The 3M Red Dot™ Resting EKG Electrode can be used on a patient undergoing an EKG diagnostic procedure while resting. These electrodes are applied to the patient's skin for short term use. These electrodes are disposable and are for single use. They are not indicated for use in long term monitoring. These electrodes will include the precaution statement: U.S.A. Federal Law restricts this device to sale by or on the order of a physician.
Device Story
3M Red Dot™ Resting EKG Electrode (2360) is a disposable, single-use electrode for resting EKG diagnostic procedures. Device consists of a conductive adhesive laminated to a conductive Ag/AgCl backing with a tab for leadwire connection. Applied to patient skin by clinical staff; facilitates electrical signal acquisition from the heart for diagnostic EKG monitoring. Benefits include improved adhesion compared to predicate device. Used in clinical settings for short-term diagnostic sessions.
Clinical Evidence
Bench testing only. Biocompatibility evaluated per ISO 10993-1. Electrical performance verified against ANSI/AAMI EC12/1991 standards for disposable ECG electrodes.
Technological Characteristics
Conductive Ag/AgCl backing; conductive adhesive; tab connector. Disposable, single-use form factor. Complies with ANSI/AAMI EC12/1991 standards.
Indications for Use
Indicated for patients undergoing resting EKG diagnostic procedures. For short-term, single-use only. Not indicated for long-term monitoring.
Regulatory Classification
Identification
An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.
Predicate Devices
3M Red Dot™ Resting EKG Electrode, Catalog Number 2330 (K932454)
Related Devices
K982521 — SKINTACT ECG ELECTRODES, S&W ECG ELECTRODES · Leonhard Lang Co. · Oct 26, 1999
K030509 — SKINTACT ECG TAB ELECTRODE · Leonhard Lang GmbH · Mar 4, 2003
K990113 — DISPOSABLE ECG MONITORING ELECTRODES, MODELS 4500,4530, 4533, 4539, 4540, AND 4560 · Maersk Medical A/S · Dec 6, 1999
K101576 — CENTERRIDGE DIAGNOSTIC ECG ELECTRODE, CENTERRIDGE MONITORING ECG ELECTRODES · Team Innovations, Inc. · Jul 22, 2010
K024247 — SKINTACT ECG ELECTRODES WITH KL 02 LIQUID GEL · Leonhard Lang GmbH · Jan 16, 2003
Submission Summary (Full Text)
{0}
JUL 26 1998
K974445
510(k) Premarket Notification Submission
3M Red Dot™ Resting EKG Electrode, 2360
Page 44 of 45
# 510(K) Summary
Name and address of device Manufacturer submitting 510(k) Notification:
3M
Medical Markets Group
3M Health Care
3M Center
St. Paul, MN 55144-1000
Regulatory Correspondent of Device Manufacturer:
Linda Johnsen
Regulatory Affairs Specialist
3M health Care
Building 275-3E-08
612 737-4376
Date Summary was prepared: November 21, 1997
Name of Device: 3M Red Dot™ Resting EKG Electrode, Catalog Number 2360
Classification: Electrocardiograph electrodes, class II, 21 CFR 870.2360
## Description:
3M Red Dot™ Resting EKG Electrode, Catalog Number 2360 has a conductive adhesive which is laminated to a conductive Ag/AgCl backing. The conductive backing has a tab end for connection to the electrocardiograph leadwire.
## Indications for Use:
The 3M Red Dot™ Resting EKG Electrode can be used on a patient undergoing an EKG diagnostic procedure while resting. These electrodes are applied to the patient's skin for short term use. These electrodes are disposable and are for single use. They are not indicated for use in long term monitoring.
## Predicate Device:
The predicate device used for the purpose of substantial equivalence and used for comparison was 3M Red Dot™ Resting EKG Electrode, Catalog Number 2330 which was filed and cleared under 510(k) Number k932454. The significant difference between the 2330 and 2360 is that the 2360 introduces a new conductive adhesive which increases the adhesion of the electrode. The electrode backing and liner of the 2360 are also different but the changes are minor when compared to the 2330.
{1}
510(k) Premarket Notification Submission
3M Red Dot™ Resting EKG Electrode, 2360
Page 45 of 45
## Safety and Efficacy:
### Biocompatibility Testing:
The biological safety of the Red Dot™ 2360 Resting EKG Electrode has been assured through the selection of materials which demonstrate appropriate levels of biocompatibility. Tests were selected on the basis of Part-1 of ISO 10993-1, “Biological Evaluation of Medical Devices”.
### Performance Testing:
The electrical performance of the 3M Red Dot™ 2360 Resting EKG Electrode has been tested and meets the voluntary standard requirements under ANSI/AAMI EC12/1991, for Disposable ECG Electrodes.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUL 26 1998
Ms. Linda Johnsen
Regulatory Affairs Specialist
3M Health Care
3M Center, Building 275-3E-08
St. Paul, MN 55114-1000
Re: K974445
3M Red Dot™ Resting EKG Electrode Model 2360
Regulatory Class: II (two)
Product Code: DRX
Dated: April 22, 1998
Received: April 27, 1998
Dear Ms. Johnsen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}
510(k) Premarket Notification Submission
3M Red Dot™ Resting EKG Electrode, 2360
Page 40 of 45
510(K) Number (If Known): _________________________________
Device Name: 3M Red Dot™ 2360 EKG Resting Electrode
Indications For Use:
The 3M Red Dot™ Resting EKG Electrode can be used on a patient undergoing an EKG diagnostic procedure while resting. These electrodes are applied to the patient's skin for short term use. These electrodes are disposable and are for single use. They are not indicated for use in long term monitoring.
These electrodes will include the precaution statement: U.S.A. Federal Law restricts this device to sale by or on the order of a physician.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-the Counter Use _________
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K974445
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.