K982521 · Leonhard Lang Co. · DRX · Oct 26, 1999 · Cardiovascular
Device Facts
Record ID
K982521
Device Name
SKINTACT ECG ELECTRODES, S&W ECG ELECTRODES
Applicant
Leonhard Lang Co.
Product Code
DRX · Cardiovascular
Decision Date
Oct 26, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2360
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Skintact ECG electrodes are intended for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Skintact ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recordings. Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin.
Device Story
Skintact ECG electrodes are disposable, non-sterile, self-adhesive sensors for ECG monitoring and diagnostic recording. Device consists of silver/silver chloride sensing element, stainless steel stud, and conductive gel on a carrier tape with pressure-sensitive medical-grade adhesive. Used in clinical settings by healthcare professionals to acquire electrical signals from patient skin for electrocardiographic procedures. Output is raw electrical signal transmitted to ECG monitoring equipment for physician interpretation. Benefits include reliable signal acquisition for patient surveillance and diagnosis.
Clinical Evidence
Bench testing only. Electrical performance verified against ANSI/AAMI EC12/1991 standards. Biocompatibility assessed per ISO 10993-1. Real-time shelf life studies conducted.
Indicated for patients requiring ECG monitoring or diagnostic recordings. For use on intact, uninjured skin only. Must be ordered by a physician.
Regulatory Classification
Identification
An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.
Predicate Devices
Skintact AG Electrodes
Related Devices
K030509 — SKINTACT ECG TAB ELECTRODE · Leonhard Lang GmbH · Mar 4, 2003
K024247 — SKINTACT ECG ELECTRODES WITH KL 02 LIQUID GEL · Leonhard Lang GmbH · Jan 16, 2003
K092291 — SKINTACT AND VARIOUS OTHER TRADENAMES · Leonhard Lang GmbH · Aug 21, 2009
K971444 — 3M RED DOT 2234 & 2266 RADIOLUCENT MONITORING ELECTRODES · 3M Medical Products Group · Jun 11, 1997
K023503 — SKINTACT ECG ELECTRODE · Leonhard Lang GmbH · Nov 14, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 26 1999
Summary ; 15. 510 (k)
长982521
Date:
Manufacturer :
Telephone: Fax :
Contact Person:
Device Trade Names:
Common Name:
Classification Name:
Regulatory Reference:
Predicate Device:
Description :
Intended Use:
October 29, 1997
Leonhard Lang GmbH Archenweg 56 A-6020 Innsbruck Austria
+43 512 33425-7 +43 512 392210
Burrhus Lang, President
Skintact ECG Electrodes S&W ECG electrodes
Disposable ECG monitoring electrode
Electrocardiograph electrode
74 DRX
Skintact AG Electrodes
Skintact ECG electrodes are self-adhesive, non-sterile, single use disposable ECG electrodes. All include a silver/silver chloride sensing element, a stainless steel stud and a conductive gel. These conductive elements are held in place on the patient's skin by a carrier tape coated with a pressure sensitive medical qrade adhesive.
Skintact ECG electrodes are intended for use in general electrocardiographic procedures where ECG monitoring is
{1}------------------------------------------------
## 1 1982521
Physical/Technical Comparison:
Performance Summary:
## Biocompatibility Testing:
Shelf Life:
deemed necessary and is ordered by a physician.
Skintact ECG electrodes are equivalent to the predicate device. Physical and technical characteristics, including, design, materials used, safety and efficacy characteristics and intended use of Skintact ECG electrodes and the predicate device are either identical or comparable.
The electrical performance of Skintact ECG electrodes meets the requirements of the voluntary standard ANSI/AAMI EC12/1991 "Disposable ECG Electrodes". In addition Skintact ECG electrodes meet the requirements of ANSI/AAMI EC12/1991 for labeling, shelf life, packaging and safety.
The biological safety of Skintact ECG electrodes has been assured through the selection of materials which demonstrate appropriate levels of biocompatibility. The tests were selected on the basis of ISO 10993-1 "Biological Evaluation of Medical Devices - Partl: Guidance on selection of tests".
Data obtained in real time shelf life studies was reviewed and found to substantiate the claimed shelf life.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three curved lines that form the shape of a person's head and torso.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 26 1999
Leonhard Lang GmbH c/o Carole Stamp Third Party Official TÜV Product Service 1775 Old Highway 8 New Brighton, MN 55112-1891
K982521 Re: Skintact™ ECG Electrodes, S&W ECG Electrodes Regulatory Class: II (two) Product Code: DRX October 8, 1999 Dated: October 12, 1999 Received:
Dear Ms. Stamp:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Ms. Carole Stamp
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
VCAloteli
Witten, Ph.D., M.D. lia M. Actinq Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
10. Statement of Indications for Use
510(k) Number (if known):
Device Name: Skintact ECG electrodes
## Indications For Use:
Skintact ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recordings.
Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin.
| (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE QN ANOTHER PAGE IF NEEDED) | | | | | and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the commen | | | | |
|----------------------------------------------------------------------------|--|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
|----------------------------------------------------------------------------|--|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| Concurrence of CRRH, Office of Device Evaluation | |
|-------------------------------------------------------------------|-------------------------|
| (Division Sign-Off) | |
| Division of Cardiovascular, Respiratory, and Neurological Devices | |
| 510(k) Number | K183521 |
| Prescription Use (Per 21 CFR 801.109) | OR Over-The-Counter Use |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.