LIFE SUPPORT FOR TRAUMA AND TRNSPORT (LSTAT) G6 LITE

K082256 · Integrated Medical Systems, Inc. · MWI · Sep 26, 2008 · Cardiovascular

Device Facts

Record IDK082256
Device NameLIFE SUPPORT FOR TRAUMA AND TRNSPORT (LSTAT) G6 LITE
ApplicantIntegrated Medical Systems, Inc.
Product CodeMWI · Cardiovascular
Decision DateSep 26, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The LSTAT G6 Lite is a portable unit intended to supply ICU functionality for adult and pediatric patients. The LSTAT G6 Lite combines the following medical device capabilities into a single platform: Physiological monitoring (electrocardiogram, invasive pressure monitoring, non-invasive blood pressure monitoring, temperature, pulse rate, blood oxygen saturation, and heart rate), low rate and high rate infusion pumps, a fluid warmer, a ventilator with carbon dioxide monitoring capabilities, and the ability to deliver oxygen to a patient. The functions of the LSTAT G6 Lite are controlled from a central user interface. The LSTAT G6 Lite may be operated using either batteries or an external power source. The LSTAT G6 Lite is intended for use in hospitals, aircraft, ambulances, field hospitals, and extended care facilities.

Device Story

LSTAT G6 Lite is a portable, integrated ICU platform for adult and pediatric patients. It consolidates physiological monitoring (ECG, invasive/non-invasive BP, temperature, pulse, SpO2, heart rate), mechanical ventilation (with CO2 monitoring), infusion pumps (low/high rate), and fluid warming into a single unit with a central user interface. Operated by clinicians in hospitals, ambulances, aircraft, and field settings; powered by battery or external source. Centralized display allows providers to monitor patient status and manage multiple therapies simultaneously. Integrated design facilitates patient transport and emergency care, potentially improving outcomes by maintaining ICU-level support during transit or in austere environments.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on design and functional comparison to the predicate device.

Technological Characteristics

Integrated portable ICU platform combining physiological monitoring, ventilation, infusion, and fluid warming. Features a central user interface for control of all modalities. Power options include internal battery or external source. Connectivity and specific material standards are not detailed in the provided documentation.

Indications for Use

Indicated for adult and pediatric patients requiring intensive care unit (ICU) functionality, including physiological monitoring, mechanical ventilation, infusion therapy, and fluid warming, in hospital, transport (aircraft, ambulance), field hospital, or extended care settings.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo with a medical symbol on the left and the letters "IMS" in bold, sans-serif font on the right. The medical symbol appears to be a caduceus, featuring a staff with two snakes coiled around it. The letters "IMS" are large and prominent, suggesting they are the primary identifier of the brand or organization represented by the logo. The image quality is somewhat degraded, with some pixelation and artifacts visible. 2 6 2008 Integrated Medical Systems, Inc. 1984 Obispo Avenue Signal Hill, CA 90755 ## 510(k) SUMMARY The following summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. All data included in this document and are accurate and complete to the best of Integrated Medical System's knowledge. | Applicant: | Integrated Medical Systems, Inc.<br>1984 Obispo Avenue<br>Signal Hill, CA 90755<br>(562) 498-1776 | |-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Renate A. MacLaren, Ph.D.<br>Director, Regulatory Affairs | | Date: | August 5, 2008 | | Device Identification: | Common Name:<br>Intensive Care Unit (ICU) Platform<br>Physiological Monitor<br>Ventilator<br>Infusion Pumps | | Trade Name: | Life Support for Trauma and Transport (LSTAT) G6 Lite | | Classification Name(s): | Cardiac Monitor (21 CFR 870.2300, Product Code DRT)<br>Continuous Ventilator (21 CFR 858.5895, Product Code CBK)<br>Infusion Pump (21 CFR 880.5725, Product Code FRN) | The Integrated Medical Systems LSTAT G6 Lite is Substantial Equivalence: substantially equivalent to the Integrated Medical Systems LSTAT 9602 (K965117) since the basic features, design and intended uses are the same or similar. The differences in design, dimensions, and features between the LSTAT G6 Lite and the predicate devices raise no new issues of safety and effectiveness, as these differences have no effect on the performance, function, or intended use of the device. The Integrated Medical Systems LSTAT G6 Lite is a portable intensive Device Description: care unit (ICU) for adult and pediatric patients. The LSTAT G6 Lite combines the medical device capabilities of a physiological monitor, a ventilator, and infusion pumps into a single platform with a central display. ﺮ ﺍﻟﻤﺮﺍﺟﻊ {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for IMS, which is a medical services company. To the left of the letters "IMS" is a medical symbol. The medical symbol is a staff with a snake wrapped around it. Intended Use: The LSTAT G6 Lite is a portable unit intended to supply ICU functionality for adult and pediatric patients. The LSTAT G6 Lite combines the following medical device capabilities into a single platform: Physiological monitoring (electrocardiogram, invasive pressure monitoring, non-invasive blood pressure monitoring, temperature, pulse rate, blood oxygen saturation, and heart rate), low rate and high rate infusion pumps, a fluid warmer, a ventilator with carbon dioxide monitoring capabilities, and the ability to deliver oxygen to a patient. The functions of the LSTAT G6 Lite are controlled from a central user interface. The LSTAT G6 Lite may be operated using either batteries or an external power source. The LSTAT G6 Lite is intended for use in hospitals, aircraft, ambulances, field hospitals, and extended care facilities. Renate A. MacLaren, Ph.D. Director, Regulatory Affairs Signed: 082256 72/2 {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" around the perimeter. Inside the circle is an abstract image of an eagle. SEP 2 6 2008 ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Integrated Medical Systems, Inc. c/o Renate A. MacLaren, Ph.D. Director, Regulatory Affairs 1984 Obispo Avenue Signal Hill, Ca 90755 ## Re: K082256 Trade/Device Name: Life Support for Trauma and Transport (LSTAT) G6 Lite Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: MWI Dated: August 5, 2008 Reccived: August 8, 2008 Dear Dr. MacLaren: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations afficcting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with {3}------------------------------------------------ Page 2 - Renate A. MacLaren, Ph.D. all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html. Sincerely yours, Bram D/Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows a logo with the word "WIMS" in a stylized font. The letters are bold and slightly distorted, giving them a textured appearance. The background is a grid-like pattern, which adds to the overall visual complexity of the logo. The logo is in black and white. Integrated Medical Systems, Inc. 1984 Obispo Avenue Signal Hill, CA 90755 ## Indications for Use 510(k) Number (if known): ## Life Support for Trauma and Transport (LSTAT) G6 Lite Device Name: Indications for Use: The Life Support for Trauma and Transport (I.STAT) G6 Lite System is a portable unit intended to supply intensive care unit (ICU) functionality for adult and pediatric patients. The LSTAT G6 Lite combines the following medical device capabilities into a single platform: Physiological monitoring (electrocardiogram, invasive pressure monitoring, non-invasive blood pressure monitoring, temperature, pulse rate, blood oxygen saturation, and heart rate), low rate and high rate infusion pumps, a fluid warmer, a ventilator with carbon dioxide monitoring capabilities, and the ability to deliver oxygen to a patient. The functions of the LSTAT G6 Lite are controlled from a central user interface. The G6 Lite may be operated using either battery power or an external power source. The G6 Lite is intended to be used in hospitals, aircraft, ambulances, field hospitals, and extended care facilitics. | Prescription Use | X | |-----------------------------|---| | (Part 21 CFR 801 Subpart D) | | AND/OR Over-the-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |------------------------------------|--| | Division of Cardiovascular Devices | | | 510(k) Number | k082256 | |---------------|---------| |---------------|---------|
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...