Sapphire II PRO

K200269 · Orbusneich Medical Trading, Inc. · LOX · Mar 4, 2020 · Cardiovascular

Device Facts

Record IDK200269
Device NameSapphire II PRO
ApplicantOrbusneich Medical Trading, Inc.
Product CodeLOX · Cardiovascular
Decision DateMar 4, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.5100
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sapphire® II PRO Balloon Dilatation Catheter (Ø1.0-1.25mm configurations) is indicated for: • balloon pre-dilatation of a stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis) for the purpose of improving myocardial perfusion The Sapphire® II PRO Balloon Dilatation Catheter (Ø1.5-4.0mm configurations) is indicated for: • balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion • balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction The Sapphire® II PRO Balloon Dilatation Catheter is also indicated for: • percutaneous transluminal angioplasty in the peripheral vasculature, including renal, femoral, popliteal, infra-popliteal, tibial, and peroneal arteries.

Device Story

Over-the-wire balloon dilatation catheter; used for percutaneous transluminal coronary angioplasty (PTCA) and peripheral vascular procedures. Device features proximal Y-type hub, distal balloon, and internal lumen for 0.014-inch guidewire. External lumen allows inflation with dilute contrast media. Radiopaque platinum marker band aids positioning. Hydrophilic coating on distal section. Operated by physicians in clinical settings. Device advanced to stenosis; balloon inflated to dilate vessel, improving myocardial perfusion or treating occlusion. Benefits include restoration of blood flow in stenotic coronary or peripheral arteries.

Clinical Evidence

Bench testing only. No clinical data. Performance testing included dimensional verification, balloon rated burst pressure, fatigue, compliance, bond strength, tip pull strength, radiopacity, coating integrity, shaft burst, inflation/deflation time, flexibility, kinking, torque strength, and particulate evaluation. Biocompatibility testing included cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity, pyrogenicity, hemocompatibility (hemolysis, complement activation, PTT, platelet/leukocyte counts), and genotoxicity. All results met acceptance criteria.

Technological Characteristics

Over-the-wire balloon catheter; semi-compliant balloon; 1.0-4.0mm diameters; 5-15mm lengths; 14 atm rated burst pressure; 6 atm nominal pressure; hydrophilic lubricious coating; platinum radiopaque marker band; EO sterilization; compatible with 5F+ guiding catheters and 4F+ guiding sheaths.

Indications for Use

Indicated for patients with coronary artery or bypass graft stenosis (≥70% for 1.0-1.25mm; symptomatic ischemia for 1.5-4.0mm), acute myocardial infarction, or peripheral vascular disease (renal, femoral, popliteal, infra-popliteal, tibial, peroneal arteries).

Regulatory Classification

Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b) *Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2) *Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 04, 2020 OrbusNeich Medical Trading Inc. John Pazienza Senior Director, Engineering 5363 NW 35th Avenue Fort Lauderdale, Florida 33309 # Re: K200269 Trade/Device Name: Sapphire II PRO Balloon Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (Ptca) Catheter Regulatory Class: Class II Product Code: LOX, LIT Dated: January 31, 2020 Received: February 4, 2020 Dear Mr. John Pazienza: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200269 Device Name Sapphire II PRO Balloon Dilatation Catheter Indications for Use (Describe) The Sapphire® II PRO balloon dilatation catheter (1.0-1.25mm configurations) is indicated for: · balloon pre-dilatation of a stenotic portion of a coronary artery or bypass graft stenosis (270% stenosis) for the purpose of improving myocardial perfusion. The Sapphire® II PRO balloon dilatation catheter (1.5-4.0mm configurations) is indicated for: · balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. · balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction. The Sapphire® II PRO balloon dilatation catheter is also indicated for: · percutaneous transluminal angioplasty in the peripheral vasculature, including renal, fomoral, popliteal, tibial, and peroneal arteries. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary This 510(k) Summary is submitted in accordance with 21 CFR 807.92(c). | Submitter: | OrbusNeich Medical Trading, Inc.<br>5363 NW 35th Avenue<br>Fort Lauderdale, FL 33309<br>Phone: 954.730.0711<br>Fax: 954.730.7601 | |--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | John D. Pazienza | | Date Prepared: | January 31, 2020 | | Trade Name: | Sapphire II PRO Balloon Dilatation Catheter | | Common Name: | Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter | | Classification Name: | Catheters, transluminal coronary angioplasty, percutaneous<br>21 CFR 870.5100(a) | | Product Code: | LOX (primary)<br>LIT (secondary | | Device Class: | Class II | | Predicate Device: | OrbusNeich Sapphire II PRO (K180921; LOX, LIT; cleared 28-Jun-<br>2018) | | Reference Devices: | Abbott Vascular Mini Trek OTW (K110617; LOX; cleared 02-Jun-<br>2011)<br>Boston Scientific Emerge OTW (K130391; LOX; cleared 10-Jul-2013)<br>Medtronic Sprinter Legend OTW (K103095; LOX; cleared 22-Oct-<br>2010) | | Device Description: | The Sapphire II PRO Balloon Dilatation Catheter is now also available<br>as an over-the-wire balloon catheter with a working length of 150cm.<br>The semi-compliant balloons are available in diameters from 1.0-<br>1.25mm and lengths from 5-15mm with a rated burst pressure of 14<br>atmospheres. The catheter consists of proximal section with a Y-type<br>hub and distal section with a balloon near the distal tip. The straight<br>port of the hub is the guidewire entrance and the side port is used to<br>inflate and deflate the balloon. The external lumen provides for<br>inflation of the balloon with dilute contrast media solution.<br>The<br>internal lumen permits the use of a standard 0.014 inch guidewire to<br>facilitate advancement of the catheter to and through the stenosis to be<br>dilated. Hydrophilic lubricious coatings are applied to the distal<br>section. One radiopaque platinum marker band is located within the<br>balloon segment. The catheter is compatible with 5F or larger guiding<br>catheters and 4F or larger guiding sheaths. Two marked sections<br>(3mm in length) are located on the proximal shaft to indicate catheter<br>position relative to the tip of either guiding catheter or guiding sheath.<br>The design of this dilatation catheter does not incorporate a lumen for<br>distal dye injections or distal pressure measurements. | | Intended Use: | The Sapphire® II PRO Balloon Dilatation Catheter (Ø1.0-1.25mm<br>configurations) is indicated for:<br>• balloon pre-dilatation of a stenotic portion of a coronary artery or<br>bypass graft stenosis (≥70% stenosis) for the purpose of improving<br>myocardial perfusion<br>The Sapphire® II PRO Balloon Dilatation Catheter (Ø1.5-4.0mm<br>configurations) is indicated for:<br>• balloon dilatation of the stenotic portion of a coronary artery or<br>bypass graft stenosis in patients evidencing coronary ischemia for<br>the purpose of improving myocardial perfusion<br>• balloon dilatation of a coronary artery occlusion for the treatment<br>of acute myocardial infarction<br>The Sapphire® II PRO Balloon Dilatation Catheter is also indicated<br>for:<br>• percutaneous transluminal angioplasty in the peripheral<br>vasculature, including renal, femoral, popliteal, infra-popliteal,<br>tibial, and peroneal arteries. | | Technological Characteristics: | At a high level, the subject and predicate devices are based on the same<br>technological elements:<br>• the same indications for use<br>• identical semi-compliant balloon<br>• nominal pressure of 6 atm<br>• rated burst pressure of 14 atm<br>• identical distal shaft<br>• hydrophilic coating<br>• 0.014" guidewire compatibility<br>• EO sterilization<br>The following technological differences exist between the subject and | {4}------------------------------------------------ - predicate device: over-the-wire catheter shaft design ● - specific materials selected for proximal shaft - . exact dimensions of proximal shaft components and catheter {5}------------------------------------------------ Performance Data: Testing was leveraged from the predicate device (K180921) when applicable: - Shelf-Life (Partial) ● - Performance Testing (Partial) ● - Dimensional Verification (Partial) O - o Balloon Rated Burst Pressure (Partial) - Balloon Fatigue O - Balloon Compliance O - Catheter Bond Strength (Partial) O - Tip Pull Strength O - Radiopacity o - Coating Integrity o Additional testing was performed to support the use of the Sapphire II PRO balloon dilatation catheter over-the-wire shaft design: - Sterilization ● - . Performance Testing - o Visual Inspection - o Dimensional Verification - Balloon Preparation, Deployment, and Retraction O - o Balloon Rated Burst Pressure - o Shaft Burst - Balloon Inflation and Deflation Time O - o Catheter Bond Strength - Flexibility and Kinking O - Torque Strength O - Particulate Evaluation O - Biocompatibility - Cytotoxicity O - Intracutaneous Reactivity O - Sensitization O - Acute Systemic Toxicity o - Pyrogenicity o - O Hemocompatibility - · Hemolysis - Complement Activation - Partial Thromboplastin Time . - Platelet and Leukocyte Counts . - o Genotoxicity The Sapphire II PRO balloon dilatation catheter test results met all acceptance criteria and were similar to the predicate device. Conclusion: This information supports a determination of substantial equivalence between the Sapphire II PRO balloon dilatation catheter and the predicate device described above.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...