SKINTACT ECG TAB ELECTRODE

K030509 · Leonhard Lang GmbH · DRX · Mar 4, 2003 · Cardiovascular

Device Facts

Record IDK030509
Device NameSKINTACT ECG TAB ELECTRODE
ApplicantLeonhard Lang GmbH
Product CodeDRX · Cardiovascular
Decision DateMar 4, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2360
Device ClassClass 2

Indications for Use

Skintact ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin.

Device Story

Skintact ECG Tab Electrodes are disposable, self-adhesive, non-sterile electrodes for diagnostic resting ECG. Device inputs are electrical signals from patient skin; conductive gel (KH 06) facilitates signal acquisition. Electrodes consist of PET tape, carbon ink, and Ag/AgCl ink. Used in clinical settings for ECG surveillance and diagnostic recording; operated by healthcare professionals. Output is a reliable, high-quality ECG signal tracing for physician interpretation to support clinical decision-making. Benefits include consistent signal quality for accurate cardiac monitoring.

Clinical Evidence

Bench testing only. Conformance to ANSI/AAMI EC 12:2000 demonstrated via AC impedance, DC offset voltage, defibrillation overload recovery, combined offset instability, internal noise, and bias current tolerance testing. Shelf-life validated via real-time and accelerated aging (3 months at 40°C). Three ECG tracings (>10 seconds each) provided to demonstrate consistent high-quality signal output.

Technological Characteristics

Materials: PET tape, carbon ink, Ag/AgCl ink, KH 06 conductive gel. Sensing principle: Ag/AgCl electrode for ECG signal acquisition. Form factor: Tab electrode, 10 electrodes per siliconized PET card. Non-sterile, single-use. Testing standards: ANSI/AAMI EC 12:2000.

Indications for Use

Indicated for patients requiring ECG monitoring or diagnostic recording in general electrocardiographic procedures. For use on intact, uninjured skin only.

Regulatory Classification

Identification

An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAR 0 4 2003 ## 510(k) Summary Submitted on behalf of: Leonhard Lang GmbH Company Name: Address: Archenweg 56 Innsbruck 6010 Austria Telephone: ++ 43 / 512 / 33 4 25 7 ++ 43 / 512 / 39 22 10 Fax: Paladin Medical, Inc. by: PO Box 560 Stillwater, MN 55082 Telephone: 715-549-6035; fax: 715-549-5380 Contact person: Elaine Duncan Date prepared: February 13, 2003 Trade Name: Skintact® ECG Tab Electrode Common Name: Disposable ECG Electrodes Classification Name: Electrocardiograph (ECG) electrode SUBSTANTIALLY EQUIVALENT TO: Skintact® ECG Tab Electrodes with KH 06 gel are substantially equivalent to the Skintact® ECG Tab Electrodes with LecTec LT 4900 gel (the manufacterer's predicate - K023920). The additional carbon ink layer added for performance, is substantially equivalent to the MSB electrode (now distributed as Nikotab) originally cleared under K944260. DESCRIPTION of the DEVICE: Skintact® ECG Tab Electrodes (and also to be sold under various private label tradenames) with KH 06 gel. Skintact® ECG Tab Electrodes with KH 06 gel are self-adhesive, non-sterile, single use disposable electrodes for diagnostic resting ECG. ECG Tab electrodes are composed of a PET tape, carbon ink and Ag/AgCl ink and a conductive gel. These are configured as 10 (ten) electrodes applied to a siliconized transparent PET card, ten cards per pouch. INDICATIONS FOR USE: Skintact ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin. (No change to original indication for use.) SUMMARY of TESTING: Biocompatibility testing confirms the materials are biocompatible and the change does not introduce new risks. Skintact ECG Tab Electrodes with KH 06 gel were tested in accordance with ANSI/AAMI EC 12:2000. A certification to conformance to this standard has been provided. The testing conducted was: AC impedance; DC offset voltage; Defibrillation overload recovery; Combined offset instability and internal noise; Bias current tolerance. Shelf life of the tab electrodes with KH 06 gel was tested in real-time aging and in accelerated aging. For accelerated aging the electrodes were in an incubator for a time of 3 months with an increased temperature of 40°C. In accelerated shelf life testing the electrodes are subjected to a controlled environment in which one or more extrinsic factors (e.g., temperature, humidity, gas atmosphere, light) is maintained at a higher than normal level. Leonhard Lang has experience for about 20 years of using the current packaging and this ensures all requirements for the 24 months shelf-life of the electrodes. No differences were required for packaging of the solid adhesive electrodes compared to the predicate electrode. For clinical data, three ECG tracings using KH 06 gel electrodes were made. Each tracing contains more than 10 seconds of data. The data demonstrate that ECG Tab electrodes with KH 06 gel provide a reliable signal tracing of consistently high quality. {1}------------------------------------------------ Image /page/1/Picture/2 description: The image shows the address for the Food and Drug Administration. The address is listed as 9200 Corporate Boulevard in Rockville, MD. The zip code is 20850. MAR 0 4 2003 Leonhard Lang, GmbH c/o Ms. Elaine Duncan, M.S.M.E., RAC President Paladin Medical, Inc. P.O. Box 560 Stillwater. MN 55082-0560 Re: K030509 Trade Name: Skintact® ECG Tab Electrodes Regulation Number: 21 CFR 870.2360 Regulation Name: Electrocardiograph Electrode Regulatory Class: Class II (two) Product Code: DRX Dated: February 13, 2003 Received: February 19, 2003 Dear Ms. Duncan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 – Ms. Elaine Duncan, M.S.M.E., RAC Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bram D Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known) KO30509 Device Name: Skintact® ECG Tab Electrodes Indications for Use: Skintact® ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin. > (Please Do Not Write Below This Line-Continue On Another Page If Needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use __ OR Over -The-Counter Use ________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) (Division Signvascular Devices **510(k) Number*****_****_*
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...