Star 60

K151512 · Skanray Technologies Private Limited · MWI · Feb 14, 2016 · Cardiovascular

Device Facts

Record IDK151512
Device NameStar 60
ApplicantSkanray Technologies Private Limited
Product CodeMWI · Cardiovascular
Decision DateFeb 14, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesPediatric

Indications for Use

Star 60 multi-parameter Patient Monitoring system is intended to monitor a single Adult, Pediatric or Neonatal patient's vital signs at the bedside or during intra hospital transport along with the appropriate accessories mentioned / supplied with the unit. Vital signs parameters include ECG (3L / 5L), SpO2, Respiration, Temperature, NIBP, IBP and Capnography (CO2). It can also display the digital values of HR/PR, SpO2, RR, Non-Invasive Blood Pressure (Systolic, Diastolic and Mean), Invasive Blood Pressure (Systolic and Mean), Temperature, EtCO2, FiCO2 readings. The user, responsible to interpret the monitored data made available, will be a professional health care provider. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The monitor is not intended for home use.

Device Story

Multi-parameter patient monitor; inputs include ECG (3L/5L), SpO2, respiration, temperature, NIBP, IBP, and capnography (CO2) via patient-connected sensors. Device processes signals to display real-time numerical values and waveforms for HR/PR, SpO2, RR, blood pressure, temperature, and CO2 levels. Operated by professional healthcare providers in clinical settings (bedside/transport). Features capacitive sensing buttons; adjustable alarm limits; visible/audible alarm signals. Output allows clinicians to interpret patient status, adjust care, and monitor physiological trends. Benefits include continuous patient surveillance and timely detection of vital sign deviations.

Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through electrical, EMC, and safety testing per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 80601-2-30, IEC 60601-2-34, IEC 60601-2-49, ISO 80601-2-55, ISO 80601-2-56, ISO 80601-2-61, and ANSI/AAMI standards.

Technological Characteristics

Multi-parameter monitor with capacitive sensing buttons. Modules: ECG/Respiration/Temp, SpO2, NIBP, CO2, IBP. Connectivity: external equipment communication (analog/system outputs). Standards: IEC 60601-1, 60601-1-2, 60601-1-8, 60601-2-27, 80601-2-30, 60601-2-34, 60601-2-49, ISO 80601-2-55, 2-56, 2-61, ANSI/AAMI EC 13, SP10. Software lifecycle per IEC 62304.

Indications for Use

Indicated for monitoring vital signs in adult, pediatric, or neonatal patients at the bedside or during intrahospital transport. Not for home use.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 14, 2016 Skanray Technologies Private Limited % Yolanda Smith Consultant Smith Associates 1468 Harwell Ave Crofton, Maryland 21114 Re: K151512 Trade/Device Name: Star 60 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: January 7, 2016 Received: January 7, 2016 Dear Yolanda Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Mitchell Stein for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K151512 Device Name STAR 60 Indications for Use (Describe) Star 60 multi-parameter Patient Monitoring system is intended to monitor a single Adult, Pediatric or Neonatal patient's vital signs at the bedside or during intrahospital transport along with the appropriate accessories mentioned / supplied with the unit. Vital signs parameters include ECG (3L / 5L), SpO2, Respiration, Temperature, NIBP, IBP, IBP and Capnography (CO2). It can also display the digital values of HR/PR, SpO2, RR, Non-Invasive Blood Pressure (Systolic, Diastolic and Mean), Invasive Blood Pressure (Systolic and Mean), Temperature, EtCO2, FiCO2 readings. The user, responsible to interpret the monitored data made available, will be a professional health care provider. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The monitor is not intended for home use. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary (As Per 21 CFR 807.92) | SPONSOR | | |----------------------|----------------------------------------------------------------------------| | Company Name: | Skanray Technologies PVt Ltd<br>Healthcare Division | | Company Address: | #360, KIADB Industrial Area,<br>Hebbal, Mysore – 570018, Karnataka, India. | | Telephone: | 91-821-2407202 | | Fax: | 92-821-2407001 | | Contact Person: | J. Mahadevan | | Summary Prepared | May 19, 2015 | | Trade Name: | Star 60 | | Common/Usual Name: | Patient Monitoring System | | Classification Name: | Physiological Patient Monitor (without arrhythmia detection or<br>alarms) | | 510(k) Number | K151512 | | Product Code: | MWI | | Device Class: | Class II | | Regulation Number: | 21 CFR 870.2300 | | Predicate Device | | | Company | Product | 510(k) # | |-------------------------|--------------------------|----------| | Welch Allyn | Spot Vital Signs LXi | K101680 | | Philips Medical Systems | SureSigns VS4 | K120132 | | Suntech Medicals | Cycle BP Monitor & Pulse | | | | Oximeter, Model 1060 | K060820 | | L&T Medical & Systems | Stellar 300 | K103763 | {4}------------------------------------------------ ## Device Description STAR 60 is a patient monitor with a wide range of communication options. Capacitive sensing buttons in keyboard is introduced. It fits in clearly to a new age of patient monitoring. - ECG+RESPIRATION+TEMPERATURE MODULE: This module is divided in area and functionality between the main board and the ECG child card. The Temperature section and the DC-DC section of the ER2T lie in the main board. - SpO2: This module is inbuilt inside the monitor on the main board. It is used to measure the partial pressure of oxygen in the human body. - . NIBP: This module is inbuilt inside the monitor on to the chassis. It is used to non-invasively measure the systolic, diastolic & mean blood pressure. - CO2: This module is inbuilt inside the monitor on to the chassis. It is used to measure the level of CO2 in the blood (EtCO2 & FiCO2). - . IBP: STAR 60 can support IBP modules. For ease of understands we shall name as IBP1 & IBP 2. IBP 1/2 module is inbuilt on the main board. It is used to invasively measure the systolic and diastolic blood pressure. ## Indications for Use Star 60 multi-parameter Patient Monitoring system is intended to monitor a single Adult, Pediatric or Neonatal patient's vital signs at the bedside or during intra hospital transport along with the appropriate accessories mentioned / supplied with the unit. Vital signs parameters include ECG (3L/5L), SpO2, Respiration, Temperature, NIBP, IBP and Capnography (CO2). It can also display the numerical values of HR/PR, SpO2, RR, Non-Invasive Blood Pressure (Systolic, Diastolic and Mean), Invasive Blood Pressure (Systolic and Mean), Temperature, EtCO2, FiCO2 readings. The user, responsible to interpret the monitored data made available, will be a professional health care provider. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The monitor is not intended for home use. # Summary of Technological Characteristics Star 60 is designed and developed with reference to our previous 510K cleared product Star55 with same product code as Star60 - MWI. These 2 products are same in terms of: - ECG, - · Respiration, - · Temperature, - IBP, - · NIBP, - SpO2 and - CO2 - Display type - Alarm systems - Thermal recorder / printer - Equipment classification, applicable standards {5}------------------------------------------------ Star60 is compared with MP70 for advanced features that are included in the Star 60 and include: - Communication with external equipment including analog and system outputs - Display size and resolution. Differences between Star 60 and Star 55 The differences between the two devices includes advanced features that include: - Number of waveform traces displayed - Waveform display options / selections The inclusion of the advanced features have been validated and verified to ensure that the new features do not impact performance and the safety and efficacy of the device. #### Non Clinical Testing Electrical and EMC Testing for the Star 60 included the following | IEC 60601-1-2 | EMC Test Report | |----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | IEC 60601-1 | Medical electrical equipment - Part 1: General requirements for basic safety and<br>essential performance | | IEC 60601-1-6 | Medical electrical equipment - Part 1-6: General requirements for basic safety<br>and essential performance -Collateral standard: Usability | | IEC 60601-1-8 | Medical electrical equipment -- Part 1-8: General requirements for basic safety<br>and essential performance --Collateral standard: General requirements, tests and<br>guidance for alarm systems in medical electrical equipment and medical electrical<br>systems | | IEC 60601-2-27 | Medical electrical equipment - Part 2-27: Particular requirements for the basic<br>safety and essential performance of electrocardiographic monitoring equipment | | IEC 80601-2-30 | Medical electrical equipment -- Part 2-30: Particular requirements for the basic<br>safety and essential performance of automatic cycling non-invasive blood<br>pressure monitoring equipment. | | IEC 60601-2-34 | Medical electrical equipment -- Part 2-34: Particular requirements for the basic<br>safety and essential performance of invasive blood pressure monitoring<br>equipment. | | IEC 60601-2-49 | Medical electrical equipment - Part 2-49: Particular requirements for the basic<br>safety and essential performance of multifunction patient monitoring equipment | | ISO 80601-2-55 | Medical electrical equipment -- Part 2-55: Particular requirements for the basic<br>safety and essential performance of respiratory gas monitors | | ISO 80601-2-56 | Medical electrical equipment -- Part 2-56: Particular requirements for basic<br>safety and essential performance of clinical thermometers for body temperature<br>measurement | | ISO 80601-2-61 | Medical electrical equipment -- Part 2-61: Particular requirements for basic<br>safety and essential performance of pulse Oximeter equipment | {6}------------------------------------------------ # Additional Performance Testing: | ANSI/AAMI EC 13 | Cardiac Monitors, Heart rate Meter and Alarms | |-----------------|--------------------------------------------------------------------------------| | ANSI/SSMI SP10 | Manual, Electronic or Automated Sphygmomanometers | | IEC 62304 | Medical device software – software life cycle processes (Software/Informatics) | | ISO 14971 | Application of risk management to medical devices | #### Substantial Equivalence The Star 60 is substantially equivalent to the predicate device in Indications for Use, Materials and Design. The conclusions drawn from the non-clinical tests demonstrate that the Star 60 is as safe, as effective, and performs as well as or better than the predicate device. {7}------------------------------------------------ # Indications for Use 510(k) Number (if known) K151512 Device Name STAR 60 Indications for Use (Describe) Star 60 multi-parameter Patient Monitoring system is intended to monitor a single Adult, Pediatric or Neonatal patient's vital signs at the bedside or during intra hospital transport along with the appropriate accessories mentioned / supplied with the unit. Vital signs parameters include ECG (3L / 5L), SpO2, Respiration, Temperature, NIBP, IBP and Capnography (CO2). It can also display the digital values of HR/PR, SpO2, RR, Non-Invasive Blood Pressure (Systolic, Diastolic and Mean), Invasive Blood Pressure (Systolic, Diastolic and Mean), Temperature, EtCO2, FiCO2 readings. The user, responsible to interpret the monitored data made available, will be a professional health care provider. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The monitor is not intended for home use. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...