K103618 · Ev3, Inc. · MCW · Jan 5, 2011 · Cardiovascular
Device Facts
Record ID
K103618
Device Name
TURBOHAWK PEROPHERAL PLAQUE EXCISION SYSTEM
Applicant
Ev3, Inc.
Product Code
MCW · Cardiovascular
Decision Date
Jan 5, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Intended Use
The TurboHawk Peripheral Plaque Excision System is intended for use in atherectomy of the peripheral vasculature. The TurboHawk Catheter is NOT intended for use in the coronary, carotid, iliac, or renal vasculature.
Device Story
TurboHawk Peripheral Plaque Excision System is a catheter-based device for peripheral artery atherectomy. System comprises a flexible catheter tracking over a 0.014" guidewire and a handheld, battery-driven ev3 Cutter Driver. Distal end features a rotating inner cutter within a tubular housing. Physician operates device by engaging cutter via thumb switch and advancing catheter across lesion; rotating cutter shaves occlusive material, which is captured and stored in the device tip. Process is repeated to achieve desired plaque excision. Used in clinical settings for treatment of atherosclerotic lesions. Output is physical removal of plaque, allowing physician to restore vessel patency. Benefits include reduction of arterial blockage.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing only. In vitro tests included effective length, guidewire loading, cutter height, cycle/life, torque/tensile strength of joints, trackability, cut depth, cut mass per pass, and tissue removal cycles. Testing leveraged from predicate K093301 for spin percentage, biocompatibility, packaging, and sterilization.
Technological Characteristics
Catheter-based atherectomy system. Flexible shaft tracks over 0.014" guidewire. Distal assembly features rotating inner cutter in tubular housing. Powered by handheld, battery-driven cutter driver. Mechanical cutting principle. No software or electronic algorithms described.
Indications for Use
Indicated for atherectomy of peripheral vasculature to treat de novo and restenotic calcified and non-calcified atherosclerotic lesions. Contraindicated for use in coronary, carotid, iliac, or renal vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
TurboHawk™ Peripheral Plaque Excision System (K093301)
SilverHawk™ Peripheral Plaque Excision System (K061188)
Related Devices
K093301 — TURBOHAWK PERIPHERAL PLAQUE EXCISION SYSTEM · Ev3, Inc. · Nov 6, 2009
K170191 — SilverHawk Peripheral Plaque Excision System, TurboHawk Peripheral Plaque Excision System · Medtronic Vascular, Inc. · Jun 16, 2017
K111723 — TURBOHAWK PERIPHERAL PLAQUE EXCISION SYSTEM · Ev3, Inc. · Oct 27, 2011
K141801 — HAWKONE DIRECTIONAL ATHERECTOMY SYSTEM · Covidien, LLC · Oct 16, 2014
K061188 — SILVERHAWK PERIPHERAL PLAQUE EXCISION SYSTEM · Foxhollow Technologies, Inc. · Oct 23, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for EV3, which is a robotics platform developed by LEGO. The logo consists of the letters "ev3" in a bold, sans-serif font, with a small cluster of three circles above the letters. The circles are arranged in a triangular shape, with two circles on top and one circle below, creating a visual element that is simple and recognizable.
Your endovascular company."
# JAN - 5 2011
K1036/8
### 510(k) Summary
## TurboHawk™ Peripheral Plaque Excision System
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R § 807.92.
#### 1. Submitter Information
| Applicant | ev3 Inc.<br>3033 Campus Drive<br>Plymouth, MN 55441-2651<br>Tel: 763-398-7000<br>Fax: 763-591-3248 |
|-----------------------|----------------------------------------------------------------------------------------------------|
| Contact Person | Brenda Johnson<br>Principal Regulatory Affairs Specialist |
| Date Prepared | December 3, 2010 |
| 2. Subject Device | |
| Device Trade Name | TurboHawk™ Peripheral Plaque Excision System |
| Device Common Name | Catheter, Peripheral, Atherectomy |
| Classification Name | Intraluminal Artery Stripper<br>21 CFR 870.4875, Product Code MCW |
| Classification Panel | Cardiovascular |
| 3. Predicate Devices | |
| Device Trade Name | TurboHawk™ Peripheral Plaque Excision System;<br>SilverHawk™ Peripheral Plaque Excision System |
| 510(k) Number | K093301; K061188 |
| 510(k) Clearance Date | November 6, 2009; October 23, 2006 |
#### 4. Device Description
The TurboHawk Peripheral Plaque Excision System (TurboHawk Catheter and ev3 Cutter Driver) is designed for the treatment of de novo and restenotic calcified and non-calcified atherosclerotic lesions located in native peripheral arteries. The TurboHawk Catheter consists of a flexible shaft designed to track over a 0.014" guidewire. At the distal end of the TurboHawk Catheter is a small cutting assembly comprised of a rotating inner cutter contained within a
{1}------------------------------------------------
tubular housing. The proximal end of the TurboHawk Catheter contains a connector and cutter positioning lever (thumb switch) designed to fit into the ev3 Cutter Driver. The ev3 Cutter Driver is a handheld, disposable, battery-driven unit (Catalog No: 02550) which powers the system.
The TurboHawk Peripheral Plaque Excision System has two switches: 1) the SilverHawk Cutter Driver main power switch and 2) the TurboHawk Catheter thumb switch. The ev3 Cutter Driver main power switch supplies power to the device when turned ON. The TurboHawk Catheter thumb switch activates the drive shaft and engages the cutter when pulled proximally to the ON position. With the cutter engaged, the TurboHawk Catheter is slowly advanced across the lesion, shaving occlusive material from the artery. The excised tissue is captured and stored in the tip of the device. The cutting process is completed by advancing the TurboHawk Catheter thumb switch distally deactivating the drive shaft and disengaging the cutter. The TurboHawk Catheter thumb switch is fully advanced distally to the OFF position in order to pack the excised plaque into the tip. This cutting sequence is repeated as necessary to achieve the desired degree of plaque excision.
#### 5. Indications for Use
The TurboHawk Peripheral Plaque Excision System is intended for use in atherectomy of the peripheral vasculature. The TurboHawk Catheter is NOT intended for use in the coronary, carotid, iliac, or renal vasculature.
#### 6. Comparison of Technological Characteristics
The TurboHawk Peripheral Plaque Excision System has the following similarities to the predicate device:
- . Identical indications for use
- . Identical intended use as the predicate TurboHawk Peripheral Plaque Excision System
- . Similar fundamental scientific technology
- . Similar operating principle
- Similar materials .
#### 7. Performance Testing Summary
To demonstrate substantial equivalence of the subject device, the TurboHawk Peripheral Plaque Excision System to the predicate device, the technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Analysis procedures, the following in vitro tests were performed:
- Effective length .
- Guidewire loading ●
- Cutter height ●
- Cycle and life
- Carbide edge attachment
- Proximal drive shaft joint torque test ●
- � Distal drive shaft assembly torque test
- Distal drive shaft assembly tensile test �
- � Tip, hinge, and distal torque shaft joints tensile test
- Cutter stop tensile .
- Trackability .
- Cut depth ●
{2}------------------------------------------------
- . Cut mass per pass
- . Guidewire lumen zip
- . Embolization
- . Tissue removal cycles
Testing leveraged from the predicate TurboHawk Device included spin percentage, biocompatibility, packaging and sterilization. Test results met the specified acceptance criteria and were included in K093301.
The results from these tests demonstrate that the technological characteristics and performance criteria of the TurboHawk Peripheral Plaque Excision System are comparable to the predicate device and that the TurboHawk Peripheral Plaque Excision System performs in a manner equivalent to the predicate device currently on the market for the same intended use.
#### 8. Conclusions
Based on the intended use, technological characteristics, safety and performance testing included in this submission, ev3 considers the TurboHawk Peripheral Plaque Excision System to be substantially equivalent to the predicate TurboHawk Peripheral Plaque Excision System (K093301) and the predicate SilverHawk Peripheral Plaque Excision System (K061188).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing the wings and a wavy line representing the body.
Food and Drug Administration 1 0903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
ev3, Inc. c/o Mr. Mark Job Reviewer Regulatory Technology Services, LLC 1394 25th Street NW Buffalo, MN 55313
Re: K103618
Trade/Device Name: TurboHawk Peripheral Plaque Excision System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal artery stripper Regulatory Class: Class II (two) Product Code: MCW Dated: December 9, 2010 Received: December 10, 2010
#### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
JAN - 5 2011
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register ning your device in the Federal Icegister. The first for
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
{4}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Duna R. Palmer
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
JAN - 5 2011
Page 1 of 1
### Indications for Use Statement
510(k) Number (if known): K103618
Device Name: TurboHawk Peripheral Plaque Excision System Indications for Use:
The TurboHawk Peripheral Plaque Excision System is intended for use in atherectomy of the peripheral vasculature. The TurboHawk Catheter is NOT intended for use in the coronary, carotid, iliac, or renal vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Duma R. Lumer
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number K103618
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.