MINI TREK RX CORONARY DILATATION CATHETER (INCLUDING 1.20 MM SIZE), MINI TREK OTW CORONARY DILATATION CATHETER (INCLUDIN

K123279 · Abbott Vascular · LOX · Jan 30, 2013 · Cardiovascular

Device Facts

Record IDK123279
Device NameMINI TREK RX CORONARY DILATATION CATHETER (INCLUDING 1.20 MM SIZE), MINI TREK OTW CORONARY DILATATION CATHETER (INCLUDIN
ApplicantAbbott Vascular
Product CodeLOX · Cardiovascular
Decision DateJan 30, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5100
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MINI TREK RX, MINI TREK OTW and MINI TREK II OTW Coronary Dilatation Catheters (balloon models 1.50 mm - 2.00 mm) are indicated for: · Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion · Balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction · Balloon dilatation of a stent after implantation (balloon models 2.00 mm only) · Balloon dilatation of de novo chronic total coronary occlusions (CTO) The MINI TREK RX, MINI TREK OTW and MINI TREK II OTW Coronary Dilatation Catheters (balloon models 1.20 mm) are indicated for: · Initial balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis (≥ 70% stenosis) · Balloon dilatation of de novo chronic total coronary occlusions (CTO)

Device Story

MINI TREK series are co-axial coronary dilatation catheters (RX and OTW designs) featuring a distal balloon. Catheters are advanced over a guide wire to a coronary stenosis; the balloon is inflated with contrast fluid to dilate the lesion. Radiopaque markers under the balloon assist in positioning. Used in clinical settings by interventional cardiologists. Output is mechanical dilation of the vessel, visualized via angiography. Benefits include improved myocardial perfusion, restored coronary flow in STEMI, and successful pre-dilatation of CTOs. The device facilitates stent placement or vessel recanalization.

Clinical Evidence

Clinical data from the EXPERT CTO trial (N=88) evaluated the MINI TREK catheter for CTO pre-dilatation. Primary endpoint: successful pre-dilatation (delivery, inflation/deflation, absence of significant complications, final TIMI 3 flow). Success achieved in 93.2% (95% CI: 85.7%, 97.5%). Secondary endpoints included device success (100%) and procedural success (97.7%). No clinically significant perforations occurred; in-hospital MACE rate was 2.27%.

Technological Characteristics

Co-axial catheter design with distal balloon. Materials tested per ISO 10993-1:2003. Features include radiopaque markers, hypotube proximal shaft, and inflation/guide wire lumens. Connectivity: None (mechanical). Sterilization: Not specified. Algorithm: None.

Indications for Use

Indicated for patients requiring balloon dilatation of coronary artery or bypass graft stenoses, coronary artery occlusions in STEMI, stent post-dilatation, or de novo chronic total coronary occlusions (CTO).

Regulatory Classification

Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b) *Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2) *Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## SECTION 2 – 510(k) SUMMARY – K123279 The 510(k) Summary is submitted in accordance with 21 CFR 807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990. | 1. SUBMITTER'S NAME | Abbott Vascular | |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------| | 2. SUBMITTER'S ADDRESS | • PO Box 9018, Temecula, CA 92590-9018<br>• 26531 Ynez Road, Temecula, CA 92591 | | 3. TELEPHONE | (951) 914-3243 | | 4. FAX | (951) 914-0339 | | 5. CONTACT PERSON | Suzanne Redman | | 6. DATE PREPARED | October 19, 2012 | | 7. DEVICE TRADE NAME | • MINI TREK™ RX Coronary Dilatation Catheter<br>• MINI TREK™ OTW Coronary Dilatation Catheter<br>• MINI TREK™ II OTW Coronary Dilatation Catheter | | 8. DEVICE COMMON NAME | • Coronary Dilatation Catheter<br>• Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter | | 9. DEVICE CLASSIFICATION NAME | PTCA Catheter, LOX, Class II, 21 CFR 870.5100 | | 10. PREDICATE DEVICE NAME | • MINI TREK™ RX Coronary Dilatation Catheter<br>• MINI TREK™ OTW Coronary Dilatation | - Catheter 1 of 99 JAN 3 0 2013 هــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ of {1}------------------------------------------------ K123279 Page 2 of 9. ### DEVICE DESCRIPTION 11. ### MINI TREK RX Coronary Dilatation Catheter 11.1 The MINI TREK RX Coronary Dilatation Catheter is a rapid exchange co-axial design with a balloon at the distal tip. Table 2-1 provides a matrix of the balloon diameters and lengths available for the MINI TREK RX family of devices. | Balloon<br>Diameter | Balloon Length | | | | | | | |---------------------|----------------|-----|------|------|------|------|------| | (mm) | 6mm | 8mm | 12mm | 15mm | 20mm | 25mm | 30mm | | 1.20 | X | X | X | X | X | | | | 1.50 | X | X | X | X | X | | | | 2.00 | X | X | X | X | X | X | X | MINI TREK RX Size Matrix Table 2-1 The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The co-axial shaft consists of a tubular inner and outer member. The inner member permits the use of a guide wire to facilitate the advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. The proximal shaft consists of a hypotube with a hub on the proximal end, a tapered distal section ending distal to the guide wire notch junction, along with brachial and femoral markers. #### 11.2 MINI TREK OTW Coronary Dilatation Catheter The MINI TREK OTW Coronary Dilatation Catheter is an over-the-wire (OTW) co-axial design with a balloon at the distal tip. Table 2-2 provides a matrix of the balloon diameters and lengths available for the MINI TREK OTW family of devices. | Table 2-2 | MINI TREK OTW Size Matrix | | | | | | | |-----------------------------|---------------------------|-----|------|------|------|------|------| | Balloon<br>Diameter<br>(mm) | 6mm | 8mm | 12mm | 15mm | 20mm | 25mm | 30mm | | 1.20 | X | X | X | X | X | | | | 1.50 | X | X | X | X | X | | | | 2.00 | X | X | X | X | X | X | X | The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The co-axial shaft consists of a tubular inner and outer member. The inner lumen permits the use of a guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. Along the proximal portion of the shaft are brachial and femoral markers to aid in gauging the catheter's position relative to the guiding catheter tip when introducing the {2}------------------------------------------------ K123279 Page 3 of 9 catheter through the guiding catheter. An adaption arm is located at the proximal end to provide access to the inflation lumen and guide wire lumen and allows connection with an inflation device. ### MINI TREK II OTW Coronary Dilatation Catheter 11.3 The MINI TREK II OTW Coronary Dilatation Catheter is an over-the-wire (OTW) co-axial design with a balloon at the distal tip. Table 2-3 provides a matrix of the balloon diameters and lengths available for the MINI TREK OTW family of devices. | Balloon<br>Diameter<br>(mm) | Balloon Length | | | | | |-----------------------------|----------------|-----|------|------|------| | | 6mm | 8mm | 12mm | 15mm | 20mm | | 1.20 | X | X | X | X | X | | 1.50 | X | X | X | X | X | | 2.00 | X | X | X | X | X | Table 2-3 MINI TREK II OTW Size Matrix · The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The co-axial shaft consists of a tubular inner and outer member. The inner lumen permits the use of a guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. Along the proximal portion of the shaft are brachial and femoral markers to aid in gauging the catheter's position relative to the guiding catheter tip when introducing the catheter through the guiding catheter. An adaption arm is located at the proximal end to provide access to the inflation lumen and guide wire lumen and allows connection with an inflation device. ### INDICATIONS FOR USE 12. The MINI TREK RX, MINI TREK OTW and MINI TREK II OTW Coronary Dilatation Catheters (balloon models 1.50 mm - 2.00 mm) are indicated for: - · Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion - Balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary . flow in patients with ST-segment elevation myocardial infarction - · Balloon dilatation of a stent after implantation (balloon models 2.00 mm only) - · Balloon dilatation of de novo chronic total coronary occlusions (CTO) The MINI TREK RX, MINI TREK OTW and MINI TREK II OTW Coronary Dilatation Catheter s(balloon models 1.20 mm) are indicated for: - · Initial balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis (≥ 70% stenosis) 3 of 9 {3}------------------------------------------------ K123279Page 4 of 9 · Balloon dilatation of de novo chronic total coronary occlusions (CTO) ### TECHNOLOGICAL CHARACTERISTICS 13. Comparisons of the new and predicate devices show that the technological characteristics such as product performance, design and intended use are substantially equivalent to the current marketed predicate devices. #### PERFORMANCE DATA 14. ### 14.1 Summary of Biocompatibility Testing Biocompatibility testing, previously conducted, included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, and complement activation according to the recommendations of Guidance for Industry and FDA Staff - Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010 and ISO 10993-1:2003, Biological evaluation of medical devices - Part 1: Evaluation and testing. #### 14.2 Summary of In Vitro Bench Testing The MINI TREK RX, MINI TREK OTW and MINI TREK II OTW Coronary Dilatation Catheters were previously subjected to the following in vitro bench tests according to the requirements of Guidance for Industry and FDA Staff- Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010: - Catheter Preparation . - Balloon Crossing Profile ● - Refolded Balloon Profile - . Balloon Inflation / Balloon Deflation - . Balloon Fatigue - . Balloon Fatigue Within a Stent - Balloon Rupture . - . Balloon Rupture Within a Stent - Balloon Compliance . - . Catheter Shaft Fatigue - Catheter Shaft Rupture - . Soft Tip to Inner Member Tensile - Proximal Balloon Seal Tensile . - . Distal Catheter Tensile - Catheter Coating Particulate . - . Catheter Coating Integrity - . Catheter Coating Friction - Kink and Flexibility ● - Torque . {4}------------------------------------------------ - Radiopacity . #### 14.3 Summary of Clinical Data The primary objective of the EXPERT CTO clinical trial is to assess the safety and effectiveness of the XIENCE V Everolimus Eluting Coronary Stent System, XIENCE nano Everolimus Eluting Coronary Stent System and the XIENCE PRIME LL Everolimus Eluting Coronary Stent System for the treatment of chronic total coronary occlusions. Another key objective of this trial is to assess the safety and effectiveness of the MINI TREK Coronary Dilatation Catheter in predilatation of chronic total occlusions. Assessment of the MINI TREK related objective of the trial has been completed. Assessment of the MINI TREK related objective was to be performed in at least the initial 60 subjects with successful guide wire crossing, identified as confirmation of the guide wire in the distal true lumen. However, evaluable data (including core laboratory ascertained angiography data immediately following pre-dilatation with the MINI TREK Coronary Dilatation Catheter) was not available in all 60 subjects, and therefore a total analysis population of 88 subjects was reached that included 65 subjects with evaluable data for the MINI TREK related primary analysis, and 23 subjects that did not have core laboratory ascertained angiography data immediately following pre-dilatation with the MINI TREK Coronary Dilatation Catheter. In both the total analysis population (N=88) and the evaluable population (N=65), the MINI TREK Coronary Dilatation Catheter was used at first attempt for pre-dilatation of the CTO. Results are presented below for the total analysis population, as this set of results best represents the performance of the MINI TREK Coronary Dilatation Catheter in the entire intended CTO study population of the EXPERT CTO trial. Abbott Vascular notes that assessment of the MINI TREK related objective of the EXPERT CTO trial involved analysis of only the MINI TREK Coronary Dilatation Catheter cohort (subpopulation of the full EXPERT CTO study), for only data derived from the index procedure through the in-hospital visit. Subjects with signs and/or symptoms considered typical of ischemic heart disease attributed to a CTO in a native coronary artery, who were suitable for a percutaneous revascularization. were included. Subjects with evidence of acute MI within 72 hours of the intended treatment were excluded. The target lesion was a de novo lesion with at least one target segment in a native coronary vessel meeting the definition of chronic total occlusion. Only one target lesion was allowed to be treated. ### Primary Endpoint The angioplasty pre-dilatation related primary endpoint was successful pre-dilatation of the CTO defined as follows: (1) successful delivery of the MINI TREK Coronary Dilatation Catheter to and across the target lesion and; (2) successful inflation and deflation of the MINI TREK Coronary Dilatation Catheter and; (3) absence (as determined by independent 5 of 9 <sup>&#</sup>x27; A total of 65 subjects with evaluable data were enrolled in the MINI TREK Coronary Dilatation Catheter cohort of the EXPERT CTO due to ongoing enrollment of the cohor that occurred in parallel with analyses to determine when at least 60 subjects with evaluable data had been enrolled. {5}------------------------------------------------ angiographic core laboratory assessment) of clinically significant vessel perforation, flowlimiting vessel dissection, reduction in thrombolysis in myocardial infarction (TIMI) from baseline and clinically significant arrhythmias requiring medical treatment or device intervention following dilatation with MINI TREK and; (4) achievement of final TIMI flow 3 for the target lesion at the conclusion of the index procedure (i.e. after stent implantation). K123279 Pogo 6 ### Baseline Subject Characteristics: The 88 subjects in the total analysis population for the angioplasty predilatation-related endpoint had a mean (±SD) age of 61.52 (±10.37) years and 76.1% (67/88) were men and 23.9% (21/88) were women. A total of 94.3% (83/88) of the subjects were dyslipidemic; 90.8% (79/87) were hypertensive and 39.8% (35/88) were diabetics with 31.4% (11/35) of the diabetic subjects requiring insulin. Cardiac history revealed prior MI in 31.0% (26/84) of the subjects and previous percutaneous coronary intervention in 45.5% (40/88). In addition, 32.2% (28/87) of the subjects had history of smoking within the last month prior to enrollment. ### Baseline Target Lesion Characteristics: A total of 88 target lesions were treated in 88 subjects in the total analysis population. Angiographic core lab data immediately post-MINI TREK pre-dilatation were available for 65 target lesions (evaluable population subset of the total analysis population), and angiographic core lab data at the end of the index procedure (i.e. after stent implantation) were available for all 88 target lesions in the total analysis population. Of the target lesions treated in the total analysis population, 35.2% (31/88) were located in the LAD artery, 14.8% (13/88) were located in the LCX, and 50.0% (44/88) were located in the RCA. The mean (±SD) lesion length was 36.68 (±17.86) mm: one lesion (1.1%, (1/88)) was < 10 mm, 14.8% (13/88) of the lesions were 10-19.9 mm and 84.1% (74/88) of them were > 20 mm. The mean (±SD) occlusion length was 13.82 (±8.67) mm. Assessment of the target lesions in the total analysis population at baseline revealed moderate calcification in 20.5% (18/88) and severe calcification in 14.8% (13/88) of the lesions, and 9.2% (8/87) were eccentric and 1.1% (1/88) had thrombus. Assessment by the angiographic core laboratory for the total analysis population included mean (±SD) pre-procedure reference vessel diameter (RVD) of 2.59 (±0.45) mm. mean (±SD) pre-procedure MLD of 0.01 (±0.03) mm, and mean (±SD) pre-procedure % diameter stenosis (DS) of 99.77 (±1.04) %. Pre-procedure TIMI flow of 0 was noted in 95.5% (84/88), and pre-procedure TIMI flow of 1 was noted in 4.5% (4/88) of the lesions in the total analysis population. ### Primary Endpoint Results: In the total analysis population (N=88), the angioplasty pre-dilatation related primary endpoint was assessed using site-reported data for the third component of the endpoint where angiographic core laboratory data was not available. This analysis yielded numerically 6 ºf 9 {6}------------------------------------------------ K123279 Page 7 of 9 similar primary endpoint results as the primary endpoint analysis of the evaluable population only.2 The primary endpoint, successful pre-dilatation of the CTO, was achieved in 93.2% (82/88) of the 88 subjects in the total analysis population with a 95% confidence interval (CI) of [85.7%, 97.5%]. The individual criteria for successful pre-dilatation included: - 1. successful delivery of at least one MINI TREK balloon to and across the target lession in 96.6% (85/88) of the total analysis population, - 2. successful inflation and deflation with at least one MINI TREK balloon in 100.0% . (85/85)3 of the total analysis population, - 3. absence of clinically significant vessel perforation, flow-limiting vessel dissection, reduction in TIMI from baseline or clinically significant arrhythmias requiring medical treatment or device intervention following the dilatation with the MINI TREK balloon in 97.7% (86/88) of the total analysis population (with site-reported data if core lab data not available immediately following MINI TREK pre-dilatation), and - 4. final TIMI 3 flow at the conclusion of the index procedure (i.e. after stent implantation) in 98.9% (87/88) of the total analysis population. | | Balloon<br>Success | 95%CI | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|----------------| | Angioplasty predilatation-related Endpoint | 93.2% (82/88) | [85.7%,97.5%] | | Successful delivery of MINI TREK Coronary<br>Dilatation Catheter to and across target lesion1 | 96.6% (85/88) | [90.4%,99.3%] | | Successful inflation and deflation with MINI<br>TREK Coronary Dilatation Catheter2 | 100.0%<br>(85/85) | [95.8%,100.0%] | | Absence of clinically significant vessel<br>perforation, flow limiting vessel dissection,<br>reduction in TIMI flow from baseline or<br>clinically significant arrhythmias requiring<br>medical treatment or device intervention<br>following dilatation with MINI TREK3 | 97.7% (86/88) | [92.0%,99.7%] | | Achievement of final TIMI flow 3 for the target<br>lesion at the conclusion of the index procedure4 | 98.9% (87/88) | [93.8%,100.0%] | #### Primary Endpoint Results - Total Analysis Population Table 2-4 7 of 9 <sup>2</sup> When assessed among the evaluable subjects only (N=65), the primary endpoint of successful pre-dilatation of the CTO was achieved in 93.8% (61/65) of the evaluable subjects with a 95% Cl of [85.0%, 98.3%]. <sup>3</sup> The denominator includes only subjects in whom inflation of the MINI TREK was attempted (i.e. subjects with successful delivery of the MINI TREK across the target lesion). A subject was considered successful for this component if at least one MINI TREK dilatation catheter was successfully inflated and deflated per the clinical site. {7}------------------------------------------------ 1 Data source: site-reported. A subject was considered successful for this component if at least one MINI TREK dilatation catheter was successfully delivered to and across the target lesion per the clinical site. K123279 2 Data source: site-reported. The denominator includes only subjects in whom inflation and deflation of the MINI TREK was attempted (i.e. subjects with successful delivery of the MINI TREK across the target lesion). A subject was considered successful for this component if at least one MINI TREK dilatation catheter was successfully inflated and deflated per the clinical site 5 Data source: angiographic core laboratory for vessel perforation, flow limiting vessel dissection and reduction in TIMI from baseline (site-reported data was used if angiographic core laboratory data was not available); site-reported for clinically significant arrhythmia. 4 Data source: angiographic core laboratory. ### Secondary Endpoint Results: Device success, defined as attainment of < 50% residual stenosis of the target lesion using the assigned study device, was achieved in 100.0% (88/88) of the lesions in the total analysis population. Procedure success, defined as device success and absence of in-hospital MACE (per the protocol definition), was achieved in 97.7% (86/88) of the total analysis population. In the total analysis population (N=88), the mean (±SD) procedural time was 77.13 (±40.17) minutes. The mean (±SD) contrast volume used during the procedure was 251.27 (±113.14) mL, and the mean (±SD) fluoroscopy duration was 32.88 (±21.05) minutes. Procedural success evaluated according to crossing technique included success in: - 98.7% (77/78) of the total analysis population using Anterograde Only crossing . technique - 100% (4/4) of the total analysis population using Retrograde Only crossing technique . - 66.7% (2/3) of the total analysis population using Combined Anterograde and . Retrograde crossing technique - 100% (3/3) of the total analysis population using multiple crossing techniques . - Other crossing methods than listed above were not used . No clinically significant perforations occurred in the total analysis population, In the total analysis population (N=88), 2.27% (2/88) of the subjects experienced in-hospital MACE events (per WHO-based protocol MI definition), with 1.14% (1/88) experiencing clinically indicated TLR, 1.14% (1/88) experiencing MI, and no in-hospital death. Immediately following pre-dilatation of the target lesion with the MINI TREK balloon, the evaluable population (N=65) had mean MLD (±SD) of 0.44 (±0.37) mm with mean (±SD) change in MLD of 0.39 (±0.35) mm. In addition, 21.5% (14/65) of the lesions had TIMI flow of 0, 20.0% (13/65) of the lesions had TIMI flow of 1, 6.2% (4/65) of the lesions had TIMI flow of 2, and 52.3% (34/65) of the lesions had TIMI flow of 3 following pre-dilatation. The mean (±SD) change in TIMI flow was 1.72 (±1.28). Assessment of these parameters for the entire total analysis population could not be performed, since angiography immediately following MINI TREK pre-dilatation was not available for angiographic core lab assessment for the remaining subjects in the total analysis population. 8 of {8}------------------------------------------------ # K123279 Page ### Conclusions: One of the main objectives of the EXPERT CTO trial was to assess the safety and effectiveness of the MINI TREK Coronary Dilatation Catheter in pre-dilatation of chronic total occlusions. In the total analysis population, with site-reported data where angiography core laboratory data immediately following MINI TREK pre-dilatation were not available, the primary endpoint of successful pre-dilatation of the CTO was achieved in 93.2% (82/88) of the subjects with a 95% CI of [85.7%, 97.5%]. These success rates compare favorably with historical averages for this complex lesion class, and are in accordance with the recent trend towards significantly higher success rates among experienced operators utilizing advanced device technology and procedural techniques. These favorable outcome data demonstrate the MINI TREK to be both effective and safe in pre-dilatation of chronic total occlusions. The effectiveness profile of the MINI TREK is further supported by the increase in TIMI flow and MLD observed immediately following pre-dilatation with the MINI TREK as well as at the conclusion of the procedure. The low rates of procedural complications associated with the use of the MINI TREK, including low in-hospital MACE (per WHO-based protocol MI definition) and low inhospital TLR rates in the total analysis population, further support the safety profile of the-MINI TREK Coronary Dilatation Catheter. The data emerging from the EXPERT CTO study compare favorably to published rates of procedural success and in-hospital complications for the recanalization of CTOs, a notoriously difficult lesion class. These data support the safety and effectiveness of the MINI TREK Coronary Dilatation Catheter for the treatment of chronic total occlusions. #### 14.4 Performance Testing - Conclusion The previously conducted biocompatibility testing, previously conducted in vitro bench testing and new clinical study results demonstrated that the MINI TREK RX, MINI TREK OTW, and MINI TREK II OTW Coronary Dilatation Catheters met all acceptance criteria and performed similarly to the predicate devices. No new safety or effectiveness issues were raised during the testing program and, therefore, these devices may be considered substantially equivalent to the predicate devices. 9 of 9 {9}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized graphic of what appears to be an abstract human form or symbol, composed of curved lines. The logo is black and white. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 JAN 3 0 2013 Abbott Vascular, Inc. C/O Ms. Suzanne Redman 26531 Ynez Road Temecula, CA 92589 Re: K123279 R12277 Coronary Dilatation Catheters Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty Catheter Regulatory Class: Class II Product Code: LOX Dated: December 20, 2012 Received: December 21, 2012 Dear Ms. Redman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {10}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 70 a ttp://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Matthew G. Hillebrenner for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {11}------------------------------------------------ ### SECTION 1 - INDICATIONS FOR USE # 510(k) Number (if known): K1232.79 MINI TREK™ RX Coronary Dilatation Catheter Device Names: MINI TREKTM OTW Coronary Dilatation Catheter MINI TREK™ II OTW Coronary Dilatation Catheter The MINI TREK RX, MINI TRETK OTW and MINI TREK II OTW Indications for Use: Coronary Dilatation Catheters (balloon models 1.50 mm - 2.00 mm) are indicated for: · Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion · Balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction · Balloon dilatation of a stent after implantation (balloon models 2.00 mm only) · Balloon dilatation of de novo chronic total coronary occlusions (CTO) The MINI TREK RX, MINI TREK OTW and MINI TREK II OTW Coronary Dilatation Catheters (balloon models 1.20 mm) are indicated for: · Initial balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis (≥ 70% stenosis) Balloon dilatation of de novo chronic total coronary occlusions . (CTO) Prescription Use X (Per 21 CFR 801.109) Over-The-Counter (Optional Format 1-1-96) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH_Office of Device Evaluation (ODE) CWA Kblebe OR (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K123219
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...