The COOK® MBC PTCA balloon dilatation catheter is indicated for balloon dilatation of hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
Device Story
The COOK® MBC PTCA balloon dilatation catheter is a double-lumen catheter with a distal balloon, used by interventional cardiologists during percutaneous transluminal coronary angioplasty (PTCA). The device is inserted over a 0.014-inch guide wire under fluoroscopic guidance. The balloon, constructed of high-density polyethylene, is inflated using a liquid medium via an external indeflator to dilate stenotic coronary lesions or bypass grafts. Two radiopaque marker bands facilitate positioning. The device serves as either definitive therapy or as a predilatation tool prior to stent placement. By restoring vessel patency, it improves myocardial perfusion and relieves ischemia. It is intended for single-use in a hospital setting with immediate access to surgical backup.
Clinical Evidence
Multicenter, prospective registry study of 150 patients (mean age 62, 65% male) undergoing elective PTCA. Primary endpoints included procedural success and 30-day incidence of major adverse cardiac events (MACE: death, Q-wave MI, CABG, or repeat PTCA). Results: 137 patients treated with the MBC balloon; 22 balloon-only, 115 balloon-plus-stent. 30-day MACE rate was 0.9% (95% CI: 0.0, 3.7). No clinical data for long-term outcomes beyond 30 days. Bench testing confirmed balloon burst strength, fatigue resistance, and inflation/deflation times.
Technological Characteristics
Double-lumen catheter, 135 cm length, 3.5 Fr proximal/3.0 Fr distal OD. Balloon material: high-density polyethylene. Radiopaque marker bands for fluoroscopy. Compatible with 0.014-inch guide wires. Sterilization: Ethylene Oxide (10^-6 SAL). No electronic components or software.
Indications for Use
Indicated for patients with hemodynamically significant coronary artery or bypass graft stenosis evidencing coronary ischemia, requiring balloon dilatation to improve myocardial perfusion. Contraindicated in patients with unprotected left main coronary artery or coronary artery spasm without significant stenosis.
Regulatory Classification
Identification
Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.
Related Devices
K230705 — POT PTCA Balloon Dilatation Catheter · Brosmed Medical Co., Ltd. · Apr 11, 2023
{0}
P990008
# Summary of Safety and Effectiveness Data
COOK® MBC PTCA Balloon Dilatation Catheter
## Table of Contents
I. General Information 2
II. Indications for Use 2
III. Contraindications 2
IV. Warnings and Precautions 2
V. Device Description 3
VI. Alternative Practices and Procedures 4
VII. Marketing History 4
VIII. Adverse Effects of the Device on Health 4
IX. Summary of Non-clinical Studies 4
Table 1. Mean Burst, Rated Burst and Maximum Recommended Pressures 5
Table 2. Pressures Required for Nominal Diameter and Diameter at Rated Burst Pressure 6
Table 3. Balloon Compliance (Diameter vs Pressure) for the COOK® MBC PTCA Catheters 7
Table 4. Deflated Balloon Profile 9
X. Summary of the Clinical Investigation 11
Table 5. Baseline Demographics and Clinical Characteristics (N=150) 14
Table 6. Baseline Angiographic Characteristics of the Culprit Lesion Prior to the Initial PTCA Procedure 15
Figure 1. Patient Flow Diagram 18
Table 7. Status of the Culprit Lesion after Using the MBC Balloon 19
Table 8. Angiographic Core Lab QCA Analysis of the Culprit Lesion 20
Figure 2. Cumulative percent of patients versus Minimum Lumen Diameter 21
Figure 3. Cumulative percent of patients versus Percent Diameter Stenosis 22
Table 9. Principal Effectiveness and Safety Results 23
XI. Conclusions Drawn From Studies 25
XII. Panel Recommendations 26
XIII. FDA Decision 26
XIV. Approval Specifications 26
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{1}
# Summary of Safety and Effectiveness Data
COOK® MBC PTCA Balloon Dilatation Catheter
COOK® INCORPORATED
## I. General Information
Generic Name of the Device: Percutaneous transluminal coronary angioplasty catheter
Device Trade Name: COOK® MBC PTCA Balloon Dilatation Catheter
PMA Number: P990008
Name/Address of Applicant: COOK® INCORPORATED
925 South Curry Pike
P.O. Box 489
Bloomington, IN 47402
## II. Indications for Use
The COOK® MBC PTCA balloon dilatation catheter is indicated for balloon dilatation of hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
## III. Contraindications
- Unprotected left main coronary artery.
- Coronary artery spasm in the absence of a significant stenosis thought to be of hemodynamic significance.
## IV. Warnings and Precautions
### A. Warnings
- This device is intended for one time use only. Do NOT resterilize and/or reuse it.
- Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.
- The inflated diameter of the balloon should approximate the diameter of the coronary artery just proximal and distal to the stenosis. The danger of coronary dissection increases as the balloon to artery ratio increases past 1-to-1.
- The catheter system should be used only by physicians thoroughly trained in the performance of percutaneous transluminal coronary angioplasty.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{2}
- When the catheter is in the body, it should be manipulated while under high-quality fluoroscopic observation. Do not advance or retract the catheter unless the balloon is fully deflated under vacuum. If strong resistance is met during manipulation, discontinue the procedure and determine the cause of the resistance before proceeding.
- Balloon pressure should not exceed the rated burst pressure. Use of a pressure monitoring device is recommended to prevent over-pressurization.
- PTCA should only be performed at hospitals where emergency coronary artery bypass graft surgery can be immediately performed in the event of a potentially injurious or life-threatening complication. A cardiac surgery team must be on alert when a PTCA procedure is being performed.
- Use only the appropriate balloon inflation medium. Never use air or any gaseous medium to inflate the balloon.
- Use the catheter prior to the “Use Before” date specified on the package.
### B. Precautions
- Before insertion of the dilatation catheter, administer appropriate anticoagulant, antiplatelet, and coronary vasodilator therapy.
- Caution should be taken not to over tighten a Tuohy-Borst type hemostatic adapter around the dilatation catheter shaft as lumen constriction may occur, affecting inflation/deflation of the balloon.
### V. Device Description
The COOK® MBC PTCA balloon dilatation catheter is a percutaneous transluminal coronary angioplasty (PTCA) balloon catheter. This device is provided sterile and is intended for one-time use. This device is used in conjunction with conventional PTCA equipment, including, but not limited to, a vascular access set, arterial sheath, guiding catheter, guide wire and indeflator device.
The COOK® MBC PTCA balloon catheter is a double lumen catheter with a balloon near the distal tip. The catheter features a minimally compliant balloon constructed from high density polyethylene material. The balloon is designed to expand to a specified diameter and length at a specific pressure as labeled. The balloon catheter is provided in an overall length of 135 cm.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{3}
The catheter is compatible with 0.014-inch standard PTCA guide wires. Two radiopaque marker bands located at the proximal and distal ends of the balloon segment facilitate fluoroscopic visualization of the balloon during use.
## VI. Alternative Practices and Procedures
Alternative treatments for coronary artery disease are medical therapy, coronary atherectomy, coronary laser, coronary stent, coronary endarterectomy, coronary artery bypass graft surgery, and other commercially available PTCA catheters.
## VII. Marketing History
The COOK® MBC PTCA balloon dilatation catheter has not been commercially available.
## VIII. Adverse Effects of the Device on Health
Possible adverse effects associated with percutaneous transluminal coronary angioplasty include, but are not limited to:
- coronary artery dissection, perforation, rupture or other injury
- conduction disturbance
- acute myocardial infarction
- unstable angina
- arteriovenous fistula
- coronary artery spasm
- total occlusion of the coronary artery or bypass graft
- hemorrhage or hematoma
- embolism
- infection
- restenosis of the dilated artery
- hypo/hypertension
- death
## IX. Summary of Non-clinical Studies
A. Summary of the Nonclinical Laboratory Studies (In-vitro)
Balloon Minimum Burst Strength Test
Fifteen (15) balloon catheters of each size were tested to determine the rated burst
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{4}
pressure.
The burst pressure data were analyzed using factors for one-sided tolerance to determine with 95% confidence that 99.9% of these balloons would not burst at or below the calculated rated burst pressure. The results are presented in Table 1. All balloon catheters had rated burst pressures of ≥ 12 atm. To provide an additional margin of safety, all balloons are labeled with a recommended pressure of 11 atm.
Table 1. Mean Burst, Rated Burst and Maximum Recommended Pressures
| Balloon Catheter Size (mm) | Mean Burst Pressure (atm) | Rated Burst Pressure (atm) | Recommended Pressure (atm) |
| --- | --- | --- | --- |
| MBC-2.5-15 | 16.8 | 14 | 11 |
| MBC-2.5-20 | 16.2 | 12 | 11 |
| MBC-2.5-30 | 16.2 | 14 | 11 |
| MBC-2.5-45 | 15.7 | 12 | 11 |
| MBC-3.0-15 | 15.9 | 13 | 11 |
| MBC-3.0-20 | 15.3 | 12 | 11 |
| MBC-3.0-30 | 15.4 | 13 | 11 |
| MBC-3.0-45 | 15.7 | 13 | 11 |
| MBC-3.5-15 | 16.4 | 13 | 11 |
| MBC-3.5-20 | 15.9 | 12 | 11 |
| MBC-3.5-30 | 15.6 | 12 | 11 |
| MBC-3.5-45 | 16.0 | 13 | 11 |
| MBC-4.0-15 | 15.3 | 12 | 11 |
| MBC-4.0-20 | 15.0 | 12 | 11 |
| MBC-4.0-30 | 14.9 | 12 | 11 |
| MBC-4.0-45 | 14.8 | 12 | 11 |
## Balloon Distensibility Test
Fifteen (15) balloon catheters of each size were tested to determine how the balloon diameter varies with increasing balloon inflation pressure.
Table 2 presents the pressure at which nominal diameter is achieved, the rated burst pressure, the diameter at the rated burst pressure, and the ratio of the diameter at rated burst pressure to the nominal diameter.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{5}
Table 2. Pressures Required for Nominal Diameter and Diameter at Rated Burst Pressure
| Balloon Catheter Size (mm) | Mean Pressure for Nominal Diameter (atm) | Rated Burst Pressure (atm) | Diameter at Rated Burst Pressure (mm) | Ratio of Diameter at Rated Burst Pressure to Nominal Diameter |
| --- | --- | --- | --- | --- |
| 2.5X15 | 6.5 | 14 | 2.84 | 1.14 |
| 2.5X20 | 5.8 | 12 | 2.79 | 1.11 |
| 2.5X30 | 6.4 | 14 | 2.83 | 1.13 |
| 2.5X45 | 6.4 | 12 | 2.77 | 1.11 |
| 3.0X15 | 6.6 | 13 | 3.33 | 1.11 |
| 3.0X20 | 6.8 | 12 | 3.29 | 1.10 |
| 3.0X30 | 6.9 | 13 | 3.34 | 1.11 |
| 3.0X45 | 7.5 | 13 | 3.31 | 1.10 |
| 3.5X15 | 6.4 | 13 | 3.90 | 1.11 |
| 3.5X20 | 7.1 | 12 | 3.83 | 1.09 |
| 3.5X30 | 7.4 | 12 | 3.82 | 1.09 |
| 3.5X45 | 7.5 | 13 | 3.87 | 1.11 |
| 4.0X15 | 7.1 | 12 | 4.36 | 1.09 |
| 4.0X20 | 7.5 | 12 | 4.35 | 1.09 |
| 4.0X30 | 7.5 | 12 | 4.35 | 1.09 |
| 4.0X45 | 7.6 | 12 | 4.33 | 1.08 |
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{6}
Table 3 presents balloon compliance (diameter versus inflation pressure) for all sizes of the COOK® MBC PTCA balloon dilatation catheter.
Table 3. Balloon Compliance (Diameter vs Pressure) for the COOK® MBC PTCA Catheters
| Balloon Diameter (mm): | | 2.5 | | | | 3.0 | | | | 3.5 | | | | 4.0 | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Balloon Length (mm): | | 15 | 20 | 30 | 45 | 15 | 20 | 30 | 45 | 15 | 20 | 30 | 45 | 15 | 20 | 30 | 45 |
| Inflation Pressure (atm) | 2 | 2.31 | 2.34 | 2.32 | 2.31 | 2.76 | 2.74 | 2.74 | 2.67 | 3.22 | 3.18 | 3.12 | 3.15 | 3.61 | 3.55 | 3.52 | 3.46 |
| | 3 | 2.35 | 2.37 | 2.36 | 2.34 | 2.81 | 2.79 | 2.79 | 2.75 | 3.27 | 3.22 | 3.21 | 3.21 | 3.68 | 3.63 | 3.62 | 3.64 |
| | 4 | 2.39 | 2.42 | 2.39 | 2.38 | 2.85 | 2.83 | 2.83 | 2.79 | 3.33 | 3.27 | 3.27 | 3.27 | 3.75 | 3.70 | 3.70 | 3.70 |
| | 5 | 2.43 | 2.45 | 2.43 | 2.43 | 2.90 | 2.89 | 2.88 | 2.85 | 3.39 | 3.34 | 3.33 | 3.32 | 3.82 | 3.77 | 3.77 | 3.78 |
| | 6 | 2.46 | 2.50 | 2.47 | 2.48 | 2.96 | 2.94 | 2.94 | 2.90 | 3.46 | 3.41 | 3.40 | 3.39 | 3.91 | 3.86 | 3.86 | 3.86 |
| | *7 | 2.51 | 2.55 | 2.52 | 2.53 | 3.02 | 3.01 | 3.00 | 2.97 | 3.53 | 3.48 | 3.47 | 3.46 | 3.99 | 3.95 | 3.95 | 3.95 |
| | 8 | 2.57 | 2.60 | 2.58 | 2.58 | 3.08 | 3.08 | 3.07 | 3.03 | 3.60 | 3.56 | 3.55 | 3.54 | 4.07 | 4.05 | 4.04 | 4.04 |
| | 9 | 2.62 | 2.66 | 2.63 | 2.63 | 3.14 | 3.14 | 3.13 | 3.10 | 3.67 | 3.63 | 3.62 | 3.61 | 4.15 | 4.13 | 4.13 | 4.12 |
| | 10 | 2.67 | 2.70 | 2.67 | 2.68 | 3.19 | 3.19 | 3.19 | 3.16 | 3.73 | 3.70 | 3.69 | 3.68 | 4.22 | 4.20 | 4.21 | 4.19 |
| | **11 | 2.71 | 2.74 | 2.71 | 2.72 | 3.24 | 3.24 | 3.24 | 3.21 | 3.79 | 3.76 | 3.75 | 3.75 | 4.29 | 4.27 | 4.27 | 4.26 |
| | 12 | 2.75 | 2.78 | 2.75 | 2.76 | 3.29 | 3.29 | 3.29 | 3.26 | 3.84 | 3.82 | 3.82 | 3.81 | 4.36 | 4.34 | 4.34 | 4.33 |
| | 13 | 2.79 | 2.82 | 2.79 | 2.80 | 3.34 | 3.34 | 3.34 | 3.31 | 3.89 | 3.88 | 3.87 | 3.87 | 4.42 | 4.40 | 4.40 | 4.39 |
| | 14 | 2.82 | 2.86 | 2.82 | 2.84 | 3.38 | 3.38 | 3.39 | 3.36 | 3.95 | 3.94 | 3.93 | 3.93 | 4.47 | 4.47 | 4.46 | 4.46 |
| | 15 | 2.86 | 2.89 | 2.86 | 2.87 | 3.42 | 3.42 | 3.43 | 3.40 | 4.00 | 3.99 | 3.98 | 3.98 | | | | |
* Applying 7 atm inflation pressure will produce near nominal diameter for all balloons.
** 11 atm is the labeled rated burst pressure.
Note: Values of balloon diameter printed were measured up to 25% above the rated burst pressure (shaded area) for the specified balloon.
## Balloon Inflation/Deflation Time Test
Two (2) catheters of each size were tested to show that the inflation and deflation times for these catheters are within clinically acceptable limits.
All the balloons inflated in no more than four (4) seconds and deflated in less than fourteen (14) seconds. Inflation times for all 16 balloon sizes were similar, averaging 3.1 ± 0.4 seconds. Deflation times were more closely related to balloon size, with larger diameter, longer balloons deflating more slowly than smaller balloons.
These inflation/deflation times are similar to those of devices currently on the market and are adequate for the intended use.
Summary of Safety and Effectiveness, COOK®MBC PTCA Balloon Dilatation Catheter, P990008
{7}
# Balloon Fatigue Test
Thirty (30) catheters of each size were tested to determine with 95% confidence that at least 90% of the balloons will sustain 40 repeated inflations to the maximum recommended pressure (11 atm).
All test samples successfully passed this fatigue test. The COOK® MBC PTCA balloon catheter should have adequate fatigue resistance when used according to the Instructions for Use.
# Bond Strength Test
Twenty (20) catheters of various sizes were tested to determine the strength of bonds in the catheter.
Two bond sites were tested for this report. Test loads exceeded the strength of the 3 Fr bulk material before either bond site failed. Therefore, the strength of the bond sites is considered adequate for the intended use.
# Catheter Diameter and Balloon Profile
All PTCA catheters of the current design have a shaft that is 3.5 Fr OD (0.044 ± 0.0015 inch) at its proximal end and 3.0 Fr OD (0.04 ± 0.0015 inch) along its distal most 25 cm. Catheters of each size were tested to determine the largest outside diameter along the distal half of the working length of the balloon and catheter tip when the catheter is in its deflated state. Table 4 presents the results of this test.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{8}
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
9
Table 4. Deflated Balloon Profile
| Balloon Catheter (mm) | Largest Outer Diameter (in) |
| --- | --- |
| 2.5X15 | 0.044 |
| 2.5X20 | 0.043 |
| 2.5X30 | 0.042 |
| 2.5X45 | 0.044 |
| 3.0X15 | 0.047 |
| 3.0X20 | 0.047 |
| 3.0X30 | 0.047 |
| 3.0X45 | 0.048 |
| 3.5X15 | 0.057 |
| 3.5X20 | 0.057 |
| 3.5X30 | 0.056 |
| 3.5X45 | 0.058 |
| 4.0X15 | 0.059 |
| 4.0X20 | 0.061 |
| 4.0X30 | 0.058 |
| 4.0X45 | 0.060 |
## Over-the-Arch Torque Strength Test
Ten (10) catheters of various sizes were tested to determine the torque strength of the catheter when its distal tip is not free to rotate. With the distal tip prevented from rotating, a high number of catheter rotations (in excess of 70) is needed to cause the catheter to fail. It is unlikely the catheter would be subjected to this number of rotations in clinical use.
## Over-the-Arch Torque Response Test
Five (5) catheters of different sizes were tested to evaluate the torque response characteristics of the catheter by analyzing the degree of distal tip rotation for a given proximal end rotation.
Ten complete rotations of the proximal end of the catheter produced no rotation of the distal end. The balloon catheter did not transmit enough torque to cause the distal end to rotate when torque was applied to the proximal end.
## B. Animal Testing (In-vivo)
Animal testing was not required on this device.
{9}
# C. Biocompatibility Testing
Appropriate compliance with the criteria specified for material and biocompatibility toxicity testing was undertaken in accordance with the recommendations of the PTCA Catheter System Testing Guidance Document and Tripartite Biocompatibility Guidance for Medical Devices prepared by the Toxicology Sub-Group of the Tripartite Subcommittee on Medical Devices in September of 1986. The tests performed to establish biocompatibility were conducted by an independent laboratory (NamSA®, Northwood, OH) and included:
1. Delayed Contact Sensitization (Guinea Pig Maximization Method)
2. Irritation (U.S.P. Intracutaneous Toxicity Study in the Rabbit)
3. Cytotoxicity (U.S.P. MEM Elution Test)
4. Systemic Toxicity (U.S.P. Mouse Systemic Toxicity Test)
5. Hemolysis (Extraction Method)
6. Implantation (7 Day Rabbit Muscle Implant Test)
7. Mutagenicity (Ames Salmonella Mutagenicity Assay with Saline and DMSO Extracts)
The results of these tests provided no evidence to suggest that the device is not biocompatible.
# D. Sterilization, Packaging and Shelf-Life Testing
COOK® Incorporated sterilizes devices using a sterilization cycle that has been validated to a $10^{-6}$ sterility assurance level. This process is routinely monitored, calibrated and operated in adherence with the Guidance for Industrial Ethylene Oxide Sterilization of Medical Devices: Association for the Advancement of Medical Instruction. COOK® Incorporated has tested to determine the resistance data (D-value) for the product in the package. The design and raw materials of the balloon catheter as well as its proposed packaging raise no new questions regarding sterility of the device when subjected to the validated sterilization cycles. The COOK® MBC PTCA balloon catheter is packaged in a peel-pouch system that is currently used for previously approved, commercially available devices. The shelf life studies reveal that the package configuration and materials provides an acceptable bacterial barrier and that two years is an appropriate sterility expiration period.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{10}
# X. Summary of the Clinical Investigation
This study was intended to collect data regarding procedural success and the incidence of major adverse cardiac events in patients treated for myocardial ischemia with the COOK® MBC PTCA Balloon Catheter. This study was designed as a multicenter, prospective registry in which 150 patients eligible for elective coronary angioplasty who met the inclusion/exclusion criteria were enrolled from 6 investigative sites. Recognizing there are numerous PTCA balloon catheters available for performing conventional angioplasty, none is considered the “gold standard.” Therefore, no contemporary control group was justifiable for performing a randomized, controlled study.
## A. Structured Abstract
**Title:** A Multicenter Registry Evaluating the COOK® MBC PTCA Balloon Catheter for Dilating the Stenotic Portion of a Coronary Artery or Bypass Graft for the Purpose of Improving Myocardial Perfusion
**Investigators:** Six (6) U.S. sites participated. In addition to the 5 principal investigators, 23 other physicians served as operators using the MBC PTCA Balloon Catheter during this study.
**Purpose:** To evaluate procedural success and the incidence of major adverse cardiac events including death, Q-wave MI, CABG or repeat PTCA within 30 days of the procedure in which the COOK® MBC PTCA Balloon Dilatation Catheter was used.
**Design:** A multicenter, prospective registry study to collect: 1) qualitative and quantitative angiographic data recorded by both the investigators during the procedure and by post-procedure independent core lab analysis; and 2) data to assess the incidence of clinically significant ischemic events including death, Q-wave MI, CABG, and repeat PTCA of the culprit vessel.
**Demographics:** One hundred fifty (150) patients were enrolled in the MBC Balloon Catheter Clinical Study. Mean age = 62 ± 11 years, 65% were male, 33% were diabetic, 29% were obese, 64% were hypertensive, 67% had high cholesterol, 12% had prior PTCA to the culprit lesion, 16% had prior CABG, 42% had prior MI, 63% had multiple vessel disease, and a mean LVEF=50%. The culprit vessels were LAD (43%), LCX
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{11}
(17%), RCA (34%), RAMUS (2%), and the SVG (4%). Key lesion characteristics included lesion class of B₂ and C (41%), moderate or severe tortuosity (28%), diffuse disease (24%), calcification (6%), and branch point (11%). Mean lesion length was 9.97±4.97 mm.
**Method:** Demographics, clinical and procedural angiographic data were collected from 150 patients consecutively enrolled. QCA was performed by the Cardiovascular Angiography Analysis Lab at The Methodist Hospital Center in Houston, Texas.
**Conclusions:** Of the 150 patients enrolled in the COOK® MBC PTCA Balloon Catheter registry, 13 patients were not treated with the MBC balloon catheter and 137 underwent dilatation with the MBC balloon catheter. Twenty-two (22) patients were treated with the MBC balloon catheter only. In 115 patients, the MBC balloon was used to predilate the vessel prior to placing a stent. The 30-day combined incidence of death, Q-wave MI and target lesion revascularization was 0.9% [95% CI = 0.0, 3.7] for all patients enrolled in the clinical study (N=150). These results indicate that the COOK® MBC PTCA Balloon Catheter can be successfully used as a definitive therapy when appropriate, or to predilate a vessel prior to elective stent placement. The low incidence of adverse events suggests that no new safety issues are raised by this device.
## B. Subject Selection and Exclusion Criteria
Patients were entered into the study if they met all of the following inclusion criteria and none of the exclusion criteria.
### Inclusion Criteria:
Patients must meet all the following inclusion criteria:
1. Patient must be eligible to undergo elective PTCA of a *de novo* or restenosed lesion(s) in a native coronary artery or graft.
2. Normal reference vessel diameter (RVD) must be 2.5 mm to 4.0 mm.
3. All lesions requiring intervention must be considered to be treatable with the COOK® MBC PTCA balloon catheter as the first and only interventional device.
4. Patient or legal guardian must have signed the informed consent document.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008 12 2.0
{12}
# Exclusion Criteria:
Patients must be excluded if any of the following conditions are true:
1. Patient will not agree to be available for follow-up.
2. Intolerance to contrast agent or aspirin.
3. Severely impaired left ventricular function (LVEF < 35%).
4. Presentation in cardiogenic shock.
5. Candidate is not acceptable for coronary artery bypass surgery.
6. Acute myocardial infarction or an MI within past 3 days.
7. Previous diagnosis of coronary artery spasm.
8. Totally obstructed coronary arteries.
9. Left main coronary artery disease.
10. Diffuse disease.
11. Calcified stenosis.
12. Prior intervention has been performed during this procedure with any other approved or investigative device (e.g., rotational atherectomy, directional atherectomy, another PTCA balloon, etc.).
## C. Study Population
One hundred fifty (150) patients were enrolled in the COOK® MBC PTCA Balloon Catheter Clinical Study. Table 5 presents baseline patient demographics and clinical characteristics for the 150 patients enrolled in the clinical study.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{13}
Table 5. Baseline Demographics and Clinical Characteristics (N=150)
| | n | % | 95% CI |
| --- | --- | --- | --- |
| Number of Males | 97 | 64.7 | 56.4, 72.2 |
| Smoking Status | | | |
| Current Smoker | 30 | 20.0 | 14.1, 27.5 |
| Quit Smoking | 61 | 40.7 | 32.8, 49.0 |
| Never Smoked | 52 | 34.7 | 27.2, 42.9 |
| Not Reported | 7 | 4.7 | 2.1, 9.8 |
| Diabetes | 50 | 33.3 | 26.0, 41.6 |
| Obesity | 43 | 28.7 | 21.7, 36.7 |
| Hypertension | 96 | 64.0 | 55.7, 71.6 |
| Hypercholesterolemia | 101 | 67.3 | 59.1, 74.6 |
| Previous PTCA to Lesion Site | 18 | 12.0 | 7.5, 18.6 |
| Previous CABG | 24 | 16.0 | 10.7, 23.1 |
| Previous MI | 63 | 42.0 | 34.1, 50.3 |
| Multi-Vessel Disease | 95 | 63.3 | 55.0, 70.9 |
| Pre Procedure Angina Class 0 | 9 | 6.0 | 3.0, 11.4 |
| 1 | 18 | 12.0 | 7.5, 18.6 |
| 2 | 27 | 18.0 | 12.4, 25.3 |
| 3 | 45 | 30.0 | 22.9, 38.1 |
| 4 | 28 | 18.7 | 13.0, 26.0 |
| Not Reported | 23 | 15.3 | 10.2, 22.3 |
| Target Lesion Vessel LAD | 64 | 42.7 | 34.7, 51.0 |
| LCX | 26 | 17.3 | 11.8, 24.6 |
| RCA | 51 | 34.0 | 26.6, 42.2 |
| RAMUS | 3 | 2.0 | 0.4, 5.7 |
| SVG | 6 | 4.0 | 1.6, 8.9 |
| Age (years) | | | |
| Mean±SD | 62±11 | | |
| Range | 30-82 | | |
| N | 150 | | |
| Ejection Fraction (%) | | | |
| Mean±SD | 49.7±10.22 | | |
| Range | 10-74 | | |
| N | 133 | | |
95%CI=95% upper and lower confidence intervals
Table 6 presents baseline angiographic characteristics of the culprit lesion site prior to the PTCA procedure. Data from the Case Report Forms (CRF) and the angiographic core lab analysis are presented for comparison.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{14}
Table 6. Baseline Angiographic Characteristics of the Culprit Lesion Prior to the Initial PTCA Procedure
| | Data from CRF (N=150) | | | Core Lab Analysis (N=149) | | |
| --- | --- | --- | --- | --- | --- | --- |
| Average Reference Vessel Diameter (mm) | | | | | | |
| Mean ±SD | 3.08±0.51 | | | 3.09±0.52 | | |
| Range | 1.65 - 4.38 | | | 1.97 - 4.27 | | |
| N | 137 | | | 148 | | |
| Minimum Lumen Diameter (mm) | | | | | | |
| Mean ±SD | 0.75±0.52 | | | 0.71±0.39 | | |
| Range | 0.03 - 2.90 | | | 0.18 - 1.99 | | |
| N | 137 | | | 148 | | |
| Lesion Length (mm) | | | | | | |
| Mean ±SD | 14.6±6.2 | | | 9.97±4.97 | | |
| Range | 5.0 - 40.0 | | | 3.59 - 36.64 | | |
| N | 110 | | | 148 | | |
| | n | % | 95% CI | n | % | 95% CI |
| Lesion Classification | | | | | | |
| A | 14 | 9.3 | 5.4, 15.5 | 24 | 16.1 | 10.8, 23.2 |
| B_{1} | 64 | 42.7 | 34.7, 51.0 | 64 | 43.0 | 35.0, 51.3 |
| B_{2} | 54 | 36.0 | 28.5, 44.3 | 54 | 36.2 | 28.7, 44.6 |
| C | 16 | 10.7 | 6.4, 17.0 | 7 | 4.7 | 2.1, 9.8 |
| Not Reported | 2 | 1.3 | 0.2, 4.7 | 0 | 0.0 | 0.0, 2.5 |
| Proximal Tortuosity | | | | | | |
| None | 45 | 30.0 | 22.9, 38.1 | 32 | 21.5 | 15.4, 29.1 |
| Mild | 75 | 50.0 | 41.8, 58.2 | 76 | 51.0 | 42.7, 59.2 |
| Moderate | 23 | 15.3 | 10.2, 22.3 | 36 | 24.2 | 17.7, 32.0 |
| Severe | 3 | 2.0 | 0.4, 5.7 | 5 | 3.4 | 1.2, 8.1 |
| Not Reported | 4 | 2.7 | 0.7, 6.7 | 0 | 0.0 | 0.0, 2.5 |
| Diffuse Disease | 28 | 18.7 | 13.0, 26.0 | 36 | 24.2 | 17.7, 32.0 |
| Any Calcification | 26 | 17.3 | 11.8, 24.6 | 9 | 6.0 | 3.0, 11.5 |
| Angulation > 45° | 12 | 8.0 | 4.4, 13.9 | 5 | 3.4 | 1.2, 8.1 |
| Eccentric | 124 | 82.7 | 75.5, 88.2 | 88 | 59.1 | 50.7, 67.0 |
| Branch Point | 22 | 14.7 | 9.6, 21.6 | 17 | 11.4 | 7.0, 17.9 |
| Ostial Location | 6 | 4.0 | 1.6, 8.9 | 6 | 4.0 | 1.7, 8.9 |
| Other Stenoses >50% in Target Vessel | 21 | 14.0 | 9.1, 20.8 | 18 | 12.1 | 7.5, 18.7 |
95% CI=95% upper and lower confidence intervals
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{15}
D. Study Period
Patients were enrolled from January 14, 1998 through October 5, 1998.
E. Safety and Effectiveness Data
The results of this study must be interpreted keeping in mind current practice in interventional cardiology. In particular, angioplasty balloons alone are rarely used as definitive therapy. The vast majority of patients treated for coronary artery disease by interventional cardiologists now receive stents on an elective basis. That is, the use of a stent is planned in advance and is rarely determined by the outcome of the balloon dilatation procedure. Thus, the “success” of the balloon dilatation procedure can no longer be defined in isolation in traditional terms (i.e., <50% residual diameter stenosis with at least a 20% reduction). The balloon is frequently used simply to adequately open the lesion so that a stent can be placed. Thus, a “successful” angioplasty procedure may now rightly be considered one in which either a balloon alone produces a residual diameter stenosis <50%, or a balloon provides initial dilatation permitting a stent to be placed thereby producing a residual diameter stenosis <50%, with no major adverse cardiac events in-hospital.
The results of this study are presented with these considerations in mind. The outcome following use of the COOK® MBC PTCA balloon catheter are presented both for those patients in whom the balloon was used alone and those patients who received stents following dilatation with the MBC balloon. The procedural outcome was assessed based upon the entire procedure whether the MBC balloon was used alone or to predilate the vessel to permit stent placement.
Figure 1 presents the flow diagram for patients enrolled in this study. Twenty-two (22) patients were treated with the MBC balloon catheter alone. In 115 patients, the MBC balloon was used to predilate the vessel prior to placing a stent. In 13 patients, the MBC balloon was not used. However, these 13 cases fall into 3 categories, not all of which should be considered a failure of the balloon. In 5 cases, another brand of balloon of at least equal size to the original MBC balloon was used successfully without changing the ancillary equipment. These 5 cases (3.3%) can be considered technical failures of the MBC balloon to perform successfully. In 6 cases, a balloon of another brand was used
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{16}
that was smaller in either diameter or length than the MBC balloon originally attempted. In these 6 cases, a smaller MBC balloon was either not available or attempted before changing to another brand. These cases should not be considered technical failures of the balloon since the optimum size MBC balloon was not attempted before changing to a different device. In 2 cases, ancillary equipment was changed and another brand balloon was used without using an MBC balloon with the new ancillary equipment. These 2 cases should not be considered technical failures of the balloon since changing the ancillary equipment may have altered the environment such that an MBC balloon would have been successful had it been attempted.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
17
25
{17}
Figure 1. Patient Flow Diagram
| All Patients Enrolled in MBC Clinical Study (N = 150) | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| MBC Balloon Catheter Used (N = 137) | | | | | MBC Balloon Catheter Attempted, But Not Used (N = 13) | | | |
| Balloon Only
N = 22 | | Balloon and Stent
N = 115 | | | Ancillary equipment was not changed.
Another brand of balloon was used with a diameter that was < the diameter of the MBC balloon.
N = 5 | Ancillary equipment was not changed.
Another brand of balloon having a smaller diameter or length than the initially attempted MBC balloon was used.
N = 6 | Ancillary equipment was changed.
Another brand of balloon was used without attempting the MBC balloon again.
N = 2 | |
| In-Hospital Events | | In-Hospital Events | | | | | | |
| | n | % | | n | % | | | |
| Death | 0 | 0.0 | Death | 0 | 0.0 | | | |
| Q-wave MI | 0 | 0.0 | Q-wave MI | 0 | 0.0 | | | |
| Non Q-MI | 0 | 0.0 | Non Q-MI | 0 | 0.0 | | | |
| TLR-CABG | 0 | 0.0 | TLR-CABG | 0 | 0.0 | | | |
| TLR-PTCA | 0 | 0.0 | TLR-PTCA | 0 | 0.0 | | | |
| Thrombosis | 0 | 0.0 | Thrombosis | 0 | 0.0 | | | |
| CVA | 0 | 0.0 | CVA | 0 | 0.0 | | | |
| Cumulative Events
0-30 Days | | Cumulative Events
0-30 Days | | | | | | |
| | n | % | | n | % | | | |
| Death | 0 | 0.0 | Death | 0 | 0.0 | | | |
| Q-wave MI | 0 | 0.0 | Q-wave MI | 0 | 0.0 | | | |
| Non Q-MI | 0 | 0.0 | Non Q-MI | 3 | 2.6 | | | |
| TLR-CABG | 0 | 0.0 | TLR-CABG | 0 | 0.0 | | | |
| TLR-PTCA | 0 | 0.0 | TLR-PTCA | 1 | 0.9 | | | |
| Thrombosis | 0 | 0.0 | Thrombosis | 0 | 0.0 | | | |
| CVA | 0 | 0.0 | CVA | 0 | 0.0 | | | |
| MACE* | 0 | 0.0 | MACE* | 1 | 0.9 | *MACE defined as Death, Q-wave MI, TLR-CABG or TLR-PTCA | | |
Table 7 presents qualitative and quantitative angiographic characteristics of the culprit lesion site after the PTCA procedure with the COOK® MBC balloon catheter. The data are analyzed separately for those patients in whom only the MBC balloon was used and those patients who were subsequently stented after using the MBC balloon. (Note: Data are from the Case Report Forms.)
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{18}
Table 7. Status of the Culprit Lesion after Using the MBC Balloon
| | MBC Balloon Only (N=22) | | | MBC Balloon+Stent (Prior to Stenting) (N=115) | | | All Patients Treated with MBC Balloon (N=137) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | n | % | 95% CI | n | % | 95% CI | n | % | 95% CI |
| Dissection Yes | 4 | 18.2 | 5.2, 40.3 | 33 | 28.7 | 20.8, 38.0 | 37 | 27.0 | 19.95, 35.38 |
| Dissection Type | | | | | | | | | |
| No Dissection | 18 | 81.8 | 59.7, 94.8 | 82 | 71.3 | 62.0, 79.2 | 100 | 73.0 | 64.62, 80.05 |
| Linear | 3 | 13.6 | 2.9, 34.9 | 27 | 23.5 | 16.3, 32.5 | 30 | 21.9 | 15.48, 29.93 |
| Spiral | 0 | 0.0 | 0.0, 15.4 | 2 | 1.7 | 0.2, 6.1 | 2 | 1.5 | 0.18, 5.17 |
| Not Reported | 1 | 4.5 | 0.1, 22.8 | 4 | 3.5 | 1.0, 8.7 | 5 | 3.6 | 1.35, 8.74 |
| Dissection Grade | | | | | | | | | |
| No Dissection | 18 | 81.8 | 59.7, 94.8 | 82 | 71.3 | 62.0, 79.2 | 100 | 73.0 | 64.62, 80.05 |
| A | 2 | 9.1 | 1.1, 29.2 | 13 | 11.3 | 6.4, 18.9 | 15 | 10.9 | 6.47, 17.71 |
| B | 1 | 4.5 | 0.1, 22.8 | 14 | 12.2 | 7.1, 19.9 | 15 | 10.9 | 6.47, 17.71 |
| C | 0 | 0.0 | 0.0, 15.4 | 4 | 3.5 | 1.0, 8.7 | 4 | 2.9 | 0.80, 7.31 |
| D | 0 | 0.0 | 0.0, 15.4 | 2 | 1.7 | 0.2, 6.1 | 2 | 1.5 | 0.18, 5.17 |
| E | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 3.2 | 0 | 0.0 | 0.00, 2.66 |
| F | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 3.2 | 0 | 0.0 | 0.00, 2.66 |
| Not Reported | 1 | 4.5 | 0.1, 22.8 | 0 | 0.0 | 0.0, 3.2 | 1 | 0.7 | 0.02, 4.00 |
| Dissection Length (mm) | 8.0 ± 2.8 | | | 10.4 ± 8.2 | | | 10.3 ± 7.9 | | |
| Mean±SD | 6.0 - 10.0 | | | 2.0 - 35.0 | | | 2.0 - 35.0 | | |
| Range | 2 | | | 26 | | | 28 | | |
| N | | | | | | | | | |
| Intimal Flap | 2 | 9.1 | 1.1, 29.2 | 15 | 13.0 | 7.7, 20.9 | 17 | 12.4 | 7.61, 19.39 |
| Thrombus Present | 0 | 0.0 | 0.0, 15.4 | 4 | 3.5 | 1.0, 8.7 | 4 | 2.9 | 0.80, 7.31 |
| Spasm | 0 | 0.0 | 0.0, 15.4 | 7 | 6.1 | 2.7, 12.6 | 7 | 5.1 | 2.26, 10.64 |
| ST Segment Changes | 1 | 4.5 | 0.1, 22.8 | 10 | 8.7 | 4.5, 15.8 | 11 | 8.0 | 4.28, 14.25 |
| Side Branch Closure | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0,3.2 | 0 | 0.0 | 0.00, 2.66 |
| Angina | 2 | 9.1 | 1.1, 29.2 | 11 | 9.6 | 5.1, 16.8 | 13 | 9.5 | 5.35, 15.99 |
| TIMI Flow | | | | | | | | | |
| 0 | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 3.2 | 0 | 0.0 | 0.00, 2.66 |
| 1 | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 3.2 | 0 | 0.0 | 0.00, 2.66 |
| 2 | 0 | 0.0 | 0.0, 15.4 | 3 | 2.6 | 0.5, 7.4 | 3 | 2.2 | 0.45, 6.27 |
| 3 | 22 | 100.0 | 84.6, 100.0 | 110 | 95.7 | 89.7, 98.4 | 132 | 96.4 | 91.26, 98.65 |
| Not Reported | 0 | 0.0 | 0.0, 15.4 | 2 | 1.7 | 0.2, 6.1 | 2 | 1.5 | 0.18, 5.17 |
95% CI=95% upper and lower confidence intervals
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{19}
Table 8 presents quantitative angiographic characteristics of the culprit lesion site prior to (Baseline) and after the PTCA procedure (Post-PTCA). The data are analyzed separately for those patients in whom only the MBC balloon was used, those patients who were subsequently stented after using the MBC balloon, those patients not treated with the MBC balloon and all patients enrolled in the study. (Note: All data are from the angiographic core lab analysis.)
Table 8. Angiographic Core Lab QCA Analysis of the Culprit Lesion
| | MBC Balloon Only (N=22) | MBC Balloon+Stent (Prior to Stenting) (N=115) | Patients Not Treated with the MBC Balloon (N=13) | All Patients (N=150) |
| --- | --- | --- | --- | --- |
| Baseline | | | | |
| Average Reference Vessel Diameter (mm) | | | | |
| Mean±SD | 2.67±0.46 | 3.22±0.48 | 2.59±0.32 | 3.08±0.52 |
| Range | 2.08 - 3.80 | 1.97 - 4.27 | 1.97 - 3.12 | 1.97 - 4.27 |
| N | 21 | 114 | 13 | 148 |
| Minimum Lumen Diameter (mm) | | | | |
| Mean±SD | | | | |
| Range | 0.56±0.74 | 0.75±0.39 | 0.49±0.34 | 0.70±0.39 |
| N | 0.26 - 1.91 | 0.18 - 1.99 | 0.26 - 1.58 | 0.18 - 1.99 |
| | 21 | 114 | 13 | 148 |
| Percent Diameter Stenosis (%) | | | | |
| Mean±SD | | | | |
| Range | 79.8±11.8 | 76.8±10.3 | 81.7±10.3 | 77.7±10.6 |
| N | 49.7 - 89.7 | 51.6 - 94.1 | 49.4 - 89.7 | 49.4 - 94.1 |
| | 21 | 114 | 13 | 148 |
| Post-PTCA | | | | |
| Average Reference Vessel | | | | |
| Diameter (mm) | | | | |
| Mean±SD | 2.61±0.45 | 3.18±0.47 | 2.46±0.27 | 3.04±0.53 |
| Range | 1.92 - 3.76 | 1.96 - 4.22 | 2.05 - 2.91 | 1.92 - 4.22 |
| N | 21 | 113 | 12 | 146 |
| Minimum Lumen Diameter (mm) | | | | |
| Mean±SD | 1.89±0.49 | 1.99±0.56 | 1.62±0.37 | 1.94±0.54 |
| Range | 1.18 - 3.11 | 0.33 - 3.58 | 1.15 - 2.27 | 0.33 - 3.58 |
| N | 21 | 113 | 12 | 146 |
| Percent Diameter Stenosis (%) | | | | |
| Mean±SD | 27.9±9.53 | 37.4±14.2 | 34.5±11.0 | 35.8±13.7 |
| Range | 3.29 - 43.8 | 1.68 - 88.6 | 20.7 - 47.7 | 1.7 - 88.6 |
| N | 21 | 113 | 12 | 146 |
Figures 2 and 3 present the cumulative frequency distribution curves for MLD and %DS, respectively, with separate curves for patients who were treated with the MBC balloon only, with the MBC balloon followed by stenting, and for patients in whom the MBC balloon was not used.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{20}

Group:
- MBC Only (n = 22)
- MBC not used (n = 13)
- MBC + Stent (n = 115)
Figure 2. Cumulative percent of patients versus Minimum Lumen Diameter in patients receiving the MBC PTCA Balloon only, the MBC balloon followed by discretionary stenting, and in patients in whom the MBC balloon was not used.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{21}

Group:
- MBC Only (n = 22)
- MBC not used (n = 13)
- MBC + Stent (n = 115)
Figure 3. Cumulative percent of patients versus Percent Diameter Stenosis in patients receiving the MBC PTCA Balloon only, the MBC balloon followed by discretionary stenting, and in patients in whom the MBC balloon was not used.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{22}
F. Principal Effectiveness and Safety Results
The clinical outcome for the patients enrolled is summarized in Tables 9 and 10. This section provides additional information with respect to adverse events that occurred subsequent to the PTCA procedure.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008 23 31
{23}
Table 9. Principal Effectiveness and Safety Results
| All Patients Enrolled in the MBC Study (N = 150) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | 0-Discharge | | | 0-30 Days | | |
| | n | % | 95% CI | n | % | 95% CI |
| Clinical Efficacy Measures: | | | | | | |
| Death | 0 | 0.0 | 0.0, 2.4 | 0 | 0.0 | 0.0, 2.4 |
| Q-MI* | 0 | 0.0 | 0.0, 2.5 | 0 | 0.0 | 0.0, 2.5 |
| Non-Q Wave MI* | 2 | 1.3 | 0.2, 4.8 | 3 | 2.0 | 0.4, 5.8 |
| CABG | 0 | 0.0 | 0.0, 2.4 | 0 | 0.0 | 0.0, 2.4 |
| PTCA | 0 | 0.0 | 0.0, 2.4 | 1 | 0.9 | 0.0, 3.7 |
| Death/Q-MI/CABG/PTCA* | 0 | 0.0 | 0.0, 2.5 | 1 | 0.9 | 0.0, 3.7 |
| Death/Any MI/CABG/PTCA* | 2 | 1.3 | 0.2, 4.8 | 3 | 2.0 | 0.4, 5.8 |
| Safety Measures: | | | | | | |
| Thrombosis* | 0 | 0.0 | 0.0, 2.5 | 0 | 0.0 | 0.0, 2.5 |
| CVA* | 0 | 0.0 | 0.0, 2.5 | 0 | 0.0 | 0.0, 2.5 |
| Vascular Event Requiring Surgery* | 1 | 0.7 | 0.0, 3.7 | 1 | 0.7 | 0.0, 3.7 |
| Bleeds or Vascular Complications* | 0 | 0.0 | 0.0, 2.5 | 1 | 0.7 | 0.0, 3.7 |
| Patients treated with the MBC Balloon Only (N = 22) | | | | | | |
| | 0-Discharge | | | 0-30 Days | | |
| | n | % | 95% CI | n | % | 95% CI |
| Clinical Efficacy Measures: | | | | | | |
| Death | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
| Q-MI | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
| Non-Q Wave MI | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
| CABG | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
| PTCA | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
| Death/Q-MI/CABG/PTCA | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
| Death/Any MI/CABG/PTCA | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
| Safety Measures: | | | | | | |
| Thrombosis | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
| CVA | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
| Vascular Event Requiring Surgery | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
| Bleeds or Vascular Complications | 0 | 0.0 | 0.0, 15.4 | 0 | 0.0 | 0.0, 15.4 |
95% CI=95% upper and lower confidence intervals
*The percentage and 95% CI were based upon N=12 and N=149 because data were not available for one patient.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
{24}
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008
## Other Reported Events
For the patients enrolled in the MBC study (N=149), the following additional events within 30 days of the procedure, were reported: 3 PTCA procedures to other lesions, 12 incidences of chest pain, 1 incidence of ventricular tachycardia, 5 episodes of bradycardia, 3 episodes of hypotension, 4 hematomas, 1 groin bleed and 2 transfusions.
## G. Gender Bias Analysis
The patient population enrolled in this study was 64.7% (97/150) male. Patients were either treated with the MBC Balloon only (n = 22; 9 female, 13 male), predilated with the MBC Balloon followed by stent placement (n = 115; 37 female, 78 male), or not treated with the MBC Balloon (n = 22; 7 female, 6 male). This distribution of outcomes is not significantly different (p = 0.253 by Pearson Chi-square).
With regard to clinical outcomes, there was only one (0.7%) MACE event within 30 days of the procedure in the entire 150 patient population. This event is described in detail in Appendix III of the PMA amendment submitted on July 19, 1999. This event did occur in a female patient. Thus, the distribution of MACE events was 0.0% in males and 1.9% in females. This distribution is not significantly different (p = 0.353 by Fisher's exact test).
In summary, acute procedure success rates for all patients and for patients with eligible lesions did not differ significantly by patient gender nor did MACE rates.
## XI. Conclusions Drawn From Studies
The results of the laboratory studies demonstrate that there are no biocompatibility issues and that the COOK® MBC PTCA Balloon Dilatation Catheter has the appropriate physical and performance characteristics for its intended use. The COOK® MBC PTCA Balloon Dilatation Catheter was evaluated *in vivo* and successfully demonstrate that the catheter system would function safely and effectively in actual use in the coronary anatomy.
{25}
The results of the clinical study indicate that the COOK® MBC PTCA Balloon Dilatation Catheter is safe and effective for the treatment of patients with coronary artery disease. Therefore, it is reasonable to conclude that the benefits of use of the device for the target population outweigh the risk of illness or injury when used as indicated in accordance with the directions for use.
## XII. Panel Recommendation
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory System Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by the panel.
## XIII. FDA Decision
FDA issued a PMA approval letter to Cook, Inc., on SEP 27 1999. FDA also performed an inspection of the manufacturing facilities and found the applicant in compliance with the Quality System Regulation (21 CFR Part 820).
## XIV. Approval Specifications
Direction for use: See Final Draft Labeling (Information for Use).
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings and Precautions, and Adverse Events in the Final Draft Labeling. Information for Use
Post-approval Requirements and Restrictions: See Approval Order.
Summary of Safety and Effectiveness, COOK® MBC PTCA Balloon Dilatation Catheter, P990008 26
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.