The Parafleet SC 014 / Parafleet SC 018 / Parafleet SC 035 is intended to dilate stenoses in the iliac, femoral, iliofemoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent dilatation post-deployment in the peripheral vasculature.
Device Story
Over-the-wire (OTW) peripheral balloon dilatation catheter; used for percutaneous transluminal angioplasty (PTA). Device comprises outer/inner tube lumen shaft, Y-hub manifold with luer lock fittings, balloon, and marker bands. One lumen for balloon inflation; second lumen for guidewire insertion (0.014, 0.018, or 0.035 inch). Operated by physicians in clinical settings. Physician inserts catheter over guidewire to target stenosis; inflates balloon to dilate vessel or post-deploy stent. Output is mechanical dilation of vessel/stent. Benefits include restoration of blood flow in peripheral arteries and dialysis fistulae.
Clinical Evidence
Bench testing only. No clinical data presented. Performance testing included dimensional verification, simulated use, balloon rated burst pressure, fatigue, compliance, inflation/deflation time, bond strength, tip pull strength, flexibility, kinking, torque, radiopacity, coating friction/integrity, particulate evaluation, and catheter body burst pressure. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, hemocompatibility, pyrogenicity).
Technological Characteristics
OTW PTA balloon catheter; semi-compliant balloon; 1.25-12.0mm diameter; 10-300mm length; 40-200cm working length. Materials include catheter shaft, balloon tubing, and coating. Sterilization via Ethylene Oxide (EtO). Compatible with 0.014/0.018/0.035 inch guidewires and 4-7F sheaths. Nominal pressure 6 atm; RBP 12-16 atm.
Indications for Use
Indicated for patients requiring dilation of stenoses in iliac, femoral, iliofemoral, popliteal, infrapopliteal, and renal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. Also indicated for post-deployment stent dilatation in peripheral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 27, 2025
BrosMed Medical Co., Ltd. Crystal Lee Registration and Clinical Medicine Department Manager 15th Building, SMEs Venture Park SongShan Lake Hi-Tech Industrial Development Zone Dongguan, Guangdong 523808 China
Re: K243704
Trade/Device Name: Parafleet SC 014 PTA Balloon Dilatation Catheter; Parafleet SC 018 PTA Balloon Dilatation Catheter; Parafleet SC 035 PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: November 28, 2024 Received: November 29, 2024
Dear Crystal Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ARIEL G. ASH- Digitally signed by ARIEL
SHAKOOR -S G. ASH-SHAKOOR -S
Date: 2025.02.27 10:00:55
-05'00'
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243704
Device Name Parafleet SC 014 PTA Balloon Dilatation Catheter; Parafleet SC 018 PTA Balloon Dilatation Catheter; Parafleet SC 035 PTA Balloon Dilatation Catheter
## Indications for Use (Describe)
The Parafleet SC 014 / Parafleet SC 035 is intended to dilate stenoses in the iliac. femoral, iliofemoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae.
This device is also indicated for stent dilatation post-deployment in the peripheral vasculature.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-family: Arial, sans-serif;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
#### 1. GENERAL INFORMATION
#### 1.1 Submitter
BrosMed Medical Co., Ltd. 15th Buildings, SMEs Venture Park SongShan Lake Hi-Tech Industrial Development Zone Dongguan 523808, China Office: +86 (769) 2289 2018 Fax: +86 (769) 2289 2016
#### 1.2 Contract person
Crystal Lee, Email: crystallee@brosmed.com Office: +86 (769) 2289 2018
1.3 Date of Preparation January 24, 2025
#### 2. NAME OF THE DEVICE
- 2.1.1 Trade/Proprietary Name Parafleet SC 014 PTA Balloon Dilatation Catheter; Parafleet SC 018 PTA Balloon Dilatation Catheter; Parafleet SC 035 PTA Balloon Dilatation Catheter;
### 2.1.2 Common/Usual Name Percutaneous Transluminal Angioplasty (PTA) Catheter
- 2.1.3 Classification Information
| Classification Name: | Percutaneous Catheter |
|----------------------------|-----------------------|
| Classification Regulation: | 21 CFR 870.1250 |
| Device Class: | Class II |
| Product Code: | LIT |
| Review Panel: | Cardiovascular |
#### 3. PREDICATE DEVICE AND REFERENCE DEVICES
- . Primary Predicate Device: Polux, Minerva, Atropos PTA Balloon Catheter (K160256, cleared on July 29, 2016)
- Reference Devices: . Ultraverse 014 and 018 PTA Balloon Dilatation Catheters (K192318, cleared on October 3, 2019) Ultraverse 035 PTA Balloon Dilatation Catheter (K142261, cleared on September 24, 2014) Sleek OTW PTA Catheter (K102035, cleared on December 1, 2010)
#### DESCRIPTION OF THE DEVICE 4.
The Parafleet balloon dilatation catheter family consists of Parafleet SC 014, Parafleet SC 018 and Parafleet SC 035 PTA balloon catheter. The PTA (OTW type) device is an over the wire (OTW) peripheral balloon catheter, specially designed for Percutaneous Transluminal Angioplasty (PTA). The
{5}------------------------------------------------
Parafleet SC catheter is available in working lengths of 40cm, 90cm, 120cm, 150cm and 200cm, with balloon diameters ranging from 1.25mm to 12.0mm. The Parafleet features an outer tube and inner tube lumen shaft ending in a Y-hub manifold with luer lock fittings. One lumen is used for inflation of the balloon and accessed via the side leg port. The second lumen, starting at the straight entry port, allows access to the distal tip of the catheter for a standard 0.014 inch / 0.018 inch / 0.035 inch guide wire insertion. The catheter is sterilized with EO and for single use only. The design of this dilatation catheter does not incorporate a lumen for distal dye injections or distal pressure measurements.
#### న. INDICATION FOR USE
The Parafleet SC 014 / Parafleet SC 018 / Parafleet SC 035 is intended to dilate stenoses in the illiac, femoral, iliofemoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
This device is also indicated for stent dilatation post-deployment in the peripheral vasculature.
#### COMPARISION SUMMARY WITH PREDICATE DEVICES 6.
The information of subject devices and predicate devices summarized as below Table 1.
# Table 1 Subject Device & Predicate Devices
| Trade Name | Parafleet SC<br>014, Parafleet<br>SC 018,<br>Parafleet SC<br>035 PTA<br>Balloon<br>Dilatation<br>Catheter<br>- Proposed<br>device | Polux, Minerva,<br>Atropos PTA<br>Balloon<br>Dilatation<br>Catheter<br>- Predicate<br>device | Ultraverse 014<br>and 018 PTA<br>Balloon<br>Dilatation<br>Catheters<br>- Reference<br>device | Ultraverse 035<br>PTA Balloon<br>Dilatation<br>Catheter<br>- Reference<br>device | Sleek OTW<br>PTA Catheter<br>-Reference<br>device |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|---------------------------------------------------|
| 510(K) Number | K243704 | K160256 | K192318,<br>K121856,<br>K093965 | K142261 | K102035 |
| Owner/Operator | BrosMed<br>Medical Co.,<br>Ltd | BrosMed<br>Medical Co.,<br>Ltd | Bard Peripheral<br>Vascular, Inc | Bard Peripheral<br>Vascular, Inc | Clear Stream<br>Technologies,<br>Ltd |
The subject devices have the following similarities to the predicate devices:
- Same principle of operation
- Same mechanism of action ●
- Same indication for use
- Same sterilization method
- Same balloon size
- Similar materials
- . Similar device components
Please see below Table 2 and Table 3 for the comparison between Subject devices and Predicate devices.
{6}------------------------------------------------
| Element of | Proposed | Predicate | Reference | Reference | Reference | Judgement |
|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Comparison | device | device | device | device | device | |
| Product<br>Name | Parafleet SC<br>014, Parafleet<br>SC 018,<br>Parafleet SC<br>035 | Polux,<br>Minerva,<br>Atropos | Ultraverse<br>014 and 018 | Ultraverse<br>035 | Sleek | / |
| Product<br>Code | LIT | LIT | LIT | LIT | LIT | Same |
| Regulation<br>Number | 870.1250 | 870.1250 | 870.1250 | 870.1250 | 870.1250 | Same |
| Device<br>Name | Catheter,<br>Angioplasty,<br>Peripheral,<br>Transluminal | Catheter,<br>Angioplasty,<br>Peripheral,<br>Translumina | Catheter,<br>Angioplasty,<br>Peripheral,<br>Transluminal | Catheter,<br>Angioplasty,<br>Peripheral,<br>Transluminal | Catheter,<br>Angioplasty,<br>Peripheral,<br>Transluminal | Same |
| Target Area | Peripheral<br>arteries and<br>arteriovenous<br>dialysis<br>fistulae | Peripheral<br>arteries and<br>arteriovenous<br>dialysis<br>fistulae | Peripheral<br>arteries and<br>arteriovenous<br>dialysis<br>fistulae | Peripheral<br>arteries and<br>arteriovenous<br>dialysis<br>fistulae | Peripheral<br>arteries and<br>arteriovenous<br>dialysis<br>fistulae | Same |
| Intended<br>use | The Parafleet<br>SC 014/<br>Parafleet SC<br>018/ Parafleet<br>SC 035 is<br>intended to<br>dilate<br>stenoses in<br>the iliac,<br>femoral,<br>iliofemoral,<br>popliteal,<br>infrapopliteal<br>, and renal<br>arteries, and<br>for the<br>treatment of<br>obstructive<br>lesions of<br>native or<br>synthetic<br>arteriovenous<br>dialysis<br>fistulae. This<br>device is also<br>indicated for<br>stent<br>dilatation | Polux,<br>Minerva,<br>Atropos are<br>indicated to<br>dilate<br>stenoses in<br>the iliac,<br>femoral,<br>iliofemoral,<br>popliteal,<br>infrapopliteal,<br>and renal<br>arteries, and<br>for the<br>treatment of<br>obstructive<br>lesions of<br>native or<br>synthetic<br>arteriovenous<br>dialysis<br>fistulae. This<br>device is also<br>indicated for<br>dilatation of<br>post-<br>deployed<br>stent in the | Ultraverse<br>014 and<br>Ultraverse<br>018 PTA<br>Balloon<br>Dilatation<br>Catheters are<br>recommended<br>for use in<br>percutaneous<br>trasluminal<br>angioplasty<br>(PTA) of the<br>renal,<br>popliteal,<br>tibial,<br>femoral, and<br>peroneal<br>arteries.<br>These<br>catheters are<br>not for use in<br>coronary<br>arteries. | The<br>Ultraverse<br>035 PTA<br>Dilatation<br>Catheter is<br>intended to<br>dilate<br>stenoses in<br>the peripheral<br>arteries, to<br>treat<br>obstructive<br>lesions of<br>native or<br>synthetic AV<br>fistulae<br>and/or re-<br>expand<br>endoluminal<br>stent graft<br>elements in<br>the iliac<br>arteries. This<br>device is also<br>recommended<br>for post-<br>dilatation of<br>balloon<br>expandable | Balloon<br>dilatation of<br>the femoral,<br>popliteal, and<br>infra-<br>popliteal<br>arteries.<br>These<br>catheters are<br>not designed<br>to be used in<br>the coronary<br>arteries. | Same with<br>Polux,<br>Minerva,<br>Atropos |
| post-<br>deployment<br>in the<br>peripheral<br>vasculature. | peripheral<br>vasculature. | and self-<br>expanding<br>stents in the<br>peripheral<br>vasculature.<br>This catheter<br>is not for use<br>in coronary<br>arteries. | | | | |
| | | | | | | |
# Table 2 General and Intended Use Comparison
{7}------------------------------------------------
{8}------------------------------------------------
| Technological<br>Characteristic | Proposed Device<br>Parafleet SC 014, Parafleet<br>SC 018, Parafleet SC 035 | Predicate/ Reference Devices<br>Polux, Minerva, Atropos<br>(K160256), Ultraverse 014 and<br>018 (K192318, K121856,<br>K093965), Ultraverse 035<br>(K142261), and Sleek (K102035) | Judgement |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Design and Performance Specification | | | |
| General Catheter Design | Consists of tip, marker<br>bands, balloon, catheter<br>shaft, strain relief, hub.<br>Over-the-Wire (OTW)<br>design | Consists of tip, marker bands,<br>balloon, catheter shaft, strain<br>relief, hub.<br>Over-the-Wire (OTW) design | Same |
| Compatible Guidewire<br>(in) | 0.014, 0.018, 0.035 | 0.014, 0.018, 0.035 (Polux,<br>Minerva, Atropos);<br>0.014, 0.018 (Ultraverse 014<br>and 018)<br>0.035 (Ultraverse 035) | Same |
| Compatible Sheath (F) | 4, 5, 6, 7 | 4, 5, 6, 7 (Polux, Minerva,<br>Atropos);<br>4, 5, 6 (Ultraverse 014 and 018)<br>5, 6, 7 (Ultraverse 035) | Same |
| Balloon Compliance<br>Type | Semi-Compliant | Semi-Compliant | Same |
| Balloon Diameter Range<br>(mm) | 1.25-12.0 | 1.5-10.0 (Polux, Minerva,<br>Atropos);<br>1.5-9.0 (Ultraverse 014 and 018)<br>3.0-12.0 (Ultraverse 035)<br>1.25-5.0 (Sleek) | Same |
| Balloon Length Range<br>(mm) | 10-300 | 5-200 (Polux, Minerva,<br>Atropos);<br>20-300 (Ultraverse 014 and 018)<br>20-300 (Ultraverse 035) | Similar |
| Catheter Working<br>Length (cm) | 40-200 | 40-150 (Polux, Minerva,<br>Atropos);<br>75-200 (Ultraverse 014 and 018)<br>75-130 (Ultraverse 035) | Similar |
| Nominal Pressure (atm) | 6 | 6 | Same |
| Rated Burst Pressure<br>(atm) | 12-16 | 14 (Polux, Minerva, Atropos);<br>11-16 (Ultraverse 014 and 018)<br>10-14 (Ultraverse 035) | Similar |
| Radio-detectability | Yes | Yes | Same |
| Biological Characteristics | | | |
| Tip | Parafleet SC 014 and Parafleet SC 018: Same as predicate<br>device<br>Parafleet SC 035: Similar to predicate device | | Substantially<br>Equivalent |
| Balloon Tubing | Same as predicate device | | Same |
| Catheter Shaft | Similar to predicate device | | Substantially<br>Equivalent |
| Coating | Same as predicate device | | Same |
| Sterilization Method | Ethylene Oxide (EtO) | | Same |
Table 3 Technological Comparison of Subject Device and Predicate Devices
{9}------------------------------------------------
#### 7. PERFORMANCE TESTING SUMMARY
Standard bench performance tests were conducted to confirm substantial equivalence between the Subject Device and the Predicate Devices.
#### > Bench Testing
The in vitro performance tests were conducted on subject device in accordance with FDA guidance 'Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions, Guidance for Industry and FDA Staff', including:
- Dimensional verification
- Simulated Use ●
- Balloon Rated Burst Pressure ●
- Balloon Fatigue
- Balloon Compliance
- Balloon Inflation and Deflation Time ●
- Catheter Bond Strength
- Tip Pull Strength
- Flexibility and Kinking
- Torque Strength
- . Radiopacity
- Coating Friction ●
- Coating Integrity
- Particulate Evaluation ●
- Catheter Body Bust Pressure
- Balloon Rated Burst Pressure (in stent)
- . Balloon Fatigue (in stent)
- A Biocompatibility Testing
The biocompatibility testing, conducted in accordance "Use of International Standard ISO 10993-1. Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", and International Standard "ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process", as recognized by FDA, included:
- . Cytotoxicity
- Sensitization ●
- Intracutaneous Reactivity ●
- Systemic Toxicity
- Hemocompatibility ●
- Hemolysis
- Thrombosis ●
- . Complement Activation
- Pyrogenicity ●
{10}------------------------------------------------
#### > Sterilization Packaging and Shelf Life
The test results met all acceptance criteria, were same or similar to the predicate devices, and ensure that the Parafleet SC 014 / Parafleet SC 018 / Parafleet SC 035 PTA Balloon Dilatation Catheter design and construction are suitable for its intended use.
#### CONCLUSION 8.
The information presented in this 510(k) submission demonstrates that the Parafleet SC 014, Parafleet SC 018, and Parafleet SC 035 are Substantially Equivalent (SE) to the marketed predicate devices described above.