XENIUM XPM 110, XENIUM XPM 130, MODELS M25649A, M25650A, XENIUM XPM 150, XENIUM XPM 170

K093120 · Baxter Healthcare Corporation - Renal Division · KDI · Feb 17, 2010 · Gastroenterology, Urology

Device Facts

Record IDK093120
Device NameXENIUM XPM 110, XENIUM XPM 130, MODELS M25649A, M25650A, XENIUM XPM 150, XENIUM XPM 170
ApplicantBaxter Healthcare Corporation - Renal Division
Product CodeKDI · Gastroenterology, Urology
Decision DateFeb 17, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.5860
Device ClassClass 2
AttributesTherapeutic

Intended Use

Hemodialysis with Xenium XPM dialyzers is indicated for patients with acute or chronic renal failure when conservative therapy is judged to be inadequate. It also may be indicated in the treatment of patients intoxicated with poisons or drugs.

Device Story

Xenium XPM is a single-use hemodialyzer; utilizes polyethersulfone fiber membrane for blood filtration; available in six sizes based on membrane surface area. Used in clinical settings for hemodialysis procedures to manage renal failure or intoxication. Operates by facilitating solute and fluid removal from blood via dialysis. Healthcare providers use the device as part of a standard hemodialysis circuit to support patient renal function or remove toxins.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing and risk analysis per ISO 14971.

Technological Characteristics

Polyethersulfone fiber dialyzer; single-use; six sizes differentiated by membrane surface area. High permeability hemodialysis system. Class II device.

Indications for Use

Indicated for patients with acute or chronic renal failure requiring hemodialysis when conservative therapy is inadequate, and for patients requiring treatment for drug or poison intoxication.

Regulatory Classification

Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ” (2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,” (3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,” (4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and (5) “Guidance for Hemodialyzer Reuse Labeling.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K093120 PAGE 1 OF 2 Special 510(k) Premarket Notification Xenium XPM Section 5 510(k) Summary ## 5. 510(k) SUMMARY September 24, 2009 # OWNER: Baxter Healthcare Corporation One Baxter Parkway -Deerfield, Illinois-60015- ## CONTACT PERSON: Diane Rennpferd Manager, Global Regulatory Affairs 1620 Waukegan Road McGaw Park, IL 60085 Telephone: (847) 473-6293 Fax: (847) 785-5116 # DEVICE NAME: Trade name: Xenium XPM ### Table 5-1. Product Codes for Xenium XPM | M25649A | Xenium XPM 110 | |---------|----------------| | M25650A | Xenium XPM 130 | | M25651A | Xenium XPM 150 | | M25652A | Xenium XPM 170 | | M25653A | Xenium XPM 190 | | M25654A | Xenium XPM 210 | # FEB 1 7 2010 {1}------------------------------------------------ Special 510(k) Premarket Notification Xenium XPM K093120 PAGE 2 OF 2 Section 5 510(k) Summary #### Common name: Hemodialyzer Classification name: Dialyzer, high permeability with or without sealed dialysate system #### PREDICATE DEVICE: | Table 5-2. | |------------------| | Previous 510(k)s | | Device | Company | Previous 510(k) | Clearance date | |------------|-------------------------------------|-----------------|----------------------| | Xenium XPH | Baxter<br>Healthcare<br>Corporation | K083778 | February 20,<br>2009 | ### DESCRIPTION OF THE DEVICE: Xenium XPM dialyzers are Polyethersulfone fiber dialyzers and will be labeled for single use only. The dialyzers are available in six (6) sizes, which are differentiated by membrane surface area. #### STATEMENT OF INTENDED USE: Hemodialysis with Xenium XPM dialyzers is indicated for patients with acute or chronic renal failure when conservative therapy is judged to be inadequate. It also may be indicated in the treatment of patients intoxicated with poisons or drugs. ## TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE: The Xenium XPM is substantially equivalent to Baxter's current legally marketed Xenium XPH dialyzer cleared February 20, 2009 (K083778). #### DISCUSSION OF NONCLINICAL TESTS: Baxter Healthcare Corporation conducts risk analyses, per ISO 14971, and design verification tests based on the result of these analyses. All test results meet the acceptance criteria, and support that the devices are appropriately designed for their intended use. #### CONCLUSION: The Xenium XPM is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized representation of a caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it, topped by a pair of wings. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G60 Silver Spring, MD 20993-0002 Ms. Diane Rennpferd Manager, Global Regulatory Affairs Baxter Health Care Corporation Renal Division 1620 Waukegan Road MCGAW PARK IL 60085 FEB 1 7 2010 Re: K093120 Trade/Device Name: Xenium XPM Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: January 11, 2010 Received: January 19, 2010 Dear Ms. Rennpferd: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {3}------------------------------------------------ Page 2 - device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure . {4}------------------------------------------------ # INDICATIONS FOR USE KO93120 510(k) Number (if known): Device Name: Xenium XPM Indications for Use: Hemodialysis with Xenium XPM dialyzers is indicated for patients with acute or chronic renal failure when conservative therapy is judged to be inadequate. It also may be indicated in the treatment of patients intoxicated with poisons or drugs. AND/OR Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) 9 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) (Division Sign-Off) Division of Reproductive, Abdominal and Division oral Devices 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...