K163660 · Biotronik, Inc. · LOX · Jan 26, 2017 · Cardiovascular
Device Facts
Record ID
K163660
Device Name
Pantera LEO
Applicant
Biotronik, Inc.
Product Code
LOX · Cardiovascular
Decision Date
Jan 26, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pantera LEO is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion and for post dilatation of coronary stents.
Device Story
Pantera LEO is a sterile, single-use, intravascular balloon catheter used in coronary angioplasty. It features a tapered soft tip for lesion crossing, a foldable balloon for low profile, and radiopaque markers for fluoroscopic positioning. The proximal section consists of a stainless steel hypotube with a Luer port for inflation/deflation. The device is coated with hydrophobic silicone on the hypotube and balloon, and a hydrophilic polymer on the distal shaft. Used by physicians in clinical settings (e.g., cath lab) under fluoroscopic guidance to dilate stenotic segments or expand stents. The physician manually inflates the balloon to a specific diameter and pressure based on a compliance chart to restore vessel patency, thereby improving myocardial perfusion. The device is a mechanical tool; it does not involve electronic processing or software.
Clinical Evidence
No clinical data submitted. Substantial equivalence is based on in-vitro bench testing, including dimensional verification, balloon fatigue, burst pressure, compliance, bond strength, torque, radiopacity, coating integrity, and particulate evaluation, performed in accordance with FDA guidance for PTCA catheters.
Technological Characteristics
Intravascular balloon catheter; stainless steel hypotube; hydrophobic silicone coating (proximal/balloon); hydrophilic polymeric-based coating (distal); radiopaque markers. Biocompatibility per ISO 10993-1 for limited contact (≤ 24 h) with circulating blood. Mechanical inflation/deflation via Luer port. No electronic or software components.
Indications for Use
Indicated for balloon dilatation of stenotic coronary arteries or bypass grafts to improve myocardial perfusion and for post-dilatation of coronary stents.
Regulatory Classification
Identification
Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus-like design, with three figures representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 26, 2017
Biotronik, Inc. Jon Brumbaugh Vice President, Regulatory Affairs And Compliance 6024 Jean Road Lake Oswego, Oregon 97035
Re: K163660
Trade/Device Name: Pantera LEO Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: December 22, 2016 Received: December 27, 2016
Dear Jon Brumbaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando
Fernando Aguel-S
Aguel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163660
Device Name Pantera LEO
Indications for Use (Describe)
The Pantera LEO is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion and for post dilatation of coronary stents.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10px;">☑</span> Research Use (Part 21 CFR 201.Subpart B) | ☐ Over-The-Counter Use (21 CFR 201.66) |
|----------------------------------------------------------------------------------|----------------------------------------|
|----------------------------------------------------------------------------------|----------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary per 21 CFR 807.92
| Date Prepared: | January 25, 2017 |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name and Address of<br>Applicant: | BIOTRONIK, Inc.<br>6024 Jean Road<br>Lake Oswego, OR 97035 |
| 510(k) Contact Person<br>and Phone Number: | Jon Brumbaugh<br>Vice President, Regulatory Affairs and Compliance<br>Phone: (888) 345-0374<br>Fax: (503) 635-9936<br>jon.brumbaugh@biotronik.com |
| Proprietary Name: | Pantera LEO |
| Common Name: | Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter |
| Device Classification: | Class II (special controls) |
| Regulation Number: | 870.5100 |
| Classification Name: | Catheter, transluminal coronary angioplasty, percutaneous |
| Product Code: | LOX |
| Predicate Device(s): | BIOTRONIK Pantera Pro (K160985, cleared October 7, 2016) |
| Device Description: | The Pantera LEO PTCA catheter is a sterile, single-use, intravascular balloon<br>catheter for the dilatation of stenotic segments in coronary arteries or bypass<br>grafts. The dilatation balloon is designed to inflate to a known diameter and<br>length at a specific inflation pressure consistent with the compliance chart, which<br>is included in the Instructions for Use (IFU), Compliance Data Card (CDC), and<br>on the labels.<br>The dilatation catheter has a soft tip that is tapered at the distal end of the<br>catheter to facilitate advancement of the catheter. The balloon is folded to<br>achieve a low crossing profile and is connected to the tip. A radiopaque marker<br>is located at each end of the balloon cylindrical section to facilitate fluoroscopic<br>visualization and positioning of the balloon catheter to and across the lesion.<br>The proximal section of the catheter is a single lumen stainless steel hypotube<br>with a single Luer port for the inflation / deflation of the balloon. The<br>Pantera LEO has two coatings; a hydrophobic silicone coating on the outer<br>surface of the hypotube (proximal outer shaft) and balloon, and a hydrophilic<br>polymeric-based coating on the outer surface of the distal outer shaft.<br>The dilatation catheter is compatible with guide wire and guiding catheter sizes<br>according to the recommendations on the label. |
{4}------------------------------------------------
| Pantera LEO Available Sizes<br>(Usable Catheter Length = 145 cm) | | | | | |
|------------------------------------------------------------------|---------------------|----|----|----|----|
| | Balloon length [mm] | | | | |
| Balloon Ø [mm] | 8 | 12 | 15 | 20 | 30 |
| 2.0 | X | X | X | X | X |
| 2.25 | X | X | X | X | X |
| 2.5 | X | X | X | X | X |
| 2.75 | X | X | X | X | X |
| 3.0 | X | X | X | X | X |
| 3.25 | X | X | X | X | X |
| 3.5 | X | X | X | X | X |
| 3.75 | X | X | X | X | X |
| 4.0 | X | X | X | X | X |
| 4.5 | X | X | X | X | X |
| 5.0 | X | X | X | X | X |
| X | Available sizes | | | | |
Indications For Use: The Pantera LEO is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion and for post dilatation of coronary stents.
Purpose of Submission: BIOTRONIK submits this 510(k) for clearance of Pantera LEO Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter. The predicate device is the Pantera Pro PTCA Dilatation Catheter (K160985, cleared October 7, 2016). The modifications described in this submission are limited to the balloon compliance and materials of construction, which are utilized in other BIOTRONIK devices.
Comparison of Pantera LEO has the same intended use as the predicate device. The Technological indications for use are similar, with only minor differences in the wording for post-Characteristics dilatation. The fundamental scientific technology and principles of operation of the Pantera LEO are identical to the predicate device Pantera Pro. There are only minor technological differences between the Pantera LEO and the Pantera Pro. which can be summarized as differences in materials of construction of a few device components. However, these differences in material do not affect the principles of operation or intended use. The safety and effectiveness of Pantera LEO are supported by product testing conducted according to the BIOTRONIK design controls.
{5}------------------------------------------------
## Summary of Non-Clinical Data
Testing according to BIOTRONIK design control and verification/validation processes demonstrates that the Pantera LEO is comparable to the predicate device. Modifications to technological characteristics do not raise new or different questions of safety and effectiveness.
### Design Verification in-vitro testing:
The following in-vitro bench tests were completed on the Pantera LEO in accordance with the requirements of Guidance for Industry and FDA Staff -Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010, and verify that it meets the required performance specifications.
●
- . Dimensional Verification
- Balloon Preparation. Deployment . and Retraction
- Balloon Rated Burst Pressure .
- . Balloon Fatique (Repeat Balloon Inflations)
- . Balloon Compliance (Diameter vs. Pressure)
- . Tip Pull Test
- Flexibility and Kink Test ●
Catheter Bond Strength
- Torque Strength ●
- Radiopacity ●
- Coating Integrity ●
- Particulate Evaluation .
- Shelf Life testing to support a 3-year shelf life
- . Balloon Inflation and Deflation Time
Additional Tests for Catheters Intended for In-Stent Restenosis or for Stent Expansion following Stent Deployment:
- Balloon Rated Burst Pressure in Stent .
- Balloon Fatique / Repeat Balloon Inflation in Stent .
## Biocompatibility Testing:
The Pantera LEO PTCA Catheter is intended for short-term use in direct contact with circulating blood. Therefore, it is classified in accordance with ISO 10993-1 "Biological evaluation of medical devices" as an external communicating device with limited (≤ 24 h) contact duration with circulating blood. Biocompatibility testing was conducted to demonstrate that the requirements of ISO 10993-1 and the 2016 FDA Guidance for ISO 10993 (#1811) have been met for the Pantera LEO PTCA balloon catheter. There was no evidence of adverse biological effects, indicating that there is no biological risk caused by the materials/device or possible leachable substances. Therefore, it can be concluded that the Pantera LEO balloon catheters are biologically safe for their intended use.
The Pantera LEO met all specified design and performance requirements. No new safety or effectiveness issues were raised during the testing. The bench validation testing and biocompatibility testing demonstrated that the Pantera LEO is substantially equivalent in terms of safety and effectiveness to the predicate device.
#### Summary of Clinical The determination of substantial equivalency on this subject device does not rely Data: upon the clinical data. There is no clinical data submitted in this application.
{6}------------------------------------------------
### Conclusion:
The intended use of the subject device, Pantera LEO, as described in the labeling, is identical to that of the predicate device, Pantera Pro (K160985). In addition, the indications for use are similar, with only minor differences in the wording for post-dilatation, and the fundamental scientific technology has not changed. BIOTRONIK believes that the minor differences proposed in this submission do not raise new or different questions regarding safety and effectiveness and that substantial equivalence for safety and effectiveness can be determined using existing design controls. Summaries of performance testing, showing Pantera LEO performs equivalently or better than the predicate, have been provided in this Special 510(k). The biocompatibility testing shows the materials of construction are biologically safe for the intended use. Therefore, Pantera LEO should be considered substantially equivalent to the predicate device, Pantera Pro.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.