Cercon® is indicated for all ceramic restorations for anterior and posterior locations, including crowns and 3-unit bridges in anterior and posterior regions.
Device Story
Cercon is a partially sintered yttria-stabilized tetragonal zirconia (Y-TZP) ceramic blank. Supplied to dental professionals, the blank is machined and sintered to fabricate all-ceramic restorations (crowns, 3-unit bridges). Used in dental clinics/labs; operated by dental professionals. Output is a custom-fabricated dental restoration, which may be veneered with compatible porcelains. The device provides a durable, esthetic restorative material for anterior and posterior teeth, aiding in the replacement of missing or damaged tooth structure.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including physical property evaluation per ISO 6872:2015 (flexural strength, chemical solubility, radioactivity, fracture toughness) and biocompatibility testing per ISO 10993-5. Finished restorations were evaluated for fit, margin, and surface quality against predicate Cercon ht.
Technological Characteristics
Material: Yttria-stabilized tetragonal zirconia (Y-TZP). Form factor: Millable ceramic blank. Standards: ISO 6872:2015 (Type II, Class 4). Processing: Machining followed by sintering. Biocompatibility: ISO 10993-5. No electronic components or software.
Indications for Use
Indicated for all ceramic restorations, including crowns and 3-unit bridges, in anterior and posterior regions for dental patients.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular text element and an abstract symbol. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. The abstract symbol is a stylized representation of a human figure, with three overlapping profiles suggesting a sense of community or interconnectedness.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2017
Dentsply Sirona Mr. Karl Nittinger Senior Manager, Regulatory Affairs 221 West Philadelphia Street Suite 60 York, Pennsylvania 17404
Re: K162888
Trade/Device Name: Cercon Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: January 25, 2017 Received: January 26, 2017
Dear Mr. Karl Nittinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Susan Runna, DDS, MA
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162888
Device Name Cercon®
Indications for Use (Describe)
Cercon® is indicated for all ceramic restorations for anterior and posterior locations, including crowns and 3-unit bridges in anterior and posterior regions.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street, Suite 60 York, PA 17404
Image /page/3/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a stylized leaf-like shape on the left, followed by the text "Dentsply" on the top line and "Sirona" on the bottom line. The text is in a sans-serif font and appears to be a dark gray color.
# SECTION 5. 510(k) SUMMARY
for
Cercon® (K162888)
- 1. Submitter Information: Dentsply Sirona 221 West Philadelphia Street Suite 60 York, PA 17404
| Contact Person: | Karl Nittinger |
|-------------------|----------------|
| Telephone Number: | 717-849-4424 |
| Fax Number: | 717-849-4343 |
Date Prepared: February 22, 2017
#### 2. Device Name:
| • Proprietary Name: | Cercon® |
|------------------------|-----------------------------------|
| • Common Name: | zirconia based millable blank |
| • Classification Name: | Porcelain powder for clinical use |
| • CFR Number: | 21 C.F.R. 872.6660 |
| • Device Class: | II |
| • Product Code: | EIH |
# 1. Predicate Device:
| Predicate Device Name | 510(k) | Company Name |
|-----------------------|---------|-------------------------------|
| BruxZir™ Anterior | K143330 | Prismatik Dentalcraft,<br>Inc |
| Reference Device for<br>Biocompatibility and composition | 510(k) | Company Name |
|----------------------------------------------------------|---------|---------------|
| Cercon® Base | K051462 | DeguDent GmbH |
| Cercon® ht | K112152 | DeguDent GmbH |
| Reference Device for specific<br>indications for use statement | 510(k) | Company Name |
|----------------------------------------------------------------|---------|---------------|
| CELTRA Press | K161269 | DeguDent GmbH |
{5}------------------------------------------------
#### 4. Description of Device:
The subject device, Cercon® is a partially sintered ceramic blank composed of yttria stabilized zirconium oxide, (yttria stabilized tetragonal zirconia particle, Y-TZP). The subject device. Cercon® is supplied to dental professionals as a blank and then processed by machining and subsequent sintering to fabricate all ceramic restorations. Cercon® blanks are zirconia based millable blanks (yttria-) stabilized zirconium oxide (zirconia) (Y-TZP)) that are used for all ceramic restorations for anterior and posterior locations. Specifically, Cercon® is indicated for all ceramic restorations for anterior and posterior locations, including crowns and 3-unit bridges in anterior and posterior regions.
The subject device, Cercon® is available in various shades (colors) for esthetics. Specifically, the subject Cercon® material is offered in the following shades: A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4 and white.
The sintered restorations which are fabricated using the subject Cercon® material may be veneered with compatible veneering porcelains.
- 5. Indications for Use:
Cercon® is indicated for all ceramic restorations for anterior and posterior locations, including crowns and 3-unit bridges in anterior and posterior regions.
- 6. Substantial Equivalence:
The proposed Cercon® has similar indications for use as the predicate BruxZir™ Anterior, K143330. The indications are specifically defined as crowns and 3-unit bridges in anterior and posterior regions. The proposed Cercon® meets the requirement of flexural strength for this indication in accordance with ISO 6872:2015 (Dentistry -Ceramic materials) Type II Class 4 material. See Technological Characteristics for more details.
{6}------------------------------------------------
| Proposed Device | Primary Predicate Device | Secondary Predicate Device | Differences |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cercon®<br>K162888 | BruxZir™<br>Anterior<br>K143330 | CELTRA Press<br>K161269 | |
| Indications for Use:<br>Cercon® is indicated<br>for all ceramic<br>restorations for<br>anterior and posterior<br>locations, including<br>crowns and 3-unit<br>bridges in anterior and<br>posterior regions. | Indications for Use:<br>The device is<br>indicated for use by<br>dental technicians in<br>the construction of<br>custom made all<br>ceramic restorations<br>for anterior and<br>posterior location. | Indications for Use:<br>CELTRA Press is an all-<br>ceramic system for the<br>creation of<br>Occlusal veneers<br>Thin veneers<br>Veneers<br>Inlays<br>Onlays<br>Crowns in the anterior and<br>posterior region<br>3-unit bridges in the anterior<br>and posterior region<br>3-unit bridges in the<br>premolar region up to the<br>second premolar as the<br>terminal abutment<br>Crown, splinted crown or 3<br>unit bridge up to the second<br>premolar placed on top of<br>an implant abutment | The Indications<br>for Use for the<br>proposed device,<br>Cercon® includes<br>additional clarity<br>and are<br>specifically<br>defined as<br>intended for use<br>in the fabrication<br>of crowns and 3-<br>unit bridges in<br>anterior and<br>posterior regions. |
# Table 5.1 Indications for Use Comparison
### Composition comparison
The proposed device, Cercon® is primarily composed of Zirconium Oxide (ZrO2). Though the exact composition of the predicate BruxZir™ Anterior, K143330 is not available; the 510(k) Summary for the BruxZir™ Anterior 510(k) K143330 describes the predicate as zirconia milling blanks which are the same as the proposed device, Cercon®.
The proposed device, Cercon® consists of the same base (primary) composition as previously cleared devices, Cercon® Base (K051462) and Cercon® ht (K112152). Base composition for the proposed device Cercon® as well as reference devices Cercon® Base (K051462) and Cercon® ht (K112152) is Zirconium Oxide (ZrO2).
### Technological Characteristics:
The proposed device, Cercon® was classified and tested in accordance with ISO 6872:2015 (Dentistry - Ceramic materials). Cercon® classification and clinical usage is noted below:
Type II: All other forms of ceramic products
Class 4:
Class 4a: Monolithic ceramic for three-unit prostheses involving molar restoration Class 4b: Partially or fully covered substructure for three-unit prostheses involving molar restoration
{7}------------------------------------------------
| Physical<br>Properties<br>ISO 6872 | Proposed Device<br>Cercon®<br>K162888 | Predicate Device<br>BruxZir™ Anterior<br>K143330 | ISO 6872:2015<br>Type II Class 4 | Differences |
|----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Flexural strength | Meets requirements per ISO<br>6872:2015 | Meets requirements, per<br>ISO 6872:2008 | Meets requirements per<br>ISO 6872:2015 | The proposed device Cercon®<br>is a Type II class 4 device in |
| Chemical solubility | Meets requirements per ISO<br>6872:2015 | Not listed | Meets requirements per<br>ISO 6872:2015 | accordance with ISO<br>6872:2015 standard while the<br>primary predicate BruxZir™ |
| Coefficient of thermal<br>expansion, CTE | A4<br>$10.1×10^{-6} K^{-1}$ (25-500°C)<br>$10.2×10^{-6} K^{-1}$ (25-600°C)<br>C4<br>$10.0×10^{-6} K^{-1}$ (25-500°C)<br>$10.1×10^{-6} K^{-1}$ (25-600°C) | $11 x 10^{-6}$ / K<br>(100-500°C) | $9.80±0.06 x 10^{-6} K^{-1}$<br>(25°C-500°C) | Anterior, K143330 is a Type<br>II class 6 (per 510(k)<br>summary) in accordance with<br>ISO 6872:2008 standard.<br>Both the proposed device<br>Cercon® and the predicate |
| Radioactivity | Meets requirements per ISO<br>6872:2015 | Not listed | Meets requirements per<br>ISO 6872:2015 | device, BruxZir™ Anterior<br>(K143330) meet the<br>requirements for the |
| Uniformity | Meets requirements per ISO<br>6872:2015 | Not listed | Not listed | appropriate type and class for<br>its clinical usage. The |
| Mixing and<br>Condensation<br>Properties of type I<br>ceramics | Not applicable | Not listed | Not listed | secondary predicate, CELTRA<br>Press (K161269) is a type II<br>class 4 material per ISO<br>6872:2015 and meets the |
| Shrinkage factor | Meets requirements per ISO<br>6872:2015 | Not listed | Not listed | requirements for its<br>indications for use same as the |
| Glass transition<br>temperature | Not applicable | Not listed | Not listed | proposed device, Cercon®. |
| Freedom from<br>extraneous materials | Meets requirements per ISO<br>6872:2015 | Not listed | Not listed | |
| Fracture Toughness | Meets requirements per ISO<br>6872:2015 | Not listed | Not listed | |
# Table 5.3 Physical Properties per ISO 6872 standard
{8}------------------------------------------------
Both the proposed device Cercon® and the predicate device, BruxZir™ Anterior (K143330) meet the requirements for the appropriate type and class for its clinical usage. The secondary predicate, CELTRA Press (K161269) is a type II class 4 material per ISO 6872:2015 and meets the requirements for its indications for use same as the proposed device. Cercon®. The strength of the proposed Cercon® meets the requirement for ISO 6872:2015 for the defined indications for use.
#### 7. Non-Clinical Performance Data:
# Physical properties
In accordance with ISO 6872:2015 (Dentistry - Ceramic materials), the product must meet the requirements for the appropriate type and class for its clinical usage. Both the proposed device, Cercon®, and the predicate device, BruxZir™ Anterior (K143330), meet the requirements of the applicable type and class for its clinical usage (Table 5.3).
In addition to ISO testing, verification testing on finished restorations fabricated with the Cercon material was conducted.. All restorations were milled and sintered in accordance with the directions instructions for use and visual inspection for cracks was performed. All finished dental restorations were found to be crack-free.
Comparative evaluation of finished restorations fabricated using the subject Cercon® material and the predicate Cercon ht (K112152) material was also conducted. The restorations were evaluated for fit, margin, surface quality, and overall impression of dental objects (dental restorations). Results of the evaluation met the predetermined acceptance criteria.
The performance of the proposed device, Cercon® satisfactorily met the requirements of the non-clinical bench testing conducted to support substantial equivalence.
# Biocompatibility Testing
Cytotoxicity testing according to ISO 10993-5 (Biological evaluation of medical devices) was completed to characterize the biocompatibility of the proposed device Cercon®. In addition, previously cleared devices (Cercon® Base, K051462 and Cercon® ht, K112152) were utilized to support biocompatibility for the proposed device Cercon®. Results of the testing support the biocompatibility of proposed device, Cercon®.
#### 8. Clinical Performance Data:
Not applicable. No data from human clinical studies has been included to support the substantial equivalence of the proposed device, Cercon®.
{9}------------------------------------------------
### 9. Conclusion Regarding Substantial Equivalence:
The comparative information included in this section regarding the proposed device, Cercon®, supports substantial equivalence to the predicate device, BruxZir™ Anterior cleared under premarket notification K143330 . Substantial equivalence of the subject Cercon device is supported by the following determinative factors:
- It has the same intended use and incorporates the same fundamental technology as the . primary predicate device (BruxZir™ Anterior, K143330).
- . It has the same specific indications for use as secondary predicate device (CELTRA Press, K161269).
- It has the same base (primary) composition Zirconium Oxide (ZrO2) as the reference . devices Cercon® Base (K051462) and Cercon® ht (K112152).
- . It has similar technological characteristics as the predicate. The proposed device, Cercon® as well the predicate device, BruxZir™ Anterior, K143330 meet the requirement of the related type and class for its usage.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.