The SilverHawk Peripheral Plaque Excision System is intended for use in atherectomy of the peripheral vasculature. The catheter is not intended for use in the coronary or carotid vasculature.
Device Story
System consists of SilverHawk Peripheral Catheter and SilverHawk Cutter Driver; used together for peripheral atherectomy. Device removes plaque from peripheral vasculature; not for coronary or carotid use. Provided sterile for single-use. Operated by clinicians in clinical setting. Output is physical excision of plaque; aids in revascularization. Benefits patient by removing obstructive plaque from peripheral vessels.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and performance testing.
Technological Characteristics
Peripheral atherectomy catheter system. Biocompatible per ISO-10993-1. Catheter sterilized via Ethylene Oxide (ANSI/AAMI/ISO 11135); cutter driver sterilized via Gamma (ANSI/AAMI/ISO 11137, 25kGy). Single-use.
Indications for Use
Indicated for atherectomy of the peripheral vasculature. Not indicated for coronary or carotid vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
SilverHawk Peripheral Plaque Excision System (K043553)
Related Devices
K061188 — SILVERHAWK PERIPHERAL PLAQUE EXCISION SYSTEM · Foxhollow Technologies, Inc. · Oct 23, 2006
K071432 — SILVERHAWK-R PERIPHERAL PLAQUE EXCISION SYSTEM FOR SURGICAL USE · Fox Hollow Technologies · Feb 12, 2008
K061063 — MODIFICATION TO: SILVERHAWK PERIPHERAL PLAQUE EXCISION SYSTEM, MODELS P4028/FG08300 · Foxhollow Technologies, Inc. · May 18, 2006
K103618 — TURBOHAWK PEROPHERAL PLAQUE EXCISION SYSTEM · Ev3, Inc. · Jan 5, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
### JAN 1 3 2006
K0534/60
## 510(k) SUMMARY
#### Device Name
Classification Name:
Common and Usual Name: Proprietary Name:
Catheter, Peripheral, Atherectomy 21 CFR \$870.4875, Class II Catheter, Peripheral, Atherectomy SilverHawk Peripheral Plaque Excision System
#### Predicate Device
The SilverHawk Peripheral Plaque Excision System (K043553), currently marketed by FoxHollow Technologies, Inc. (Redwood City, CA).
#### Summary
This summary of Special 510(k) safety and effectiveness is being submitted in accordance with requirements of SMDA 1990.
The SilverHawk Peripheral Plaque Excision System is intended for atherectomy of the peripheral vasculature and is not intended for use in the coronary or carotid vasculature. The SilverHawk Peripheral Plaque Excision System consists of two major components which are packaged separately, but used together during atherectomy procedures. The two components are the SilverHawk Peripheral Catheter and SilverHawk Cutter Driver.
The SilverHawk Peripheral Plaque Excision System will be provided sterile for singleuse. The catheter will be sterilized by Ethylene Oxide (ANSI/AAMI/ISO 11135), including limits for Ethylene Oxide residuals and validated to a sterility assurance level (SAL) of 10 ". The cutter driver is sterilized by Gamma Sterilization Cycle (ANSI/AAMI/ISO 11137), providing a minimum SAL of 10°, with a minimum dose of 25kGy, using the VDmax method. The device is biocompatible per ISO-10993-1.
The SilverHawk Peripheral Plaque Excision System is substantially equivalent in material of construction, overall design, intended use, and safety and efficacy to the predicate device. The subject device was shown to have substantially equivalent performance when compared to the predicate device.
The SilverHawk Peripheral Plaque Excision System with labeling modifications is considered substantially equivalent to the SilverHawk Peripheral Plaque Excision System (K043553).
Contact:
Date: January 5, 2006
Melissa Murphy Regulatory Specialist FoxHollow Technologies, Inc. 740 Bay Road Redwood City, CA 94063 (650) 421-8579
{1}------------------------------------------------
Public Health Service
Image /page/1/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
JAN 1 3 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FoxHollow Technologies, Inc. c/o Ms. Melissa Murphy Regulatory Specialist 740 Bay Road Redwood City, CA 94063
Re: K053460
SilverHawk Peripheral Plaque Excision System Regulation Number: 21 CFR 870.4875 Regulation Name: Peripheral Atherectomy Catheter Regulatory Class: Class II (two) Product Code: MCW Dated: December 9, 2005 Received: December 13, 2005
Dear Ms. Murphy:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your be determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Ms. Melissa Murphy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276- . . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Diring R. Victinel
Image /page/2/Picture/5 description: The image shows a signature. The signature is illegible, but it appears to be written in cursive. The signature is black and is on a white background. The signature is located on the left side of the image.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known): . K053460 _ .
Device Name: SilverHawk Peripheral Plaque Excision System
Indications For Use:
・・・
The SilverHawk Peripheral Plaque Excision System is intended for use in atherectomy of the peripheral vasculature. The catheter is not intended for use in the coronary or carotid vasculature.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dmn R. Volmer
(Division Sign-Off) (Division Sign-on)
Division of Cardiovascular Devices
510(k) Number
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.