TELECARE D.R. SERVER SOFTWARE SYSTEM

K033133 · Healthcare Vision, Inc. · MWI · Apr 13, 2004 · Cardiovascular

Device Facts

Record IDK033133
Device NameTELECARE D.R. SERVER SOFTWARE SYSTEM
ApplicantHealthcare Vision, Inc.
Product CodeMWI · Cardiovascular
Decision DateApr 13, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The Telecare®D.R. is intended to be used in conjunction with 510(k) certified devices that upon prescription by an authorized healthcare provider for patients provides a means of receiving and storing audio, video, and patient biometric data including vital signs information received over standard phone lines or high speed connection between patients, typically at home, and health professionals at the health care provider's site.

Device Story

Telecare D.R. is a software-based telemedicine communications module acting as an accessory to FDA-cleared vital signs measurement devices. It facilitates real-time audio/visual communication and data transmission between patients (typically at home) and healthcare professionals (at a facility). Input data includes vital signs (BP, pulse, heart rate, O2 saturation, glucose, weight, ECG, temperature, breath peak flow) and auscultation sounds (heart, lung, bowel) captured by external devices connected to a standalone videophone. The system uses HCV NurseStation software for communication and data collection, and the Telecare D.R. server for storage and browser-based review. Healthcare providers use the output to monitor patient status; the system serves as a diagnostic aid requiring clinical judgment. It is not intended for emergency intervention or as a substitute for direct medical care.

Clinical Evidence

No clinical data submitted. Substantial equivalence demonstrated via functional performance testing of the software and verification that external vital signs measurement devices operate equivalently when used with the Telecare D.R. system compared to standalone operation.

Technological Characteristics

Software-based telemedicine system; operates as an accessory to 510(k) cleared vital signs monitors. Connectivity via standard network (H.323), standard phone lines (H.324), and wireless. Utilizes standard database architecture for data management. Compatible with various external measurement devices (e.g., glucometers, NIBP, pulse oximeters, spirometers).

Indications for Use

Indicated for patients requiring remote monitoring of vital signs, biometric data, and audio/visual communication with healthcare providers. Contraindicated for patients requiring direct medical supervision or emergency intervention.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ HealthCare Vision, Inc. Telemedicine Solutions ## Section E: 510K Summary #### Submitter HealthCare Vision, Inc. 2601 Scott Avenue, Suite 600 Fort Worth, Texas 76103 Telephone: (817) 531-8992 Facsimile: (817) 531-8999 ## Contact Person Craig A. Walker HealthCare Vision, Inc. Tel: 817.531.8992 ext 320 Tel: 512.832.6710 Fax: 817.531.8999 Fax: 512.832.5521 ## Date Prepared September 29, 2003 ## Software Information Trade Name: Telecare D.R. Common Name: Telemedicine Communications Module Classification Name: Powered Communication System ## Software Description The Telecare DR System consists of multiple software and hardware components. Utilizing this system of components, healthcare professionals are empowered with the ability to conduct real-time, secure audio/visual communication and measurement collection encounters from the patient's home to the healthcare facility and presented for secure Internet browser review. Typically, the Telecare D.R. resides on a health facility's server and connects to the HCV NurseStation. Using the HCV NurseStation software component, the healthcare professional instantiates audio/visual communication with the patient using a standalone videophone system (i.e., CS-126S, CS-126, or CS-146 videophone devices) that is equipped with the capability to transmit and receive data from a variety of external vital {1}------------------------------------------------ signs and medical measurement devices. The CS 126S also includes an integrated digital stethoscope that provides for both local and remote auscultation functions and an optional accessory plug-in pulse oximetry sensor unit. Upon connection with a standalone videophone, the HCV NurseStation, located at a healthcare professional's office or facility, identifies all the external vital signs and medical measurement devices (i.e., weight scale, glucometer, vital signs monitor, NIBP, pulse oximetry and spirometer) connected to the standalone videophone. Using the HCV NurseStation, the healthcare professional may proceed to communicate using standard phone line or high speed connection through internal modems and transmit real-time video, audio, and data between them. The real-time video and audio communications allow the patient and caregiver to view and speak with each other. Using the HCV NurseStation component, the healthcare professional is capable of monitoring the data from the measurement devices in real-time during collection, as well as, storing these measurements to the centralized data repository using the Telecare D.R. server software. To terminate the encounter, the healthcare professional disconnects the video call via the HCV NurseStation software component. This event triggers the HCV Conduit software component, which completes remaining data inserts/updates, broadcasts notifications and prepares data for secure browser viewing via the secure HCV Rendezviewer software component. With existing legally marketed vital signs measurement devices attached to the patient's system, the health professional and caregiver may monitor the patient's blood pressure. pulse rate, temperature, weight, blood oxygen saturation, blood glucose level, breath peak flow, images, and transmit this data to the caregiver's system. Vital signs measurement devices used with the HCV NurseStation and Telecare D.R. are FDA approved devices and are used for the same purposes which they received 510(k) approval or are devices that are exempt under applicable 21 CFR sections. The data may be captured, displayed, and/or saved on Telecare D.R. ## Substantial Equivalence The Telecare D.R. system and the predicate devices listed above have the same intended use and very similar principles of operations and technological characteristics. The Telecare D.R. and it predicate systems have the same general use to provide the capability for health professionals to monitor remote patient vital signs and/or breath and heart sounds and store the information on a data server/web server over the Internet and over standard phone line. The Telecare D.R. has no internal medical devices and it is a software system that performs as an accessory to 510K medical devices. Communication occurs through a plural of modalities include using standard network connection (H.323), standard phone line (H.324), and wireless communications to connect to devices for recording vital sign data (i.e., blood pressure, pulse, heart rate, oxygen saturation, glucose, weight, ECG, temperature breath peak flow and auscultations of the chest and abdominal cavities). These devices function independently in accordance with their own individual specifications and operation. None of the systems are intended solely for {2}------------------------------------------------ diagnostic purposes. In general, basic operation for the Telecare D.R. consists of 1) receiving patient vital sign data, images and sounds of the heart, lung, or bowel cavities 2) presenting the data in a graphical and visual format, and 3) providing Internet access to the data by assigned health professionals. In addition, the following devices can be used with the Telecare D.R. system to provide the external vital signs monitoring functions: Criticare Vital Signs Monitor (K884216) by Criticare Systems Inc., NIBP Monitor UA-767PC (K982481) by A&D Engineering, Inc., Digital Weight Scale UC-321 (exempt under 21 CFR 800.2700) by A&D Engineering. Inc., Ferraris KOKO Peak Pro (K013489) by PDS Healthcare Products Inc., Freestyle Blood Glucose Monitor (K000582) by Therasense. With regards to the transmission of patient vital signs information from external measurement devices, it is equivalent to the Carecompanion Nurse Station/Carecompanion Patient Station (K020584). With respect to the pulse oximetry accessory, it is equivalent to the Zymed Easi-View Telemetry System (K001308). The main difference from the predicate devices and platforms is that Telecare D.R. is a software component to the predicate devices and systems. Using off-the-shelf hardware and patented technology, Telecare D.R. provides more interface options with approved 510(k) vital sign devices and systems. In addition, the Telecare D.R. uses standard database architecture for managing the vital sign data received from the devices. Telecare D.R. and its predicate devices and systems have the same general use: to provide the capability for health care professionals to monitor blood pressure, pulse, heart rate, oxygen saturation, glucose, weight, ECG, temperature breath peak flow and auscultation sounds of their patients from remote locations. The main functional difference between the systems are the Telecare D.R. is the data server/Web server that permits the storage, retrieval, and presentation of patient-specific vital sign data using standard phone lines and high speed connections over the Internet. #### Intended Use The Telecare D.R. is intended to be used upon prescription of an authorized health care provider by patients to provide two-way audio/video communications, data transmission & communications, transmission of vital signs information over standard telephone lines and high speed telecommunications between the patient, typically at home, and a health professional at the health care professional's site. The information includes heart, lung, and bowel sounds, blood oxygen saturation, pulse rate, blood pressure, temperature, blood glucose, weight, and breath peak flow measurements. The information is collected upon the request and direction of the health care provider. The software system is a diagnostic aid that is an accessory to 510(k) devices. Clinical judgment and experience are required to check and interpret the information transmitted. The software system is not intended as a substitute for medical care. The software is {3}------------------------------------------------ contraindicated for patients requiring direct medical supervision or emergency intervention. ## Brief Description of Non-Clinical Testing Testing was performed to validate the functional performance of the Telecare D.R. Specifically, performance testing of the software was conducted to show that performance exceeds and thus meets the substantial equivalency of the predicate devices. Further testing was performed with each vital signs measurement device to show that they operate equivalently when used with the Telecare D.R. as when operated as indenendent devices. ## Brief Desciption of Clinical Testing Clinical study information was not submitted for the purpose of demonstrating substantial equivalence to the predicate devices. ## Conclusion The indications for use of the Telecare D.R. software as an accessory of 510(k) devices is consistent with that in labeling for telemedicine communications modules legally marked in the United States. The results of testing Telecare D.R. with approved 510(k) devices for monitoring or recording vital sign data (i.e., blood pressure, pulse, heart rate, oxygen saturation, glucose, weight, ECG, temperature breath peak flow and auscultations of the chest and abdominal cavities) indicate that the software is substantially equivalent to its predicates and does not raise any new questions of safety or effectiveness. {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/4/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized image of a bird. APR 1 3 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Craig Walker Vice President of Public Policy & Development HealthCare Vision, Inc. 2601 Scott Avenue Suite 600 Fort Worth, TX 76103 Re: K033133 Trade Name: Telecare D.R. Server Software System Regulation Number: 21 CFR 870.2300 Regulation Name: Patient Physiological Monitor (Without Arrhythmia Detection or Alarms) Regulatory Class: Class II (two) Product Code: MWI Dated: January 28, 2004 Received: January 29, 2004 Dear Mr. Walker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially cquivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. Craig Walker Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed prodicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Duana R. Luchner (A) Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ 510K Pending File: K033133 Healthcare Vision, Inc. Telecare D.R. Server Software System # Section D: Indications For Use - Revised 3-30-2004 510(k) Number (if known): K033133 Device Name: Telecare® D.R. Indications For Use: The Telecare®D.R. is intended to be used in conjunction with 510(k) certified devices that upon prescription by an authorized healthcare provider for patients provides a means of receiving and storing audio, video, and patient biometric data including vital signs information received over standard phone lines or high speed connection between patients, typically at home, and health professionals at the health care provider's site. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use __________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) Dana R. Vachner (Division Sign-Off) Jivision of Cardiovascular Devices 25 510(K) Number_K0 331
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...