GONOTEC CALIBRATION SOLUTION 300 MOSMOL/KG FOR OSMOMETER

K032608 · Gonotec Gesellschaft Fuer Mess-Und Regeltechnik MB · JIS · Oct 21, 2003 · Clinical Chemistry

Device Facts

Record IDK032608
Device NameGONOTEC CALIBRATION SOLUTION 300 MOSMOL/KG FOR OSMOMETER
ApplicantGonotec Gesellschaft Fuer Mess-Und Regeltechnik MB
Product CodeJIS · Clinical Chemistry
Decision DateOct 21, 2003
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

The primary calibrator “Calibration Solution for Osmomat 030/Auto” is used to calibrate the Osmometers “Osmomat 030” and “Osmomat Auto” which both are freezing point osmometers for use in clinical chemistry. The Osmolality of the calibrator is equal to the Osmolality of body fluids (isotonic solution), providing that the osmometer will be calibrated correctly for the fluids to be measured for medical purposes.

Device Story

Calibration solution for freezing-point osmometers (Osmomat 010/030/auto); aqueous solution containing sodium chloride. Used in clinical chemistry laboratories to calibrate osmometers; ensures accurate measurement of osmolality in patient body fluids. Osmometers function by measuring freezing-point depression; dissolved substances in water depress freezing point; device calculates osmolality based on this physical effect. Calibrator provides known reference value for standard curve. Benefits patient by ensuring accurate diagnostic testing of body fluids (e.g., urine) to assess kidney function and physiological status.

Clinical Evidence

No clinical data. Bench testing only; quality of ingredients (sodium chloride and distilled water) controlled via U.S. Pharmacopeia and European Pharmacopeia standards. Traceability established via certificate of analysis confirming quantity by weight.

Technological Characteristics

Aqueous solution of sodium chloride; primary reference material. Designed for use with freezing-point osmometers. No active energy source or software; chemical calibrator only.

Indications for Use

Indicated for use as a primary calibrator for Osmomat 030 and Osmomat auto freezing-point osmometers in clinical chemistry settings to ensure accurate measurement of body fluid osmolality.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Gonotec, 510(k) submission for calibration solution for Osmomat 010 / 030 / au 21 CFR part 807.87(h), A 510(k) summary as descri 807.92 or a 510(k) statement as described in Sec. 807.93. SUMMARY: Sec. 807.92, Content and format of a 510(k) summary. 1.6.1 (1) The submitter's name, address, telephone number, a contact person, 1.6.1.1 and the date the summary was prepared; ESTABLISHMENT (GONOTEC Gesellschaft fuer Mess- und REGISTRATION-NUMBER: 3003591759 Regeltechnik mbH, Berlin, Germany) Klaus Noack OWNER-ID: 904 8267 Taunusstr. 12 14193 Berlin 01149 (0)30 826 36 92 CONTACT PERSON:(Official Correspondent:) Thomas Bock Celeste Managementberatung Schoenhauser Str.. 73 C 13158 Berlin, Germany Tel: 01149 (0)30 912 085 94 Fax: 01149 (0)30 912 085 91 e-mail: t.bock@celeste-berlin.de ## (2) The name of the device, including the trade or proprietary name if 1.6.1.2 applicable, the common or usual name, and the classification name, if known; IN VITRO DIAGNOSTIC PRIMARY CALIBRATOR AS DESCRIBED IN 21 CFR PART 862.1150, | TRADE NAME | CALIBRATION SOLUTION FOR OSMOMAT 010 / 030 /<br>auto | |---------------------|------------------------------------------------------| | REGULATION NUMBER: | 862.1150 | | PRODUCT CODE: | JIS | | COMMON NAME: | CALIBRATOR, PRIMARY; CALIBRATION SOLUTION | | ADVISORY COMMITTEE: | CLINICAL CHEMISTRY, (75), CH | ## 1.6.1.3 (3) An identification of the legally marketed device to which the submitter claims equivalence. PREDICATE DEVICE: | 510(k)-number | Regulation number | Product Code | decision date | Applicant name | Device name | |---------------|-------------------|--------------|---------------|-------------------------------------------------|---------------------------------------| | K931834 | 862.1150 | JIS | 08-25-1993 | King Diagnostics,<br>registration #:<br>1827622 | Sodium/Potassium<br>standard modified | 5 x {1}------------------------------------------------ - 1.6.1.4 (4) A description of the device that is the subject of the premarket notification submission, such as might be found in the labeling or promotional material for the device, including an explanation of how the device functions, the scientific concepts that form the basis for the device, and the significant physical and performance characteristics of the device, such as device design, material used, and physical properties; A FREEZING-POINT OR CRYOSCOPIC OSMOMETER IS A DEVICE TO MEASURE THE NUMBER OF MOLECULES OR IONS DISSOLVED IN A FLUID, IN THIS CASE, WATER. THE PHYSICAL EFFECT THE MEASUREMENT IS BASED ON IS THAT DISSOLVED SUBSTANCES IN WATER DEPRESS THE FREEZING POINT OF THE SOLUTION. THIS DEPRESSION IN THE FREEZING TEMPERATURE ITSELF, MEASURED EXACTLY, GIVES AN EXACT MEASURE FOR HOW MANY MOLECULES OR IONS ARE DISSOLVED IN THE AQUEQUS CALIBRATOR OR SPECIMEN. THIS EFFECT IS PHYSICAL, WHICH MEANS THAT THE CHEMICAL PROPERTIES OF THE CALIBRATOR OR SPECIMEN REMAIN UNCHANGED. AS ANY DISSOLVABLE SUBSTANCE CAUSES THIS EFFECT, THE RESULT OF THE MEASUREMENT GIVES NO INFORMATION ABOUT THE IDENTITY AND/OR COMPOSITION OF THE SUBSTANCES DISSOLVED, IF THERE SHOULD BE SEVERAL. THE MEASURED VALUE EXPRESSES IF THE RANGE OF OSMOLALITY OF THE SPECIMEN IS TO BE EXPECTED AS BEING NORMAL FOR A SPECIMEN FROM A HEALTHY HUMAN BEING. OR THE OSMOLALITY SHOWS THAT SOMETHING IS WRONG WITH THE PATIENT. FURTHER MEDICAL EXAMINATIONS WOULD FOLLOW. IN THE CALIBRATOR THE ONLY SUBSTANCE DISSOLVED IS SODIUM CHLORIDE. SO THE VALUE OF THE OSMOLALITY ALSO REPRESENTS THE CONCENTRATION OF SODIUM CHLORIDE. TO MEASURE THE FREEZING-POINT-DEPRESSION. THE SAMPLE IS FROZEN CAREFULLY. THE EXACT TEMPERATURE OF THE FLUID IS MEASURED AT ANY TIME WHILE THE SAMPLE IS COOLED DOWN. THE TEMPERATURE AT WHICH THE FLUID FREEZES TO ICE IS DETERMINED EXACTLY AND THE INTERNAL CALCULATOR OF THE OSMOMETER CALCULATES AND DISPLAYS THE EQUIVALENT OSMOLALITY OF THE LIQUID. IN MATRICES LIKE URINE THE MEASUREMENT SHOWS IF THE KIDNEYS WORK PROPERLY ASSUMABLY. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines suggesting movement or progress. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 OCT 2 1 2003 GONOTEC GmbH c/o Mr. Thomas Bock CELESTE MANAGEMENTBEARTUNG SCHOENHAUSER STRASSE 73C BERLIN GERMANY 14193 Re: k032608 > Trade/Device Name: Calibration Solution for Osmomat 010/030/auto Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIS Dated: August 19, 2003 Received: August 25, 2003 Dear Mr. Brock: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ ## Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## 4.2 Indications for Use IVD DEVICE CALIBRATION IS MOST COMMONLY PERFORMED USING CALIBRATORS (REFERENCE MATERIALS) SPECIFICALLY INTENDED TO BE USED AS A STANDARD CURVE OR CUT-OFF POINT FOR AN ASSAY. A CALIBRATOR HAS AN ASSIGNED VALUE THAT IS ESTABLISHED BY THE MANUFACTURER BY A REFERENCE METHOD. CALIBRATORS EXIST IN A VARIETY OF MATRICES SUCH AS SIMULATED AQUEOUS, SERUM. PLASMA OR OTHER TYPES OF SPECIMENS. PRIMARY REFERENCE CALIBRATORS ARE HIGHLY PURIFIED CHEMICALS THAT CAN BE DIRECTLY WEIGHED OR MEASURED TO PRODUCE A SOLUTION OF KNOWN CONCENTRATION. ALTERNATIVELY, THEY MAY BE MORE COMPLEX BIOLOGICAL MATERIALS HAVING RECEIVED A VALUE ASSIGNMENT USING REFERENCE (STANDARD) METHODOLOGY. THEY ARE SUPPLIED WITH A CERTIFICATE OF ANALYSIS FOR EACH LOT (FOR EXAMPLE, STANDARD REFERENCE MATERIALS (SRMS) FROM THE U.S. NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY (NIST)). THE PRIMARY CALIBRATOR "CALIBRATION SOLUTION FOR OSMOMAT 010 / 030 / auto" IS USED TO CALIBRATE THE OSMOMETERS "OSMOMAT 030" AND "OSMOMAT AUTO", WHICH BOTH ARE FREEZING-POINT OSMOMETERS FOR USE IN CLINICAL CHEMISTRY. THE OSMOLALITY OF THE CALIBRATOR IS EQUAL TO THE OSMOLALITY OF BODY-FLUIDS (ISOTONIC SOLUTION), PROVIDING THAT THE OSMOMETER WILL BE CALIBRATED CORRECTLY FOR THE FLUIDS TO BE MEASURED FOR MEDICAL PURPOSES. Jean Cooper Division Sign-Off Office of In Vitro Discussio Device Evaluation and Safe 510(k) K03 2608
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...