CLEANER ROTATIONAL THROMBECTOMY SYSTEM

K060904 · Rex Medical, L.P. · MCW · Apr 28, 2006 · Cardiovascular

Device Facts

Record IDK060904
Device NameCLEANER ROTATIONAL THROMBECTOMY SYSTEM
ApplicantRex Medical, L.P.
Product CodeMCW · Cardiovascular
Decision DateApr 28, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4875
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Cleaner™ Rotational Thrombectomy System is designed for mechanical declotting of synthetic dialysis access grafts.

Device Story

Percutaneous mechanical thrombectomy device; restores patency to occluded synthetic dialysis access grafts. System utilizes rotational guidewire to perform mechanical declotting. Design includes flexible safety core within rotational guidewire and lubricious coating on macerating element. Operated by clinicians in clinical setting. Device functions as mechanical tool; no software or automated analysis involved. Benefits patient by restoring graft function without surgical intervention.

Clinical Evidence

No clinical data; bench testing only. Safety and effectiveness supported by risk assessment and verification testing of design changes based on internal criteria and ISO standards.

Technological Characteristics

Mechanical thrombectomy catheter; features rotational guidewire with internal flexible safety core and lubricious coating on macerating element. Class II device (Product Code: MCW).

Indications for Use

Indicated for mechanical declotting of synthetic dialysis access grafts in patients requiring restoration of graft patency.

Regulatory Classification

Identification

An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Exhibit #1 APE 28 11 Special 510(k): Device Modification Cleaner™ Rotational Thrombectorny System Image /page/0/Picture/3 description: The image shows the logo for REX MEDICAL. The logo consists of the text "REX MEDICAL" in a bold, sans-serif font. A curved graphic element is placed to the left of the text, partially encircling the "R" in "REX". The logo is in black and white. # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is: __ K 06 0904 ### 1. Submitter's Identification: Rex Medical, LP 1100 East Hector Street, Suite 245 Conshohocken, PA 19428 Contact: Mr. John Leedle Date Summary Prepared: February 9, 2006 ### 2. Name of the Device: Cleaner™ Rotational Thrombectomy System #### 3. Predicate Device Information: K031610, Cleaner™ Rotational Thrombectomy System K040252, Cleaner™ Rotational Thrombectomy System Kit ### 4. Device Description: The Rex Medical Cleaner™ Rotational Thrombectomy System is percutaneous mechanical thrombectomy device which provides an effective means to restore patency to occluded synthetic dialysis access grafts. The device is used in exactly the same manner as the predicate devices. #### 5. Intended Use: The Cleaner™ Rotational Thrombectomy System is designed for mechanical declotting of synthetic dialysis access grafts. {1}------------------------------------------------ ### 6. Comparison to Predicate Devices: The Cleaner™ Rotational Thrombectomy System is identical to the predicate Cleaner™ Rotational Thrombectomy Systems with some design improvements added to the rotational guidewire. A flexible safety core was added to the inside of the rotational guidewire and a lubricious coating was added over the macerating element. ## 7. Discussion of Non-Clinical Tests Performed: All testing performed on the Cleaner™ Rotational Thrombectomy System was derived from the risk assessment which evaluated the safety and effectiveness of the design changes to the guidewire. Test methodology and acceptance criteria were derived from within Rex Medical and from related ISO standards for evaluation of this device. #### 8. Discussion of Clinical Tests Performed: Not Applicable #### ல் Conclusions: The subject device, Cleaner™ Rotational Thrombectomy System, has identical indications for use as the predicate Cleaner™ Rotational Thrombectomy Systems (K040252 and K031610). The verification testing contained in our submission demonstrates that there are no differences in their technological characteristics due to the device modification, thereby not raising any new issues of safety or effectiveness. The Cleaner™ Rotational Thrombectorny System is therefore substantially equivalent to the predicate Cleaner™ Rotational Thrombectomy Systems (K040252 and K031610). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes entwined around it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 2 8 2005 Rex Medical c/o Ms. Susan Goldstein-Falk mdi Consultants, Inc. 55 Northern Blvd. Suite 200 Great Neck, NY 11021 Re: K060904 > Trade Name: Rex Medical Cleaner™ Rotational System Regulation Number: 21CFR §870.4875 Regulation Name: Thrombectomy Catheter Regulatory Class: II (two) Product Code: MCW Dated: February 9, 2006 Received: April 03, 2006 Dear Ms. Goldstein-Falk: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA mav publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Susan Goldstein-Falk Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Duna R. Luchner Image /page/3/Picture/5 description: The image contains a signature and the words "Br." and "Di.". The signature is a stylized symbol with curved lines and sharp angles. The letters "Br." and "Di." are positioned to the right of the signature, suggesting they might be initials or an abbreviation related to the signature. Bram D. Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Exhibit B # Indications for Use of of of 1 a Page 1 510(k) Number (if known): K060904 Device Name: Cleaner ™ Rotational Thrombectomy System Indications For Use: The Rex Medical Cleaner™ Rotational Thrombectomy System is designed for mechanical declotting of synthetic dialysis access grafts. Prescription Use _ X (Per 21 CFR 801 Subpart D) OR Over-The Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Dougher 12 - Va (Division Sign-Off) (Division Sign-On) Division of Cardiovascular Devices 510(k) Number 166904
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...