DRAKE WILLOCK SYSTEM 1000 DIALYSATE DELIVERY SYSTEM ALTRATOUCH 1000 DIALYSATE DELIVERY SYSTEM
K964922 · Althin Medical AB an Affiliate of Baxter Intl · KDI · May 21, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K964922
Device Name
DRAKE WILLOCK SYSTEM 1000 DIALYSATE DELIVERY SYSTEM ALTRATOUCH 1000 DIALYSATE DELIVERY SYSTEM
Applicant
Althin Medical AB an Affiliate of Baxter Intl
Product Code
KDI · Gastroenterology, Urology
Decision Date
May 21, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The intended use of the device is to provide hemodialysis treatments in the acute and chronic setting including high flux hemodialysis. - The device is intended to be used by trained operators when prescribed by a physician. - The device is intended to be used in conjunction with a hollow fiber or parallel plate dialyzer. - The auto start option is intended to allow up to a three day delayed start of the fluid path heat clean or rinse cycle.
Device Story
Dialysate delivery system for hemodialysis; mixes concentrate with water; regulates temperature and ionic concentration; removes fluid from blood; functions as total artificial kidney with dialyzer and blood pump. Modification adds 'AutoStart' feature allowing clinicians to schedule/delay fluid path heat clean or rinse cycles by up to 3 days. Used in clinical settings by trained operators. Output consists of prepared dialysate for patient treatment; AutoStart output is automated initiation of disinfection/rinse cycles. Benefits include operational flexibility for facility maintenance schedules.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Hemodialysis proportioning system. Fluid path and hardware identical to predicates. Software modification for delayed disinfection/rinse scheduling. No changes to safety systems or materials.
Indications for Use
Indicated for patients requiring hemodialysis in acute and chronic settings, including high flux hemodialysis. Intended for use by trained operators under physician prescription.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
K013562 — AURORA SYSTEM 1000 SERIES HOME SELF-CARE SINGLE PATIENT DELIVERY SYSTEM; MODELS- SYS 1000, L3 WITH N100 ARM, 4B SOFTWARE · Baxter Healthcare Corp · Dec 12, 2002
K242155 — DBB-06 PRO Hemodialysis Delivery System (DBB-06 PRO) · NIKKISO CO., LTD. · May 15, 2025
K970253 — COBE CENTRYSYSTEM 3 DIALYSIS CONTROL UNIT/COBE CARTRIDGE BLOOD TUBING SET · Gambro Healthcare · Sep 26, 1997
K103220 — VOYAGER PERITONEAL DIALYSIS SYSTEM · Deka Research & Development Corp. · Jul 29, 2011
Submission Summary (Full Text)
{0}
MAY 21 1997
page 1/2
# 510(k) SUMMARY
**Date:**
December 4, 1996
**Name:**
Althin Medical Inc.
Drake Willock® dialysis equipment
**Address:**
13520 S.E. Pheasant Court
Portland, Oregon 97222-1298
Phone Number: 503-659-3355
Fax Number: 503-652-0225
Contact person: Thomas D. Kelly
**Trade Names:**
Drake Willock® System 1000® Dialysate Delivery System with AutoStart Disinfection Option
AltraTouch™ 1000 Dialysate Delivery System with AutoStart Disinfection Option
**Common Name:**
Dialysate Delivery System
**Classification Name:**
Hemodialysis System and accessories, Class III Device.
**Product code:**
78KDI
**Equivalence Information / Comparison to Predicate Device:**
The AutoStart feature is a modification to predicate System 1000® and AltraTouch™ 1000 Dialysate Delivery Systems [510(k) numbers K910215 and K954987.] The System 1000® and AltraTouch™ 1000 single patient hemodialysis delivery systems with the AutoStart option are built with the same fluid path and hardware components as the predicate System 1000® and AltraTouch™ 1000 single patient hemodialysis delivery systems. The only difference is a software change to delay the start of heat clean or fluid path rinse. The machine and patient safety systems are the same as the predicate devices.
Althin Medical, Inc. believes that the design and testing of this modified System 1000®/AltraTouch™ 1000 Dialysate Delivery System demonstrates that it is safe and effective.
Pages of this 510(k) that refer to the new AutoStart option are marked with a star (★) in the right hand margin.
{1}
page 2/2
## Device Description:
The Drake Willock® System 1000® Single Patient Dialysate Delivery System is a dialysate proportioning system for hemodialysis. The system fulfills the following functions:
- Mixes concentrate with water in the appropriate proportions to produce dialysate
- Delivers dialysate at the appropriate temperature and ionic concentration to the dialyzer,
- Removes the appropriate amount of liquid from the patient’s blood
- Along with the dialyzer and blood pump acts as a total artificial kidney.
## The following option is the subject of this submission that will be implemented with FDA approval.
The fundamental modification is to provide the patient care provider with a means to delay the start of fluid path rinse or heat clean up to 3 calendar days.
## Intended Use:
### Intended use of the System 1000® and AltraTouch™ 1000 single patient hemodialysis delivery systems
The intended use of the device is to provide hemodialysis treatments in the acute and chronic setting including high flux hemodialysis.
- The device is intended to be used by trained operators when prescribed by a physician.
- The device is intended to be used in conjunction with a hollow fiber or parallel plate dialyzer.
### Intended use System 1000® and AltraTouch™ 1000 single patient hemodialysis delivery systems with the Auto Start option *
The intended use of the device is to provide hemodialysis treatments in the acute and chronic setting including high flux hemodialysis.
- The device is intended to be used by trained operators when prescribed by a physician.
- The device is intended to be used in conjunction with a hollow fiber or parallel plate dialyzer.
## Technological Characteristics:
The System 1000® and AltraTouch™ 1000 single patient hemodialysis delivery systems with the Auto Start ★ option are built with the same fluid path and hardware components as the predicate System 1000® and AltraTouch™ 1000 single patient hemodialysis delivery systems. The only difference is a software change to delay the start of heat clean or fluid path rinse. The machine and patient safety systems are the same as the predicate devices.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
MAY 21 1997
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Mr. Thomas D. Kelly
Senior Manager of Marketing and
Regulatory Affairs
Althin Medical, Inc.
13520 S.E. Pheasant Court
Portland, Oregon 97222-1298
Re: K964922
Auto Start Disinfection Option - modification of the
Altra Touch™ and Drake Willock® System 1000®
Dialysate Delivery Machines
Dated: March 21, 1997
Received: March 24, 1997
Regulatory class: III
21 CFR §876.5860/Product code: 78 KDI
Dear Mr. Kelly:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{3}
510(k) Number K964922
Page 1 of 1
Device Name: Drake Willock® System 1000® Dialysis Delivery System with Auto Start Disinfection Option
AltraTouch™ 1000® Dialysis Delivery System with Auto Start Disinfection Option
Indications For Use:
Indications for use of the Drake Willock® System 1000® Dialysis Delivery System with Auto Start Disinfection Option and AltraTouch™ 1000® Dialysis Delivery System with Auto Start Disinfection Option:
The indications for use of the device are to provide hemodialysis treatments in the acute and chronic setting including high flux hemodialysis.
- The device is intended to be used by trained operators when prescribed by a physician.
- The device is intended to be used in conjunction with a hollow fiber or parallel plate dialyzer.
- The auto start option is intended to allow up to a three day delayed start of the fluid path heat clean or rinse cycle.
(Please do not write below this line continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964922
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over the Counter Use
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.