The PM-8000 Patient Monitor is a vital signs monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PM-8000 Patient Monitor has many features and functions, yet is easy to use through an integrated keypad, knob and an intuitive menu system. The patient parameters that can be monitored by the PM-8000 Patient Monitor are: ECG (3-lead or 5-lead selectable), Heart Rate derived from selected source (SPO2, ECG), Respiration Rate (derived from ECG), Non-Invasive Blood Pressure (NIBP), Saturation of Pulse Oxygen (SpO2), Temperature (TEMP) and invasive pressure (IBP). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs. The PM-8000 Patient Monitor is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment. The PM-8000 Patient Monitor is not recommended for use in a patient's home or residence, during the patient transport, or when it has not been ordered by a physician.
Device Story
Battery or line-powered patient monitor; acquires physiological signals (ECG, RESP, NIBP, SpO2, TEMP, IBP). Signals converted to digital data; processed for alarm conditions; presented on color TFT display. Optional built-in recorder provides hard copies of waveforms and trends. Used in healthcare facilities by qualified medical personnel. Operator adjusts alarm settings via keypad/knob/menu; audible/visual alerts notify of excursions. Output assists clinical decision-making regarding patient vital signs. Modification adds new ECG cables and leadwires.
Clinical Evidence
Bench testing (electrical safety, function/performance, environmental) and clinical testing were conducted. Results demonstrate the device meets design specifications and performs as well as the predicate device.
Technological Characteristics
Battery or line-powered patient monitor. Features color TFT display, integral keypad, and knob. Supports 3-lead or 5-lead ECG, NIBP, SpO2, temperature, and IBP monitoring. Includes optional built-in recorder. Connectivity via ECG cables and leadwires.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in healthcare facilities for monitoring ECG, heart rate, respiration rate, NIBP, SpO2, temperature, and invasive pressure. Not for home use, patient transport, or use without physician order.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
#### 510 (k) Summary
#### Exhibit #1 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
The assigned 510(k) number is: K04 3348
## Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 2658 2888
Fax: +86 755 2658 2680
#### Contact Person: ●
Li Dongling Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
#### ● Date Prepared:
November 15, 2004
### Name of the device:
- Trade/Proprictary Name: PM-8000 Patient Monitor (ECG Cables and Leadwires for the PM-8000 Patient Monitor)
- Common Name: Patient Monitor
#### ● Classification
| 21 CFR 870.2300 | Cardiac monitor (including cardiotachometer and rate alarm) | Class II |
|-----------------|---------------------------------------------------------------------|----------|
| 21 CFR 870.1130 | Non-Invasive blood pressure measurement System | Class II |
| 21 CFR 870.1110 | Blood pressure computer | Class II |
| 21 CFR 880.2910 | Clinical Electronic Thermometer -<br>Temperature Monitor with Probe | Class II |
| 21 CFR 870.2700 | Oximeter, Pulse | Class II |
{1}------------------------------------------------
# Legally Marketed Predicate Device:
K032733 PM-8000 Patient Monitor
# Description:
The PM-8000 Patient Monitor is a battery or line-powered patient monitor. The PM-8000 Patient Monitor acquires the physiological signals such as ECG, respiration (RESP), non-invasive blood pressure (NIBP), Saturation of Pulse Oxygen (SpO2), temperature (TEMP) and invasive pressure (IBP). These physiological signals are converted into digital data and processed. The PM-8000 Patient Monitor examines the data for alarm conditions and presents them on the color TFT display.
The optional built-in recorder provides hard copies of all digital data and waveforms as well as Tabular and Graphic Trend Information.
#### Statement of intended Use:
The PM-8000 Patient Monitor is a vital signs monitor used on a human patient. The target populations are adult, pediatric and neonatal patients. The PM-8000 Patient Monitor has many features/and functions, yet is easy to use through an integral keypad, knob and an intuitive menu system.
The patient parameters that can be monitored by PM-8000 are: ECG (3-lead or 5-lead selectable), Heart Rate derived from selected source (SpO2, ECG), Respiration Rate (derived from ECG), Non-invasive blood pressure (NIBP), Saturation of Pulse Oxygen (SpO2), Temperature (TEMP) and Invasive pressure (IBP). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The PM-8000 Patient Monitor is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.
The PM-8000 Patient Monitor is not recommended for use in a patient's home or residence, during the patient transport, or when it has not been ordered by a physician.
### Comparison of Technological Characteristics:
The PM-8000 Patient Monitor employs the same functional technology as the predicate device.
.
{2}------------------------------------------------
# Test Summary:
The device modification involved the addition of Shenzhen Mindray ECG cables and leadwires for the PM-8000 Patient Monitor. The purpose of adding these 6 new ECG cables and lead wires is a business decision. The subject 6 new ECG cables and lead wires are selectable by customers; we will not discontinue marketing the previously cleared ECG cables and lead wires.
Laboratory testing was conducted to validate and verify that the PM-8000 Patient Monitor met all design specifications and was substantially equivalent to predicate device.
The following quality assurance measures were applied to the development of the device:
- Risk Analysis
- Electrical Safety Testing
- . Function and Performance Testing
- Environmental Testing
- Clinical Testing
## Conclusion:
The conclusions drawn from clinical and laboratory testing of the PM-8000 Patient Monitor (ECG Cables and Leadwires for the PM-8000 Patient Monitor) demonstrates that the device is as safe, as effective, and performs as well as the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/11 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three heads, representing the department's focus on health, human services, and science. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 6 2005
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. c/o Ms. Susan D. Goldstein-Falk mdi Consultants, Inc. 55 Northern Boulevard, Suite 200 Great Neck, NY 11021
Re: K043348
Trade/Device Name: PM-8000 Patient Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Physiological Patient Monitor Regulatory Class: Class II Product Code: MWI Dated: November 15, 2004 Received: December 6, 2004
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Susan D. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bermuda for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Exhibit B
Page 1
510(k) Number (if known): _ KO4 3348
Device Name: PM-8000 Patient Monitor
#### Indications For Use:
The PM-8000 Patient Monitor is a vital signs monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PM-8000 Patient Monitor has many features and functions, yet is easy to use through an integrated keypad, knob and an intuitive menu system.
The patient parameters that can be monitored by the PM-8000 Patient Monitor are: ECG (3-lead or 5-lead selectable), Heart Rate derived from selected source (SPO2, ECG), Respiration Rate (derived from ECG), Non-Invasive Blood Pressure (NIBP), Saturation of Pulse Oxygen (SpO2), Temperature (TEMP) and invasive pressure (IBP). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The PM-8000 Patient Monitor is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.
The PM-8000 Patient Monitor is not recommended for use in a patient's home or residence, during the patient transport, or when it has not been ordered by a physician.
Prescription Use
Over-The Counter Use
(Per 21 CFR 801 Subpart D)
OR (21 CFT 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Duna R. de Luna
(Division Simi-Off) Division of Cardiovascular Devices
510(k) Number 1-04-3348
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.