrel but is<br>not provided with an<br>introducer | | Method of Supply | Sterile, single use | Sterile, single use | Identical | | Sterilization | Ethylene oxide,<br>single patient use | Ethylene oxide,<br>single patient use | Identical | | Biocompatibility | Biocompatible per ISO 10993-1 | Biocompatible per ISO 10993-1 | Identical | There are no significant technological differences between the subject device and predicate device. {6}------------------------------------------------ #### VII. PERFORMANCE DATA Non- clinical tests were performed to demonstrate safety and substantial equivalence. Bench testing performed is summarized in Table 2. Biocompatibility testing performed is summarized in Table 3. | Test | Test Method Summary | Results | |------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------| | Tensile Strength | This test measures the tensile strength<br>of the catheter bonds using a tensile<br>tester and are pulled to failure. | Device met acceptance<br>criteria and has tensile<br>strength similar to the<br>predicate. | | Shaft Flexibility<br>(stiffness) | This test measures the bending stiffness<br>of the distal and proximal catheter shaft<br>segments. | Device met acceptance<br>criteria and has a similar<br>shaft flexibility as the<br>predicate. | | Shape Retention | This test measures the ability of the<br>catheter tip to form and retain a steam<br>shape using conventional catheter lab<br>shaping techniques. | Device met acceptance<br>criteria and has a similar<br>shape retention as the<br>predicate. | | Kink Resistance | This test measures the distal and<br>proximal catheter shaft resistance to<br>kinking. | Device met acceptance<br>criteria and has kink<br>resistance similar to the<br>predicate. | | Static Burst | This test measures the catheters<br>resistance to burst failure by using a<br>high pressure injector to pressurize the<br>lumen while the distal tip is occluded. | Device met acceptance<br>criteria and has<br>static burst<br>pressure similar<br>to the predicate. | | Simulated Use | The device was used in accordance to<br>the Instructions for Use. | Device met acceptance<br>criteria. | | Particulate | This test assesses the coating integrity<br>by measuring the quantity and size of<br>particles generated during simulated<br>use of the device in an anatomical<br>model. | The number and size<br>of particles were<br>similar to the predicate<br>device. | | Coating Friction and<br>Durability | This test measures the lubricity of the<br>coating and the durability after repeated<br>abrasion cycles. | Device met acceptance<br>criteria. | | Torque Testing | This test measures the torque strength of<br>the catheter when turned to failure in a<br>tortuous path model while the catheter<br>tip is fixed. | Device met acceptance<br>criteria and has torque<br>strength similar to the<br>predicate. | {7}------------------------------------------------ | Test | Extract/Test System | Results | |-----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Cytotoxicity (MEM<br>Elution) | The test extract, a positive control, and<br>a negative control were extracted at<br>37°C for 24 hours in MEM solution<br>(5% serum supplemented cell culture<br>medium) and exposed to mouse<br>fibroblast cells. | Non-cytotoxic.<br>The test article is<br>considered non-<br>cytotoxic to cells. | | Hemocompatibility<br>(ISO In Vitro<br>Hemocompatibility<br>– Direct Contact) | Blood samples from three human<br>donors were pooled and diluted. The<br>test article is added to aliquots of<br>human blood and incubated at 37 °C<br>for a minimum of 3 hours. | Non-hemolytic. There<br>were no differences<br>between the hemolytic<br>index of the test article<br>and the negative control. | | Hemocompatibility<br>(Hemolysis –<br>Indirect) | Test samples were extracted in<br>phosphate buffered saline at 37°C for<br>72 hours. The test article extract was<br>incubated at 37°C for a minimum of 3<br>hours. | Non-hemolytic<br>There were no<br>significant<br>differences between the<br>test article extract/solid<br>and negative control<br>article results. | | Hemocompatibility<br>(ISO Complement<br>Activation C3 and<br>SC5b-9 Test –<br>Direct Contact) | The test article, predicate, negative<br>control, and positive control (latex)<br>were added to the serum pooled from<br>three human blood samples. All were<br>incubated at 37 °C for 30, 60, and 90<br>minutes. | C3a and SC5b-9<br>complement proteins<br>were considered to be<br>non-activated as<br>compared to the<br>negative<br>control. | | Hemocompatibility<br>(Thrombogenicity) | Devices were placed in a canine carotid<br>vessel. | Non-thrombogenic. No<br>significant thrombus<br>was observed on any of<br>the subject catheters,<br>and the device was<br>determined to not show<br>thrombogenic<br>potential | ## Table 3 - Biocompatibility Testing Summary ## VIII. CONCLUSION After a comparison of the subject CerusEndo Microcatheter (027) intended use, technological characteristics, and expected performance to the legally marketed predicate, the Headway 27 Microcatheter (K142449), Cerus Endovascular concludes that the CerusEndo Microcatheter (027) is substantially equivalent to the legally marketed predicate device. The differences between the subject device and the predicate do not raise different questions of safety and effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...