K024264 · Leonhard Lang Co. · DRX · Jan 16, 2003 · Cardiovascular
Device Facts
Record ID
K024264
Device Name
SKINTACT
Applicant
Leonhard Lang Co.
Product Code
DRX · Cardiovascular
Decision Date
Jan 16, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2360
Device Class
Class 2
Indications for Use
Skintact ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin.
Device Story
Skintact ECG electrodes are disposable, self-adhesive, single-use snap electrodes used for ECG monitoring and diagnostic recording. The device consists of an ABS sensor element coated with silver and silver chloride, utilizing KS 01 solid wet gel as the conductive medium. Electrodes are applied to intact skin by clinical staff or patients during monitoring procedures. The device functions as a passive interface, capturing electrical signals from the patient's skin and transmitting them to an external electrocardiograph for display and analysis. The electrodes facilitate continuous monitoring, including Holter monitoring, by maintaining electrical contact and adhesion for up to 48 hours. The device benefits patients by providing reliable signal acquisition for cardiac surveillance and diagnosis.
Clinical Evidence
Clinical evaluation included a comparison of ECG traces between the subject KS 01 solid wet gel electrodes and the predicate OR liquid gel electrodes, confirming equivalent performance. Three 48-hour Holter monitoring wear tests were conducted on volunteers. Results showed no deterioration in electrical performance, no gel dry-out, and no skin irritation. Adhesion remained effective during normal activity and sports.
Technological Characteristics
Disposable, non-sterile, self-adhesive snap electrodes. Construction: ABS sensor element coated with silver and silver chloride; stainless steel stud. Conductive medium: KS 01 solid wet gel. Electrical performance conforms to ANSI/AAMI EC 12:2000 (AC impedance, DC offset, defibrillation overload recovery, offset instability, noise, bias current tolerance). Shelf-life: 24 months.
Indications for Use
Indicated for patients requiring ECG monitoring or diagnostic recording. For use on intact, uninjured skin only. Prescription use only.
Regulatory Classification
Identification
An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.
Predicate Devices
Skintact® ECG Electrodes with OR liquid gel (K982521)
Related Devices
K023503 — SKINTACT ECG ELECTRODE · Leonhard Lang GmbH · Nov 14, 2002
K024247 — SKINTACT ECG ELECTRODES WITH KL 02 LIQUID GEL · Leonhard Lang GmbH · Jan 16, 2003
K982521 — SKINTACT ECG ELECTRODES, S&W ECG ELECTRODES · Leonhard Lang Co. · Oct 26, 1999
K092291 — SKINTACT AND VARIOUS OTHER TRADENAMES · Leonhard Lang GmbH · Aug 21, 2009
K030509 — SKINTACT ECG TAB ELECTRODE · Leonhard Lang GmbH · Mar 4, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K024/264
JAN 1 6 2003
# 510(k) Summary
# SUBMITTED ON BEHALF OF:
Trade Name:
Common Name:
| | Company Name:<br>Address: | Leonhard Lang GmbH<br>Archenweg 56<br>6010 Innsbruck<br>Austria | |
|-----|----------------------------------|-----------------------------------------------------------------|--|
| | Telephone: | ++ 43 / 512 / 33 4 25 7 | |
| | Fax: | ++ 43 / 512 / 39 22 10 | |
| by: | Elaine Duncan, MS.M.E., RAC | | |
| | President, Paladin Medical, Inc. | | |
| | PO Box 560 | | |
| | Stillwater, MN 55082 | | |
| | Telephone: | 715-549-6035 | |
| | Fax: | 715-549-5380 | |
| | CONTACT PERSON: | Elaine Duncan | |
| | DATE PREPARED: | December 18, 2002 | |
| | Trade Name: | Skintact® ECG Electrode | |
Classification Name: Electrocardiograph (ECG) electrode SUBSTANTIALLY EQUIVALENT TO: Skintact® ECG Electrodes with KS 01 solid wet gel
Disposable ECG Electrodes
are substantially equivalent to the Skintact® ECG Electrodes with OR liguid gel (the manufacterer's predicate device cleared under K982521)
DESCRIPTION of the DEVICE: Skintact® ECG Electrodes (and as also to be offered for sale under various private label tradenames) will be offered with KS 01 solid wet gel. Just like the OR liguid gel electrodes, KS 01 solid wet gel electrodes are self-adhesive, non-sterile, single use disposable snap electrodes. The KS 01 solid wet gel electrodes are identical in size, shape and configuration to the QR liquid gel Skintact ECG electrodes currently marketed by Leonhard Lang, GmbH.
All electrode configurations include a stainless steel stud to guarantee an unimpaired performance during the shelf-life of the product. All electrodes include an ABS sensor element coated with silver. The silver layer is either completely or partially (in the areas in contact with the conductive gel) covered with a silver chloride layer. This is the same construction as the current Skintact® ECG electrode using OR liguid gel conducting media.
## INDICATIONS FOR USE:
Skintact ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin. (NO CHANGE to ORIGINAL INDICATION for USE)
{1}------------------------------------------------
### 510(k) Summary-Continued
#### SUMMARY of TESTING:
Biocompatibility testing confirms the materials are biocompatible and the change does not introduce new risks. The following testing showed no adverse results: Cytotoxicity; Skin Irritation; Sensitization.
The ANSI/AAMI EC 12:2000 "Disposable ECG electrodes" was used to define the requirements for Skintact ECG Electrodes with KS 01 solid wet gel. All electrical tests are according to ANSVAAMI EC 12:2000. A certification to conformance EC12:2000 with this standard has been provided. The testing conducted was: AC impedance; DC offset voltage: Defibrillation overload recovery; Combined offset instability and internal noise; Bias current tolerance.
The shelf life of the electrodes with KS 01 solid wet gel was tested in real-time aging.
Leonhard Lang has experience for about 20 years of using the current packaging and this ensures all requirements for the 24 months shelf-life of the electrodes. No differences were required for packaging of the solid wet gel electrodes compared to the predicate electrode.
The results of these tests confirm that the shelf-life of Leonhard Lang Skintact ECG Electrodes with KS 01 solid wet gel is well inside the limits defined in ANSI/AAMI EC12-2000, both for the first tests and for the tests with real-time aged electrodes. Thus the conclusion that the electrical performance of the KS 01 solid wet gel electrodes will stay within the limits during their shelf-life of 24 months. The comparison with the predicate device and the data from the KS 01 solid wet gel electrodes shows similar results. The difference is negligible in the limits defined in ANSI/AAMI EC12-2000. Therefore electrical performance of the predicate device and KS 01 solid wet gel electrodes is equivalent.
Clinical data: The potential effect of material change to the conducting signal was evaluated by repeating the clinical trace testing per the FDA guidance on the KS 01 solid wet gel electrodes and determined that the KS 01 solid wet gel performs the same. Comparing the ECG traces between KS 01 solid wet gel electrodes and QR liquid gel electrodes demonstrate that KS 01 solid wet gel electrodes are equivalent to the QR liquid gel electrodes. Three wear test reports for Holter monitoring electrodes used for 48 hours were provided. A physician reviewed the performance of the wear tests and examined for skin irritation. The tests confirm that the performance of electrodes is quite good and that wearing of the electrodes does not cause any skin problems for the volunteers. During a time of 48 hours there was no deterioration of electrical performance and the gel did not The electrodes were not replaced during this time and did not show any dry out. displacement; although the volunteers worked and participated in sports as usual demonstrating the adhesion was also very good. (No change to adhesive formulation was made.)
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem that resembles an eagle or bird-like figure with three stylized lines forming its body and wings.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 6 2003
Leonhard Lang GmbH c/o Ms. Elaine Duncan Paladin Medical, Inc. P.O. Box 560 Stillwater, MN 55082-0560
Re: K024264
Skintact® ECG Tab Electrodes with KS 01 solid wet gel Regulation Number: 21 CFR 870.2360 Regulation Name: Electrocardiograph Electrode Regulatory Class: Class II (two) Product Code: DRX Dated: December 18, 2002 Received: December 23, 2002
Dear Ms. Duncan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely vours.
K. Garten
Rhona D. Zuckerman, M.D.
Bram D. Žuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known)
K024264/
Device Name:
Skintact® ECG Electrodes (with KS 01 gel)
Indications for Use:
Skintact ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording.
Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin.
> (Please Do Not Write Below This Line-Continue On Another Page If Needed) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X
OR
Over -The-Counter Use
(Optional Format 1-2-96)
K.Q. Carter
(Division Sign-Off)
Division of Cardiovascular Devices
**510(k) Number** K024264
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.