The VitalLink3 (VL3) Mobile Vital Signs System is intend for use as a portable vital signs monitor for patients who are remotely located from medical professionals. The system can be used to acquire and display vital signs from patients in remote locations and transmit the data, real time, to a medical professional at a Medical Call Center / Medical Service Provider in order to help determine the patient's transportation needs.
Device Story
VitalLink3 (VL3) is a portable, modular vital signs monitor for remote patient assessment. Inputs: 6/12 lead EKG, NiBP, SpO2, pulse, core temperature. Sensors communicate with VL3 unit via Bluetooth RF link. VL3 features integrated display for real-time data viewing, programmable soft keys for communication management, and onboard memory. Data transmitted in real-time via dialup, Ethernet, satellite phone, satellite data, cellular, or 802-11 b/g networks to a remote Medical Call Center. Clinical Interface software (installed on clinician PC) receives, decrypts, and displays data for clinician review. Output informs patient transport decisions. Used by clinicians at call centers to assess patients in remote environments (land, air, sea). Benefits include enabling remote clinical oversight and timely transport triage.
Clinical Evidence
Bench testing only. Device evaluated against recognized standards including IEC60601-1 (safety), IEC60601-1-2 (EMC), IEC60601-2-27 (EKG monitoring), DO-160 (airborne environmental conditions), IEC60529 (enclosure protection), and IEC60068-2 series (vibration/free fall). No clinical data presented.
Technological Characteristics
Modular system comprising VL3 monitor and sensor components. Connectivity via Bluetooth RF (sensors to monitor) and various WAN/LAN options (monitor to remote center). Power: battery-operated. Form factor: portable, fabric-cased. Software: proprietary Clinical Interface application for PC. Standards: IEC60601-1, IEC60601-1-2, IEC60601-2-27, DO-160, IEC60529, IEC60068-2-6, IEC60068-2-32.
Indications for Use
Indicated for patients in remote locations (land, aircraft, sea) requiring vital signs monitoring (6/12 lead EKG, NiBP, SpO2, pulse, core temperature) for transport decision-making by remote medical professionals.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the number K073094 in a handwritten font. Below the number is the text 'p. 1/4', also in a handwritten font. The number and text are in black ink and the background is white. The image appears to be a scan or photograph of a document.
Image /page/0/Picture/1 description: The image shows the logo for TeleMedic Systems. The logo consists of a square with a plus sign on top and a line graph inside. The text "TeleMedic Systems" is to the right of the square.
JAN - 4 2008
Revision A 112307
## 2.0 510 (k) Summary
Data Prepared on: 24 October 2007 Date Revised on: 23 November 2007 (Revision A)
- 1. Submitter: TeleMedic Systems Ltd TeleMedics House 10 Billetfield Taunton Somerset TA1 3NN
- 2. Contact Person
| Name: | Gerald L Buss |
|-----------|--------------------------------|
| Position: | Chief Operating Officer |
| Phone: | +44 (0) 8701 417330 |
| Fax: | +44 (0) 8701 417425 |
| Email: | gbuss@UK. TeleMedicSystems.com |
#### 3. Device Identification
Trade Name / Proprietary Name: VitalLink3 Mobile Vital Signs System Common Name: Portable vital signs monitor
| 4. Classification name and reference | |
|-----------------------------------------------------------------------------|---------------------------------------------|
| Monitor, Physiological, Patient (Without<br>Arrhythmia Detection or Alarms) | CFR 870.2300, Class II<br>Product Code: MWI |
| Transmitters and Receivers,<br>Physiological signal, radio/frequency | CFR 870.2910, Class II<br>Product Code: DRG |
- 5. Indications for use
The VitalLink3 Mobile Vital Signs Monitor is intended for use as a portable vital signs monitor for patients who are remotely located from medical professionals. The system can be used to acquire and display vital signs data from patients in remote locations and transmit that data, in real time, to
TeleMedic Systems TeleMedic House - 10 Briletfield - Taunton - Taun SNN - UK T +44 (6) 8701 417330 F +44 +0) 8701 +17425 - E info (i)telemedicaystems com - W www.telemedios.stems.co.n
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for a company called "TeleMedic Systems". The logo consists of a stylized plus sign enclosed in a square. The words "TeleMedic" are stacked on top of the word "Systems" and are located below the square and plus sign.
INTERIOR
PAINT
Revision A 112307 medical professional located at a call centre / different location in order to help determine the patients transport needs.
#### 6. Device Description
The VitalLink3 System is a modular hardware/software system for acquiring, monitoring and/or communicating patient vital signs data.
The VitalLink system acquires vital signs date from patients in remote locations (on land, in aircraft or on ships at sea) and communicates the data in real time to a medical professional at a call centre (Medical Service provider) in order to help determine the patient's need for transportation to a medical facility.
The VitalLink® System is comprised of the following discrete units:
- a. The VitalLink3 (VL3) monitors 4 vital sign parameters: 6/12 lead EKG, Non-invasive blood pressure (NiBP), blood oxygen saturation (SpO2), pulse rate and core temperature. The VitalLink user interface includes a built in display for viewing collected data real time and a series of programmable soft keys for choosing communication preferences and manage communications with a medical service provider. The VitalLink also has onboard memory for storing the data collected from the parameters.
The Vitall ink is delivered to users in a fabric case with integrated compartments for the VitalLink3, the sensor components, battery chargers, labelling, and consumable accessories.
- b. Parameters: The EKG and SpO2 parameters identified in "6.a" above communicate with the VitalLink3 via a Bluetooth RF link. The parameters are paired with a VitalLink3 and the RF link tested just prior to when the device is packaged for shipping.
- c. Clinical Interface: The Clinical Interface is a TeleMedic Systems developed proprietary software application that receives data from a VitalLink®, decrypts the data and displays the information for use by medical professional / clinician in making a patient transport decision. The Clinical interface application is installed on a computer at the clinician's location.
- d. Communication Options: There are six (6) possible means by which the VitalLink® can be configured to communicate with the Clinical Interface: Dialup telephone landline, direct wired connection to an IP network (Ethernet), satellite phone, satellite based data services, cellular phone and wireless 802-11 b/g networks.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for TeleMedic Systems. The logo consists of a square with rounded corners and a plus sign in the upper right corner. Below the square are the words "TeleMedic" on the first line and "Systems" on the second line. The text is in a simple, sans-serif font.
Revision A 112307
### 7. Statement of Substantial Equivalence
TeleMedic Systems believes that the VitalLinks is a modification of the VitalLink1200 (K010732) as it has the same indications for use and does not include a change in the fundamental science incorporated in the predicate device. Further, through a detailed comparison establishes substantially equivalent to the predicate device.
The INTENDED USE of the modified device as described in it labelling HAS NOT CHANGED.
- 8. Performance Standards
Section 514 performance standards have not been promulgated for this device. There are voluntary standards for this device some of which are FDA recognized. The FDA recognized standards are identified with an "4" in the following table.
| Standard Number | Standard<br>Organization | Standard Title | Rev | Date |
|-------------------------------------------------|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-----------------|
| ♦ IEC60601-1 | International<br>Electrotechnical<br>Commission | Medical Electrical Equipment.<br>Part 1: General Requirements<br>For Safety | 2 | 30-Dec-<br>1988 |
| ♦IEC60601-1-2 | International<br>Electrotechnical<br>Commission | Medical Electrical Equipment.<br>Part General Requirements For<br>Safety 2. Collateral Standard;<br>Electromagnetic Compatibility<br>Requirements and Tests | 2 | 29-Aug-<br>2005 |
| ♦ IEC60601-1<br>(Amendment1) | International<br>Electrotechnical<br>Commission | Medical Electrical Equipment.<br>Part 1: General Requirements<br>For Safety - Amendment 1 | - | 13-Nov-<br>1991 |
| ♦ IEC60601-1<br>(Amendment 2) | International<br>Electrotechnical<br>Commission | Medical Electrical Equipment.<br>Part 1: General Requirements<br>For Safety - Amendment 2 | - | 7-Mar-1995 |
| ♦ UL60601-1 (US<br>Deviations to<br>IEC60601-1) | Underwriters<br>Laboratories | Medical Electrical Equipment.<br>Part 1: General Requirements<br>For Safety | 1 | 25-Apr-<br>2003 |
| ♦ IEC60601-2-27 | International<br>Electrotechnical<br>Commission | Medical Electrical Equipment.<br>Part 2: Particular requirements<br>for the safety, including essential<br>performance, of<br>electrocardiographic monitoring<br>equipment | 2 | 29-Aug-<br>2005 |
| DO-160 | RTCA | Environmental Conditions and<br>Test Procedures for Airborne<br>Equipment | E | 9-Dec-2004 |
| IEC60529 | International<br>Electrotechnical<br>Commission | Degrees of Protection Provided<br>by Enclosures | 2.1 | 27-Feb-<br>2001 |
{4}------------------------------------------------
K073094 p 4/4
Image /page/3/Picture/1 description: The image shows the logo for TeleMedic Systems. The logo consists of a square with a plus sign in the upper right corner. Inside the square is a graphic that appears to be a medical chart. Below the square are the words "TeleMedic Systems".
| | Revision A 112307 | | | |
|-------------------------|-------------------------------------------------|-------------------------------------------------------------------------------|-----|-----------------|
| Standard Number | Standard<br>Organization | Standard Title | Rev | Date |
| IEC60068-2-6 | International<br>Electrotechnical<br>Commission | Environmental testing - Part 2:<br>Tests - Test Fc: Vibration<br>(sinusoidal) | 6 | 31-Mar-<br>1995 |
| IEC60068-2-32 | International<br>Electrotechnical<br>Commission | Environmental testing - Part 2:<br>Tests - Test Ed: Free fall | 2 | 1-Jan-1975 |
| IEC60068-2-32<br>(amd2) | International<br>Electrotechnical<br>Commission | Amendment 2 - Environmental<br>testing. Part 2: Tests. Test Ed:<br>Free fall | 2 | 31-Oct-<br>1990 |
{5}------------------------------------------------
Public Health Service
Image /page/4/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
JAN - 4 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
TeleMedic Systems Ltd c/o Mr. Gerald L. Buss Chief Operating Officer 10 Billetfield Taunton, Somerset UNITED KINGDOM TA1 3NN
Re: K073094
VitalLink Mobile Vital Signs System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (two) Product Code: MWI Dated: November 23, 2007 Received: December 10, 2007
Dear Mr. Buss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Gerald L. Buss
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Brimmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Revision A 112307
# 1.0 Indications for use Statement
510(K) Number: _______________________________________________________________________________________________________________________________________________________________
VitalLink3 Mobile Vital Signs System Device Name:
Indications for use:
The VitalLink3 (VL3) Mobile Vital Signs System is intend for use as a portable vital signs monitor for patients who are remotely located from medical professionals. The system can be used to acquire and display vital signs from patients in remote locations and transmit the data, real time, to a medical professional at a Medical Call Center / Medical Service Provider in order to help determine the patient's transportation needs.
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| Prescription Use | XXX |
|-------------------------|-----|
| OR Over-the Counter Use | |
| (Per 21 CRF 801.109) | |
*Division Sign-off*
| 510 (k) Number | K073094 |
|----------------|---------|
|----------------|---------|
Proprietary and Confidential - TeleMedic Systems Ltd. Unauthorized copying or distribution prohibited
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.