CEREC Cercon 4D™ Abutment Blocks, CEREC Cercon 4D™ Abutment System
Applicant
Dentsply Sirona
Product Code
NHA · Dental
Decision Date
May 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
CEREC Cercon 4D™ Abutment System is intended for use in partially or fully edentulous mandibles and maxillae in support of single cement-retained restorations. The system comprises three parts: CEREC Cercon 4D™ Abutment Block, TiBase, CAD/CAM system. The CEREC Cercon 4D™ ceramic structure cemented to the TiBase is recommended for two-piece hybrid abutments for single tooth restorations and hybrid abutment crowns, used in conjunction with endosseous dental implants.
Device Story
Device consists of Yttria-doped zirconia blocks used to fabricate ceramic structures for two-piece hybrid abutments (mesostructure and crown). Blocks feature pre-drilled screw access channels and anti-rotation features. Used with TiBase (titanium base) and CAD/CAM system. Operated by dental professionals in clinical or lab settings. CAD/CAM software designs the restoration; user selects TiBase, which defines gingival height and post height. Final ceramic structure is cemented to TiBase. System ensures structural integrity for single-tooth restorations. Benefits include patient-specific aesthetic and functional restoration of edentulous sites.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including flexural strength (ISO 6872:2015) and fatigue testing (ISO 14801:2016) on worst-case combinations of angulation, platform size, and gingival height.
Technological Characteristics
Yttria-doped zirconia blocks. Designed for CAD/CAM milling. Non-sterile. Requires steam sterilization post-assembly. Meets ISO 6872 (flexural strength >1,100 MPa) and ISO 14801 (fatigue). Biocompatible per ISO 10993.
Indications for Use
Indicated for partially or fully edentulous patients requiring single cement-retained restorations supported by endosseous dental implants. Compatible with various Dentsply Sirona, BioHorizons, Nobel Biocare, Straumann, Thommen Medical, Osstem/Hiossen, Zimmer/Biomet, MIS, and Altatec Camlog implant systems.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
CEREC Tessera Abutment Block, CEREC Tessera Abutment System (K221402)
Reference Devices
Sirona Dental CAD/CAM System (inCoris ZI meso material) (K111421)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 7, 2024
Dentsply Sirona Rebecca Sporer Principal Regulatory Affairs Specialist 221 West Philadelphia Street Suite 60W York, Pennsylvania 17401
Re: K234018
Trade/Device Name: CEREC Cercon 4D™ Abutment Blocks, CEREC Cercon 4D™ Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA. PNP Dated: December 19, 2023 Received: April 5, 2024
Dear Rebecca Sporer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K234018
### Device Name
CEREC Cercon 4D™ Abutment Blocks, CEREC Cercon 4D™ Abutment System
### Indications for Use (Describe)
CEREC Cercon 4D™ Abutment System is intended for use in partially or fully edentulous mandibles and maxillae in support of single cement-retained restorations.
The system comprises three parts:
- CEREC Cercon 4D™ Abutment Block
- TiBase
- CAD/CAM system
The CEREC Cercon 4D™ ceramic structure cemented to the TiBase is recommended for two-piece hybrid abutments for single tooth restorations and hybrid abutment crowns, used in conjunction with endosseous dental implants.
Implant Systems:
• Dentsply Sirona: PrimeTaper EV, OmniTaper EV, AstraTech OsseoSpeed TX, Frialit / XiVE, AstraTech Implant EV, Ankylos
- · BioHorizons: Internal connection
· Nobel Biocare: Replace Select, Nobel Active, NobelReplace Conical Connection, Brânemark, NobelSpeedy Groovy
- · Straumann: Tissue Level, Bone Level
- · Thommen Medical: Element, Contact
- · Osstem/Hiossen: Osstem TS, USA: Hiossen ET
- · Zimmer/Biomet: External hex, Certain, Tapered Screw-Vent
- · MIS: C1 Conical Connection, V3 Conical Connection, SEVEN internal hex, M4 internal hex
- · Altatec Camlog
CAD/CAM Systems: ·Sirona Dental CAD/CAM System
Manufacturer: Dentsply Sirona Implant Implant: Primetaper EV Platform: S; TiBase AT EV 3.6 GH 1 S; Reference: 6586312; Size: S Platform: M; TiBase AT EV 4.2 GH 1 L; Reference: 6586320; Size: L Platform: L; TiBase AT EV 4.8 GH 1 L; Reference: 6586338; Size: L
Manufacturer: Dentsply Sirona Implant
Implant: AstraTech OsseoSpeed TX
Platform 3.5 / 4.0; TiBase AT TX 3.5/4.0 GH 1 L; Reference: 6598093; Size: L Platform 4.5 / 5.0; TiBase AT TX 4.5/5.0 GH 1 L; Reference: 6598101; Size: L Platform 3.5 / 4.0; TiBase AT OS 3.5/4.0 GH 1 L; Reference: 6282532; Size: L Platform 4.5 / 5.0; TiBase AT OS 4.5/5.0 GH 1 L; Reference: 6282540; Size: L
Manufacturer: Dentsply Sirona Implant Implant: Frialit / XiVE Platform: 3.4; TiBase FX 3.4 GH 1 S; Reference: 6282433; Size: S
{4}------------------------------------------------
Platform: 3.8; TiBase FX 3.8 GH 1 S; Reference: 6282441; Size: S Platform: 4.5; TiBase FX 4.5 GH 1 L; Reference: 6282458; Size: L Platform: 5.5; TiBase FX 5.5 GH 1 L; Reference: 6282466; Size: L
Manufacturer: Dentsply Sirona Implant Implant: AstraTech Implant EV Platform: XL; TiBase AT EV 5.4 GH 1 L; Reference: 66586346; Size: L
Manufacturer: Dentsply Sirona Implant
Implant: Ankylos Platform: C/X; TiBase ANK C/ GH 1 S; Reference: 6586528; Size: S Platform: C/X; TiBase ANK C/ GH 2 S; Reference: 6586536; Size: S Platform: C/X; TiBase ANK /X GH 1 S; Reference: 6586544; Size: S Platform: C/X: TiBase ANK /X GH 2 S; Reference: 6586551; Size: S
Manufacturer: BioHorizons
Implant: Internal Connection
Platform: 3.0; TiBase BH 3.0 GH 1 S; Reference: 6532779; Size: S
Platform: 3.5; TiBase BH 3.5 GH 1 L; Reference: 6532894; Size: L
Platform: 4.5; TiBase BH 4.5 GH 1 L; Reference: 6532951; Size: L
Platform: 5.7; TiBase BH 5.7 GH 1 L; Reference: 6536242; Size: L
Manufacturer: Nobel Biocare
Implant: Replace, Replace Select
Platform: NP; TiBase NB RS 3.5 GH 1 L; Reference: 6282474; Size: L Platform: RP; TiBase NB RS 4.3 GH 1 L; Reference: 6282482; Size: L Platform: WP; TiBase NB RS 5.0 GH 1 L; Reference: 6282490; Size: L Platform: 6.0; TiBase NB RS 6.0 GH 1 L; Reference: 6282508; Size: L
Implant: Nobel Active Platform: NP; TiBase NB A 4.5 GH 1 L; Reference: 6308188; Size: L
Implant: NobelReplace Conical Connection Platform: RP; TiBase NB A 5.0 GH 1 L; Reference: 6308253; Size: L
Implant: Brånemark Platform: NP, TiBase NB B 3.4 GH 1 L; Reference: 6282516; Size: L
Implant: NobelSpeedy Groovy Platform: RP; TiBase NB B 4.1 GH 1 L; Reference: 6282524; Size: L
Manufacturer: Straumann Implant: Tissue Level Platform: RN (4.8 mm): TiBase SSO 4.8 GH 1 L: Reference: 6284249; Size: L Platform: WN (6.5 mm); TiBase SSO 6.5 GH 1 L; Reference: 6284256; Size: L
Implant: Bone Level Platform: NC (3.3 mm); TiBase S BL 3.3 GH 1 L; Reference: 6308154; Size: L Platform: RC (4.1 mm / 4.8 mm); TiBase S BL C 4.1 GH 1 L; Reference: 6308337; Size: L
Manufacturer: Thommen Medical Implant: Element, Contact Platform: 3.5; TiBase TM 3.5 GH 1 S; Reference: 6531854; Size: S Platform: 4; TiBase TM 4 GH 1 S; Reference: 6532829; Size: S
{5}------------------------------------------------
Platform: 4.5; TiBase TM 4.5 GH 1 S; Reference: 6532837; Size: S Platform: 5; TiBase TM 5 GH 1 S; Reference: 6544360; Size: S Platform: 6; TiBase TM 6 GH 1 S; Reference: 6544378; Size: S
Manufacturer: Osstem / Hiossen Implant: Osstem TS (US Hiossen ET) Platform: Mini; TiBase O TS 3.5 GH 1 L; Reference: 6527035; Size: L Platform: Regular; TiBase O TS 4.0 GH 1 L; Reference: 6527043; Size: L
Manufacturer: Zimmer / Biomet Implant: External hex Platform: 3.4; TiBase B O 3.4 GH 1 L; Reference: 6282557; Size: L Platform: 4.1; TiBase B O 4.1 GH 1 L; Reference: 6282565; Size: L Platform: 5.0; TiBase B O 5.0 GH 1 L; Reference: 6282573; Size: L
Implant: Certain
Platform: 3.4; TiBase B C 3.4 GH 1 S; Reference: 6308048; Size: S Platform: 4.1; TiBase B C 4.1 GH 1 L; Reference: 6308097; Size: L Platform: 5.0; TiBase B C 5.0 GH 1 L; Reference: 6308121; Size: L
Implant: Tapered Screw-Vent
Platform: 3.5; TiBase Z TSV 3.5 GH 1 L; Reference: 6282581; Size: L Platform: 4.5; TiBase Z TSV 4.5 GH 1 L; Reference: 6282599; Size: L Platform: 5.7; TiBase Z TSV 5.7 GH 1 L; Reference: 6282607; Size: L
Manufacturer: M.I.S. Implants
Implant: C1 Conical Connection
Platform: NP; TiBase CN-TB001 C1 NP GH 0.5; Reference: CN-TB001; Size: L Platform: NP: TiBase CN-B015 C1 NP GH 1.5; Reference: CN-TB015; Size: L
Implant: V3 Conical Connection
Platform: NP; TiBase VN-TB001 V3 NP GH 0.5; Reference: VN-TB001; Size: L Platform: NP; TiBase VN-TB015 V3 NP GH 1.5; Reference: VN-TB015; Size: L
Implant: V3 Conical Connection / C1 Conical Connection Platform: SP; TiBase CS-TB001 SP GH 0.5; Reference: CS-TB001; Size: L Platform: SP; TiBase CS-TB015 SP GH 1.5; Reference: CS-TB015; Size: L Platform: SP; TiBase CS-TB030 SP GH 3; Reference: CS-TB030; Size: L
Implant: C1 Conical Connection
Platform: WP; TiBase CW-TB001 C1 WP GH 0.5; Reference: CW-TB001; Size: L Platform: WP; TiBase CW-TB015 C1 WP GH 1.5; Reference: CW-TB015; Size: L Platform: WP; TiBase CW-TB030 C1 WP GH 3; Reference: CW-TB030; Size: L
Implant: SEVEN internal hex, M4 internal hex
Platform: NP; TiBase MN-TB001 INT HEX NP GH 0.5; Reference: MN-TB001; Size: L Platform: NP; TiBase MN-TBC15 INT HEX NP GH 1.5; Reference: MN-TBC15; Size: L Platform: SP; TiBase MD-TB001 INT HEX SP GH 0.5; Reference: MD-TB001; Size: L Platform: SP; TiBase MD-TBC15 INT HEX SP GH 1.5; Reference: MD-TBC15; Size: L Platform: SP; TiBase MD-TBC30 INT HEX SP GH 3; Reference: MD-TBC30; Size: L Platform: WP; TiBase MW-TB001 INT HEX WP GH 0.5; Reference: MW-TB001; Size: L Platform: WP; TiBase MW-TBC15 INT HEX WP GH 1.5; Reference: MW-TBC15; Size: L Platform: WP; TiBase MW-TBC30 INT HEX WP GH 3; Reference: MW-TBC30; Size: L
{6}------------------------------------------------
Manufacturer: Straumann Implant: Tissue Level Platform: NNC (3.5 mm); TiBase NNC Variobase C 3.5 GH 1; Reference: 220.018; Size: S Platform: RN (4.8 mm): TiBase RN Varionbase C 3.5 GH 1; Reference: 220.019; Size: L Platform: WN (6.5 mm); TiBase WN Variobase C GH 1; Reference: 220.020; Size: L
Implant: Bone Level Platform: NC (3.3 mm); TiBase NC Variobase C 3.3 GH 1; Reference: 220.043; Size: S Platform: RC (4.1 mm / 4.8 mm); TiBase RC Variobase C 4.1 GH 1; Reference: 220.044; Size: L
Manufacturer: Altatec GmbH-Camlog
Implant: Camlog
Platform: 3.3; CAMLOG Titanium Base CADM/CAM, for Ø 3.3 mm GH 0.4; Reference K2244.3348; Size S Platform: 3.8; CAMLOG Titanium Base CAD/CAM, for Ø 3.8 mm GH 0.3; Reference: K2244.3848; Size: S Platform: 4.3; CAMLOG Titanium Base CAD/CAM, for Ø 4.3 mm GH 0.3; Reference: K2244.4348; Size: S Platform: 5.0; CAMLOG Titanium Base CAD/CAM, for Ø 5.0 mm GH 0.3; Reference K2244.5048; Size: L Platform: 6.0; CAMLOG Titanium Base CAD/CAM, for Ø 6.0 mm GH 0.3; Reference K2244.6048; Size: L
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{7}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
Image /page/7/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a blue abstract shape on the left and the words "Dentsply Sirona" in blue on the right. The abstract shape is made up of two curved lines that form an S shape. The text is in a sans-serif font.
# 510(k) SUMMARY
### for
### CEREC Cercon 4D™ Abutment Blocks, CEREC Cercon 4D™ Abutment System (K234018)
- 1. Submitter Information: Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
| Contact Person: | Rebecca Sporer |
|-------------------|---------------------------------|
| Telephone Number: | 717-849-4793 |
| Email: | Corporate-RA@dentsplysirona.com |
| Date Prepared: | 02 May 2024 |
- 2. Device Name:
- CEREC Cercon 4D™ Abutment Blocks, CEREC Cercon 4D™ Proprietary Name: ●
- Abutment System
- Classification Name: Endosseous dental implant abutment
- 872.3630 ● CFR Number:
- . Device Class: Class II
- Primary Product Code: NHA .
- . Secondary Product Code: PNP
#### 3. Predicate/Reference Devices:
| Predicate Device Name | 510(k) | Company Name |
|----------------------------------------------------------------|---------|-----------------|
| CEREC Tessera Abutment Block, CEREC Tessera<br>Abutment System | K221402 | Dentsply Sirona |
| Reference Device Name | 510(k) | Company Name |
|------------------------------------------------------------------------------|---------|-----------------|
| Technological characteristics / Performance / Composition / Biocompatibility | | |
| Sirona Dental CAD/CAM System (inCoris ZI meso material) | K111421 | Dentsply Sirona |
| Material Composition / Biocompatibility Only | | |
| Cercon | K162888 | Dentsply Sirona |
| Reverse Engineering Compatibility Testing | | |
| Sirona Dental CAD/CAM System with inLab Software | K200191 | Dentsply Sirona |
| Sirona Dental CAD/CAM System with CEREC Chairside Software | K193408 | Dentsply Sirona |
| Sirona Dental CAD/CAM System | K181520 | Dentsply Sirona |
| Sirona Dental CAD/CAM System | K100152 | Dentsply Sirona |
{8}------------------------------------------------
| Compatible Dental Implants | 510(k) | Company Name |
|-----------------------------------------------------------------------------------------|---------------------------------------------|-----------------|
| PrimeTaper EV | K210610 | Dentsply Sirona |
| OmniTaper EV | K221094 | Dentsply Sirona |
| AstraTech OsseoSpeed TX | K053384 | Dentsply Sirona |
| Frialit/ XiVE | K013867 | Dentsply Sirona |
| AstraTech Implant EV | K120414 | Dentsply Sirona |
| Ankylos | K083805 | Dentsply Sirona |
| Internal Connection | K143022,<br>K071638,<br>K093321,<br>K042429 | BioHorizons |
| Replace, Replace Select | K020646 | Nobel Biocare |
| Nobel Active and NobelReplace Conical Connection | K071370 | Nobel Biocare |
| Branemark and NobelSpeedy Groovy | K022562 | Nobel Biocare |
| Straumann Tissue Level and Bone Level | K151324 | Straumann |
| Element, Contact | K093615,<br>K090154 | Thommen Medical |
| Osstem TS (Hiossen ET) | K121585 | Osstem/Hiossen |
| External Hex, Certain and Tapered Screw-Vent | K061410,<br>K061629,<br>K014235 | Zimmer/Biomet |
| C1 Conical Connection, V3 Conical Connection,<br>SEVEN internal hex and M4 internal hex | K191152 | Dentsply Sirona |
| Camlog | K083496 | Altatec GmbH |
## 4. Description of Device:
The CEREC Cercon 4D Abutment Blocks, which are used for fabrication of a ceramic structure, two-piece hybrid abutments (meso-structure and crown) and abutment crowns, that are cemented to a TiBase (titanium base) used with dental implant systems. The CEREC Cercon 4D Abutment Blocks are not provided as the finished, fully assembled dental implant medical devices. The abutment blocks are materials supplied to dental professionals that must be further processed/manufactured using CAD/CAM technology and they are not intended to be reused as in the context of direct patient-applied devices and materials.
CEREC Cercon 4D™ Abutment Block are Yttria-doped zirconia blocks suitable for chairside and lab side use in fabrication of single cement-retained restorations. CEREC Ceron 4D™ Abutment Block are designed with a pre-drilled screw access channel and anti-rotation feature. The design allows for fabrication of a ceramic structure, two-piece hybrid abutments (mesostructure and crown) and abutment crowns, that are cemented to the TiBase (Titanium base) used with dental implant systems. The following patient-specific ranges for CAD/CAM fabrication of the subject abutments are:
- . Maximum angulation of 200
- . Minimal wall thickness of 0.5 mm
- Gingival height ranges from 0.3-2.0 mm ●
- Minimum abutment post height (i.e. length above the gingival height) of ≥ 4 mm .
{9}------------------------------------------------
- 5. Indications for Use:
CEREC Cercon 4D™ Abutment System is intended for use in partially or fully edentulous mandibles and maxillae in support of single cement-retained restorations.
The system comprises three parts:
- CEREC Cercon 4D™ Abutment Block
- TiBase
- CAD/CAM system
The CEREC Cercon 4D™ ceramic structure cemented to the TiBase is recommended for twopiece hybrid abutments for single tooth restorations and hybrid abutment crowns, used in conjunction with endosseous dental implants.
Implant Systems:
- Dentsply Sirona: PrimeTaper EV, OmniTaper EV, AstraTech OsseoSpeed TX, . Frialit / XiVE , AstraTech Implant EV , Ankylos
- BioHorizons: Internal connection .
- Nobel Biocare: Replace, Replace Select, Nobel Active, NobelReplace Conical . Connection, Brånemark, NobelSpeedy Groovy
- Straumann: Tissue Level, Bone Level ●
- Thommen Medical: Element, Contact
- Osstem/Hiossen: Osstem TS, USA:Hiossen ET
- Zimmer/Biomet: External hex, Certain, Tapered Screw-Vent ●
- MIS: C1 Conical Connection, V3 Conical Connection, SEVEN internal hex, M4 ● internal hex
- . Altatec - Camlog
CAD/CAM Systems:
- Sirona Dental CAD/CAM System .
{10}------------------------------------------------
# Compatible TiBase and Implant Systems
| Implant System | | Titanium Base | | |
|----------------------------------------|----------------------|-------------------------------------------------------|------------|------|
| Manufacturer / line | Platform | Name | REF | Size |
| Dentsply Sirona Implants | S | TiBase AT EV 3.6 GH 1 S | 6586312 | S |
| Primetaper EV | M | TiBase AT EV 4.2 GH 1 L | 6586320 | L |
| | L | TiBase AT EV 4.8 GH 1 L | 6586338 | L |
| | 3.5 / 4.0 | TiBase AT TX 3.5/4.0 GH 1 L | 6598093 | L |
| AstraTech<br>OsseoSpeed TX | 4.5 / 5.0 | TiBase AT TX 4.5/5.0 GH 1 L | 6598101 | L |
| | 3.5 / 4.0 | TiBase AT OS 3.5/4.0 GH 1 L | 6282532 | L |
| | 4.5 / 5.0 | TiBase AT OS 4.5/5.0 GH 1 L | 6282540 | L |
| Frialit / Xive | 3.4 | TiBase FX 3.4 GH 1 S | 6282433 | S |
| | 3.8 | TiBase FX 3.8 GH 1 S | 6282441 | S |
| | 4.5 | TiBase FX 4.5 GH 1 L | 6282458 | L |
| | 5.5 | TiBase FX 5.5 GH 1 L | 6282466 | L |
| AstraTech Implant<br>EV | XL | TiBase AT EV 5.4 GH 1 L | 6586346 | L |
| Ankylos | C/X | TiBase ANK C/ GH 1 S | 6586528 | S |
| | | TiBase ANK C/ GH 2 S | 6586536 | S |
| | | TiBase ANK /X GH 1 S | 6586544 | S |
| | | TiBase ANK /X GH 2 S | 6586551 | S |
| BioHorizons | | | | |
| Internal connection | 3.0 | TiBase BH 3.0 GH 1 S | 6532779 | S |
| | 3.5 | TiBase BH 3.5 GH 1 L | 6532894 | L |
| | 4.5 | TiBase BH 4.5 GH 1 L | 6532951 | L |
| | 5.7 | TiBase BH 5.7 GH 1 L | 6536242 | L |
| Nobel Biocare | | | | |
| Replace, Replace<br>Select | NP | TiBase NB RS 3.5 GH 1 L | 6282474 | L |
| | RP | TiBase NB RS 4.3 GH 1 L | 6282482 | L |
| | WP | TiBase NB RS 5.0 GH 1 L | 6282490 | L |
| | 6.0 | TiBase NB RS 6.0 GH 1 L | 6282508 | L |
| Nobel Active | NP | TiBase NB A 4.5 GH 1 L | 6308188 | L |
| NobelReplace<br>Conical Connection | RP | TiBase NB A 5.0 GH 1 L | 6308253 | L |
| Brånemark | NP | TiBase NB B 3.4 GH 1 L | 6282516 | L |
| NobelSpeedy<br>Groovy | RP | TiBase NB B 4.1 GH 1 L | 6282524 | L |
| Straumann | | | | |
| Tissue Level | RN (4.8 mm) | TiBase SSO 4.8 GH 1 L | 6284249 | L |
| | WN (6.5 mm) | TiBase SSO 6.5 GH 1 L | 6284256 | L |
| Bone Level | NC (3.3 mm) | TiBase S BL 3.3 GH 1 L | 6308154 | L |
| | RC (4.1 mm / 4.8 mm) | TiBase S BL 4.1 GH 1 L | 6308337 | L |
| Thommen Medical | | | | |
| Element, Contact | 3.5 | TiBase TM 3.5 GH 1 S | 6531854 | S |
| | 4 | TiBase TM 4 GH 1 S | 6532829 | S |
| | 4.5 | TiBase TM 4.5 GH 1 S | 6532837 | S |
| | 5 | TiBase TM 5 GH 1 S | 6544360 | S |
| | 6 | TiBase TM 6 GH 1 S | 6544378 | S |
| | Implant System | Titanium Base | | |
| Manufacturer / line | Platform | Name | REF | Size |
| Osstem / Hiossen | | | | |
| Osstem TS | Mini | TiBase O TS 3.5 GH 1 L | 6527035 | L |
| (USA: Hiossen ET) | Regular | TiBase O TS 4.0 GH 1 L | 6527043 | L |
| Zimmer /Biomet | | | | |
| | 3.4 | TiBase B O 3.4 GH 1 L | 6282557 | L |
| External hex | 4.1 | TiBase B O 4.1 GH 1 L | 6282565 | L |
| | 5.0 | TiBase B O 5.0 GH 1 L | 6282573 | L |
| | 3.4 | TiBase B C 3.4 GH 1 S | 6308048 | S |
| Certain | 4.1 | TiBase B C 4.1 GH 1 L | 6308097 | L |
| | 5.0 | TiBase B C 5.0 GH 1 L | 6308121 | L |
| | 3.5 | TiBase Z TSV 3.5 GH 1 L | 6282581 | L |
| Tapered Screw-Vent | 4.5 | TiBase Z TSV 4.5 GH 1 L | 6282599 | L |
| | 5.7 | TiBase Z TSV 5.7 GH 1 L | 6282607 | L |
| M.I.S | | | | |
| C1 Conical | NP | CN-TB001 C1 NP GH 0.5 | CN-TB001 | L |
| Connection | | CN-TB015 C1 NP GH 1.5 | CN-TB015 | L |
| V3 Conical | | VN-TB001 V3 NP GH 0.5 | VN-TB001 | L |
| Connection | NP | VN-TB015 V3 NP GH 1.5 | VN-TB015 | L |
| V3 Conical | | CS-TB001 SP GH 0.5 | CS-TB001 | L |
| Connection, C1 | SP | CS-TB015 SP GH 1.5 | CS-TB015 | L |
| Conical Connection | | CS-TB030 SP GH 3 | CS-TB030 | L |
| | | CW-TB001 C1 WP GH 0.5 | CW-TB001 | L |
| C1 Conical<br>Connection | WP | CW-TB015 C1 WP GH 1.5 | CW-TB015 | L |
| | | CW-TB030 C1 WP GH 3 | CW-TB030 | L |
| | | MN-TB001 INT HEX NP GH 0.5 | MN-TB001 | L |
| | NP | MN-TBC15 INT HEX NP GH 1.5 | MN-TBC15 | L |
| | | MD-TB001 INT HEX SP GH 0.5 | MD-TB001 | L |
| SEVEN internal hex,<br>M4 internal hex | SP | MD-TBC15 INT HEX SP GH 1.5 | MD-TBC15 | L |
| | | MD-TBC30 INT HEX SP GH 3 | MD-TBC30 | L |
| | WP | MW-TB001 INT HEX WP GH 0.5 | MW-TB001 | L |
| | | MW-TBC15 INT HEX WP GH 1.5 | MW-TBC15 | L |
| | | MW-TBC30 INT HEX WP GH 3 | MW-TBC30 | L |
| Straumann | | | | |
| | NNC (3.5 mm) | NNC Variobase C 3.5 GH 1 | 220.018 | S |
| Tissue Level | RN (4.8 mm) | RN Variobase C GH 1 | 220.019 | L |
| | WN (6.5 mm) | WN Variobase C GH 1 | 220.020 | L |
| Bone Level | NC (3.3 mm) | NC Variobase C 3.3 GH 1 | 220.043 | S |
| | RC (4.1 mm / 4.8 mm) | RC Variobase C 4.1 GH 1 | 220.044 | L |
| Camlog | | | | |
| | 3.3 | CAMLOG® Titanium base CAD/CAM, for Ø 3.3 mm GH<br>0.4 | K2244.3348 | S |
| Camlog | 3.8 | CAMLOG® Titanium base CAD/CAM ,for Ø 3.8 mm GH<br>0.3 | K2244.3848 | S |
| | 4.3 | CAMLOG® Titanium base CAD/CAM, for Ø 4.3 mm GH<br>0.3 | K2244.4348 | S |
| | 5.0 | CAMLOG® Titanium base CAD/CAM, for Ø 5.0 mm GH<br>0.3 | K2244.5048 | L |
| | 6.0 | CAMLOG® Titanium base CAD/CAM, for Ø 6.0 mm GH<br>0.3 | K2244.6048 | L |
{11}------------------------------------------------
{12}------------------------------------------------
#### Substantial Equivalence: 6.
The indications for use of the proposed device do not constitute a new intended use. The proposed CEREC Cercon 4D™ Abutment Block has the same intended use and similar indications for use compared to the predicate, CEREC Tessera Abutment Block, CEREC Tessera Abutment System (K221402) and reference device Sirona Dental CAD/CAM System (inCoris ZI meso material) (K111421) in that they are used with TiBases, Implant Systems and CAD/CAM technology. The main difference is the compatible Implant systems and titanium bases. The block material and TiBase make up an a 2-piece hybrid abutment which is used in conjunction with endosseous dental implants to restore the function and aesthetics in the oral cavity. Table 6.2 compare the Intended Use and Indications for Use when compared to the predicate and reference devices.
| Table 6.1-Comparison of Intended Use | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| CEREC Cercon 4D™ Abutment<br>Blocks, CEREC Cercon 4DTM<br>Abutment System (K234018) | Predicate Device<br>CEREC Tessera Abutment Block,<br>CEREC Tessera Abutment System<br>(K221402) | Reference Device<br>Sirona Dental CAD/CAM System<br>(inCoris ZI meso material)<br>(K111421) | Discussion | |
| CEREC Cercon 4D™ Abutment<br>Blocks are a ceramic intended for<br>fixed dental prosthetic restorations. | CEREC Tessera Abutment Blocks are a<br>ceramic intended for fixed dental<br>prosthetic restorations. | inCoris ZI meso blocks are intended for use<br>in manufacturing individually designed<br>mesostructures, which are glued to a fitting<br>titanium base after milling and sintering. | The proposed and predicate<br>device have the same Intended<br>Use.<br>The Intended Use of the<br>reference device is similar to<br>Intended Use of the proposed<br>and predicate devices in that the<br>mesostructure is cemented to the<br>titanium base as part of a two-<br>piece hybrid abutment. | |
| Table 6.2 - Comparison of Indications for Use | | | | |
| CEREC Cercon 4D™ Abutment<br>Blocks, CEREC Cercon 4DTM<br>Abutment System (K234018) | Predicate Device<br>CEREC Tessera Abutment Block,<br>CEREC Tessera Abutment System<br>(K221402) | Reference Device<br>Sirona Dental CAD/CAM System<br>(inCoris ZI meso material)<br>(K111421) | Discussion | |
| CEREC Cercon 4D™ Abutment<br>System is intended for use in<br>partially or fully edentulous<br>mandibles and maxillae in support<br>of single cement-retained<br>restorations.<br>The system comprises three parts:<br>• CEREC Cercon 4D™ Abutment<br>Block<br>• TiBases<br>• CAD/CAM system | CEREC Tessera Abutment System is<br>intended for use in partially or fully<br>edentulous mandibles and maxillae in<br>support of single cement-retained<br>restorations.<br>The system comprises three parts:<br>- CEREC Tessera Abutment Block<br>- TiBase<br>- CAD/CAM system | The Sirona Dental CAD/CAM System is<br>intended for use in partially or fully<br>edentulous mandibles and maxillae in<br>support of single or multiple-unit cement<br>retained restorations. The system consists of<br>three major parts: TiBase, InCoris<br>mesostructure, and CAD/CAM software. | Same as predicate and<br>reference devices.<br>The reference device is<br>additionally indicated for<br>multiple-unit cement<br>retained restorations which<br>pertains to other materials<br>used within the CAD/CAM<br>system. | |
| The CEREC Cercon 4D™ ceramic<br>structure cemented to the TiBase is<br>recommended for two-piece hybrid<br>abutments for single tooth<br>restorations and hybrid abutment<br>crowns, used in conjunction with<br>endosseous dental implants. | The CEREC Tessera ceramic structure<br>cemented to the TiBase is recommended for<br>two-piece hybrid abutments for single tooth<br>restorations and hybrid abutment crowns,<br>used in conjunction with endosseous dental<br>implants. | Specifically, the InCoris mesostructure and<br>TiBase components make up a two-piece<br>abutment which is used in conjunction with<br>endosseous dental implants to restore the<br>function and aesthetics in the oral cavity. The<br>InCoris mesostructure may also be used in<br>conjunction with the Camlog Titanium base<br>CAD/CAM (types K2244.xxxx) (K083496)<br>in the Camlog Implant System. The<br>CAD/CAM software is intended to design<br>and fabricate the InCoris mesostructure. | The proposed and predicate<br>devices are the same.<br>The reference device is<br>similar with an additional<br>use with Camlog Implant<br>System and Titanium base.<br>The proposed block is also<br>indicated for use with the<br>Camlog (3rd party products).<br>Fatigue testing to support the<br>use with 3rd party implant<br>systems and TiBases is<br>included in the Performance<br>Bench Testing section. | |
| Table 6.2 - Comparison of Indications for Use | | | | |
| CEREC Cercon 4D™ Abutment<br>Blocks, CEREC Cercon 4D™<br>Abutment System (K234018) | Predicate Device<br>CEREC Tessera Abutment Block,<br>CEREC Tessera Abutment System<br>(K221402) | Reference Device<br>Sirona Dental CAD/CAM System<br>(inCoris ZI meso material)<br>(K111421) | Discussion | |
| Implant Systems:<br>• Dentsply Sirona: PrimeTaper EV,<br>OmniTaper EV, AstraTech<br>OsseoSpeed TX, Frialit / XiVE,<br>AstraTech Implant EV, Ankylos<br>• BioHorizons: Internal connection<br>• Nobel Biocare: Replace, Replace<br>Select, Nobel Active,<br>NobelReplace Conical<br>Connection , Brånemark,<br>NobelSpeedy Groovy<br>• Straumann: Tissue Level, Bone<br>Level<br>• Thommen Medical: Element,<br>Contact<br>• Osstem/Hiossen: Osstem TS<br>, USA:Hiossen ET<br>• Zimmer/Biomet: External hex,<br>Certain, Tapered Screw-Vent<br>• MIS: C1 Conical<br>Connection, V3 Conical<br>Connection, SEVEN | The compatible Implant systems, titanium<br>bases and CAD/CAM systems are shown<br>below:<br>Implant Systems:<br>• Dentsply Sirona: AstraTech OsseoSpeed<br>TX (K053384), XiVE (K013867),<br>AstraTech Implant EV (K120414),<br>Ankylos (K083805), PrimeTaper EV<br>(K210610)<br>• MIS: C1 Conical connection (K172505<br>NP, K180282 WP), V3 Conical connection<br>(K163349), SEVEN internal hex<br>(K112162), M4 internal hex (K112162)<br>CAD/CAM Systems:<br>• Sirona Dental CAD/CAM System<br>(K193408, K200191)<br>Titanium Bases:<br>• Refer to Table 6.2a for the<br>compatible Titanium Bases | The InCoris mesostructure and TiBase two-<br>piece abutment is compatible with the<br>following implant systems:<br>• Nobel Biocare Replace (K020646)<br>• Nobel Biocare Branemark (K022562)<br>• Friadent Xive (K001386)<br>• Biomet 3i Osseotite (K980549)<br>• Astra Tech Osseospeed (K091239)<br>• Zimmer Tapered Screw-Vent (K061410)<br>• Straumann SynOcta (K061176)<br>• Straumann Bone Level (K053088)<br>• Biomet 3i Certain (K014235)<br>• Nobel Biocare Active (K071370) | Proposed device does not<br>include implant system or<br>titanium bases in the<br>Indications for Use<br>statement. Instead, a<br>compatibility table for the<br>Implant Systems and<br>TiBases is included within<br>the Instructions for Use<br>(IFU). | |
| Table 6.2 - Comparison of Indications for Use | | | | |
| CEREC Cercon 4D™ Abutment<br>Blocks, CEREC Cercon 4D™<br>Abutment System (K234018) | Predicate Device<br>CEREC Tessera Abutment Block,<br>CEREC Tessera Abutment System<br>(K221402) | Reference Device<br>Sirona Dental CAD/CAM System<br>(inCoris ZI meso material)<br>(K111421) | Discussion | |
| internal hex, M4 internal hex | | | | |
| • Altatec – Camlog | | | | |
| CAD/CAM Systems: | | | | |
| • Sirona Dental CAD/CAM System | | | | |
| Refer to compatible Titanium Bases | | | | |
| and Implant Systems Table above | | | | |
| in the Indications for Use-Section 5. | | | | |
| Table 6.2a- Compatibility Table included in the Predicate Device Indications for Use Statement above |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.